MHC Medical Products Easy Touch User manual

EasytTouch
®
Glucose Monitoring System
User Manual

See our other products at
www.EasyTouchProducts.com
EasytTouch
®
Glucose Monitoring System

Important Information ..................................................................................3
Introduction ...................................................................................................7
About the Product ........................................................................................8
About the EasyTouch® Meter ................................................................9
About the EasyTouch® Meter Display ................................................. 11
About the EasyTouch® Test Strip ........................................................13
About the EasyTouch® Test Strip Vial .................................................14
Installing/Replacing the Battery ..........................................................15
Setting Up the Meter ...................................................................................16
Setting Up the Date, Time ..................................................................16
Setting Up Alarm ................................................................................17
Unit Selecton ......................................................................................20
Table of Contents

The Control Solution Test ......................................................................... 21
Testing Your Blood Glucose Level ........................................................... 25
Preparing the Test Strip ..................................................................... 26
Obtaining a Blood Sample................................................................. 27
Applying Blood Sample to the EasyTouch® Test Strip ....................... 28
Memory Features ....................................................................................... 31
Transfer Test Results to a Computer ....................................................... 33
Care and Storage ....................................................................................... 34
Display Messages...................................................................................... 35
............................................................................................ 42
Limitations of the Procedure.............................................................. 44
Service and Warranty ................................................................................ 44
Customer Service ...................................................................................... 46

3
There are limitations for doing AST. Please
consult your healthcare professional before
you do AST.
Forearm
Palm
Available Alternate Sites Testing : Palm and Forearm.
Alternate Site Testing (AST) means you can use other parts of the body other
1. The EasyTouch® Glucose Monitoring System allows you to test on the
Important Information

4
alternative site if:
CAUTION:
samples from the forearm and other alternate sites. Rub the alternate test
sites about 20 seconds before lancing. If you are testing for hypoglycemia
(low blood glucose), or if you suffer from hypoglycemia unawareness, we

5
more often than you do now. For people with diabetes, more frequent
testing is a good thing. Just remember: any time you want to be sure of an
ONLY in the following intervals:
DO NOT use AST if:

6
medical condition. If your blood glucose reading is unusually low or high,
or if you do not feel the way your reading indicates, repeat the test with
a new test strip. If your reading is not consistent with your symptoms or
if your blood glucose result is less than 60 mg/dL (3.3 mmol/L) or higher
than 240 mg/dL (13.3 mmol/L) you should contact your
healthcare professional.
® blood
glucose test results without the consent and advice of a physician or
healthcare professional is not recommended.
results. If you believe you are suffering from severe dehydration, consult
your physician immediately.

7
® Glucose Monitoring System, the
latest advancement in biosensor technology.
®
Glucose Monitoring System is designed to process accurate results utilizing
only a small amount of blood for in vitro diagnostics. The EasyTouch®
Glucose Monitoring System is intended for use in the home and in
professional settings to monitor whole blood glucose levels obtained from the
Introduction

8
The EasyTouch® Glucose Meter & Monitoring System is available in two
different package options with the following contents:
A. Your EasyTouch® Glucose Monitoring System contains:
® Meter
B. Your EasyTouch® Institutional (Meter Only) Box contains:
® Meter
Test Strips, Control Solution, & Lancing Device can be purchased individually.
About the Product

9
About the EasyTouch® Meter
S/N:M05B13286
BLOOD GLUCOSE METER
EasytTouch
®
For In Vitro Diagnostic Use Only
Made in Taiwan to Specs of
MHC Medical Products, LLC
8695 Seward Road
Faireld, OH 45011
Toll #: 877.358.4342
%)t3FWEBUF

10 1. Display
screen guides you through the test by using numbers and symbols.
2. M Button The M (mode) button is used to set time, alarm, enters the
memory mode to show the consecutive 7/14/30-day average results and
turns the Meter on and off.
3. S Button The S (select/set) button is used to select year, month, day,
time, memory of results and control solution mode.
4. Test Port This is where you insert the Test Strip into the Meter.
5. Unit Switch Switch for mmol/L or mg/dL.1 mmol/L =18 mg/dL
6. Battery Cover
7. Label
8. Reset Button Press this button to re-set the system.
9. Data Port

11
1. Date
2. Time
3. Control Solution Test
Appears when doing a control
result as a control solution
test.
4. Memory Symbol
Indicates in memory mode.
5. Units of Measurement
Unit of mmol/L or mg/dL will
appear with the test result.
About the EasyTouch® Meter Display
5
2
7
10
13
6
8
11
3
1
4
9
12

12 6. Battery Symbol
Appears when the battery is low or must be replaced.
7. Blood Drop Symbol
Flashes when it is ready to apply sample to Test Strip.
8. Warning Symbol
Appears when result is out of the range of your setting reference value.
9. Alarm Function Symbol
10. Test Strip Symbol
Appears to tell you when the meter is ready for test.
11. Day Average Symbol
Shows the average value of test results.
12. Test Result Area
Display test results.
13. Temperature Message
Appears if your meter is out of operating temperature range.

