MEWOI MEWOI9000 User manual

Leakage Current Monitoring Recorder
MEWOI9000,MEWOI9000C
User Manual
www.mewoi.com


CONTENT
Warning................................................................................................ 1
I.Brief Introduction ............................................................................. 3
II.Electrical Symbol............................................................................ 4
III.Series Model.................................................................................. 4
IV. Technical Specifications.................................................................. 4
V. Instrument Structure......................................................................... 6
VI. Instrument Operation...................................................................... 7
1 Turning on and off...................................................................... 7
2. Language Selection.................................................................. 7
3. Changeover Display Modes...................................................... 7
4. Selecting Power-saving Mode.................................................. 8
5. Setting Time.............................................................................. 8
6. Data Inquiry............................................................................... 9
7. Deleting Data .......................................................................... 10
8. Leakage/current measurement............................................... 10
9. Real-time Monitoring............................................................... 13
10. Data Download ..................................................................... 13
VII. How to operate the monitor software .......................................... 14
1. Home Page............................................................................. 14
2. Real-time Monitoring Menu..................................................... 15
3. Alarm Prompt .......................................................................... 15
4. Storing Waveform ................................................................... 16
5. Historical Inquiry Display ........................................................ 16
6. Reading Historical Data.......................................................... 17
7. Historical Data Queuing inAscending Order.......................... 18
8. Historical Data Queuing in Descending Order ....................... 18
9. Historical Data Storage........................................................... 19
VIII. Replacing Battery ....................................................................... 20
Ⅸ.Accessories................................................................................... 21

-1-
- 2-
Warning
Thank you for buying MEWOI9000 Leakage/Current Monitoring Recorder
manufactured by our company. Please keep to the following requirements in order
to use the product properly:
——Read the User Manual carefully.
——keep to the safety rules and precaution given in the manual strictly.
Pay special attention to safety when using this instrument in any case.
Observe the text and symbols attached to the front and rear panels of the
instrument.
Please replace the battery with a new one when the batter voltage is so low
that the LCD display is dim.
This instrument can not be turned off automatically. Please turn off it after
use.
Don not use the instrument to measure a line with a voltage higher than
600V.
It is forbidden to use the instrument when neither rear cover of the instrument
nor the batter cover is in place.
Never use the instrument when the instrument case or measuring wire is
broken and therefore any metallic part is exposed.
Don not place the instrument at a environment with high temperature, high
humidity, condensation and direct sunshine for a long time
It is necessary to maintain the instrument and clamp periodically to keep
them clean. Don not wipe the clamp by using corrosive medium or abrasive
object.
Avoid impact upon the current clamp, especially its mating surface
Please observe the correct polarity when replacing the battery. Remove the
battery from the instrument if you expect not to use the instrument for a long
time.
Operation, disassembly and maintenance of the instrument must be carried

-2-
out only by a authorized and qualified person.
When there is an instrument fault, never use it because continual use may
result in danger. In this case, isolate the instrument immediately and
delivered to an authorized agency for dealing with it.
The user must follow the safety instructions preceded by “ ”warning symbol
on the instrument and manual.
The user must strictly follow the instructions preceded by “ ”danger symbol
on the instrument and manual.

