1 Overview
The miPositon water-resistant pendant is small, light and easy to wear,
washable and provides advanced functionality including an
accelerometer to establish movement and an algorithm to maximize
battery life.
The pendant is available in the following configurations:
Note: For UL 2560 installations, refer to the miPosition Gateway
Installation Instructions
1.a miAlert Contact Information
If you have any problems with this procedure, contact miAlert technical
services:
• Email: [email protected]
•Toll Free: 855-382-8999
1.b Maximum Number of Repeaters for a UL 2560
Installation
To achieve the 99.99% alarm message reliability required for UL 2560
compliance, system installations must operate within the following limits for
end device and repeater counts.
1.c Pendant Components
Figure 1 Pendant components (water resistant pendant with watch
band shown)
2.a Installation Notes
• These products are designed to be maintained by certified
technicians.
•Products are intended for indoor use.
•Pendant is shipped with the coin cell battery installed. But it will be
in storage mode which will not affect battery life, to take Pendant
out of this mode please hold down the Activation button for 5
seconds.
•All products should be manually tested monthly. “Test the
Transmitter/Receiver section”.
Part # Check-In Location Wearable
miPos2 - W 60 minutes North America Watch/Neck
miPos2 - C 60 minutes North America
End Devices Maximum
Repeaters
128 256
256 512
512 1028
1028 1028
A Activation button B TX/RX alarm LED and Low Battery LED
C Battery Replace remove back cover 4 screws
2 Installation and Startup
2.b Coin Cell Battery Installation / Replacement
Turn the Pendant with the back facing up and unscrew the back cover,
place the battery in the battery compartment, ensuring that the positive
terminal (+) faces up.
Watch/Neck
www.mialert.com
Toll Free: 855-382-8999
Email: [email protected]
27 Armthorpe Rd, Brampton, L6T 5M4
A
B
C
2.c Register the Pendant
The Pendant sends a check-in message to the Repeater via the Gateway
every 60 minutes for the purposes of supervision. The Pendant must be
registered and supervised, the Pendant should be tested after registration
to ensure operation. The Repeater will send the Pendant an
aknowledgement signal to confirm check-in supervion has been received.
Caution: The Pendant must have a supervision window of no more
than 12 hours to avoid not discovering a missing device within 24 hours.
2.d Wear the Pendant
The pendant can be worn in three ways:
• Around the neck with the neck lanyard, which is included with the
water resistant pendant with neck lanyard.
•On the arm with the wristband, which comes attached on the
water resistant pendant with wristband.
•On the belt, using the optional belt loop attachment kit.
Model Name: mi Pos 2Pendant
FCC I D: 2AZU7- MI POS2PENDNT
miAlert 2021 - www.miAlert.com 2
3 Operate the Pendant TX/RX
3.a Send an Alarm
To send an alarm: Press the activation button for at least one second.
• When activated, the pendant LED will be red to indicate
Alarm being sent
• The pendant will receive an acknowledgment form Repeater and
1. Place the alarm clearance key-chain tag over the activation button, the
red transmission LED will cease flashing and the blue alarm clear LED
will flash to indicate the alarm has cleared.
2. The alarm can be cleared remotely by sending a Clear Alarm signal
from server to Repeater via Gateway
the LED will flash blue for 0.5 seconds.
Note: If the pendant does not receive an acknowledgment from the
Repeater it will keep trying.
3.b Clear an Alarm
There are two ways to clear an alarm:
Using the Alarm Clearance Tag
Note: The Pendant alarm clearance key-chain tag is an optional
accessory sold separately.
4.c Take the Device out of Storage Mode
To remove the device from storage mode:
1. Press the Pendat alarm activation button twice.
2. Press the activation button for at least one second the second time
• The red LED will turn on and the pendant to indicate that an alarm has
been sent. This serves as confirmation that the pendant is no longer in
storage mode.
•When the button is released, an LED will flash briefly Blue to indicate
the aknowledgement from the Repeater was received.
5 Cleaning and Care
5.a Cleaning
Cleaning should be performed by hand using a damp cloth and mild soap,
or disinfectant wipes designed for household use.
Do Not Use
The following products should never be used to clean the Pendants:
• Strong cleaning agents such as ammonia, bleach, alcohol or
quaternary disinfectant.
• Abrasive or powder cleansers.
• Alcohol-based hand sanitizers.
The following equipment and processes must never be used:
•Dishwasher or washing machine.
•Soaking or suspending the pendant in water or other liquids.
3.c Coin Cell Low Battery Alert
When a low battery is detected, the Pendant transmitter will send a
low battery alert flag with the check-in supervision, and start a thirty
day countdown. The server will report the low battery condition and
escalate the alarm to the appropriate party via alarm alert or email on
the low battery report.
4 Storage Mode
Storage mode is an ultra low battery state designed to protect the life of
the internal battery for 12 months. The pendant is shipped from miAlert in
storage mode. TThe device will also automatically go into storage mode if
in an alarm condition or if check-in transmission as not been
aknowledged for 12 hours.
Note: When the pendant is in storage mode, it will not transmit alarms or
check-in messages.
4.a Extended Storage
If a pendant will not be deployed to an end user for more than 12 months
from the date of manufacture that appears on the device or device
packaging, you must install the coin cell battery and press the button before
the 12 months elapses to ensure proper operation.
If the device has remained idle for more than 6 months, we advise a 72
hour charging period before deploying the pendant to an end user.
If the device will remain unused after installing the coin cell battery, it will
operate for approximately three years, assuming no button presses.
Wearable Recommendations
The pendant transmitter housing is water resistant up to the IP65 standard
when the back-plate is properly seated and secured by the 4 screws. The
Pendant is designed for incidental/momentary submersion in water, but
not for extended submersion. As such, we offer the following
recommendations to the end user:
* The pendant is able to successfully transmit an alarm or check-in
message when exposed intermittently to water in a shower setting.
However, when the pendant is fully submerged, the water absorbs the radio
signal such that messages may not be reliably detected by the receiver.
Do not spray water directly on the
Pendant at distances of less than
12 inches; avoid spraying water at
extremely high pressure.
Usage Situation OK to Use? Cautions
Shower Yes*
Bath Yes* Momentary submersion is
permissible at water temperatures
less than 107°F. Extended
submersion of the Pendant is not
