Maxtec Handi+ User manual

R218M12 REV. M
Instructions for Use
ENGLISH
Handi+
MEDICAL

ENGLISH 2WWW.MAXTEC.COM • (800) 748-5355
CLASSIFICATION
Classification....................................................................................................Class II medical device
Protection against electric shock.................................................... Internally powered equipment
Protection against water...............................................................................................................IPX4
Mode of Operation .............................................................................................................Continuous
Sterilization.................................................................................................................. See section 6.0
Flammable anesthetic mixture .....................................................Not suitable for use in presence
of a flammable mixture
Product Disposal Instructions:
The sensor, batteries, and circuit board are not suitable for regular trash dis-
posal. Return sensor to Maxtec for proper disposal or dispose according to local
guidelines. Follow local guidelines for disposal of other components.
WARRANTY
Maxtec, warrants the Handi+ to be free from defects of workmanship or materials for a period
of two-(2) years from the date of shipment from Maxtec, under normal operating conditions
and provided that the Handi+ is properly operated and maintained in accordance with Maxtec's
operating instructions. Should Handi+ fail prematurely, the replacement Handi+ is warranted
for the remainder of the original analyzer warranty period. Based on Maxtec's product evalua-
tion, Maxtec's sole obligation under the foregoing warranty is limited to making replacements,
repairs or issuing credit for equipment found to be defective. This warranty extends only to
the buyer purchasing the equipment directly from Maxtec or through Maxtec's designated
distributors and/or agents as new equipment. Routine maintenance items are excluded from
this warranty. Maxtec shall not be liable to the purchaser or other persons for incidental or
consequential damages or equipment that has been subject to abuse, misuse, misapplica-
tion, alteration, negligence or accident. THESE WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL
OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTY OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE.
WARNING
Indicates a potentially hazardous situation, if not avoided, could result in death or serious
injury.
◆ Before use, all individuals who will be using this product must become thoroughly familiar
with the information contained in this Operation Manual. Strict adherence to the operat-
ing instructions is necessary for safe and effective product performance. This product will
perform only as designed and only if installed and operated in accordance with the manu-
facturer's operating instructions.
◆ Although the sensor of this device has been tested with various anesthesia gases including
nitrous oxide, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane and found to
have acceptably low interference, the device in entirety (including electronics) is not suit-
able for use in the presence of a flammable anesthetic mixture with air or with oxygen or
nitrous oxide. Only the threaded sensor face, flow diverter, and “T” adapter may be allowed
to contact such a gas mixture.
◆ This device does not contain automatic barometric pressure compensation.
DO NOT use near any type of flame or flammable/explosive substances, vapors or
atmosphere.
◆ Not for use in an MRI environment.
◆ Improper use of this devicecan cause inaccurate oxygen readings which can lead to improper
treatment, hypoxia or hyperoxia. Follow the procedures outlined in this user manual.
CAUTION
Indicates a potentially hazardous situation, if not avoided, could result in minor or moderate
injury and property damage.
◆ The Maxtec MAX-250 oxygen sensor is a sealed device containing a weak acid electrolyte,
lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste constituents and
should be disposed of properly, or returned to Maxtec for proper disposal or recovery.
◆ The Handi+ is not intended for steam, ethylene oxide or radiation sterilization.
DO NOT autoclave or expose the sensor to high temperatures.
DO NOT immerse the Handi+ oxygen analyzer in any cleaning solution.
◆ The flow diverter provided with the Handi+ is for use with flowing gases only.
DO NOT use the diverter when performing static sampling (e.g., in incubators, oxygen
tents, oxygen hoods).
DO NOT attempt any repairs or procedures, which are not described in this Operation
Manual. Maxtec cannot warrant this product from damage resulting from misuse, unau-
thorized repair or improper maintenance of this product.
◆ The Federal (USA) law restricts this device to sale by or on the order of a physician.
◆ There are no internal user-serviceable parts.
◆ Never install the sensor in a location that will expose the sensor to patient’s exhaled breath
or secretions.
SYMBOL GUIDE
The following symbols and safety labels are found on the Handi+:
Follow instructions for use On/O Button
Warning Calibration Button
Meets ETL standards
Do not throw away.