-3-
I.Brief Introduction
MEWOI9000 series Current/Leakage Monitoring Recorder is well
designed and manufactured for online measurement, monitoring and recording of
AC leakage / current. It is composed of a tester, monitoring software, current
clamp and communication cable, etc. This instrument has been widely applied in
many fields, such as electrical power, telecommunication, meteorology, railway, oil
fields, construction, metering, research agencies, universities, mines and
enterprises.
Both Chinese and English MMI is built in the instrument. Therefore, it is
suitable for use by engineers and technicians of different counties. The instrument
has a large LCD display with blue background and a large storage space. 4200
sets of data can be stored in the instrument. In addition, the instrument can
automatically store data at an interval, which is adjustable from 1 to 60 min.
MEWOI leakage/current monitor software provides online and real-time
monitoring and historical data inquiry. With the software developed by our
technical department, the instrument can display leakage current dynamically,
display current waveform, maximum value, minimum value and average value.
The users can preset the alarm limit. The instrument will generate an alarm when
the measured current exceeds the limit. In addition, the users can read, inquire,
store, analyze, process, sequence historical data and generate and print a fitting
curve and report.
The current clamp is made of a special alloy, adopts the latest CT technique
and magnetic shielding technique and is not almost interfered by external
magnetic field. This ensures high accuracy, high stability and high reliability of the
values which are measured continuously.
The Leakage/Current Monitor Recorder is suitable for measuring AC lines up
to AC600V. With the instrument, the users can easily find out the cause of
insulation fault and circuits with an insulation fault, monitor intermittent electric
leakage, find out the time when an insulation fault occurs and repair lines.

-4-
II.Electrical Symbol
Extremely dangerous!The operators must keep to the safety
rules strictly, otherwise, electric shock will result in death or injury.
Dangerous!The operators must keep to the safety rules strictly,
otherwise, electric shock will result in will result in death or injury.
Warning!The operators must keep to the safety rules strictly,
otherwise, personal injury or equipment damage will occur..
Double insulation
Alternating current (AC)
Direct current (DC)
III.Series Model
Model
Range
Resolution
Accuracy
Clamp size
MEWOI9000
0.0A-500A
0.1A
±2%±5dgt
35mm×40mm
MEWOI9000C
0.00mA-20A
0.01mA
±2%±5dgt
25mm×30mm
IV. Technical Specifications
Functions
Measuring AC leakage current and AC current, on-line
monitoring and recording, locating insulation faults and
repairing electrical lines.
Power source
DC6V, alkaline battery LR6 1.5V×4
Testing
Method
Clamp CT, true RMS
Display
Method
LCD:128dots×64dots
Instrument
Dimension
Tester(L×W×H): 75mm ×170mm ×30mm
Current clamp(L×W×H): 170mm ×70mm ×38mm
Weight
Tester: 240g (battery included)
Current clamp: 170g

-5-
LCD
Dimension
Display field: 44mm×27mm
Shield Type
Type B: Double shielding
Sampling
Rate
2 samples/s
Data storage
4200 sets (data loss does not occur after power failure
or replacing the battery)
Time Setting
Recording internal adjustable from 1 to 60 min, If set to
0 min, the auto-store function is turned off
Recording
Time
Continuously operate for about 10 days at the
power-saving mode
Line Voltage
Line with a voltage up to AC600V
Overflow
Display
The “OL” symbol is displayed when a measured value is
beyond the measurement range
Battery
Voltage
A Low-Battery symbol is displayed and the user should
replace the battery with a new one when the battery
voltage drops to 4.6V to 4.8V. In this case, the
measured values are still accurate.
Rated Power
about 5mA at the power-saving mode, maximum current
consumption: 20mA
Lead Length
2 m
Temperature
and Humidity
Working: -10℃~40℃;below 80%rh
Storage: -10℃~60℃;below 70%rh
Insulation
Resistance
Higher than 100MΩ, 1000V
Safety Codes
IEC1010-1, IEC1010-2-032, Pollution Class 2, CAT
Ⅲ(600V), IEC61326(EMC standard)

-6-
V. Instrument
Structure
1. Current clamp interface
2. LCD(128dots×64dots, Chinese and English MMI)
3. Up, down, left, right keys and MEM control key
4. Instrument body
5. POWER key (for turning on and turning off the instrument)
6. RS232 interface for data download
7. Round-end current clamp
8. Current clamp signal output plug

