MEDENUS DF 100 Manual

© 09.2015
Cellular Gas Filters
DF 100
Operating and Maintenance Instructions

Subject to technical modifications! Reprint prohibited!

© 09.2015
Table of Contents
1 General Information 4
1.1 Warranty and Liability 4
1.2 Symbols, Notes 5
1.3 Terms, Abbreviations 5
2 Safety Instructions 6
2.1 Hazards of Handling the Device 6
2.2 Personnel Requirements 6
2.3 Country-Specific Regulations 6
2.4 Handover of the Operating and Maintenance Instruction 6
2.5 Safety in operation 7
2.6 What to do in case of danger 7
3 Responsibility of the Operator 7
4 Transport, Storage and Packaging 8
4.1 Transport 8
4.2 Storage 8
4.3 Packaging 8
5 Mounting and Commissioning 9
5.1 Safety Instructions and Preparation 9
5.2 Installation 9
5.3 Leakage Test (Test for External Leakage ) 10
5.4 Initial Commissioning / Recommissioning 10
5.5 Decommissioning 10
6 Maintenance 11
6.1 Maintenance Plan 11
6.2 Maintenance Procedure 11
6.3 Table Screw Tightening Torques MA 13
6.4 Lubricants Table 13
7 Troubleshooting 13
8 Replacement and Disposal 13
9 Spare Parts 14
9.1 Spare Parts Drawing DF 100 - DN 25 14
9.2 Spare Parts Drawing DF 100 - DN 50 / DN 80 / DN 100 15
9.3 Ersatzteilzeichnung DF 100 - DN 150 / DN 200 16
9.4 Spare Parts Drawing Differential Pressure Measuring Instrument 17
9.5 Parts for Maintenance Work 18
10 Notes 19

4
© 09.2015
1 General Information
The personnel entrusted with installation, operation or maintenance of the cellular gas filter must have completely read
and understood beforehand the following documents:
• Cellular Gas Filter DF 100 Product Information
The product information contains technical data, dimensions and a description of the design and the mode of
operation.
• Cellular Gas Filter DF 100 Operating and Maintenance Instructions
This document allows safe and efficient handling of the device and contains information on assembly,
commissioning, maintenance, troubleshooting and repair according to regulations.
It is an integral part of the scope of delivery of the device and must be kept in close proximity of the device and
readily accessible to personnel at any time.
The basic requirement of safe operation is compliance with all safety instructions and guidelines specified in
these instructions. Accordingly, the information and instructions must be observed when working on the device
or on the gas line. In addition, the local occupational safety regulations and the general safety regulations for
the application range of the device shall apply.
The figures in these instructions are provided for basic understanding and may differ from the actual design.
The contents in these instructions are protected by copyright. They may be used as part of operating the
device. Any other use and / or reproduction is not permitted without prior authorization by MEDENUS Gas-
Druckregeltechnik GmbH.
1.1 Warranty and Liability
Claims under warranty or liability for personal injury and material damage are generally void,
if one or several of the following conditions are not observed:
• Work on the device during the warranty period may only be performed in consultation with the manufacturer
• Designated use of the device must take place in accordance with the defined conditions of use
• Proper installation, commissioning, operation and maintenance of the device
• Operation of the device using only properly installed and functioning safety devices
• Operating and maintenance instructions of the device or of the system
• Compliance with the maintenance regulations
• Properly performed repairs
• Supply lines are free of defects
• The use of original MEDENUS© spare parts and lubricants specified in these instructions
or
• force majeure
It is generally prohibited
• to perform constructive modifications on the device and
• to keep using the device despite the detection of a defect

5© 09.2015
1.2 Symbols, Notes
The instructions contain safety instructions marked with symbols to indicate possible consequences in case of non-
observance:
1.3 Terms, Abbreviations
Terms and abbreviations are explained below:
DN Nominal width
MATightening torque
MOP Maximum operating pressure in a system
SRV Safety relief valve
DANGER
Note
NOTICE
This combination of symbol and signal word indicates a potentially hazardous
situation, which may result in light injuries, in damage to the device, in the
breakdown of the system and in material or environmental damage if not
avoided.
This combination of symbol and signal word indicates an imminent hazardous
situation, resulting in death or serious injuries if not avoided.
This signal word highlights useful tips, recommendations and information for
efficient and trouble-free operation.

