MicroAire 7800 User manual

REF 7800 Battery Sternum Saw
Instructions for Use

IM-7800 Rev G 2021-05 2 of 16
TABLE OF CONTENTS
Model REF 7800 Battery Sternum Saw
REF 7800-002 Sternum Guard & Accessories
Instruction Manual
Intended Use and Introduction ................................................................3
General Warnings.......................................................................................3-4
Markings........................................................................................................5-6
List of Compatible Accessories..................................................................7
Environmental Parameters .........................................................................7
Technical Data and Handpiece Safety Mechanisms......................7-8
Handpiece Features...................................................................................8-9
System Setup, Assembly, and Operation........................................ 9-10
Troubleshooting .......................................................................................... 10
Instrument Cleaning Instructions ......................................................... 11
Sterilization Parameters ................................................................. 11
Warranty, Service and Repair .................................................................. 12
Battery Information.................................................................................... 13
Electromagnetic Compatibility ........................................................14-15

IM-7800 Rev G 2021-05
3 of 16
APPLICABLE PARTS
REF 7800
INTENDED USE
Multi-purpose, modular, powered instrument system intended for use in general, orthopedic, plastic and
trauma surgery.
INTRODUCTION
This manual has been written to help describe the procedures required to keep the MicroAire REF 7800 Battery
Sternum Saw system operating properly. Throughout the manual, the following terms are used to identify tips and
precautions that will help avoid accidental injury to patients or personnel, or prevent damage to the system.
NOTE: Used to point out the easiest means of carrying out techniques.
WARNING: Used to indicate that the safety of the patient and hospital personnel could be involved.
CAUTION: Used to point out special procedures or precautions that must be followed to avoid damaging the
system/instruments.
GENERAL WARNINGS
WARNING: The REF 7800 Battery Sternum Saw must never be used without the REF 7800-002 sternum guard
securely in place and properly oriented with the MicroAire REF ZR-032M or ZR-033M sternum saw
blade.
WARNING: Replace battery pack only with a MicroAire REF 6640-710, REF 7505-710, or REF 7500-620 battery.
WARNING: Explosion Hazard. Not suitable for use in the presence of flammable anesthetics or oxygen.
WARNING: Electric Shock. Do not remove cover, return to factory for servicing.
WARNING: Medical electrical equipment may be aected by electromagnetic interference. It should be installed
and used in accordance with the electromagnetic compatibility information provided herein.
WARNING: Portable and mobile RF communications equipment can aect medical electrical equipment.
WARNING: Prior to use, system components should be inspected and operated to detect any damage or
malfunction. Do not use if damage is apparent.
WARNING: Always point the handpiece away from the patient and surgical team when testing.
WARNING: If the handpiece will not securely hold the saw blade or sternum guard, discontinue use and return
the instrument to the factory for service.
WARNING: Check Safety Lock and Trigger before operating the handpiece. If the trigger should stick in the
depressed “ON” position, do not use the handpiece. Immediately remove the battery. Return the
instrument to the factory for service. Always use the Safety Lock when the instrument is not running.
WARNING: Check that the battery is fully seated and locked.
WARNING: Disconnect battery in the event of continuous running.
WARNING: No modication of this equipment is allowed.
WARNING: Operating Duty Cycle is 1 minute on, 2 hours off.
WARNING: Irrigation must be applied to the cutting accessory when cutting bone in order to keep bone
temperature within safe limits. Cutting without irrigation may cause unsafe temperatures leading to
bone necrosis.
WARNING: Use of the REF 7800 Sternum Saw adjacent to or stacked with other equipment should be avoided be
cause it could result in improper operation. If such use is necessary, the REF 7800 Sternum Saw and
the adjacent equipment should be observed to verify that they are operating normally.

