
| 2| CS100®/CS100i®| Abbreviated Operator’s Guide |
OPERATING MODES
AUTO Operation Mode
All aspects of pump Operation are automated.
Most appropriate trigger source is automatically selected.
Inflation and deflation timing is automatically set.
Lead I, II, III, or External ECG source is automatically selected.
System searches for the next best trigger source in the event of a loss of
trigger source.
Timing adjusts automatically if trigger source changes, heart rate changes, rhythm
changes.
With sustained unpredictable rhythms, R-wave deflation is automatically selected.
The message “Auto R-Wave Deflate” will be displayed. The deflation indicator
will automatically be repositioned to the far right, “R-Trac” will be displayed in the
IAB deflation indicator. The system will return to predictive timing once the rhythm
becomes predictable.
SEMI AUTO Operation Mode
Operator selects the most appropriate trigger source.
Operator establishes initial timing and thereafter, software algorithms
automatically track changes in patient heart rate or rhythm and adjusts
timing accordingly.
Changing the trigger source will cause the pump to go to Standby. Pumping will
resume when the START key is pressed.
Lead I, II, III, AVR, AVL, AVF, V or External ECG source can be selected.
With sustained unpredictable rhythms and a valid ECG Trigger, R-wave
deflation is automatically selected. The message “Auto R-Wave Deflate” will be
displayed. The deflation indicator will automatically be repositioned to the far right,
“R-Trac” will be displayed in the IAB deflation indicator. The system will return to
predictive timing once the rhythm becomes predictable.
Loss of trigger causes an alarm and pumping will stop.
Manual Operation Mode
Trigger source and timing of IAB inflation and deflation is determined by
the operator.
Changing the trigger source will cause the pump to go to Standby. Pumping will
resume when the START key is pressed.
Lead I, II, III, AVR, AVL, AVF, V or External ECG source can be selected.
Timing must be readjusted by the operator if heart rate or rhythm changes.
Loss of trigger causes an alarm and pumping will stop.
Manual timing is typically used for pediatric IABP patients.
WARNING: The CS100 must be in Semi-Auto Operational mode whenever no aortic
pulse is present, and IAB assist is desired. For example, whenever circulatory bypass or
a laminar flow, left ventricular assist device is in use.