6
© 09.2015
2 Safety Instructions
National accident prevention regulations and the system operator‘s safety regulations are not superseded by these
operating and maintenance instructions and must be taken into consideration with priority (in Germany, see, among
others, DVGW Code of Practice G 600, G 459/II, G 491 and G 495)
When performing work on the device, the current general and specific safety regulations must be observed.
The application limits of the device in terms of the medium, operating pressure and operating temperature are to be
found on the type plate affixed to the device or on the acceptance test certificate.
Using the device under different operating conditions must be agreed upon in consultation with the MEDENUS©Gas-
Druckregeltechnik GmbH.
The mechanical components of the device do not have any potential ignition sources of their own nor any hot surfaces
and are thus not covered by the scope of ATEX 95 (94/9/EC). The electronic accessories used comply with the ATEX
requirements.
2.1 Hazards of Handling the Device
MEDENUS©devices conform with current standardsand directives, the recognized technical rules and the recognized
safety rules.
However, improper use can result in hazards to the user or to third parties. Likewise, this can result in damage to the
device or to the system.
This is why the device may only be used:
• in accordance with its designated use
• in perfect condition
• while observing the notes given in these operating and maintenance instructions and inspection and
maintenance regulations, which apply to the functioning and safety of the overall system
Malfunctions or faults must be eliminated immediately.
2.2 Personnel Requirements
The device may only be mounted by qualified personnel.
Only authorized personnel in possession of the required qualification is allowed to perform settings or repairs on the
device.
2.3 Country-Specific Regulations
The rules and regulations applicable at the place of use, with respect to:
• gas lines, installation of the gas system
• gas supply
• work on the gas system
• accident prevention, must be observed and complied with
2.4 Handover of the Operating and Maintenance Instruction
The supplier of the system shall hand over these operating and maintenance instructions to the operator of the system
no later than during commissioning and training of the operating personnel with the reminder to carefully store these
instructions.

7© 09.2015
2.5 Safety in operation
The device may only be used when all protective devices on the device or in the system are fully functional.
At least once a year, the device must be inspected by a representative of the manufacturer or by a qualified person for
externally visible damage and for proper functioning.
A more frequent inspection may become necessary, depending on the system conditions.
2.6 What to do in case of danger
What to do in case of danger and in case of accidents can be found in the respective operator‘s or specialist companies‘
work instructions.
3 Responsibility of the Operator
Operator An operator is a person who operates the device himself for commercial or economic purposes or hands
it over to a third party for use / application and is legally responsible for the safety of the user, personnel
or third parties.
The device is used in the industrial sector. Accordingly, the operator of the device is subject to the legal
obligations concerning occupational safety. In addition to the safety instructions contained in these
instructions, the established maintenance intervals must be observed, taking into account the respective
national standard (alarm and hazard prevention plan).
In particular, the following applies:
• The operator is obliged to perform work on the MEDENUS© devices during the warranty period
only after consultation with the manufacturer. Otherwise the claims under warranty will become
void.
• The operator must obtain information on the current occupational safety regulations and
determine additional hazards resulting from the special work conditions at the place of use of
the device in a risk assessment. The operator must implement them in the form of operating
manuals for operating the device.
• During the entire time of use of the device, the operator must check whether the operating
manuals drawn up by him conform to the current state of the regulations and, if necessary,
adapt them.
• The operator must clearly regulate and define the responsibilities for installation, operation,
troubleshooting, maintenance and cleaning.
• The operator must ensure that all persons handling the device must have read and understood
these instructions. In addition to that, he must train the personnel at regular intervals and
inform it about hazards.
• The operator must make available to the personnel the required protective equipment and
oblige it to wear the required protective equipment.
• Moreover, the operator is responsible for the device always being in technically perfect condition.
This is why the follow applies:
• The operator must make sure that the maintenance intervals described in these instructions
are observed.
• The operator must have all safety devices checked regularly for functioning and completeness.
Operator‘s
obligations