13
About the EasyTouch® Test Strip
The EasyTouch® Test Strip is used to perform
EasyTouch® Meter.
Each strip can be used only once.
The Test Strip consists of the following parts:
has been applied to the target area.
2. Target Area
This is where blood is drawn into the Test Strip.
3. Contact Bar
This end of the Test Strip is inserted into the Test
Port to activate the Meter. (white side face up).
3
1
2

14
WARNING: Re-cap immediately &
always close lid tightly.
Store at 39-86°F (4-30°C).
Do not refrigerate.
Always store test strips inside vial.
Discard Date: ______________
Discard 3 months after opening.
Made in Taiwan to Specications of
MHC Medical Products, LLC
8695 Seward Rd, Faireld, OH 45011
%)t3FWEBUF
For testing of glucose in capillary whole blood.
For in vitro diagnostic use.
EXP
50
TEST STRIPS
Glucose Test Strips For use with &BTZt5PVDI® Meter.
Easyt5PVDI
®
About the EasyTouch® Test Strip Vial
Control Solution Range
Lot Number
Expiration Date
VIAL LABELTEST STRIP VIAL
WARNING: Re-cap immediately &
always close lid tightly.
Store at 39-86°F (4-30°C).
Do not refrigerate.
Always store test strips inside vial.
Discard Date:
______________
Discard 3 months after opening.
Made in Taiwan to Specications of
MHC Medical Products, LLC
8695 Seward Rd, Faireld, OH 45011
%)t3FWEBUF
For testing of glucose in capillary whole blood.
For in vitro diagnostic use.
EXP
50 TEST STRIPS
Glucose Test Strips For use with &BTZt5PVDI® Meter.
Easyt5PVDI®

15
is shown, the Meter will provide accurate results for
Installing/Replacing the Battery
How to Replace the Battery
1 2 3
BLOOD GLUCOSEMETER
EasytTouch
®
ForIn Vitro Diagnostic Use Only
Made in Taiwanto Specs of
MHC Medical Products, LLC
8695 Seward Road
Faireld, OH 45011
Toll#: 877.358.4342
%)t3FWEBUF
BLOOD GLUCOSEMETER
EasytTouch
®
ForIn Vitro Diagnostic Use Only
Made in Taiwanto Specs of
MHC Medical Products, LLC
8695 Seward Road
Faireld, OH 45011
Toll#: 877.358.4342
%)t3FWEBUF
BLOOD GLUCOSEMETER
EasytTouch
®
ForIn Vitro Diagnostic Use Only
Made in Taiwanto Specs of
MHC Medical Products, LLC
8695 Seward Road
Faireld, OH 45011
Toll#: 877.358.4342
%)t3FWEBUF
BLOOD GLUCOSE METER
EasytTouch
®
For In Vitro Diagnostic Use Only
Made in Taiwan to Specs of
MHC Medical Products, LLC
8695 Seward Road
Faireld, OH 45011
Toll #: 877.358.4342
%)t3FWEBUF

16 Setting Up the Meter
Setting Up the Date & Time
1. Press the M button to enter this mode.
Press and release S
mode, press M button and the year setting
will appear. Press and release the S button
until the correct year appears. (Fig.1)
2. To set the year and move on to choose the
month, press the M button. To change the
month, press the S button. (Fig.2)
3. To set the month and move on to choose the
day, press the M button. To change the day,
press the S button. (Fig.3)
(Fig.1)
(Fig.2)
(Fig.3)

17
4. To set the day and move on to choose the hour setting,
press the M button. (Fig.3)
5. To change the hour, press the S button. To set the hour
and move on to change the minutes, press the M button.
(Fig.4)
6. To change the minutes, press the S button. To set the
minutes and move on to choose the alarm setting, press
the M button. (Fig.4)
M button (yes). If you
press S button (no), it will be moved on to health
management alarm setting (step 11). To change the hour,
press S button. To enter the hour and move on to change
the minutes, press M button. To change the minutes,
press S button. To enter the minutes and move on to
choose the second alarm setting, press the M button.
(Fig.5)
Setting Up Alarm
(Fig.4)
(Fig.5)

18 8. To set the second alarm, press the M button (yes).
If you press S button (no), it will be moved on to health
management alarm setting (step 11). To change the hour,
press S button. To set the hour and move on to change the
minutes, press M button. To change the minutes, press S
button. To set the minutes and move on to choose the third
alarm setting, press the M button. (Fig.6)
9. To set the third alarm, press the M button (yes). If you press
S button (no), it will be moved on to health management
alarm setting (step 11). To change the hour, press S button.
To set the hour and move on to change the minutes, press
M button. To change the minutes, press S button. To set the
minutes and move on to choose the fourth alarm setting,
press the M button. (Fig.7)
(Fig.6)
(Fig.7)
Other manuals for Easy Touch
4
Table of contents
Other MHC Medical Products Blood Glucose Meter manuals

MHC Medical Products
MHC Medical Products Easy Touch User manual

MHC Medical Products
MHC Medical Products Easy Touch User manual

MHC Medical Products
MHC Medical Products EasyTouch BluLink User manual

MHC Medical Products
MHC Medical Products EasyTouch HealthPro User manual

MHC Medical Products
MHC Medical Products Easy Touch User manual

MHC Medical Products
MHC Medical Products Easy Touch User manual