MELAG MELAdem 55 User manual

Technical details are subject to modification without prior notice
Operating
Manual
Reverse
Osmosis
Water
Purifier
MELAdem®55
Dear Doctor,
Thank you for placing your trust in this reverse osmosis water purifier.
Since 1951, MELAG has specialised in the production of sterilisation equipment for medical
practice. During this period, our family-owned and operated company has succeeded in becoming
a leading manufacturer of sterilisation equipment. More than 450,000 MELAG appliances world-
wide testify to the exceptional quality of our sterilizers.
Before using the MELAdem55 please study this operating manual carefully.
In order to ensure a long working life for your reverse osmosis water purifier and to preserve its
value, please make sure that it is serviced regularly.
MELAG management and personnel.

Contents
CONTENTS
1Installation .............................................................................................................................. 3
2Information on components of the system .............................................................................. 3
3Placing into operation for the first time.................................................................................... 5
4Connection of the MELAdem system...................................................................................... 6
4.1 Connection of the MELAdem system to a Cliniklav 25 autoclave......................................................................... 6
4.2 Connection of the MELAdem system to an autoclave Type 25............................................................................ 7
5Maintenance instructions........................................................................................................ 8
5.1 Replacement of the Pre-filter................................................................................................................................ 8
5.2 Replacement of the activated carbon filter........................................................................................................... 9
5.3 Replacement of the mixed-bed resin cartridge in the ion-exchanger.................................................................... 9
5.4 Quality test for the permeate................................................................................................................................ 9
6Technical Data.......................................................................................................................10

Operating
Manual
MELAdem55
Page 3
1 Installation
The MELAdem 55 demineralises water using the principle of reverse osmosis. Incoming water at mains pressure is
passed through a semi-permeable membrane contained in a reverse-osmosis module. The mineral content of the water
which permeates the membrane is reduced by approximately 95 %.
The incoming water is thus split up into two parts:
•a demineralised permeate, and
•a part with increased salt concentration (concentrate), which is disposed of.
In order to ensure that even poor quality mains water is purified sufficiently for the operation of medical equipment e.g.
an autoclave, the permeate water from the reverse-osmosis module is also passed through an ion exchange unit. This
contains an ion-exchange resin which is able to reduce any residual mineral contents of the permeate to a minimum. A
storage container is provided in which the purified water can then be stored. It can be connected by a pipe directly to the
autoclave. When the storage container is full, the water inlet valve closes automatically. All operations are controlled by
water pressure.
2 Information on components of the system
1. Mains water supply
connection
2. Pre-filter
3. Activated carbon filter
4. Ion exchanger
5. Reverse osmosis module
6. Storage container
7. Concentrate outflow pipe
1
3
4
5
7
Sensor
feed water
to feed water
intake of the
Concentrate out flow pipe
Illustration
1 Components of the MELAdem 55
2

Page 4
1: Mains water supply connection
The water purifier is connected to the mains supply via a 3/4’’ stop tap, which is normally left on at all times.
The stop tap should be fitted with a non-return valve and an air-release valve in order to prevent
water flowing back into the mains supply. It should only be closed if the system is not going to be used for a
period of time (see Section 5“Maintenance instructions“ here), or during servicing and maintenance work. We
also recommend fitting a Water Stop (MELAG Art. No.: 01056), in order to prevent damage due to leaks.
2: Pre-filter
A pre-filter in an exchangeable filter cartridge helps to protect the sensitive reverse-osmosis membrane. It
removes suspended particles, rust and other undesirable matter from the incoming mains water. The
cartridge should be exchanged at least every 6 months, or more frequently if the water supply contains
particularly high levels of impurities.
3: Activated carbon filter
Further protection is provided for the reverse osmosis membrane by an activated carbon filter, again in an
exchangeable cartridge. The activated carbon filter is intended primarily to remove free chlorine,
which would damage the reverse-osmosis membrane. The activated carbon cartridge should also be
replaced every 6 months, or more often if the chlorine level of the water supply is high.
4: Ion exchanger
The permeate from the reverse osmosis module flows into an ion exchanger, where it is
demineralized (conductivity < 1µs). The ion exchange resin in the storage container element should be
renewed when it is exhausted.
5: Reverse osmosis module
The reverse osmosis module is the key element in the water purifier. It consists of a thin, wound polyamide
membrane. In addition to colloids, it also holds back many organic substances such as
hydrocarbons, chloro-hydrocarbons, phenols, pesticides etc. The retention capability of the reverse osmosis
filter is approx.
95%. The water which has passed through the filter, known as the permeate, flows on to the ion exchanger.
The residual concentrate is disposed of.
6: Water storage tank
After passing through the ion exchanger, the demineralised water is led into a plastic storage
container where it is stored until needed.
The water storage tank is equipped with a shut-off valve that stops the water feed when the water level
reaches the maximum of approx. 15 litres. When this valve is closed, the shut-off valve of the reverse-
osmosis unit also closes, which then stops the flow of untreated water.
An electrical float switch provides a signal to the autoclave type 25 if there is not enough water in the water
tank (< 3 litres).
The permeate storage tank (for the output water) is directly connected to the autoclave if it is equipped for
automatic water replenishment feed (see Section 4“Connection of the MELAdem system“ here). In such
cases, the autoclave will intake demineralised water from the lower-lying water storage tank.
7: Connection of the drainage hose for used water
The concentrate from the reverse osmosis is led away for disposal through the concentrate outflow pipe. If
the system is connected to a Cliniklav25 autoclave or an autoclave type 25, the drainage hose for used
water is connected to the drainage hose of the autoclave itself.