8
© 09.2015
4 Transport, Storage and Packaging
4.1 Transport
The device is delivered with flange protective caps. They must be removed prior to installation.
Make sure that the device is transported horizontally using suitable lifting gear. The device must be
handled carefully and secured against impacts and knocks.
In case of transport damage, we will require the following information from the type plate affixed to
the device:
• Type of device
• Device model
• Year of construction / fabrication number
Note
4.2 Storage
Equipment and spare parts must be stored under the following conditions:
• Do not store outdoors.
• Store in a dry and dust-free location.
• Store on a flat surface.
• Do not expose to aggressive media.
• Do not expose to ozone or ionizing radiation.
• Do not store adjacent to direct heat sources.
• Avoid mechanical vibrations.
• Storage temperature: 0 to 25 °C.
• Relative humidity: < 55 %.
Spare parts:
• Components susceptible to corrosion must be provided with a suitable preservative.
• Do not store O-rings and seals for more than 7 years even if stored properly.
• Spare parts must be stored in their original packaging until use.
Storage period for devices:
• Store the device for up to one year:
Store the cellular gas filter in its original packaging and original condition at the time of supply. All protective
caps must remain mounted.
• Storage of the device for more than 1 year (e.g. as reserve device):
Store the device in its original packaging and original condition at the time of supply and check it for damage
once a year. Check the housing surface for dirt, damage and corrosion. If necessary, clean all external parts.
After 7 years, all O-rings and seals must be replaced.
4.3 Packaging
• The individual packaged items have been packaged with a view to the transport conditions to be expected.
• The symbols on the packaging must be observed during transport and storage.
• For the packaging, only environmentally-friendly materials were used.
• The packaging is designed for protecting the individual components until mounting from transport damage,
corrosion and other damage. This is why the packaging must not be destroyed and only removed just prior to
mounting.

9© 09.2015
5 Mounting and Commissioning
5.1 Safety Instructions and Preparation
Prior to starting work on pressurized components:
• Close all connections to the gas line.
• Depressurize all pressurized components. Also discharge residual energies.
• Defective components charged with pressure in operation must be replaced immediately by
a suitable qualified person.
Prior to starting work, ensure sufficient clearance for mounting.
Before installing the device, check whether the performance date (type plate) and the scope of delivery
coincide with the order or the system data, i.e., make sure that the provided device are suitable
for their intended purpose. In particular, the inlet pressure of the system must be lower than the
maximum allowable pressure of the device.
A direct contact of gas valves and fittings, i.e., the control system, with hardening masonry, concrete
walls or floors, is not allowed. Provide suitable supports, working materials and protective equipment.
Take into account the minimum clearances for maintenance as stated in the product information.
Before installing the device in the pipeline, check whether upstream and downstream of the device
to be installed a shut-off device that interrupts the gas flow supply to the device has been mounted.
Prior to commissioning, make sure that all installation work has been carried out and completed
in accordance with the data and information given in these instructions and that no unauthorized
persons stay in the danger zone.
5.2 Installation
Remove packaging and protection of the connection flange surfaces.
Install the device without twisting the pipeline.
Make sure to observe the direction of flow, i.e., the arrow on the housing must point in the direction
of flow.
In other installation positions than the horizontal, the MEDENUS©Gas-Druckregeltechnik GmbH
must be consulted.
Tighten the screws crosswise. Observe the prescribed screw tightening torques
(see chapter 6.3 - page 13).
DANGER
Note
Note
DANGER
Note

