RHTEMP1000IS
Intrinsically Safe Temperature
and Humidity Data Logger
PRODUCT
USER GUIDE
To view the full MadgeTech product line, visit
our website at madgetech.com.
TABLE OF CONTENTS
2Product Overview
2Installation Guide
2Device Operation
3Device Maintenance
4Need Help?
Product User Guide | 2
RHTemp1000IS
PRODUCT USER GUIDE
Product Overview
The RHTemp1000IS carries hazardous location, intrinsically
safe certification in accordance with the latest issue of:
FM3600, FM3610, CAN/CSA-C22.2 No. 60079-0:15,
CAN/CSA-C22.2 No. 60079-11:14
Certified Intrinsically Safe for:
• Class 1 Division 1 Group ABCD
• Class 1 Division 2 Group ABCD
• Temperature Class: T4A
Operational Warnings
• When used in hazardous locations, the RHTemp1000IS
is to be installed prior to the location becoming
hazardous and removed only after the area is no longer
hazardous.
• The maximum allowed ambient temperature for the
RHTemp1000IS (under any circumstances) is 80 °C. The
minimum rated operating temperature is -40 °C.
• The RHTemp1000IS is approved for use only with the
Tadiran TL-2150/S battery. Replacement with any other
battery will void the safety rating.
• Batteries are user replaceable, but are to be removed or
replaced only in locations known to be non-hazardous.
• Tampering or replacement of non-factory components
may adversely affect the safe use of the product, and is
prohibited. Except for replacement of the battery, the
user may not service the RHTemp1000IS. MadgeTech,
Inc. or an authorized representative must perform all
other service to the product.
Ordering Information
• 902213-00 — RHTemp1000IS
• 902218-00 — RHTemp1000IS-KR (key ring end cap)
• 900319-00 — IFC400
• 900325-00 — IFC406
• 901745-00 — Battery Tadiran TL-2150/S
Installation Guide
Installing the Software
The Software can be downloaded from the MadgeTech
website at madgetech.com. Follow the instructions
provided in the Installation Wizard.
Installing the USB Interface Drivers
IFC400 or IFC406 — Follow the instructions provided in
the Installation Wizard to install the USB Interface Drivers.
Drivers can also be downloaded from the MadgeTech
website at madgetech.com.
Device Operation
Connecting and Starting the Data Logger
1. Once the software is installed and running, plug the
interface cable into the docking station (IFC400 or
IFC406).
2. Connect the USB end of the interface cable into an
open USB port on the computer.
3. Place the data logger into the docking station (IFC400
or IFC406).
4. The data logger will automatically appear under
Connected Devices within the software.
5. For most applications, select Custom Start from
the menu bar and choose the desired start method,
reading rate and other parameters appropriate for the
data logging application and click Start. (Quick Start
applies the most recent custom start options, Batch
Start is used for managing multiple loggers at once,
Real Time Start stores the dataset as it records while
connected to the logger.)
6. The status of the device will change to Running or
Waiting to Start, depending upon your start method.
7. Disconnect the data logger from the interface cable
and place it in the environment to measure.
Note: The device will stop recording data when the end of memory is
reached or the device is stopped, unless user selectable memory wrap
is enabled. At this point the device cannot be restarted until it has been
re-armed by the computer.
Product User Guide | 3
Device Operation (cont’d)
Downloading Data from a Data Logger
1. Place the logger into the docking station (IFC400 or
IFC406).
2. Highlight the data logger in the Connected Devices
list. Click Stop on the menu bar.
3. Once the data logger is stopped, with the logger
highlighted, click Download.
4. Downloading will offload and save all the recorded
data to the PC.
Device Maintenance
Battery Replacement
Materials: Replacement Battery (Tadiran TL-2150/S)
1. Move the device to a non-hazardous location before
replacing battery.
2. Observe Operational Warnings when removing and
replacing the battery.
3. Unscrew the bottom of the data logger and remove the
battery.
4. Place the new battery into the logger. Caution: Observe
correct battery polarity when installing.
5. Screw the cover onto the data logger.
RHTemp1000IS
PRODUCT USER GUIDE
O-Rings
O-ring maintenance is a key factor when properly caring
for the RHTemp1000IS. The O-rings ensure a tight
seal and prevent liquid from entering the inside of the
device. Please refer to the application note “O-Rings 101:
Protecting Your Data”, found at madgetech.com, for
information on how to prevent O-ring failure.
Recalibration
Recalibration is recommended annually. To send devices
back for calibration, visit madgetech.com.
Additional Services:
Custom calibration and verification point options
available, please call for pricing.
Call for custom calibration options to accommodate specific application needs.
Prices and specifications subject to change. See MadgeTech’s terms and conditions
at madgetech.com.
To send devices to MadgeTech for calibration, service or repair, please use the
MadgeTech RMA Process by visiting madgetech.com.
Communication
To ensure desired operation of the RHTEMP1000IS, please keep the surface clear of any foreign objects
or substances. The RHTEMP1000IS’s data is downloaded through external contact with the IFC400 or
IFC406 docking station. Covering the surface with foreign objects (i.e. Calibration Labels) can prevent the
communication and/or downloading process.
RHTemp1000IS
NEED HELP?
DOC-1272036-00 REV 2 2020.08.18
6 Warner Road, Warner, NH 03278
(603) 456-2011
info@madgetech.com
madgetech.com
Product Support & Troubleshooting:
• Visit our Resources online at madgetech.com/resources.
• Contact our friendly Customer Support Team at (603) 456-2011 or support@madgetech.com.
MadgeTech 4 Software Support:
• Refer to the built-in help section of the MadgeTech 4 Software.
• Download the MadgeTech 4 Software Manual at madgetech.com.
• Contact our friendly Customer Support Team at (603) 456-2011 or support@madgetech.com.

