Before you begin:
Condition your MicroDirect pH30 electrodes by immersing it in electrode storage solution or tap
water or at least 30 minutes before use. DO NOT use de-ionized water.
pH Buffer Set Selection
Your tester features USA (pH 4.01, pH 7.00 and pH 10.01) or
NIST (pH 4.01, pH 6.86, and pH 9.18) standards. Select either one to suit your requirements.
1. While pressing the HOLD/ENT button, switch on the tester by pressing the ON/OFF button.
2. Release the HOLD/ENT button. The display will flash either USA or nSt.
3. Press CAL button to toggle between the two buffer set standards.
4. Press the HOLD/ENT button to confirm the selection of the buffer set.
pH Calibration
Calibration should be done regularly, preferably once a week. You can calibrate up to three
points using either the USA or the nSt buffer set standards.
1. Press ON/OFF button to switch unit on.
2. Dip electrode about 2 to 3 cm into the pH standard buffer solution.
3. Press the CAL button to enter calibration mode. The ‘CAL’ indicator will be shown. The upper
display will show the measured reading based on the last calibration while the lower display will
indicate the pH standard buffer solution.
Note: All testers have dual display during calibration mode
Note: To abort calibration, press the ‘CAL’ button.
4. Allow about 2 minutes for the tester reading to stabilize before pressing the HOLD/ENT button
to confirm the first calibration point. The upper display will be calibrated to the pH standard
buffer solution and the lower display will then be toggling in between readings of the next pH
standard buffer solution.
5. Repeat with other buffers if necessary. Rinse electrode in tap water before dipping into next
buffer.
Note: The calibration mode allows you to perform up to three calibration points
before returning to the measurement mode automatically. However, if you opted to
have only one or two calibration points, simply skip the remaining calibration points
by exiting to the measurement mode by pressing the CAL button.
pH Measurement
1. Press the ON/OFF button to switch the tester on.
2. Dip the electrode about 2 to 3 cm into the test solution. Stir and let the reading stabilize.
3. Note the pH value or press HOLD/ENT button to freeze the reading. To release the reading,
press HOLD/ENT again.
4. Press ON/OFF to turn off tester. If you do not press a button for 8.5 minutes, the tester will
automatically shut off to conserve batteries.
HOLD Function
This feature lets you freeze the display for a delayed observation
1. Press HOLD/ENT button to freeze the measurement. A ‘HOLD’ indicator will be displayed and
the measurement will be frozen.
2. Press HOLD/ENT again to release the measurement. The ‘HOLD’ indicator will not be displayed
anymore indicating the held measurement is released.
User Reset
You can reset the pH calibration to the factory default by using the user reset function.
Buffer set selection and temperature user calibration (pHTester30) are not affected by the user
reset function.
1. Switch off the tester.
2. While pressing the ‘CAL’ button, press and release the ON/OFF button to enter the ‘User
Reset’ selection menu. The screen will display ‘rSt’ on the bottom display with a flashing ‘nO’
selection.
3. Use the ‘CAL’ button to toggle between ‘nO’ and ‘YES’ selection.
• nO deactivates reset selection
• YES activates the reset selection
4. Press the HOLD/ENT button to confirm the selection made.
5. If you have selected ‘YES’, the unit will show ‘CO’ momentarily and proceed to the measure-
ment mode with the calibration reset back to factory default value.
6. If ‘nO’ is selected, the unit will proceed to the measurement mode without any calibration
reset performed.
Automatic Temperature Compensation (ATC)
Through its inbuilt temperature sensor, the measurement error due to the changes in electrode
sensitivity due to changes in temperature is compensated to give the actual pH reading of the
sample measured.
Temperature Calibration
From the measurement mode,
1. Press the HOLD/ENT button to bring the tester to the ‘HOLD’ mode.
2. Press the CAL button for 3 seconds to switch to the °C or °F mode setting selection screen.
Pressing the CAL button continuously for 3 seconds allows you to toggle in between the °C and
°F mode setting selection screen.
3. Release the CAL button to confirm your mode selection and the display will go to the tem-
perature calibration mode with the upper display flashing. The upper display shows the current
measured temperature reading based on the last set offset and the lower display shows the
current measured temperature reading based on factory default calibration.
4. Dip the tester into a solution of known temperature and allow time for the in built tempe-
rature sensor to stabilize.
5. Press the HOLD/ENT button to set the upper display to the temperature value of the solu-
tion.
6. Once the new temperature setting is reached, the new value is automatically confirmed and
returns to the measurement mode if no button is pressed after 5 seconds.
Notes: To exit this program without confirming the calibration, press the CAL button before the
automatic confirmation takes place.
GB
• Large Screen
• Waterproof
pH / Tempe-
rature Tester
• Double Junc-
tion
Instruction
manual
Fahrenheit
pH30
Electrode Maintenance
1. Rinse the electrode with tap water or electrode storage solution after each measurement.
Care has to be taken not to damage the sensor’s glass electrode especially while rinsing the
pHSpear penetrating tip electrode.
2. In aggressive chemicals, dirty or viscous solutions, and solutions with heavy metals or proteins,
take readings quickly and rinse electrode immediately afterward.
3. If possible, keep a small piece of paper or sponge in the electrode cap – moistened with
clean water or electrode storage solution (NOT de-ionized water) – and close the cap over the
electrode.
Changing Batteries
1. Open battery compartment lid (with attached lanyard loop).
2. Remove old batteries; replace with fresh ones. Note polarity
Self-Diagnostic Messages
Electrode Replacement
You can replace the electrode module at the fraction of the cost of a new tester.
When the tester fails to calibrate or gives fluctuating readings in calibration standards, you need
to change the electrode.
1. With dry hands, grip the ribbed tester collar with electrode facing you. Twist the collar counter
clockwise (see picture A). Save the ribbed tester collar and O-ring for later use.
2. Pull the old electrode module away from the tester.
3. Align the four tabs on the new module so that they match the four slots on the tester (see
picture B).
4. Gently push the module onto the slots to sit it in position. Push the smaller O-ring fully onto
the new electrode module. Push the collar over the module and thread it into place by firmly
twisting clockwise.
Note: It is necessary that you recalibrate your tester prior to measurement after an electrode
replacement.
Applications
Water quality testing • pools • spas • aquariums • aquaculture • hydroponics • ecology studies
• water and wastewater treatment • boilers • steam generators • car washes • sanitation plants
• labs • food sectors and more!
Warranty
The waterproof MicroDirect WP 30 are warranted to be free from manufacturing defects for
2 years and electrode module for 6 months. If repair, adjustment or replacement is necessary
and has not been the result of abuse or misuse within the time period, please return the tester
– freight prepaid – and correction will be made without charge. Out of warranty products will
be repaired on a charge basis.
Return of Items
Authorization must be obtained from your distributor before returning items for any reason.
When applying for authorization, please include information regarding the reason the item(s)
are to be returned.
Note: We reserve the right to make improvements in design, construction and appearance
of products without notice. Prices are subject to change without notice.
Accessories:
19 46 36 Replacement sensor MicroDirect pH30
Technical Specifications pH30
Range -1.00 .....15.00 pH
Resolution 0.01 pH
Accuracy 0.01 pH
Calibration Points 1-, 2-, 3-point
Buffer Solutions USA 4.01/7.00/10.01
NIST 4.01/6.86/9.18
Calibration Window (USA Buffer Set Standard) +/-1.00 pH (pH 4.01 & pH 10.01),
+/-1.50 pH (pH 7.00)
Calibration Window (NIST Buffer Set Standard) +/-1.00 pH (pH 4.01 & pH 9.18)
+/-1.25 pH (pH 6.86)
Temperature Range 0-50.0°C or 32.0-122.0°F
Automatic Temperature Compensation (ATC) Yes
Temperature Resolution 0.1 °C / °F
Temperature Accuracy 0.5 °C / 0.9 °F
Temperature Calibration Window +/- (5°C /9 °F) from default window
Auto Off 8.5 minutes
Batteries 4 x 1.5V “A 76 ” Micro Alkalie Batteries
Battery Life > 500 hours
Operation Temperature 0 – 50 °C
Dimensions (Tester) 165 x 38 mm
Weight 0,090 kg