10
© 09.2015
5.3 Leakage Test (Test for External Leakage )
The devices are subjected to a strength and leakage test ex works at the MEDENUS©Gas-Druckregeltechnik GmbH.
The leakage test in the fully assembled system must be performed prior to commissioning and following maintenance
work.
For the external leakage test in the fully assembled system, the following applies:
For Germany:
According DVGW Code of Practice G 491, the fully assembled system must be subjected to a leakage test with air or an
inert gas at the installation site, using 1.1 times the maximum operating pressure of the system (MOP).
An exception is the room between the actuator installed in the gas pressure regulator and the first shut-off valve on the
outlet side. This room must be checked using a test pressure corresponding to the maximum possible setpoint of the
gas pressure regulator. In this test, all detachable connections must be checked using a foaming agent.
For other countries:
The relevant national and international standards shall apply.
Procedure
• Close the ball valves upstream of the valves and fittings.
• Close the downstream shut-off devices (ball valves, solenoid or pneumatic valve).
• Depressurise the system.
• If the test pressure is higher than the relief pressure of the safety relief valve (SRV), the line upstream of the
SRV must be closed.
• Connect the testing device to measuring points upstream and downstream of the regulator.
• Raise the test pressure always slowly and steadily.
In doing so, please observe the following:
Pressure in the outlet chamber ≤pressure in the inlet chamber
Pressure build-up always from the inlet side (inlet chamber)
Pressure reduction always from the outlet side (outlet chamber)
• After the leakage test:
• Reopen the ball cock in the SRV line.
5.4 Initial Commissioning / Recommissioning
For commissioning, please refer to the relevant manufacturer‘s documentation of the gas pressure regulator installed
in the system and the work instructions of the system operator.
5.5 Decommissioning
For decommissioning, please refer to the relevant manufacturer‘s documentation of the gas pressure regulator installed
in the system and the work instructions of the system operator.
Note
NOTICE

11 © 09.2015
6 Maintenance
6.1 Maintenance Plan
The following sections describe the maintenance work required for optimal and trouble-free operation of the device.
If increased wear is detected during regular inspections, the required maintenance intervals must be reduced in
accordance with the actual wear.
For any questions on maintenance work and intervals, please contact the manufacturer.
The intervals for monitoring and maintenance work are strongly dependent on the operating situation and the
condition of the gas. This is why no fixed intervals can be given. For Germany, it is recommended to observe initially the
maintenance periods according to the data given in DVGW Code of Practice G 495. For each system, this must be followed
by determining the maintenance interval independently on a medium-term basis.
During maintenance work, the components must be cleaned and subjected to a thorough visual inspection. This is also
necessary if irregularities in the operating behaviour have been detected during operation or during functional tests.
The check must cover in particular the filter insert and the gaskets.
Damaged parts and O-rings dismounted during dismounting must be replaced with new ones.
The item numbers mentioned in chapter 6.2 (Maintenance Procedure) correspond to those listed in the spare parts
drawings and spare parts lists.
It is recommended having the parts listed in chapter 9.5 on page 18 ready for maintenance work.
Interval Maintenance activities Personnel
when required
Replacing the filter cartridge
qualified person
Replacing the O-ring between the cover and the housing
Replacing the O-ring for sealing the filter cartridge
Replacing the O-ring for sealing the valve seat
Replace the O-ring between the cover / spacer and the housing
Replacing the O-ring between the cover and the spacer
Replacing the sealing of the connection of the differential pressure device
Replacing the sealing of the progressive ring in the connection of the
differential pressure device
6.2 Maintenance Procedure
If components were removed, make sure they are mounted correctly, reinstall all fastening elements
and observe the screw tightening torques.
Prior to recommissioning, observe the following:
• Make sure that all maintenance work has been carried out and completed in accordance
with the data and information given in these instructions.
• Make sure that no unauthorized persons stay in the danger zone.
• Make sure that all covers and safety devices have been installed and are working properly.
NOTICE
DANGER
For recommissioning, please refer to the relevant manufacturer‘s documentation of the gas pressure regulator installed
in the system, the work instructions of the system operator, and the remaining safety regulations of the system in which
the filter has been installed.