-7-
VI. Instrument Operation
1 Turning on and off
Turn on the instrument by pressing the POWER key.
The LCD will light up. The display on the LCD is as
shown in the left figure. If the LCD display is dim, the
battery voltage may be low. If so, replace the battery with
a new one. Press the POWER key again to turn off the instrument. This
instrument can not be turned off automatically. Please turn off it after use.
2. Language Selection
Both Chinese and English MMI is built in the instrument to facilitate use by
engineers and technicians of different counties. After the instrument has
been turned on normally, select the language by pressing the LEFT and
RIGHT arrow keys, press the MEM key to confirm the selected language
and enter the measurement status.
Selecting English MMI Selecting Chinese MMI
3. Changeover Display Modes
After confirming the selected language and entering the measurement status,
press the LEFT and RIGHT arrow keys to change over between
measurement display modes. The measured values can be displayed in two
modes as shown in Figures A and B. Only the measured leakage current is
displayed in Fig. A. The measured leakage current, number of stored data
sets, date &time and preset storage interval are displayed in detail in Fig. B.
Press the RIGHT arrow key to change over to the display mode as shown in
Fig. B, and the LEFT arrow key to change over to the display mode as
shown in Fig. A.

-8-
4. Selecting Power-saving Mode
Press the DOWN arrow key at the measurement status to turn off the LCD
back-light and enter the power-saving mode. Press the UP arrow key to turn
on the backlight. The power consumption at the power-saving mode is only
20 percent of the one at the common mode (the backlight is turned on). It is
recommended that the power-saving mode be adopted for long-time online
measurement and recording.
5. Setting Time
Press the MEM key at the measurement state to access the function menu,
press the UP or DOWN arrow keys to move the cursor to the Time Setting
item, and press the MEM key again to enter the date& time setting mode. At
the date& time setting mode, press the UP/DOWN arrow keys to change
values, press the LEFT or RIGHT keys to move the cursor, press MEM key
to “confirm” or “cancel” the setting.
After the recording interval has been set, return to the measurement state.
The instrument will automatically store the measured values at the set
interval. The instrument can store up to 4200 data sets. If the memory is full,
FULL will be displayed on the LCD. The instrument can store new data only
after some of stored data is deleted.

-9-
The instrument doesn’t have automatic clock function.
The date and time will reset to the default value (08.08 AM,
Aug 8, 2008) each time the instrument is turned on.
Press the MEM key to “Cancel” or maintain the previous
setting.
The storage interval is defaulted to “00” min each time the
instrument is turned on, that is to say, no measured value
is stored.
6. Data Inquiry
At the measurement state, press the MEM key to access the function menu,
press the UP or DOWN arrow keys to move the cursor to “Data Inquiry” item,
and press the MEM key to enter the inquiry submenu. Inside the inquiry
submenu, the first left line display the current page number and total pages,
the third line displays three sets of recorded values, and the right side
displays the page number increment. It is allowed to rapidly navigate to the
desired page number. Press the UP or DOWN arrow keys to move the
cursor and select the page number increment and then press the MEM key
to confirm the selection.
Press the LEFT and RIGHT arrow keys under the inquiry submenu to move
the cursor. Press the MEM key with the cursor located at certain data set to
display the detailed information about this data set, including the set number
/stored data sets, current amplitude and recording time, etc.
Generally speaking, press the arrow keys to move the cursor and press MEM
key to confirm the selection under the inquiry submenu. The relevant display
is as shown in the following figure:

-10-
7. Deleting Data
At the measurement state, press the MEM key to access the function menu,
press the UP or DOWN arrow keys to move the cursor to “Deleting Data”
item, When press the MEM key again, a data deleting tips will pop up. Press
the MEM key with the cursor located at “OK” to delete the stored data. Press
the MEM key with the cursor at Cancel to cancel the deletion and return to
the main menu.
It is impossible to recover the deleted data. Take care to
delete any data.
The deletion operation will delete all of the stored data.
8. Leakage/current measurement
Electric shock, dangerous !The instrument can be
operated only by the trained and authorized person. The
operators must keep to the safety rules strictly;
otherwise, electric shock will result in personal injury or
damage to the equipment.
Don not use the instrument to measure the lines with a
voltage of higher than 600V and current of higher than
20A, Otherwise, electric shock will result in personal
injury or damage to the equipment.