recommended.
Rain Yes* Avoid extended exposure to rain;
wear the Pendant under a sleeve,
shirt, or jacket.
Water exercise,
swimming pool,
hot tubs steam
rooms and sauna
No Do not use the Pendant during
these activities. Please note that
the IP65 rating does not test for
exposure to liquids with a chemical
content such as chlorine or salt
water.
Water Exposure
It has been evaluated that the Pendant meets the water spray and
submersion tests required
under UL 2560, section 38.
In addition, these pendants meet IP65 certification standards for dust and
water ingress protection, ensuring that:
• Ingress of dust particles does not interfere with the satisfactory
operation of the device.
•Protected against low pressure jets if water from all directions -
limited ingress permitted
miAlert 2021 - www.miAlert.com 3
6 Device LED Indicators
LED Pattern Red Blue Status
1Blinks for 20ms
duty is set by
server
Heart beat check-in
message for location
status
OFF
2Blinks for 1s
when Pendant
button is pressed
OFF Pendant sending
Alarm message to
Repeater
3Blinks for 20ms
after Pendant
button is pressed
Pendant has
received check-in
aknowledgement
from Repeater
OFF
4OFF Blinks for 50ms
after Pendant
button is pressed
Pendant has
received Alarm
aknowledgement
from Repeater
5Blinks for 2ms
every 3sec when
Pendant in Alarm
mode
OFF Pendant is in
Alarm mode and
waiting for staff to
cancel at source
7
6Blinks for 2ms
duty cycle set by
server
Pendant sends
Repeater Position
Wake status
Blinks for 2ms
duty cycle set by
server
Pendant enters
storage mode and
is in ultra low
battery mode
Pendant enters storage mode it
will flash Red 1sec - Blue 1sec -
Red 1sec
7 Specifications
Dimensions: 36mm x 41mm x 12mm.
Weight: 16g with battery inserted.
Operating temperature: 32 to 140°F (0 to 60°C).
Operating environment: water resistant to IP65 standard.
Estimated battery life: One year (assuming five alarm activations per day
and alarm clearance within ten minutes of alarm activation).
Battery type: Panasonic CR2450.
Power requirement: 3VDC, 15µA.
Storage requirement: Ambient conditions of 69 to 73°F (20.5 to 22.8°C).
UL certification for UL 2560
FCC RF Radiation Exposure Statement
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. End users must follow the specific operating
instructions for satisfying RF exposure compliance. This transmitter must
not be co-located or operating in conjunction with any other antenna or
transmitter except in accordance with FCC multi-transmitter product
procedures.
FCC Part 15 and Industry Canada Compliance
This device complies with part 15 of the FCC Rules and Industry Canada
license-exempt RSS standard(s). Operation is subject to the following two
conditions: (1) this device may not cause interference, and (2) this device
must accept any interference, including interference that may cause
undesired operation of the device. Changes or modifications not expressly
approved by the party responsible for compliance could void the user's
authority to operate the equipment.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables
aux appareils radio exempts de licence. L'exploitation est autorisée aux
deux conditions suivantes : (1) l'appareil ne doit pas produire de brouillage,
et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique
subi, même si le brouillage est susceptible d'en compromettre le
fonctionnement.
Battery Handling
If you are installing CR2450 lithium metal coin cell batteries in multiple
devices, keep the batteries from coming into contact with each other to
prevent discharge.
Pendant Shipping
Both lithium-ion and lithium-metal battery types are subject to
transportation regulations such as International Air Transport Association
(IATA), International Maritime Dangerous Goods (IMDG), US Department
of Transportation (DOT) and UN Recommendation on the Transport of
Dangerous Goods.
To prevent accidental alarm activation due to air pressure changes,
miAlert recommends that you avoid shipping the devices by air with the
CR2450 lithium-metal coin cell battery installed in the device.
When disposing of this electronic device or depleted CR2450 lithium-metal
coin cell batteries, please do so in accordance with federal, state and local
regulations.
Note: Changes or modifications not expressly approved by the party
responsible for compliance could void the user's authority to operate the
equipment.
7.a Medical Device Interference
Under FDA and FCC guidelines, the responsibility for verifying continuous
and safe operation of medical devices such as pacemakers and implanted
defibrillators in the presence of electromagnetic interference (EMI) rests
with the manufacturer of the medical device. As such, miAlert does not
have the authority or specific device knowledge to conduct or interpret
formal tests on their behalf.
miAlert transmitter devices comply with Part 15 of the FCC Rules and
Industry Canada license-exempt RSS standards.
If you have any concerns about the use of miAlert transmitter devices in
the presence of medical devices used by a particular resident or patient, we
recommend that you consult with their physician. Another potential source
for answers is the medical device manufacturer, who can provide more
information as to their compliance with federal guidelines and how they
have addressed EMI risk.
7.b Radiation Exposure Limits
This equipment complies with ISED RSS-102 radiation exposure limits set
forth for an uncontrolled environment. End users must follow the specific
operating instructions for satisfying RF exposure compliance. This
transmitter must not be co-located or operating in conjunction with any
other antenna or transmitter.
Cet équipement est conforme avec ISED RSS-102 des limites d'exposition
aux rayonnements définies pour un environnement non contrôlé. Les
utilisateurs finaux doivent suivre le fonctionnement spécifiqueinstructions
pour satisfaire la conformité à l'exposition RF. Cet émetteur doitne pas être
colocalisé ou fonctionner conjointement avec une autre antenne ou
émetteur.
Note: IP65 certification was performed by an independent laboratory not
affiliated with UL.
Note: Users that have achieved certification and will install UL 2560
certified systems are responsible for labeling all fundamental devices with
the UL 2560 system certification mark. Use compatible Repeater and
appropriate Gateway for UL 2560 installations.
Note: miAlert supports recycling and reuse whenever possible. Please
recycle these parts using a certified electronics recycler.
Note: For UL 2560 installations, miAlert repeaters must have 20 minute
check-in times. miAlert transmitters must have a minimum of 60 minute
check-in times.
Note: In a UL 2560 installation the miAlert Pendants must be used to
complete emergency call systems for assisted living and independent
living facilities.
For UL 2560 certified system installations, the following miAlert devices
are approved for installation within maximum system configuration limits
defined in section 1.b of this document:

This manual suits for next models

2

Popular Telephone Accessories manuals by other brands

PreCise Biometrics Tactivo User reference guide

PreCise Biometrics

PreCise Biometrics Tactivo User reference guide

jotron Phontech BTS 4000 Documentation handbook

jotron

jotron Phontech BTS 4000 Documentation handbook

Ferrari electronic OM EyeSDN USB 1-BRI Quick start manual

Ferrari electronic

Ferrari electronic OM EyeSDN USB 1-BRI Quick start manual

Digitone Call Blocker owner's manual

Digitone

Digitone Call Blocker owner's manual

Aire Mate100 G user manual

Aire

Aire Mate100 G user manual

AlzaPower AGD-WCP Series user manual

AlzaPower

AlzaPower AGD-WCP Series user manual

Viking Teleguard PC-7 Technical practice

Viking

Viking Teleguard PC-7 Technical practice

i-SERV 8630 Operation manual

i-SERV

i-SERV 8630 Operation manual

Viking LM-1A Technical practice

Viking

Viking LM-1A Technical practice

Easydive Leo3 instructions

Easydive

Easydive Leo3 instructions

Honeywell TELEPHONE ACCESS MODULE W8735D Features

Honeywell

Honeywell TELEPHONE ACCESS MODULE W8735D Features

Hello Direct SUPER PRO AMPLIFIER user guide

Hello Direct

Hello Direct SUPER PRO AMPLIFIER user guide

Panasonic KX-TA308 Manual de instalación

Panasonic

Panasonic KX-TA308 Manual de instalación

Inovonics EN1221S-60 Installation and operation manual

Inovonics

Inovonics EN1221S-60 Installation and operation manual

DAC DA-113/HF user guide

DAC

DAC DA-113/HF user guide

Plantronics HL2 manual

Plantronics

Plantronics HL2 manual

Coby DP151 Series instruction manual

Coby

Coby DP151 Series instruction manual

AT&T Call Display 435 user manual

AT&T

AT&T Call Display 435 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.