Follow local guidelines
for disposal
Do Not Manufacturer
Contains acid Caution
Catalog Number Serial Number
Lot code/Batch code Ingress Protection
Rating
Federal law (USA) restricts
this device to sale by or
on order of a physician
Authorized
Representative
in the European
Community
50°C
(122°F)
-15°C
(5°F)
Storage Temperature Range
AUTHORIZED REPRESENTATIVE:
QNET BV, Kantstraat 19, NL-5076 NP Haaren, The Netherlands
NOTE: The latest edition of this operating manual can be downloaded from our website at www.maxtec.com
Maxtec
2305 South 1070 West
Salt Lake City, Utah 84119
USA
phone: (800) 748.5355
fax: (801) 973.6090
email: sales@maxtec.com
web: www.maxtec.com
EN Instructions for Use

WWW.MAXTEC.COM • (800) 748-5355 3ENGLISH
1.0 INTRODUCTION
1.1 Component Identification
LCD DISPLAY: A 3-digit display provides a
direct readout of oxygen concentration in
the range of 0 - 100%. The display is blank
when the Handi+ enters its Sleep (power
off) mode. The Handi+ will automatically
enter the Sleep mode after approximately
1.3 minutes from the last time the ON button
was pressed.
ON/OFF BUTTON: Use this button to turn the
Handi+ on and off. When the Handi+ is in the
Sleep (power off) mode, the LCD display is
blank. The analyzer turns off after 1.3 min-
utes if no buttons are pressed.
CALIBRATION BUTTON: Press the calibration (CAL) button to adjust the calibration value to
reflect the known oxygen concentration. To simplify operation, the Handi+ Analyzer automati-
cally determines the calibration gas being used as room air (20.9%) or high grade (100%).
FLOW DIVERTER: The flow diverter is designed to fit industry standard 15 mm I.D. "T" adapters.
OVER RANGE INDICATOR: The appearance of a decimal point after the first digit means that the
Handi+ is reading in excess of 99.9%.
Example: 0.0.0 = 100% 0.0.1 = 101% 0.0.2 = 102%
(If the display reads > 0.0.3 the Handi+ should be re-calibrated.)
1.2 Description
The Maxtec Handi+ oxygen analyzer is designed to monitor oxygen concentration in the patient-
breathing environment. It is one of a full line of oxygen analyzers by Maxtec. The Handi+ oxygen
analyzer utilizes the Maxtec MAX-250 oxygen sensor and is engineered for fast response, maxi-
mum reliability and stable performance. The Handi+ is designed primarily for spot-checking of
oxygen levels delivered by medical oxygen delivery equipment and respiratory care systems.
Its lightweight, compact size, extended battery life, and "auto off" feature makes this oxygen
analyzer ideal for portable oxygen analysis by qualified heath care professionals.
2.0 CALIBRATING THE HANDI+
2.1 Before You Begin
A protective film covering the threaded sensor face must be removed and the sensor allowed
to “breath” for at least 30 minutes. Next, the Handi+ Oxygen Analyzer should be calibrated.
Thereafter, Maxtec recommends calibration on a weekly basis. However, more frequent cali-
bration will not adversely affect product performance.
2.2 Calibrating the Handi+ Oxygen Analyzer
Calibration of the instrument is necessary if the temperature of the gas stream changes by
more than 3 degrees Celsius.
Changes in elevation result in calibration error of approximately 1% of reading per 250 feet. In
general, calibration of the instrument should be performed when the geographic elevation at
which the product is being used changes by more than 500 feet.
In addition, calibration is recommended if the user is unclear when the last calibration proce-
dure was performed or if the measurement value displayed is in question.
Calibrate the Handi+ to an oxygen source, which has a traceable certificate and/or USP certifi-
cation, of either 20.9% oxygen, as found in room air, or 100% oxygen. It is recommended that
the oxygen source be documented at a pressure and flow similar to your clinical application.
NOTE: Before beginning calibration the Handi+ must be in thermal equilibrium. You may also
need to be aware of other factors, which affect device calibration values. For more informa-
tion, refer to section 4.0 "Factors Influencing Calibration and Performance" of this manual.
NOTE: We recommend use of medical grade USP or >99% purity oxygen when calibrating the
Handi+.