Operating
Manual
MELAdem55
Page 5
3 Placing into operation for the first time
When placing the device into operation for the first time, please proceed as follows:
1. Remove the enclosure lid from the unit: unscrew the screws, and carefully remove the enclosure lid in an
upward direction.
2. Use the enclosed screws to mount the unit to a mounting bracket on the wall, or in a lower cabinet.
Install the unit so that it is possible at all times to exchange the filter.
3. Place the water storage tank in the lower cabinet, or next to the unit on the floor
(see Section 4 here).
4. Remove the black closure flaps from the connections on the device, and keep them in a safe place
(you will need them again when you shut down the device, or move it to another place).
5. Hook up the lines as shown in the illustration and description in Section 2 here. All the connecting lines on the unit
have been designated by adhesive labels. The hose connections are delivered with threaded fittings, clamping
rings, and coupling rings (union nuts). Do not allow these hoses to kink (twist or curl).
6. Remove the container element from the ion exchanger. This is necessary to prevent rinse water from
damaging it.
7. Open the feed for the untreated water, and check the threaded fittings to make sure that there is no leakage.
8. Disconnect the hose screw connection from the reverse osmosis unit to the internal storage tank on the latter and
place the hose end in an outflow, floor drain or a bucket.
9. Unscrew the filter housing of the subsequent ion exchanger (4), remove the mixed-bed resin cartridge and screw
on the empty filter housing to flush the unit.
10.Open the raw water tap and let the raw water run through the unit for ca. 20 minutes in order to free the unit from
preservatives and dust residue.
11.Connect the raw water tap after flushing.
12.Unscrew the filter housing of the subsequent ion exchanger. Attention: the filter housing is full to the brim with
water.
13.Pour off the water in the filter housing. Replace the mixed-bed resin cartridge in accordance with the section 5.3
and screw-on the filter housing again.
14.Screw on the house screw connection from the reverse osmosis unit to the internal storage tank to the latter
15.Open the raw water tap. Check whether the reverse osmosis unit and the hose connections are free from leaks
16.Allow the water storage tank to fill for approx. 1.5 - 2 hours, and check the function of its shut-off valve. When
the tank is completely full, the valve in the tank and in the reverse-osmosis unit will close. This means that
no concentrate will flow.
17.Do not use the contents of the first tank filling (the preservative contained in the system will pollute the first tank
filling). One tank filling will take approx. 3 - 4 h, depending on the pressure and temperature of the untreated
water.
18.Rinse the enclosure and put the container element back in its place.
19.Check the quality of the permeate with a conductivity tester.
Replace the enclosure lid onto the unit. The system will now function completely automatically.

