LORD SG-Link-200 User manual

LORD QUICK START GUIDE
SG-Link®-200
Wireless Analog Input Node
Wireless Simplicity, Hardwired Reliability™
The SG-Link®-200 is a 3-channel wireless sensor with a rugged, weatherproof
enclosure. It includes onboard PGA, filtering, and a high-resolution ADC for
precise measurement of a large range of sensor types including strain gauges,
load cells, pressure transducers, and accelerometers. The SG-Link-200
is ideal for both test and measurement and long term condition monitoring
applications.
Use this document to deploy the SG-Link-200 for data collection. This includes
electrical wiring, mounting the device, and using SensorConnect software to
configure the node, start sampling, and display data.
Figure 1. SG-Link®-200
Indicator Behavior Node Status
Device status indicator
OFF Node is OFF
Rapid green flashing on start-up Node is booting up
1 (slow) green pulse per second Node is idle and waiting for a
command
1 green blink every 2 seconds Node is sampling
Blue LED during sampling Node is resynchronizing
Red LED Built-in test error
Table 1. Indicator Behaviors

SG-Link®-200 Quick Start Guide
2
Pin # Name Description Pin Type Range
1VIN External supply voltage Power input 4.0 to 36 V
2GND Ground Power return GND
3NC No Connect -- --
4NC No Connect -- --
5NC No Connect -- --
6S3+
Channel 3 Three-wire excitation.
Sensor excitation for three wire configuration of
quarter bridge strain gauges.
Short to S3- when using two-wire configuration.
Analog Output 0 to 2.5 V
7S2- Channel 2 Sensor Input Analog Input 0 to 2.5 V
8GND Sensor Ground GND GND
9S1+
Channel 1 Three-wire excitation.
Sensor excitation for three wire configuration of
quarter bridge strain gauges.
Short to S1- when using two-wire configuration.
Analog Output 0 to 2.5 V
10 S3- Channel 3 Sensor Input Analog Input 0 to 2.5 V
11 GND Sensor Ground GND GND
12 S2+
Channel 2 Three-wire excitation.
Sensor excitation for three wire configuration of
quarter bridge strain gauges.
Short to S2- when using two-wire configuration.
Analog Output 0 to 2.5 V
13 S1- Channel 1 Sensor Input Analog Input 0 to 2.5 V
14 GND Sensor Ground GND GND
Table 3. SG-Link-200 Quarter-Bridge Pinout Descriptions
1. Pinout and Sensor Wiring
Pin # Name Description Pin Type Range Optional
Cable
1VIN External supply voltage Power input 4.0 to 36 V Red
2GND Ground Power return GND Black
3SP
Sensor Excitation Voltage On
continuously or duty cycle to the
sensor to save power
Analog Output 1.5V/2.5V
(100 mA) White
4SP
Sensor Excitation Voltage On
continuously or duty cycle to the
sensor to save power
Analog Output 1.5V/2.5V
(100 mA) White
5SP
Sensor Excitation Voltage On
continuously or duty cycle to the
sensor to save power
Analog Output 1.5V/2.5V
(100 mA) White
6S3+ Channel 3 Sensor Input +
(Full-bridge only) Analog Input 0 to 2.5 V Purple
7S2- Channel 2 Sensor Input - Analog Input 0 to 2.5 V White,
yellow stripe
8GND Sensor Ground GND GND White,
green stripe
9S1+ Channel 1 Sensor Input +
(Full-bridge only) Analog Input 0 to 2.5 V Blue
10 S3- Channel 3 Sensor Input - Analog Input 0 to 2.5 V White,
purple stripe
11 GND Sensor Ground GND GND White,
green stripe
12 S2+ Channel 2 Sensor Input +
(Full-bridge only) Analog Input 0 to 2.5 V Yellow
13 S1- Channel 1 Sensor Input - Analog Input 0 to 2.5 V White,
blue stripe
14 GND Sensor Ground GND GND White,
green stripe
Table 2. SG-Link-200 Full and Half-Bridge Pinout Descriptions with Optional
Cable Colors

SG-Link®-200 Quick Start Guide
3
Figure 4. Two Wire Quarter-Bridge*
Figure 5. Three Wire Quarter-Bridge*
* Requires optional on-board bridge completion
Figures 2 through 5 illustrate Full-Bridge (Standard), Half-Bridge, Two Wire
Quarter-Bridge, and Three Wire Quarter-Bridge wiring.
Figure 2. Full-Bridge (Standard)
Figure 3. Half-Bridge*

