BEFORE USE ....
Thank you for choosing M-System. Before use, please check
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System’s Sales Office or representatives.
■PACKAGE INCLUDES:
DC alarm (body + base socket + input resistor) ................(1)
Input resistor is provided only with current input type.
■MODEL NO.
Confirm Model No. marking on the product to be exactly
what you ordered.
■INSTRUCTION MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection and
basic maintenance procedures.
POINTS OF CAUTION
■CONFORMITY WITH EU DIRECTIVES
• This equipment is suitable for Pollution Degree 2, Meas-
urement Category II (output, transient voltage 2500V)
and Installation Category II (transient voltage 2500V).
Basic insulation (signal input to output to power input:
300V) is maintained. Prior to installation, check that the
insulation class of this unit satisfies the system require-
ments.
• Altitude up to 2000 meters.
• The equipment must be mounted inside a panel.
• The equipment must be installed such that appropriate
clearance and creepage distances are maintained to con-
form to CE requirements. Failure to observe these re-
quirements may invalidate the CE conformance.
• The actual installation environments such as panel con-
figurations, connected devices, connected wires, may af-
fect the protection level of this unit when it is integrated
in a panel system. The user may have to review the CE
requirements in regard to the whole system and employ
additional protective measures to ensure the CE conform-
ity.
• Install lightning surge protectors for those wires connect-
ed to remote locations.
■POWER INPUT RATING & OPERATIONAL RANGE
• Locate the power input rating marked on the product and
confirm its operational range as indicated below:
100 – 120V AC rating: 85 – 132V, 47 – 66 Hz, approx. 3VA
200 – 240V AC rating: 170 – 264V, 47 – 66 Hz, approx. 3VA
■GENERAL PRECAUTIONS
• Before you remove the unit from its base socket or mount
it, turn off the power supply and input signal for safety.
■ENVIRONMENT
• Indoor use.
• When heavy dust or metal particles are present in the
air, install the unit inside proper housing with sufficient
ventilation.
• Do not install the unit where it is subjected to continuous
vibration. Do not subject the unit to physical impact.
• Environmental temperature must be within -5 to +55°C
(23 to 131°F) with relative humidity within 30 to 90% RH
in order to ensure adequate life span and operation.
■WIRING
• Do not install cables close to noise sources (relay drive
cable, high frequency line, etc.).
• Do not bind these cables together with those in which
noises are present. Do not install them in the same duct.
■AND ....
• The unit is designed to function as soon as power is sup-
plied, however, a warm up for 10 minutes is required for
satisfying complete performance described in the data
sheet.
COMPONENT IDENTIFICATION
100
50
%
0
KSE
Model
K
-UNIT
Connection Diagram
Base Socket
Body
Specifications
Input Resistor
■FRONT PANEL CONFIGURATION
Hi Monitor LED
Hi Setpoint Adj. Dial
100
50 %
0
DC ALARM
(dial adjustment; single trip) MODEL KSE-x1
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-3609-A Rev.2 P. 1 / 3
INSTRUCTION MANUAL
INSTALLATION
Detach the yellow clamps located at the top and bottom of
the unit for separate the body from the base socket.
Clamp
(top & bottom)
DIN Rail
35mm wide
Spring Loaded
DIN Rail Adaptor
Shape and size of the base socket
are slightly different with various
socket types.
■DIN RAIL MOUNTING
Set the base socket so that its
DIN rail adaptor is at the bot-
tom. Hang the upper hook at
the rear side of base socket on
the DIN rail and push in the
lower. When removing the
socket, push down the DIN
rail adaptor utilizing a minus
screwdriver and pull.
■WALL MOUNTING
Refer to “EXTERNAL DI-
MENSIONS.”
TERMINAL CONNECTIONS
Connect the unit as in the diagram below or refer to the connection diagram on the top of the unit.
When an input resistor is provided with the module, attach it together with input wiring to the input screw terminals.
■EXTERNAL DIMENSIONS unit: mm (inch)
3456
2187
80 (3.15)
50 (1.97)
[3.3 (.13)]
80 (3.15)
20
(.79)
40 (1.57)
50 (1.97)
7.8 (.31)
CLAMP
(top & bottom)
DIN RAIL
35mm wide
2–4.5 (.18) dia.
MTG HOLE
15 (.59) deep
8–M3.5
SCREW
•When mounting, no extra space is needed between units.
116 (4.57)
136 (5.35)
■CONNECTION DIAGRAM
+
–
*
INPUT
3
R
4
*Input shunt resistor attached for current input.
U
V
POWER
7
8
1
2
5
6
HI
LO
COM
COM
HI OUTPUT
■TERMINAL ASSIGNMENTS unit: mm (inch)
56
2187
34
12 (.47)
INPUT RESISTOR
(model: REM)
Input shunt resistor attached
for current input.
KSE-x1
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-3609-A Rev.2 P. 2 / 3
CHECKING
1) Terminal wiring: Check that all cables are correctly con-
nected according to the connection diagram.
2) Power input voltage: Check voltage across the terminal
7 – 8 with a multimeter.
3) Input: Check that the input signal is within 0 – 100% of
the full-scale.
4) Alarm operations: Check the alarm operations referring
to the figure below.
5) Output load: Check that the output load is 380V
AC/120VA or 125V DC/30W at the maximum (≤ 0.5A for
CE marking). For use with inductive load, external pro-
tection is recommended.
Alarm Trip Operation Terminal No. in parentheses
Trip Operation in Power Failure
: Terminals 5 – 6 turn ON.
050 100
Input
(%) ▲
Setpoint
Hi Output
(5–6)ON
(1–2)ON
ADJUSTMENT PROCEDURE
■HOW TO ADJUST ALARM SETPOINTS
Use a signal source and measuring instruments of sufficient
accuracy level. Turn the power supply on and warm up for
more than 10 minutes.
1) Turn the front dial to an approximate position for the
setpoint.
2) Vary simulated input signal around the setpoint and con-
firm relay operation and set.
MAINTENANCE
Regular calibration procedure is explained below:
■CALIBRATION
Warm up the unit for at least 10 minutes.
Increase the input signal from a value lower than the set-
point and check that the trip point remains within the ac-
curacy described in the data sheet.
When the setpoints are shifted, please contact M-System’s
Sales Office or representatives.
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protector for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.
KSE-x1
5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201 Fax: +81(6)6659-8510 E-mail: info@m-system.co.jp
EM-3609-A Rev.2 P. 3 / 3