IM-7800 Rev G 2021-05 4 of 16
WARNING: Portable RF communications equipment should be used no closer than 30 cm (12 inches) away
from any part of the REF 7800 Sternum Saw. Otherwise degradation of the performance of the
REF 7800 could result.
WARNING: Use of Battery Packs other than those specied in the List of Compatible Accessories could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and
result in improper operation.
WARNING: REF 7800 Sternum Saw has been tested for use in the vicinity of High Frequency Electrosurgical /
Electro coagulation equipment. Use of such equipment in close proximity to the REF 7800 Sternum
Saw may cause unintended interruptions or unintended motion of durations up to 1 second. Place
the Instrument on a safe surface when not in use. Do not place powered Instruments on the patient
when not in use. Avoid entwining HF surgical cables with powered instruments.
WARNING: The following items should be periodically inspected for signs of damage and repaired or replaced as
needed to ensure continued safety with regard to electromagnetic disturbances over the life of the
system:
• Check that the Battery Connection base, the rear end cap and the front of the instrument
remain securely connected to the main instrument body.
• Check for damage to the battery pack housings.
CAUTION: Do not run the handpiece without the saw blade.
CAUTION: Disconnect battery from handpiece before loading or unloading the saw blade.
CAUTION: Leaving a battery pack attached to a handpiece when not in use may accelerate battery drain.
CAUTION: Make sure the tip of the sternum saw blade is inside the foot of the sternum guard and does not
strike any part of the sternum guard. Improper alignment of the sternum guard with the saw blade
can cause damage to the sternum saw or saw blade.
CAUTION: If the handpiece will not hold the saw blade or sternum guard securely, discontinue use and return to
the factory for service.
CAUTION: Check Safety Lock and Trigger before operating the handpiece. If the trigger should stick in the
depressed “ON” position, do not use the handpiece. Immediately remove the battery. Return the
instrument to the factory for service. Always use the Safety Lock when the instrument is not running.
CAUTION: Check that the battery is fully seated and locked.
CAUTION: Federal Law (U.S.A.) restricts this device to sale by or on the order of a physician (or properly licensed
practitioner).
CAUTION: Disconnect sternum guard, battery, and saw blade prior to sterilization. Batteries must be sterilized
separate from the handpiece and sternum guard.
CAUTION: DO NOT run instruments while warm. Cool by exposure to room temperature.
CAUTION: DO NOT immerse in liquid to cool.
CAUTION: Repairs or alterations to MicroAire products made by anyone other than MicroAire or an Authorized
MicroAire Repair Facility will void that product’s warranty, and the customer will be responsible for
any costs related to returning the product to working condition
NOTE: Disconnect sternum guard, battery, and saw blade prior to sterilization. Batteries must be
sterilized separately.
NOTE: The REF 7800 Sternum Saw is suitable for use in hospitals and surgery centers. These systems should
not be used near Magnetic Resonance Imaging equipment.
NOTE: The emissions characteristics of this product make it suitable for use in industrial areas and hospitals
(CISPR 11 Class A). If it is used in a residential environment (for which CISPR 11 Class B is normally
required) this equipment might not offer adequate protection to radio frequency communication
services. The user might be required to take mitigation measures, such as relocating or re-orienting
the equipment.
NOTE: Mailing address is located on back cover.

IM-7800 Rev G 2021-05
5 of 16
MARKINGS
Name Ref#
(ISO
7000)2Symbol Description Use Standard
Refer to Instruction
Manual / Booklet
ISO-7010
M002
• Indicates a MANDATORY action for the user to consult the
Instructions For Use (IFU).
• Symbol must be blue, as shown.
IEC 60601-1:20051
Consult
Instructions For
Use (IFU)
1641
Indicates the need for the user to consult the Instructions For Use
(IFU). Not required in conjunction with the Caution
symbol, if applicable.
ISO 15223-1:20121
Caution 0434A /
0434B CIndicates the need for the user to consult the Instructions For Use
(IFU) for important cautionary information such as warnings and
precautions that cannot, for a variety of reasons, be presented on
the device itself.
ISO 15223-1:20121
UL symbol N/A
E494242
MEDICAL-GENERAL MEDICAL EQUIPMENT AS TO ELECTRIC SHOCK,
FIRE, AND MECHANICAL HAZARDS ONLY. IN ACCORDANCE WITH
ANSI/AAMI ES 60601-1 (2005) + A1 (2012) + CAN/CSA C22.2 No.
60601-1 (2014) | Control Number: E494242
UL
Type B Applied
Part 5840 BIndicates a medical device complying with the specied
requirements of IEC 60601-1 to provide protection against electric
shock, particularly regarding allowable patient leakage current and
patient auxiliary current.
IEC 60601-1:20051
Locked 5569 To identify on a control that a function is locked or to show the
locked status. IEC 60878:20151
Unlocked 5570 To identify on a control that a function is not locked or to show the
unlocked status. IEC 60878:20151
Do Not Expose to
Stray Magnetic
Fields
N/A Indicates a medical device that is not to be exposed to stray
magnetic elds. N/A
Authorized
Representative
in the European
Community
N/A
Indicates the authorized representative in the European
Community. This symbol shall be accompanied by the name and
address of the authorized representative, adjacent to the symbol.
ISO 15223-1:20121
Serial # 2498
• Indicates the manufacturer’s serial number so that a specic
medical device can be identied.
• Per EN980:2008, the SN symbol may be used without
surrounding box.
ISO 15223-1:20121
Lot / Batch Code 2492 Indicates the manufacturer’s batch code so that the batch or lot
can be identied. ISO 15223-1:20121
REF (Catalog #) 2493
• Indicates the manufacturer’s catalog number so that the medical
device can be identied.
• Per EN980:2008, the REF symbol may be used without
surrounding box.
ISO 15223-1:20121
Sterile 2499
Indicates a medical device that has been subjected to a
sterilization process. Use of the Sterile symbol requires a use-by date
(see Use-By Date symbol).
ISO 15223-1:20121
Sterilized using
Irradiation
(gamma)
2502 [Indicates a medical device that has been sterilized using irradiation
(gamma). Use of the Sterile symbol requires a use-by date
(see Use-By Date symbol).
ISO 15223-1:20121
Non-Sterile 2609
• Indicates a medical device that has not been subjected to a
sterilization process. This symbol should only be used to
distinguish between identical or similar medical devices sold
in both sterile and non-sterile conditions.
• Also indicates a medical device that is provided non-sterile
but must be sterilized prior to use.
ISO 15223-1:20121
Temperature
Limitation 0632 >Indicates the temperature limits to whichthe medical device can be
safely exposed. The upper and lower limits to temperature shall be
indicated adjacent to the upper and lower horizontal lines.
ISO 15223-1:20121