2.3 In Line Calibration (Configuration A)
1. Put the Handi+ in an upright position such that you can read the product label.
2. Connect a sample supply hose to a standard "T" adapter The Maxtec "T" is
precision-tapered to insure a tight connection with O-rings of the MAX-250 oxygen
sensor diverter.
3. Insert the Handi+ in the center position of the "T" adapter.
4. Attach an open-ended reservoir to the end of the "T" adapter. Then start the
calibration flow of oxygen at 1-10 liters per minute.
NOTE: Six to 10 inches of corrugated tubing works well as a reservoir.
NOTE: A calibration oxygen flow to the Handi+ of 2 liters per minute is recommended to mini-
mize the possibility of obtaining a "false" calibration value.
5. Allow the oxygen to saturate the sensor. Although a stable value is usually
observed within 30 seconds, allow at least 2 minutes to ensure that the sensor is
completely saturated with the calibration gas.
6. If the Handi+ is not already turned on, do so now by pressing the analyzer ON
button .
7. Press the “CAL” button on the Handi+. The calibration gas value on the
analyzer display should read 20.9 or 100 depending on the gas stream used.
2.4 Direct Flow Calibration (Configuration B)
1. Attach the Barbed Adapter to the Handi+.
2. Connect the clear sampling tube to the Barbed Adapter.
3. Attach the other end of the clear sampling tube to a source of oxygen with a
known oxygen concentration value and initiate flow of the calibration gas to the
unit at a rate of 1-10 liters per minute (2 liters per minute is recommended).
4. Allow the oxygen to saturate the sensor. Although a stable value is usually
observed within 30 seconds, allow at least 2 minutes to ensure complete
saturation of the sensor with the calibration gas.
5. If the Handi+ is not already turned on, do so now by pressing the analyzer “ON”
button .
6. Press the “CAL” button on the Handi+. The calibration gas value on the
analyzer display should read 20.9 or 100 depending on the gas stream used.
Configuration A
"T" Adaptor
Configuration B
Clear Tubing
Barbed Adapter
(To gas)
Flow Diverter
LCD DISPLAY
CALIBRATION
BUTTON
"ON" BUTTON

ENGLISH 4WWW.MAXTEC.COM • (800) 748-5355
3.0 OPERATING THE HANDI+
To Check the Oxygen Concentration of a Sample Gas:
1. Maintain the Handi+ in an upright position such that you can read the product
label.
2. Place the Handi+ in the sample gas stream.
IMPORTANT: When using a standard "T" adapter, make sure that the sensor is mounted in the
adapter with the flow diverter pointing down. Make sure that there is a tight fit between the
flow diverter and the "T" adapter.
3. Start the flow of the sample gas to the sensor.
4. Allow the oxygen sensor to remain in the flow of the sample gas until stable.
5. If the Handi+ is not already turned on, do so now by pressing the analyzer ON
button .
6. Read the value displayed on the LCD.
NOTE: If the Handi+ is used to measure the oxygen concentration with equipment using a
heated or humidified gas stream, it is recommended that the Handi+ be placed upstream of
the heater and/or humidifier. For more information, refer to "Factors Influencing Calibration
and Performance" below.
For hospital and home care a new calibration is required when:
•The measured O2 percentage in 100% O2 is below 97.0% O2
•The measured O2 percentage in 100% O2 is above 103.0% O2.
For ID testing (or optimum accuracy), a new calibration is required when:
7. The measured O2 percentage in 100% O2 is below 99.0% O2.
8. The measured O2 percentage in 100% O2 is above 101.0% O2.
4.0 FACTORS INFLUENCING
ACCURATE READINGS
4.1 Elevation Changes
•Changes in elevation result in a reading error of approximately 1% of reading per
250 feet.
•A change in altitude greater than 500 ft will require sensor recalibration.
•This device does not automatically compensate for changes in barometric pressure
or altitude. If the device is moved to a location of a different altitude, it must be
recalibrated before use (see section 2.2).
4.2 Temperature Effects
The Handi+ will hold calibration and read correctly within ±3% when in thermal equilibrium
within the operating temperature range. The device must be thermally stable when calibrated
and allowed to thermally stabilize after experiencing temperature changes before readings
are accurate. For these reasons, the following is recommended:
•When used in a breathing circuit, place the sensor upstream of the heater.
