recommended for use with this instrument.
The CM500 is fuse protected to 440V 10kA. The maximum
current which could flow through this fuse in the case of a fault is
limited to 10 kA by the impedance of the test leads. There is a
fuse accessible in the battery compartment which protects the
Insulation and Continuity range. If warning of a ruptured fuse is
given during a test, and if the warning symbol is present when
attempting to carry out an insulation test with the test leads
shorted together, this fuse must be replaced by another fuse of
the same type and rating i.e. 500 mA (F) 500 V H.B.C. 10 kA.
Fuse Replacement
To replace the fuse, disconnect the test leads,switch the
instrument Off and with a screwdriver, loosen the captive screw
holding the battery compartment cover in place. Remove the
cover. Depress the fuseholder slightly, and turn a quarter turn
anti-clockwise to release. Remove the holder and replace the
fuse with one of the correct size and rating. Relocate the fuse
holder,depress and turn a quarter turn clockwise. Replace and
re-secure the cover.
Battery Replacement
When the low battery symbol appears, the cells are nearly
exhausted and should be replaced as soon as possible. When
the battery is exhausted, the instrument will not perform tests
and the cells must be replaced. Use Alkaline cells IEC LR6 (AA)
or 1,5 V nickel cadmium cells only.
To install or replace the cells, disconnect the test leads,switch
the instrument Offand loosen the captive screw holding the
battery compartment cover in place. Remove the cover,lift out
and disconnect the battery holder to access the cells. Ensure
that the replacement cells are fitted with the correct polarity,in
accordance with the symbols on the battery holder moulding.
Carefully re-connect the battery holder to the plug, replace the
battery holder in the compartment, and re-secure the cover.
Remove the cells if the instrument is not going to be used for any
extended period of time. Stored results are retained when the
battery is disconnected.
Test Leads
All test leads form part of the measuring circuit of the instrument
and must not be modified or changed in any way,or be used with
any other electrical instrument or appliance. The power cord
supplied with the CM500 is a test lead that forms part of the
measuring circuit of the instrument. The overall length of this lead
must not be altered. If the power cord plug is not suitable for your
type of socket outlets, do not use an adaptor. You may change
the plug once only by cutting the cord as close to the plug as
possible and fitting a suitable plug.
The colour code of the cord is:
Earth (Ground) Yellow/Green
Neutral Blue
Phase (Line) Brown
Note: A plug severed from the power cord must be destroyed, as
aplug with bare conductors is hazardous in a live socket outlet.
7
Incorrect battery cell polarity
can cause electrolyte leakage,
resulting in damage to the
instrument.