Popular Test Equipment manuals by other brands

Testboy TB 40 Simple operating instructions

Testboy

Testboy TB 40 Simple operating instructions

Hioki MEMORY HiCORDER 8855 instruction manual

Hioki

Hioki MEMORY HiCORDER 8855 instruction manual

STS 1660 Operator's manual

STS

STS 1660 Operator's manual

Koehler K185 Series Operation and instruction manual

Koehler

Koehler K185 Series Operation and instruction manual

Keysight ixia UHD100T32 Getting started guide

Keysight

Keysight ixia UHD100T32 Getting started guide

Klein Tools ET45 instructions

Klein Tools

Klein Tools ET45 instructions

Anritsu ME7840A Operation and maintenance manual

Anritsu

Anritsu ME7840A Operation and maintenance manual

Spin OK CLIMA instruction manual

Spin

Spin OK CLIMA instruction manual

Agilent Technologies 54831A User's quick start guide

Agilent Technologies

Agilent Technologies 54831A User's quick start guide

ELIT 1560 user guide

ELIT

ELIT 1560 user guide

Burster DIGISTANT 4420-V001 Operation manual

Burster

Burster DIGISTANT 4420-V001 Operation manual

Jeulin JO 20 manual

Jeulin

Jeulin JO 20 manual

Anritsu MG3641A Operation manual

Anritsu

Anritsu MG3641A Operation manual

Mecmesin Vortex operating instructions

Mecmesin

Mecmesin Vortex operating instructions

ACD MiniBump instruction manual

ACD

ACD MiniBump instruction manual

HT PVCHECK manual

HT

HT PVCHECK manual

Amprobe BT-120 user manual

Amprobe

Amprobe BT-120 user manual

Tektronix MDO3000 Series instructions

Tektronix

Tektronix MDO3000 Series instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.