IM-7800 Rev G 2021-05 6 of 16
Use-By Date 2607 t
Indicates the date after which the medical device is not to be used.
This symbol shall be accompanied by a date to indicate that the
medical device should not be used after the end of the month
shown. The date is expressed as YYYY-MM (e.g. 2015-11) or
YYYY-MM-DD (e.g. 2015-11-29).
ISO 15223-1:20121
Do Not Reuse 1501 2Indicates a medical device that is not to be resterilized. ISO 15223-1:20121
Do Not Use if
Package is
Damaged
2606 wIndicates a medical device that should not be used if the package
has been damaged or opened. This symbol may also mean
“Do not use if the product sterile barrier system or its packaging is
compromised”.
ISO 15223-1:20121
Sternum Saw
Guard Alignment
Arrow
N/A Indicates alignment of Sternum Saw Guard. N/A
Humidity
Limitation 2620
Indicates the range of humidity to which the medical device can be
safely exposed. The humidity limitations shall be indicated
adjacent to the upper and lower horizontal lines.
ISO 15223-1:20121
Atmospheric
Pressure
Limitation
2621
Indicates the range of atmospheric pressure to which the medical
device can be safely exposed. The atmospheric pressure
limitations shall be indicated adjacent to the upper and lower
horizontal lines.
ISO 15223-1:20121
Do Not Immerse in
any Liquid 5995 Indicates a medical device that is not to be immersed in any liquid. IEC 60335-2-15
Do Not Lubricate N/A Indicates a medical device that is not to be lubricated. N/A
Date of
Manufacture 2497
• Indicates the date when the medical device was
manufactured. The date is expressed as YYYY-MM (e.g. 2015-11)
or YYYY-MM-DD (e.g. 2015-11-29).
• If the symbol is lled (see Manufacturer symbol), both
the date of manufacture and the name/address of the
manufacturer may be combined in one symbol.
ISO 15223-1:20121
Manufacturer 3082
• Indicates the medical device manufacturer. This symbol shall be
accompanied by the name and address of the manufacturer. The
date of manufacture may be combined with this symbol.
• When using MicroAire as the manufacturer, use the MicroAire
LLC symbol.
ISO 15223-1:20121
CE Mark with NB N/A
2797
Indicates the European Conformity Mark with Notied Body
Number. 2797 is the BSI-NL-registered Notied Body.
Council Directive
93/42/EEC
Prescription N/A Caution: Federal Law (U.S.A.) restricts this device to sale by or on
the order of a physician (or properly licensed practitioner).
FDA Title 21,
Chapter 1,
Subchapter H,
Part 801.15(F)
Dispose of per
WEEE Directive
2012/19/EU
N/A
Indicates a medical device that is not to be disposed of as
unsorted municipal waste. Medical device is to be disposed of per
WEEE Directive 2012/19/EU.
Council Directive
2012/19/EU
Dispose of per
WEEE Directive
2012/19/EU
N/A
Indicates a medical device that is not to be disposed of as
unsorted municipal waste. Medical device is to be disposed of per
WEEE Directive 2012/19/EU. This symbol is used in place of the
above symbol if the product entered the market after 13 August,
2005.
Council Directive
2012/19/EU
(Symbol: European
Standard EN 50419)
Packaging is
Recyclable 1135 Indicates that the marked item or its material is part of a recovery
or recycling process. IEC 60878:2015
1 ISO 15223-1:2012 – “Medical devices – Symbols to be used with medical device labels, labelling and information to be supplied – Part 1: General requirements”
2 ISO 7000/IEC 60417 – “Graphical symbols for use on equipment – Registered symbols”