12
© 09.2015
6.2 Maintenance Procedure
As a functional test, a pressure loss measurement for the flow resistance can be carried out.
Recommended limit value for a contaminated filter cartridge: 0.8 to 1 bar, maximum 2bar
• Close ball valves upstream and downstream of the filter. Depressurize the filter.
• Dismount the cover (item 12). To do so, unscrew the Allen screws (item 13) on the cover. Exception for DN 25:
Unscrew cover (item 12).
• Remove O-ring (item 11) inside the housing or cover. Clean the O-ring groove and insert a new greased O-ring
into the groove.
• Pull out the filter cartridge (item 10) towards the top and check whether it is dirty. If necessary, the filter
cartridge must be cleaned or replaced.
• For DN 150 / DN 200: Dismount the spacer (item 17). To do so, unscrew the Allen screws (item 13) at the
spacer. Remove the O-ring (item 11) from the housing (item 1) or from the spacer.
To lower the dismounting dimension for nominal widths DN 150 and DN 200, the filter cartridge and
the spacer should be removed together.
• Clean the O-ring grooves and insert new, greased O-rings into the grooves.
• Pull out the valve seat (item 8) towards the top. (does not apply to DN 100)
• Remove the O-ring (Pos.7) for sealing the valve seat. Clean the O-ring groove and insert a new greased O-ring
into the groove. (does not apply to DN 100)
• For DN 25 / DN 100: Remove the O-ring for sealing the filter cartridge (item 9). Clean the O-ring groove and
insert a new greased O-ring into the groove.
• Dismount the lid (item 5). To do so, unscrew the Allen screws (item 6) from the cover. It may be necessary to
clean the cover.
• Remove the O-ring (item 4) from the housing (item 1). Clean the O-ring groove and insert a new greased O-ring
into the groove.
• Remount the cover (item 5). To do so, screw down the Allen screws (item 6) at the lid crosswise, using a torque
wrench.
• Press the valve seat (item 8) back into the housing passage. Place the cleaned or new filter cartridge (item
10) in the valve seat. The filter cartridges DN 25 and DN 100 comprise a centring ring at one end guaranteeing
the correct fit of the cartridge in the valve seat. During installation, the centring ring is located at the bottom
inside the filter.
• For DN 150 / DN 200: Mount the spacer (item 17). To do so, screw down the Allen screws (item 13) at the spacer
crosswise, using a torque wrench.
To lower the mounting dimension for nominal widths DN 150 and DN 200, the filter cartridge and the
spacer should be inserted together.
• Insert the cover (item 12) and screw down the Allen screws (item 13) crosswise, using a torque wrench.
Exception for DN 25: Screw in cover (item 12) and tighten it, using a torque wrench. When screwing in, exert a
slight pressure on the cover. The filter cartridge is oversized and, due to the leak tightness to be achieved, it
must be compressed somewhat.
• The maintenance parts in the differential pressure measuring instrument are only replaced when necessary.
• Open the valves and fittings upstream and downstream of the filter for leak tightness (using a foaming agent).
To guarantee smooth operation, we recommend always keeping a maintenance set in reserve.
Note
Note
Note
Note

13 © 09.2015
6.3 Table Screw Tightening Torques MA
Item
no.
Nominal width
DN 25 DN 50 DN 80 DN 100 DN 150 DN 200
3 30 Nm 30 Nm 30 Nm 30 Nm 30 Nm 30 Nm
6 35 Nm 35 Nm 35 Nm 35 Nm 35 Nm 35 Nm
12 30 Nm - - - - -
13 - 35 Nm 35 Nm 35 Nm 35 Nm 35 Nm
6.4 Lubricants Table
Components (apply a thin layer) Lubricants Article number
All O-rings Syntheso Proba 270 Syntheso Proba 270
All fastening and locking screws Anti Seize AS 450 AS-450
7 Troubleshooting
Description of the error Possible cause Corrective action Personnel
Differential pressure across the
filter too high Filter cartridge (item 10) dirty Clean filter cartridge or
replace
Qualified
person
Dust in the system Filter cartridge (item 10) defective Clean filter cartridge
O-ring (item 7 or 9) damaged Replace O-ring
Leakage towards the outside O-ring (item 4 or 11) damaged Replace O-ring
Faulty differential pressure display Pressure gauge or Reed contact defective Replace pressure
gauge* or Reed contact
8 Replacement and Disposal
After the device has reached the end of its useful life, it must be dismounted and disposed of in an environmentally
compatible manner.
During dismounting, components that may present a risk of injury by contamination, depending
on the medium, are removed. Depending on the processed medium, the components must be
properly decontaminated. Components capable of diffusion (filter cartridge, O-ring, etc.) may
have to be taken to a special disposal unit, depending on the medium.
If no return or disposal agreement has been reached, dismantled components should be recycled:
• Metals should be scrapped
• The remaining components should be disposed of sorted according to material.
For technical information, please contact our customer service:
MEDENUS Gas-Druckregeltechnik GmbH
Saßmicker Hammer 40
D-57462 Olpe
Telephone +49 (0) 2761 / 82788-0
Fax +49 (0) 2761 / 82788-9
In addition, we are always interested in information and experience resulting from the application and which can be
valuable for improving our products.
E-Mail [email protected]
Internet www.medenus.de
NOTICE
*) If the connection lines are equipped with ball valves (item 28), the pressure gauge can be replaced during operation.