Page 6
Connections seen from below
Feed water 3/4
Cooling water3/4
4 Connection of the MELAdem system
4.1 Connection of the MELAdem system to a Cliniklav 25 autoclave _______________________________
6 Wall mounted trap with double hoze
nuzzle
8 Outlet stop trap 1/2”–3/4” with air
release and non return valve with Y-
connection
9.1, 9.2, 9.3 Water stop, consisting of
control unit, solenoid valve and water
sensor
11 Pressure hose for water purification
system
12 Pressure hose for cooling water
12.1Connection for hose cooling water 3/4”
13 Pipe for wastewater DN22
13.1Connection for pipe waste water DN
22 Cliniklav25
14 Pressure hose for feed water supply
14.1 Connection for Pressure hose feed
water supply
15 Hose for wastewater for water
purification system
15.1 Connection for pressure hose waste
water MELAdem 55
16 Hose for feed water for storage tank
17 Water purifier (reverse osmosis
system
18 Storage tank for water purification
system
The MELAdem 55 reverse-osmosis system can be directly connected to autoclaves that have been prepared for
connection of a non-pressurized automatic water feed system. With such a configuration, the autoclave directly pulls
demineralised water out of the deeper-lying water storage tank.
Please follow the steps below to connect the MELAdemreverse-osmosis system to such a properly prepared autoclave:
1.Install the reverse-osmosis system as described in Section 3 “Placing into operation for the first time“
above.
2.Use the 1/2" delivered pressure hose with 3/4" fittings to connect the suction element of the permeate storage tank
to the feed-water connection of the autoclave.
3. Connect the drain hose of the water-treatment unit to the drain hose connection of the autoclave.
Illustration 2 Cliniclave 25, complete with lower cabinet
and water treatment unit (view from behind)
Wastewater Cliniclave 25
Wastewater water purifier

Operating
Manual
MELAdem55
Page 7
4.2 Connection of the MELAdem system to an autoclave Type 25________________________________________________
6 Wall mounted trap with double hoze
nuzzle
8 Outlet stop trap 1/2”–3/4” with
air release and non return
valve with Y- connection
9.1, 9.2, 9.3 Leak detector,
consisting of control unit,
solenoid valve and water
sensor
11 Pressure hose (2 pieces)
13 Wastewater hose DN22
14 Pressure hose for feed water supply
15 Wastewater hose for water
purification system
15.1 Connection for waste water hose
16 Feed water hose to storage tank
17 Water treatment unit (reverse
osmosis unit)
18 Storage tank for water
treatment unit
19 Sensor for feed water shortage
The MELAdem 55 reverse-osmosis system can be directly connected to autoclaves that have been prepared for
connection of a non-pressurized automatic water feed system. With such a configuration, the autoclave directly pulls
demineralised water out of the deeper-lying water storage tank.
Please follow the steps below to connect the MELAdemreverse-osmosis system to such a properly prepared autoclave:
1. Install the reverse-osmosis system as described in Section 3 “Placing into operation for the first time“
above.
2. Use the 1/2" delivered pressure hose with 3/4" fittings to connect the suction element of the permeate storage tank
to the feed-water connection of the autoclave.
3. Connect the drain hose of the water-treatment unit to the drain hose of the autoclave.
Illustration
3 Autoclave Typ 25, complete with lower cabinet
and water treatment unit (view from behind)

Page 8
5 Maintenance instructions
The following maintenance tasks should be carried out at the stated intervals in order to ensure trouble-free operation:
Frequency
Maintenance task
Daily
Check the permeate with
the conductivity tester (with the Cliniklav
25 autoclave or the autoclave type 25, this will take place
automatically, before every program run).
Every 6 months
Replace the pre-filter
Every 6 months
Replace the activated carbon filter
As necessary (if conductivity values are poor)
Renew the ion-exchange resin cartridge
At regular intervals
Check the tubes and threaded connections for leaks, compressions
or kinks (twists or curls)
Various measures are necessary before or after periods of inoperation:
Period of inoperation
Precautionary measures
Up to 2 weeks
Turn off water intake
More than 2 weeks
Turn off water intake, empty storage container
On resuming operation after more than 4
weeks
Swill out equipment and storage container, and install new pre-
filter, activated carbon filter and ion exchange resin
5.1 Replacement of the Pre-filter_________________________________________________________________
•Turn off the mains water supply
•Unscrew the filter cylinder (a) from the head (b)
•Remove the filter unit (1) and place the new filter unit in
the cylinder
•Take the sealing ring out of the cylinder (a), clean, give
a slight coating of silicone and replace
•Screw back the cylinder by hand (do not use
tools) Check that all parts sit tightly, with no
leaks.
•Check that all parts sit tightly, with no leaks.
Illustration 4 Pre-filter