•For best results, perform the calibration procedure at a temperature close to the
temperature where analysis will occur.
•Allow adequate time for the sensor to equilibrate to a new ambient temperature.
4.3 Pressure Effect
Readings from the Handi+ are proportional to the partial pressure of oxygen. The partial pres-
sure is equal to the concentration times the absolute pressure. Thus the readings are pro-
portional to the concentration if the pressure is held constant. Flow rate of sample gas can
affect pressure at the sensor in that back pressure at the sensing point may change. For these
reasons, the following is recommended:
•Calibrate the Handi+ at the same pressure as the sample gas.
•If sample gases flow through tubing, use the same apparatus and flow rates when
calibrating as when measuring.
•The Handi+ oxygen sensor has been validated at pressures up to 2 atmospheres
absolute. Calibration or operation above this pressure is beyond the intended use.
4.4 Humidity Effect
Humidity has no effect on the performance of the Handi+ other than diluting the gas, as long
as there is no condensation. Depending on the humidity, the gas may be diluted by as much as
4%, which proportionally reduces the oxygen concentration. The device responds to the actual
oxygen concentration rather than the dry concentration. Environments where condensation
may occur are to be avoided since condensate may obstruct passage of gas to the sensing
surface, resulting in erroneous readings and slower response time.
For this reason, the following is recommended:
•Avoid usage in environments greater than 95% relative humidity.
•When used in a breathing circuit, place the sensor upstream of the humidifier.
5.0 CALIBRATION ERRORS AND
ERROR CODES
The Handi+ analyzers have a self test feature built into the software to detect faulty calibra-
tions, oxygen sensor failures, and low operating voltage. These are listed below, and include
possible actions to take, if an error code occurs.
E03: No valid calibration data available
Make sure unit has reached thermal equilibrium. Press and hold the Calibration Button for
three seconds to manually force a new calibration.
E04: Battery below minimum operating voltage
Unit is at end of life, see "Product Disposal Instructions" for proper disposal.
CAL Err St: O2 Sensor reading not stable
Wait for displayed oxygen reading to stabilize when calibrating the device at 100% oxygen.
Wait for unit to reach thermal equilibrium (Please note that this can take up to one half hour,
if the device is stored in temperatures outside the specified operating temperature range).
CAL Err lo: Sensor voltage too low
Press and hold the Calibration Button for three seconds to manually force a new calibra-
tion. If unit repeats this error more than three times, contact Maxtex Customer Service for
possible sensor replacement.
CAL Err hi: Sensor voltage too high
Press and hold the Calibration Button for three seconds to manually force a new calibra-
tion. If unit repeats this error more than three times, contact Maxtec Customer Service for
possible sensor replacement.
CAL Err Bat: Battery voltage too low to recalibrate
Unit is at end of life, see "Product Disposal Instructions" for proper disposal.
6.0 CLEANING AND MAINTENANCE
When cleaning or disinfecting the Handi+, take appropriate care to prevent any solution from
entering the analyzer.
•The Handi+ surface may be cleaned using a cloth moistened with 65% isopropyl
alcohol/water solution or germicidal wipe.
•The Handi+ may be disinfected using standard topical disinfectants.
•The Handi+ is not intended for steam, ethylene oxide or radiation sterilization.
•Store the Handi+ in a temperature similar to its ambient environment of daily use.

WWW.MAXTEC.COM • (800) 748-5355 5ENGLISH
7.0 HANDI+ SPECIFICATIONS
7.1 Analyzer Specifications
Measurement Range:.................................................................................0.0 - 99.9% oxygen (gas)
Resolution/Display:........................................................................................................................0.1%
The three digit LCD indicates values between 0.0 - 99.9% oxygen
Over range indicated by one decimal point on display located after the first digit
Accuracy and Linearity: ...................................1% of full scale at constant temperature, R.H. and
pressure when calibrated at full scale
Total Accuracy:.......................................±3% actual oxygen level over full operating temp range
Response Time: ..........................................................< 15 seconds for 90% step change. (at 25˚C)
Warm-up Time:.............................................................................................................None required
Operating Temperature:...........................................................................15˚C - 40˚C (59˚F - 104˚F)
Storage Temperature:................................................................................-15˚C - 50˚C (5˚F - 122˚F)
Atmospheric Pressure:...............................................................................................800-1013 mBars
Humidity: .....................................................................................................0-95%(non-condensing)
Environmental:......................................................General purpose housing equivalent to NEMA 1
The Handi+ is not waterproof
Warranty:..................................................................Two (2) years in normal operating conditions
Power Requirements:................................................................................Powered by one internal,
non-replaceable Lithium battery, CR2450.