14
© 09.2015
1
2
3
15
16
5
6
12
10
11
9
7
8
4
9 Spare Parts
9.1 Spare Parts Drawing DF 100 - DN 25

15 © 09.2015
1
2
3
15
16
5
6
12
10
11
7
8
4
9
13
9.2 Spare Parts Drawing DF 100 - DN 50 / DN 80 / DN 100
shown DN 50
shown DN 100

16
© 09.2015
1
2
3
15
16
5
6
14
12
11
7
8
4
13
10
17
13
11
9.3 Spare Parts Drawing DF 100 - DN 150 / DN 200
shown DN 150

17 © 09.2015
26
26
24
22
24
22
28
9.4SparePartsDrawingDifferentialPressureMeasuringInstrument
without Reed contact
with Reed contact
Ball valve (item
28) in operating
position

18
© 09.2015
..............................................................................
..............................................................................
..............................................................................
..............................................................................
..............................................................................
..............................................................................
..............................................................................
..............................................................................
10 Notes
9.5 Parts for Maintenance Work
Item
no. Designation Number Order number
DN 25 DN 50 DN 80 DN 100 DN 150 DN 200
2 Sealing ring* 2 DR-21 DR-21 DR-21 DR-21 DR-21 DR-21
4 O-ring 1 OF-41 OF-42 OF-42 OF-45 OF-45 OF-45
7 O-ring 1 OF-71 OF-72 OF-73 - OF-75 OF-76
9 O-ring 1 OF-71 - - OF-94 - -
10 Filter cartridge* 1 FP-101 FP-102 FP-103 FP-104 FP-105 FP-106
11 O-ring 1 (2)** OF-111 OR-94 OF-113 OF-114 OF-115 OF-116
Differential pressure measuring instrument
22 Sealing ring* 2 DR-21 DR-21 DR-21 DR-21 DR-21 DR-21
24 Progressivring 2 PR-41 PR-41 PR-41 PR-41 PR-41 PR-41
Order example:
Item no. Designation Number Order number Fabrication no.
7 O-ring 1 Stück OF-72 1504-091
Alternatively to the fabrication number, you can also give us the following information:
Example:
Device type DF 100
Year of construction 2015
Nominal width DN 50
Type of gas natural gas
*) These parts are replaced during maintenance only when necessary.
**) Number given in brackets for DN 150 and DN 200

MEDENUS Gas-Druckregeltechnik GmbH
Fon +49 (0)2761 82788-0
Fax +49 (0)2761 82788-9
Saßmicker Hammer 40 / D-57462 Olpe
www.medenus.de
Other manuals for DF 100
2
Table of contents
Other MEDENUS Water Filtration System manuals
Popular Water Filtration System manuals by other brands

3M
3M Aqua-Pure AP1610SS Installation and operating instructions

Brita
Brita Purity C Series manual

North Star
North Star NSRO42C4 Installation and operation manual

Aquaporin
Aquaporin ZERO Replacement guide

Grunbeck
Grunbeck GENODOS DME 6 Operation manual

Glasco UV
Glasco UV GUV-7 Operation & maintenance manual

3M
3M Aqua-Pure AP430SS Installation and operating instructions

3M
3M 3MRO401 Installation and operating instructions for

AQUAMAXX
AQUAMAXX CTECH T-2 manual

Hyundai
Hyundai Poseidon user guide

ELECTROSEA
ELECTROSEA CLEARLINE CL-2000-X2.0 installation manual

Filtration Group
Filtration Group Pi 210 Series manual