
MODEL9-3 TechnicalManual Section2
Ludlum Measurements, Inc. Page 2-2 June 2018
Instrument Test
Note:
The Model 9-3 has two options for chamber wall voltage. An
internal jumper may be selected to allow continuous voltage on
the chamber wall with the instrument turned off. This option
allows a faster, three-minute settling time when the instrument is
frequently used. If the instrument is infrequently used, it is
recommended that this option not be used, because of battery
drain. Battery life with the continuous wall voltage on and
instrument off is approximately six months.
Without this option, the battery drain is zero with the instrument
OFF. When the instrument is turned on, a settling time of 5
minutes is required for a reading of less than 0.2 mR/hr, 10
minutes for less than 0.1 mR/hr, and several hours for readings
near zero.
The instrument will be shipped with the chamber wall voltage
turned on when the instrument is turned off.
After checking the batteries with the range switch on the ×10K position, turn
the instrument range switch to the ×1 position. Note that as the selector switch
moves from ×100 to ×10, a meter transient will occur. This transient is caused
by an internal range relay and is normal. Allow time for the ×1 scale to stabilize.
Open the beta shield slide. Switch the AUD ON OFF switch to ON. Expose
the chamber window to a check source. Observe that the audio click frequency
increases as the meter reading increases.
Note:
The RESET circuit discharges the chamber and opens the
chamber connection to the electrometer. This causes a transient
on the ×1 and ×10 scales. The meter will indicate full scale and
count-down for five seconds when the RESET switch is pressed
and released. For near zero reading on ×1 scale, allow several
more seconds for setting time after RESET is released.