-11-
1) Connect the current clamp with the instrument body properly, turn on the
instrument and enter the measurement mode.
2) Clamp the measured object with the current clamp (ensure that the clamp
end should be fully closed). Observe the reading. If an OL symbol is
displayed on the instrument LCD, the measured current exceeds the
upper measurement limit of the instrument.
3) Reference legend:
In order to measure the leakage current of an
electrical equipment, clamp the phase line and neutral
line. (take care to clamp the two lines)
In order to measure the leakage current of a ground
conductor, clamp the ground conductor. (take care to
clamp only one conductor)
In order to measure the total current of a main line,
clamp this main line. (Take care to clamp only one
line).
For the sake of safety, remove the instrument away
from the conductor with high voltage and current after
measurement has been completed correctly.
Detect electric leakage of cables embedded in wall

-12-
Detect electric leakage of equipments
Detect electric leakage of ground conductors

-13-
Measure line current
9. Real-time Monitoring
Turn on the instrument and enter the measurement state. Connect a PC with
the instrument via a RS232 communication cable supplied with the
instrument. Run the software installed in the PC. If the communication is
normal, the users can monitor current in real-time manner via the PC.
10. Data Download
Connect the instrument with the PC via the RS232 communication cable
(supplied with the instrument) and turn on the instrument. Run software,
select the Historical Inquiry and read the data. The more the data is, longer
time it takes to read them. If the instrument memory is full, it takes about
2min to read all data.

-14-
VII. How to operate the monitor software
MEWOI Leakage/Current Monitoring software needs to be installed in
Windows XP/2000 systems. The monitor software provides online and
real-time monitoring and historical data inquiry. With the MEWOI software,
the instrument can display current dynamically, display current waveform,
maximum value, minimum value and average value. The users can preset
the alarm limit. The instrument will generate an alarm when the measured
current exceeds the limit. In addition, the users can read, inquire, store,
analyze, process, sequence historical data and generate and print a fitting
curve and report. The following paragraphs will describe these functions
briefly.
1. Home Page
Enter the real-time monitoring menu by clicking “Enter” with a mouse.
Visit the company website for inquiry if the users require other services.

-15-
2. Real-time Monitoring Menu
Special notes: You can zoom the waveform by pressing the left key of
mouse and dragging the cursor upward or downward when the cursor is
located within the waveform display area.
3. Alarm Prompt

-16-
A red alarm lamp will blink when the measured current value exceeds the
preset alarm limit. The users may cancel the alarm setting value.
4. Storing Waveform
A waveform may be stored while the current is monitored in a real-time
manner. A waveform is stored in BMP format.
5. Historical Inquiry Display

-17-
The historical inquiry functions include such functions as reading, storing
and clearing historical data, generating a curve, printing data, queuing data in
ascending or descending order, data backup and page-up and page-down.
6. Reading Historical Data
This manual suits for next models
1
Table of contents
Other MEWOI Measuring Instrument manuals

MEWOI
MEWOI 8000 User manual

MEWOI
MEWOI MEWOI400 User manual

MEWOI
MEWOI MEWOI1200 User manual

MEWOI
MEWOI MEWOI-HY128 Series User manual

MEWOI
MEWOI 7800 User manual

MEWOI
MEWOI MEWOI1000 User manual

MEWOI
MEWOI MEWOI9000G User manual

MEWOI
MEWOI MEWOI8100E User manual

MEWOI
MEWOI MEWO1000T User manual

MEWOI
MEWOI MEWOI1330B User manual
Popular Measuring Instrument manuals by other brands

MR
MR MESSKO MFLOC 2.0 operating instructions

Hioki
Hioki PW3336-01 Measurement guide

Ludlum Measurements
Ludlum Measurements 9-4 user manual

Hanna Instruments
Hanna Instruments HI 8731 instruction manual

Thermo Scientific
Thermo Scientific Orion Versa Star Pro user manual

Solinst
Solinst 101 P2 Replacement manual