Operating
Manual
MELAdem55
Page 9
5.2 Replacement of the activated carbon filter ____________________________________________________
•Turn off the mains water supply.
•Unscrew the filter cylinder (a) from the head (b)
•Remove the activated carbon filter unit (1) and place the
new activated carbon unit in the cylinder (with the arrow
•pointing upwards!)
•Take the sealing ring out of the cylinder (a), clean, give a
•slight coating of silicone and replace
•Screw back the cylinder by hand (do not use tools)
•Check that all parts sit tightly, with no leaks
.
Illustration 5 Activated carbon filter
NOTE! It is important to exchange the filter unit regularly in order to avoid damage to the reverse
osmosis module which would shorten its operational life.
5.3 Replacement of the mixed-bed resin cartridge in the ion-exchanger____________________________
When the mixed-bed resin is exhausted (inadequate water
quality of the permeate) then it must be exchanged:
•Turn off the mains water supply and if necessary empty
the storage container
•Unscrew the cylinder (a) from head (b)
•Take out the container element (1), remove the lid
and pour out the exhausted resin
•Refill with 0.5 litre fresh resin (the contents of one
package), knocking the element to compact the resin, so
that the entire contents can be included
•Close the lid of the container element
•Take the sealing ring out of the cylinder (a), clean,
grease slightly and replace
•Re-insert the resin unit with the three feet on the
underside
•Tighten the cylinder by hand (do not use a tool)
•Check that all parts sit tightly, with no leaks.
5.4 Quality test for the permeate _________________________________________________________________
The quality of the permeate must be regularly checked. The Cliniklav 25 autoclave and the type 25 autoclave have an
integrated conductivity tester. In the initial operating state of the autoclave type 25, you can display the conductivity by
pressing the button “Water”. At the Cliniklav 25 autoclave the conductivity can be displayed at any time by pressing the
button . The system will automatically measure the conductivity before every program run. If the water quality is poor,
the autoclave display will show the error message “Water quality poor”, and no program will start. The quality of the
permeate should therefore be regularly tested, in order to allow prompt maintenance of the water treatment system. If the
MELAdem system is not connected to the autoclave, you should regularly check the permeate with a conductivity tester.
Illustration
6 Ion-exchanger

Page 10
Rev.: 1 – 14/0068 55 BA_eng_doc
6 Technical Data
Water source
Mains water supply
Bacteriological quality
Drinking water
Membrane
TFC-bacteria resistant, wound module
Permeate output
max. 200 litre/day (at 4.5 bar/25°C)
Retention capability RO
Rated retention capability approx. 95%
Conductivity of purified water (permeate)
- approx. 20-30 µS/cm from 600 µS intake water
- < 1 µS/cm from 600 µS intake water with ion exchanger
Water pressure min/max.
2 bar to 6 bar
Water temperature min./max.
5°C to 35°C
Water pH-value min./max.
4.0 to 10.0
Total mineral content of mains water
max. 1500 mg/l
Iron content of mains water
max. 0.1 mg/l
Yield
approx. 20 - 25 %
Pre-filter I
Fine particulate filter 10”, 5 µm
Pre-filter II
Activated carbon 10”
Post-treatment
Ion exchanger, contents approx. 0.5 litre resin
Dimensions H x W x D
approx. 445 mm x 390 mm x 130 mm
Overall weight (with filters)
approx. 12 kg
Capacity, storage container
approx. 15 litre
Dimensions, storage container H x W x D
approx. 350 mm x 270 mm x 230 mm
Length of inlet / outlet pipes
approx. 1500 mm
Table of contents
Other MELAG Water Filtration System manuals
Popular Water Filtration System manuals by other brands

Steinbach
Steinbach Speedclean Comfort 50 ORIGINAL OWNER'S MANUAL

Water Well
Water Well MB18 user manual

Honeywell Home
Honeywell Home HM600XROF1/U manual

Whirlpool
Whirlpool WHEUFF Installation and operation manual

Beko
Beko CLEARPOINT 3eco S040 Installation and operation manual

Oase
Oase BioSmart30000 operating instructions

Atlantic Ultraviolet
Atlantic Ultraviolet MIN-1 owner's manual

Vortex
Vortex 85-1448-10 (10 GPM) Installation and service manual

Propur
Propur ProMax instruction manual

Katadyn
Katadyn HFD-PS40 quick start guide

Watts
Watts PWSYS-FIL-ICE3 Installation and operation manual

Sartorius
Sartorius arium 611VF Installation and operation manual