Other M-system Measuring Instrument manuals

M-system 47 Series User manual

M-system

M-system 47 Series User manual

M-system KADS User manual

M-system

M-system KADS User manual

M-system 73VR3100 User guide

M-system

M-system 73VR3100 User guide

M-system R9WTU-ED User manual

M-system

M-system R9WTU-ED User manual

M-system TR30 User manual

M-system

M-system TR30 User manual

M-system R9EWTU User manual

M-system

M-system R9EWTU User manual

M-system 47LPT User manual

M-system

M-system 47LPT User manual

M-system 47DV User manual

M-system

M-system 47DV User manual

M-system 47 Series User manual

M-system

M-system 47 Series User manual

M-system R10-BS1 User manual

M-system

M-system R10-BS1 User manual

M-system R7CWTU User manual

M-system

M-system R7CWTU User manual

M-system M Series User manual

M-system

M-system M Series User manual

M-system AS4M User manual

M-system

M-system AS4M User manual

M-system 47LT User manual

M-system

M-system 47LT User manual

M-system 47 Series User manual

M-system

M-system 47 Series User manual

M-system 47 Series User manual

M-system

M-system 47 Series User manual

M-system 47LR User manual

M-system

M-system 47LR User manual

M-system 48NV User manual

M-system

M-system 48NV User manual

M-system 47 Series User manual

M-system

M-system 47 Series User manual

M-system 47DAC User manual

M-system

M-system 47DAC User manual

M-system 48NAV User manual

M-system

M-system 48NAV User manual

M-system R1MS-GH3 User manual

M-system

M-system R1MS-GH3 User manual

M-system 47LPA User manual

M-system

M-system 47LPA User manual

M-system R9MWTU User manual

M-system

M-system R9MWTU User manual

Popular Measuring Instrument manuals by other brands

Greenlee CSJ-100 instruction manual

Greenlee

Greenlee CSJ-100 instruction manual

Third Dimension GapGun Pro user guide

Third Dimension

Third Dimension GapGun Pro user guide

Sinar Technology 6070-BeanPro user guide

Sinar Technology

Sinar Technology 6070-BeanPro user guide

Miller Electric Selec-O-Gas H1230E owner's manual

Miller Electric

Miller Electric Selec-O-Gas H1230E owner's manual

PCB Piezotronics 355B02 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 355B02 Installation and operating manual

Chauvin Arnoux C.A 1246 quick start guide

Chauvin Arnoux

Chauvin Arnoux C.A 1246 quick start guide

SIGLENT SVA1075X user manual

SIGLENT

SIGLENT SVA1075X user manual

Keysight Technologies B2900A Series Demo guide

Keysight Technologies

Keysight Technologies B2900A Series Demo guide

Hanna Instruments HI 96710C instruction manual

Hanna Instruments

Hanna Instruments HI 96710C instruction manual

Gigahertz Solutions ME 3030B manual

Gigahertz Solutions

Gigahertz Solutions ME 3030B manual

Kleinwächter EFM 113B manual

Kleinwächter

Kleinwächter EFM 113B manual

GREISINGER electronic GMH3111 user manual

GREISINGER electronic

GREISINGER electronic GMH3111 user manual

Omega HHC280 instruction sheet

Omega

Omega HHC280 instruction sheet

Ametek Jofra CTC-155 A/C user manual

Ametek

Ametek Jofra CTC-155 A/C user manual

Viso Systems LabFlicker user manual

Viso Systems

Viso Systems LabFlicker user manual

Airmodus A11 user manual

Airmodus

Airmodus A11 user manual

ABB GLA232 Series user manual

ABB

ABB GLA232 Series user manual

Quantrol GTX005 user manual

Quantrol

Quantrol GTX005 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.