SG-Link®-200 Quick Start Guide
4
3. Node Operational Modes
Sensor nodes have three operational modes: active, sleep, and idle. When
sampling, the node is in active mode. When sampling stops, the node switches
into idle mode, which is used for configuring node settings and allows toggling
between active and sleep mode. The node will automatically enter into sleep
mode after a user-determined period of inactivity. The node will not enter into
sleep mode while active.
Figure 7. Node Operational Modes
2. Mounting Recommendations
The SG-Link-200 may be mounted using the holes sized for M8 fasteners.
Stainless steel compression limiters are rated for up to class 10.9 bolts
Recommended installation torque is 20 Nm ± 2Nm.
Figure 6. Mounting Dimensions

SG-Link®-200 Quick Start Guide
5
6. Connect to Nodes
Several methods can be used in SensorConnect to establish communication
with the nodes: the automatic node discovery on the same frequency, automatic
node discovery on a different frequency, and add node manually.
A. Automatic Node Discovery on Same Frequency
If the base and node are on the same operating frequency, the node will
populate below the Base Station listing when powering on the RTD-Link-200.
Figure 9. Node Discovered On Same Frequency
B. Automatic Node Discovery on Different Frequency
If a red circle with a number appears next to the Base Station, the node may
be operating on a separate radio channel. Select the Base Station and then
select the Nodes on Other Frequencies tile.
Figure 10. Node On Other Frequency
4. Install Software
Install the SensorConnect software on the host computer before connecting
hardware. Access the free software download on the LORD Sensing website.
http://www.microstrain.com/software
5. Establish Gateway Communication
Drivers for the USB gateways are included the SensorConnect software
installation. With the software installed, the USB gateway will detect
automatically whenever the gateway is plugged in.
1. Power is applied to the gateway through the USB connection. Verify
the gateway status indicator is illuminated, showing the gateway is
connected and powered on.
2. Open the SensorConnect™ software.
The gateway should appear in the Controller window with a communication
port assignment. If the gateway is not discovered, verify the port is active on
the host computer, then remove and re-insert the USB connector.
Figure 8. USB Gateway Communication

SG-Link®-200 Quick Start Guide
6
If the node was successfully added, two confirmation messages appear and it
will be listed under the Base Station.
Figure 13. Add Node Confirmation
If the node fails to add, a failure message appears. The node did not respond
to the base station which could indicate the node is not in idle mode or may
be on another frequency. If “Add Node Anyway” is selected, it associates that
node with the channel entered, but likely causes a communication error. If the
node was not in idle, move the base station to the frequency of the node and
issue a “Set to Idle” command.
Figure 14. Failure to Add Node
Highlight the new node being added and select Move Node to Frequency (#).
Figure 11. Move Node
C. Manually Add Node
Adding a node manually requires entering the node address and its current
frequency setting.
Figure 12. Add Node By Address

SG-Link®-200 Quick Start Guide
7
7. Configure Node
Node settings are stored to non-volatile memory and are configured using
SensorConnect. Select the Node under the Devices menu. Click on the
Configure button under the Setup heading.
Figure 15. Node Configuration Menu
The Hardware menu tab displays the current node settings.
1. To change the default Input range settings, use the dropdown menus
found under Input Range. This setting is used only when the Sensor
Type is set to Uncompensated Resistance.
2. To change the default Low Pass Filter settings, use the dropdown menu
found under Filter Cutoff.
3. Select Apply Configuration to write to node memory.
Figure 16. Hardware Configuration Tab

SG-Link®-200 Quick Start Guide
8
8. Configure Sampling Setting and Start Data Acquisition
1. Click the Base Station > Sample Network tile, and indicate the nodes to
be sampled by checking the box to the left of each node in the list.
Figure 17. Select Node To be Sampled
2. Under the Sampling column, select Sample Rate from the drop down
menu. Select Continuously to sample indefinitely.
Figure 18. Sampling Setting
3. Select Apply and Start Network. A message indicates the Network
Started and includes a shortcut link to the Data Tab.
Figure 19. Network Started with shortcut to Data Tab
4. Navigate to the Data tab and view the Empty Dashboard 1. Click the
+Add Widget button to add a Widget to Dashboard 1.
Figure 20. +Add Widget