Power on push button automatically shuts off after 80 seconds time-out.
Electronics rated general purpose; not for use in
hazardous areas or for use with flammable gases.
Sensor Type:......................................................................Maxtec MAX-250 series galvanic fuel cell
Weight:.................................................................................................................... Approx. 60 grams
7. 2 Effect of Interferent Gases and Vapors
GAS VOLUME % DRY INTERFERENCE IN O2
Nitrous Oxide 75% < 2%
Halothane 5% < 2%
Isoflurance 5% < 2%
Enflurane 5% < 2%
Sevoflurane 6% < 2%
Desflurane 15% < 2%
Carbon Dioxide 10% < 2%
Helium 70% < 2%
8.0 HANDI+ SPARE PARTS AND
ACCESSORIES
8.1 Standard Replacement Parts and
Accessories
PART NUMBER ITEM
R218M12 Operation Manual, Handi+
RP16P02 "T" Adapter
R110P10-001 Flow Diverter Fitting
R207P17 Barbed Adapter
R100P92-002 Clear Tubing
RP76P06 Lanyard
R218P09 Handi+ Cover
9.0 FREQUENTLY ASKED
QUESTIONS
After I calibrate to 20.9%, the display changes as
much as ± 1%.
•This is within the normal Handi+ operating tolerance of ± 1% when temperature
and pressure are constant.
I have calibrated to 99.9% but when I check my oxy-
gen delivery equipment, the Handi+ reads ".0.4" or
greater (Over Range Indicator).
•It is recommended that you conduct the calibration procedure again to get
another reading. The most likely cause is that the Handi+ has received a "false"
calibration value.
•Make sure that the calibration gas is connected to the Handi+ at 2 liters per minute
for a minimum of 2 minutes prior to proceeding with calibration.
•This 2-minute equilibration time is necessary to insure that the sensor is
completely saturated with the calibration gas.
I have found the reading to drift greater than ± 3%
from a know source value. What is the possible cause?
•The sensor may be at or near its useful life.
•Replace your Handi+.
•Sensor life is dependent on the oxygen concentration exposure.
For example, a sensor which is used to check flow meters once a week for 8 hours
will outlast one which is used to analyze oxygen blender performance 24 hours
per day, 5 days a week.
2.45 in
0.90 in
2.91 in
1.44 in

ENGLISH 6WWW.MAXTEC.COM • (800) 748-5355
10.0 ELECTROMAGNETIC
COMPATABILITY
The information contained in this section (such as separation distances) is in general specifi-
cally written with regard to the Handi+. The numbers provided will not guarantee faultless
operation but should provide reasonable assurance of such. This information may not be
applicable to other medical electrical equipment; older equipment may be particularly sus-
ceptible to interference.
Note: Medical electrical equipment requires special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC infor-
mation provided in this document and the remainder of the instructions for use this device.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Cables and accessories not specified within the instructions for use are not authorized. Using
other cables and/or accessories may adversely impact safety, performance and electromag-
netic compatibility (increased emission and decreased immunity).
Care should be taken if the equipment is used adjacent to or stacked with other equipment;
if adjacent or stacked use is inevitable, the equipment should be observed to verify normal
operation in the configuration in which it will be used.
ELECTROMAGNETIC EMISSIONS
This equipment is intended for use in the electromagnetic environment specified below.
The user of this equipment should assure that it is used in such an environment.
