T10 Turbidity Meter
Quick Start Guide for Benchtop Systems
•Perform a quality control check by analyzing a known standard from the calibration kit, secondary standards kit, or prepared
from a stock solution. If the result of the quality control check is more than 5% different from the expected value, re-read the
QC. If it continues to read >5% off expected, perform a new guided and free calibration.
•Ensure samples are well mixed by slowly inverting the sample containers 10-20 times. Avoid causing bubbles to form in the
samples.
•Pour samples directly into the empty cuvettes, being sure to keep track of which sample is which. You may use the top of the
cap to identify samples if preferred, but do NOT put any markings or identification directly on the cuvette glass
•Ensure the cap is screwed on tight and the cuvette is filled to the neck
•
U
se a lint-free wipe to wipe down the sides of the cuvette, then place it in the measuring chamber and close the lid
•
Press the button and the meter will start the measurement. A progress bar on the meter screen shows the analysis
proceeding. You will see the NTU result appear on screen once done.
•
Repeat the above process for all samples in the set.
What do I do if my QC checks are failing?
1. Re-pour QC standard from bottle in a different clean cuvette and read again
a. If second attempt does not pass, move to step 2
2. Re-attempt free calibration at one or several points according to your standard operating procedure
a. Then re-attempt a read of the QC standard
b. If third attempt does not pass, move to step 3
3. Re-attempt guided calibration using sealed cuvette standards
a. Then re-attempt free calibration, followed by a read of a QC standard
b. If this attempt does not pass, move to step 4
4. Purchase a new sealed calibration kit
a. The shelf life for the kits is 2 years, however this can sometimes be shorter depending on storage conditions
b. After acquiring the new kit, repeat step 3 above.
c. If the QC standard is still not passing, contact MANTECH support to send your T10 meter in for service
How often should I calibrate?
1. Guided calibrations are recommended at a minimum once every three months
2. Free calibrations are recommended at a minimum once per month
3. QC checks are recommended every day that you are analyzing samples
How do I adjust the measuring time?
1. Hold the MENU button for three seconds
2. Use the Arrows and READ buttons to navigate to Menu>Config>Instrument>Sample
3. This setting adjusts the number of measurements used to calculate each result. Higher numbers will result in longer measuring
time.
4. Use the Arrow keys to adjust the setting between 8-100.
5. Hold the SAVE button to save the new setting, then press the ESC button until the NTU reading is back on screen
6. Perform a new guided calibration and free calibration before performing sample analysis
Questions/Assistance – support@mantech-inc.com
Quotes/Product Info – info@mantech-inc.com
Phone – +1-866-763-2122