SG-Link®-200 Quick Start Guide
9
9. Radio Specifications
The SG-Link-200 employs a 2.4GHz IEEE 802.15.4 compliant radio transceiver
for wireless communication. The radio is a direct-sequence spread spectrum
radio and can be configured to operate on 16 separate frequencies ranging from
2.405 GHz to 2.480 GHz. Following the 802.15.4 standard, these frequencies
are aliased as channels 11 through 26. For all newly manufactured nodes, the
default setting is 2.425 GHz (channel 15).
SG-Link®-200
FCC ID: XJQMSLINK0011
IC ID: 8505A-MSLINK00 11
This device complies with Part 15 of the United States FCC Rules, and
Industry Canada’s license-exempt RSSs. Operation is subject to the
following two conditions: 1) This device may not cause interference, and
2) This device must accept any interference, including interference that
may cause undesired operation of the device. Changes or modifications,
including antenna changes not expressly approved by LORD Corporation
could void the user’s authority to operate the equipment.
Cet appareil est conforme à la Partie 15 des Règles de la FCC des États-Unis
et aux RSSS exempts de licence d’Industrie Canada. Le fonctionnement est
soumis aux deux conditions suivantes: 1) Cet appareil ne doit pas causer
d’interférences et 2) Cet appareil doit accepter toute interférence, y compris
les interférences pouvant entraîner un fonctionnement indésirable de
l’appareil. Les changements ou modifications, y compris les changements
d’antenne non expressément approuvés par LORD Corporation, pourraient
annuler l’autorisation de l’utilisateur d’utiliser l’équipement.
5. Select the Time Series tile to add a Time Series graph to Dashboard 1.
The Time Series Graph displays with “No channels selected”. View the
available channels and options under the Node’s
V
Menu.
Figure 21. Add Time Series Graph Channel Selection
6. Select the desired channels and options to begin data collection.
Figure 22. Time Series Graph with Channel 1 Selected

SG-Link®-200 Quick Start Guide
LORD Corporation MicroStrain®Sensing Systems
459 Hurricane Lane, Suite 102 Williston, VT 05495 USA
Phone: 802-862-6629
Copyright © 2019 LORD Corporation
OD UI8001 (Rev.0 3/19) | Document 8501-0104 Revision -. Subject to change without notice. www.microstrain.com
10. Battery Hazards
The SG-Link-200 contains internal, non-rechargeable lithium batteries. Lithium
batteries are a fire and explosion hazard. Do not store or operate the node at
temperatures above 212°F (100°C). Do not disassemble, short circuit, crush,
puncture, or otherwise misuse the battery.
Lithium batteries contain toxic chemicals that are harmful to humans and the
environment. Disposal is subject to federal and local laws. Do not discard the
battery or the node in the trash. Follow proper battery disposal protocol, or
contact LORD Sensing Technical Support for information on extracting the
battery or returning the product for proper recycling and disposal.
11. Power Supply
Apply only the input voltage range specified for the SG-Link-200. Connect
to a power source that is near the device, is accessible, and adheres to all
national wiring standards. Compliance with wiring standards is assumed in
the installation of the power source and includes protection against excessive
currents, short circuits, and ground faults. Failure to do so could result in
personal injury and permanent damage to the device.
Table of contents
Other LORD Accessories manuals

LORD
LORD Lord Fusor 117 User manual

LORD
LORD MV5-AR User manual

LORD
LORD 3DM-CV5-25 User manual

LORD
LORD MicroStain 3DM-GX5-45 User manual

LORD
LORD V-Link-200 User manual

LORD
LORD MicroStrain 3DM-GX5-15 User manual

LORD
LORD Torque-Link-LXRS User manual

LORD
LORD MicroStrain 3DM-GX5-25 User manual

LORD
LORD V-Link-200 User manual
Popular Accessories manuals by other brands

OZ Trampolines
OZ Trampolines Round In-Ground Trampoline Information sheet

CKD
CKD PPX Series manual

Clearaudio
Clearaudio TT5 user manual

Opale-Paramodels
Opale-Paramodels Pilot Ben user manual

Juniper
Juniper IDP 600 Series Installer's guide

Cooper Controls
Cooper Controls Greengate OAWC-DT-120W installation instructions