EMISSIONS
COMPLIANCE
ACCORDING
TO
ELECTROMAGNETIC ENVIRONMENT
RF Emissions
(CISPR 11)
Group 1 The Handi+ uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
CISPR Emissions
Classification
Class A The Handi+ is suitable for use in all establish-
ments other than domestic and those directly
connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
NOTE: The EMISSIONS characteristics of this equip-
ment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class
B is normally required) this equipment might not
offer adequate protection to radio-frequency
communication services. The user might need to
take mitigation measures, such as relocating or
re-orienting the equipment.
Harmonic Emissions
(IEC 61000-3-2)
Class A
Voltage Fluctuations Complies
ELECTROMAGNETIC IMMUNITY
This equipment is intended for use in the electromagnetic environment specified below.
The user of this equipment should assure that it is used in such an environment.
IMMUNITY AGAINST IEC 60601-1-2: (4TH
EDITION) TEST LEVEL
ELECTROMAGNETIC
ENVIRONMENT
Professional
Healthcare
Facility
Environment
Home
Healthcare
Environment
Electrostatic discharge,
ESD (IEC 61000-4-2)
Contact discharge: ±8 kV
Air discharge: ±2 kV,
±4 kV, ±8 kV, ±15 kV
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be kept at
levels to reduce electrostatic
charge to suitable levels.
Mains power quality should be
that of a typical commercial or
hospital environment.
Equipment which emits high
levels of power line magnetic
fields (in excess of 30A/m)
should be kept at a distance
to reduce the likelihood of
interference.
If user requires continued
operation during power
mains interruptions, ensure
that batteries are installed
and charged. Ensure that
battery life exceeds longest
anticipated power outages
or provide an additional
uninterruptible power source.
Electrical fast
transients / bursts
(IEC 61000-4-4)
Power supply lines: ±2 kV
Longer input / output
lines: ±1 kV
Surges on AC mains
lines (IEC 61000-4-5)
Common mode: ±2 kV
Diferential mode: ±1 kV
3 A/m power
frequency magnetic
field 50/60 Hz
(IEC 61000-4-8)
30 A/m
50 Hz or 60 Hz
Voltage dips and
short interruptions
on AC mains input
lines (IEC 61000-4-11)
Dip>95%, 0.5 periods
Dip 60%, 5 periods
Dip 30%, 25 periods
Dip >95%, 5 seconds
Recommended separation distances between portable and mobile
RF communications equipment and the equipment
RATED MAXIMUM
OUTPUT POWER OF
TRANSMITTER
W
Separation distance according to frequency
of transmitters in meters
150 kHz to 80 MHz
d=1.2/V1] √P
80 MHz to 800 MHz
d=1.2/V1] √P
800MHz to 2.5 GHz
d=2.3 √P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
11.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23

WWW.MAXTEC.COM • (800) 748-5355 7ENGLISH
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable
to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propogation is
affected by absorption and reflection from structures, objects, and people.
This equipment is intended for use in the electromagnetic environment
specified below. The customer or the user of this equipment
should assure that it is used in such an environment.
IMMUNITY TEST IEC 60601-1-2: 2014 (4TH
EDITION) TEST LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Professional
Healthcare Facility
Environment
Home
Healthcare
Environment
Conducted RF
coupled into lines
(IEC 61000-4-6)
3V (0.15 - 80 MHz)
6V (ISM bands)
3V (0.15 - 80 MHz)
6V (ISM &
Amateur bands)
Portable and mobile RF communications
equipment (including cables) should
be used no closer to any part of the
equipment than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter as below.
Recommended sparation distance:
d=1.2 √P
d=1.2 √P 80 MHz to 800 MHz
d=2.3 √P 800 MHz to 2.7 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey a,
should be less than the compliance
level in each frequency range b.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Radiated RF
immunity (IEC
61000-4-3)
3 V/m
80 MHz - 2.7 GHz
80% @ 1 KHz
AM Modulation
10 V/m
80 MHz - 2.7 GHz
80% @ 1 KHz
AM Modulation
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz
to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the mea-
sured field strength in the location in which the equipment is used exceeds the applicable
RF compliance level above, the equipment should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorient-
ing or relocating the equipment.

2305 South 1070 West
Salt Lake City, Utah 84119
(800) 748-5355
www.maxtec.com
Other manuals for Handi+
4
Table of contents
Other Maxtec Medical Equipment manuals