Other MadgeTech Data Logger manuals

MadgeTech QuadTemp2000 User manual

MadgeTech

MadgeTech QuadTemp2000 User manual

MadgeTech VFC2000 Specification sheet

MadgeTech

MadgeTech VFC2000 Specification sheet

MadgeTech PRTemp1000 Specification sheet

MadgeTech

MadgeTech PRTemp1000 Specification sheet

MadgeTech HiTemp140-FP-6 Specification sheet

MadgeTech

MadgeTech HiTemp140-FP-6 Specification sheet

MadgeTech RFRTDTemp2000A Specification sheet

MadgeTech

MadgeTech RFRTDTemp2000A Specification sheet

MadgeTech RHTemp2000 User manual

MadgeTech

MadgeTech RHTemp2000 User manual

MadgeTech RFCurrent2000A Specification sheet

MadgeTech

MadgeTech RFCurrent2000A Specification sheet

MadgeTech RHTemp2000 User manual

MadgeTech

MadgeTech RHTemp2000 User manual

MadgeTech CryoTemp Specification sheet

MadgeTech

MadgeTech CryoTemp Specification sheet

MadgeTech RFpHTemp101A User manual

MadgeTech

MadgeTech RFpHTemp101A User manual

MadgeTech VFC2000-MT Specification sheet

MadgeTech

MadgeTech VFC2000-MT Specification sheet

MadgeTech Cryo-Temp User manual

MadgeTech

MadgeTech Cryo-Temp User manual

MadgeTech HiTemp140-PT Specification sheet

MadgeTech

MadgeTech HiTemp140-PT Specification sheet

MadgeTech RTDTemp2000 User manual

MadgeTech

MadgeTech RTDTemp2000 User manual

MadgeTech Shock101-EB Specification sheet

MadgeTech

MadgeTech Shock101-EB Specification sheet

MadgeTech RFRTDTemp2000A Specification sheet

MadgeTech

MadgeTech RFRTDTemp2000A Specification sheet

MadgeTech AVS140-1 Specification sheet

MadgeTech

MadgeTech AVS140-1 Specification sheet

MadgeTech PR2009 Specification sheet

MadgeTech

MadgeTech PR2009 Specification sheet

MadgeTech Shock101-5 User manual

MadgeTech

MadgeTech Shock101-5 User manual

MadgeTech QuadVoltV2 Specification sheet

MadgeTech

MadgeTech QuadVoltV2 Specification sheet

MadgeTech X Series Specification sheet

MadgeTech

MadgeTech X Series Specification sheet

MadgeTech 902003-00 Specification sheet

MadgeTech

MadgeTech 902003-00 Specification sheet

MadgeTech HiTemp140-M12 Specification sheet

MadgeTech

MadgeTech HiTemp140-M12 Specification sheet

MadgeTech TC101A User manual

MadgeTech

MadgeTech TC101A User manual

Popular Data Logger manuals by other brands

SDI G-Logger 3330 quick start guide

SDI

SDI G-Logger 3330 quick start guide

DTS TSR AIR user manual

DTS

DTS TSR AIR user manual

Hartmann CML00 manual

Hartmann

Hartmann CML00 manual

GRAPHTEC midi LOGGER GL240 Q&A

GRAPHTEC

GRAPHTEC midi LOGGER GL240 Q&A

inelo TachoReader Combo Plus Tutorial

inelo

inelo TachoReader Combo Plus Tutorial

Furuno VR-7000 instruction manual

Furuno

Furuno VR-7000 instruction manual

LogTag Recorders UTREL30-16 Product user guide

LogTag Recorders

LogTag Recorders UTREL30-16 Product user guide

TFA Dostmann KlimaLogg Pro instruction manual

TFA Dostmann

TFA Dostmann KlimaLogg Pro instruction manual

Omega OM-EL-USB Series user guide

Omega

Omega OM-EL-USB Series user guide

Fourtec DaqPRO 5300 user guide

Fourtec

Fourtec DaqPRO 5300 user guide

RAWLINK 36685 instruction manual

RAWLINK

RAWLINK 36685 instruction manual

Evenmore DL-200BT user guide

Evenmore

Evenmore DL-200BT user guide

SIAP+MICROS t026 TTEPRH User manual and maintenance

SIAP+MICROS

SIAP+MICROS t026 TTEPRH User manual and maintenance

JRC JCY-1800 - instruction manual

JRC

JRC JCY-1800 - instruction manual

TESTO 580 instruction manual

TESTO

TESTO 580 instruction manual

Elma Instruments DT 178A manual

Elma Instruments

Elma Instruments DT 178A manual

Hioki 3645-20 instruction manual

Hioki

Hioki 3645-20 instruction manual

VOLTCRAFT DL-111K operating instructions

VOLTCRAFT

VOLTCRAFT DL-111K operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.