IM-7800 Rev G 2021-05
7 of 16
LIST OF COMPATIBLE ACCESSORIES
Part Number Description IFU Type B Applied Parts
REF ZR-032M Sternum Saw Blade IM-7800 Yes
REF 7800-002 Sternum Saw Guard IM-7800 Yes
REF 6640-710 NiMH Battery Pack IM-PWRBATT No
REF 7505-710 Large NiMH Battery Pack IM-PWRBATT No
REF ZR-033M Aggressive Saw Blade IM-7800 Yes
REF 7500-700 NiMH Battery Charger IM-7500-700 No
REF 7500-615 Aseptic Battery Housing IM-ASEPBATT No
REF 7500-620 Aseptic NiMH Battery Pack IM-ASEPBATT No
REF 7500-625 Charging Adapter IM-ASEPBATT No
REF 7500-630 Aseptic Transfer Shield IM-ASEPBATT No
WARNING: Use of Battery Packs other than those specied in the List of Compatible Accessories could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and
result in improper operation.
ENVIRONMENTAL PARAMETERS (REF 7800, REF 7800-002)
OPERATING CONDITIONS
This device has been tested and proven to operate within the following conditions:
Temperature Humidity Atmospheric
59° F/15° C
80.6° F/27° C
>20%
60%
70 kPA
106 kPA
SHIPPING & STORAGE CONDITIONS
This device has been tested and proven to operate after repeated exposure to the following conditions:
Temperature Humidity Atmospheric
-0.4° F/-18° C
120° F/49° C
>0%
91%
70 kPA
106 kPA
Shipping: The materials and components used in the construction of this device were selected to ensure that the
device could be shipped by any standard commercial method without special handling conditions.
TECHNICAL DATA SPECIFICATIONS
Operating Speed: 12,000 CPM (nominal)
Cutting Stroke: 0.117 in. (3.0 mm)
REF 7800 Handpiece Weight
(without battery pack) :
2.65 lb. (1211 g)
REF 6640-710 Battery Weight: 0.95 lb. (431 g)
Battery Voltage: 14.4 volts
Duty Cycle: 1 minute on, 2 hours off.
Rating: 14.4 VDC 225W

IM-7800 Rev G 2021-05 8 of 16
HANDPIECE SAFETY MECHANISMS
Trigger in Upright - “ON” Position
Trigger in “OFF” Position
(Trigger can be turned in either direction
to be turned “OFF”)
REF 7800002 STERNUM GUARD
The REF 7800 Battery Sternum Saw requires the use of the included REF 7800-002. The guard can be attached to the
handpiece in either a superior or inferior position, depending on the position at which the blade was loaded.
INSTRUMENT COMPONENTS
Features & Descriptions
1. Blade Locking Collet
Insert the REF ZR-032M or ZR-033M Sternum Saw
Blade into the blade locking collet by turning the
collet in the direction indicated by the arrow, and then
inserting the blade.
2. Sternum Guard Lock / Release Button
Depress this button to lock or release the
REF 7800-002 Sternum Guard from the handpiece.
3. Trigger / Safety Lock
The REF 7800 Battery Sternum Saw operates at
variable speeds and is controlled by depressing the
trigger. The instrument is in the OFF position when the
trigger is not depressed. As the trigger is depressed
the speed increases from 0% - 100%. The trigger
also serves as the safety lock for the handpiece (see
previous page). After loading the saw blade and
attaching the sternum guard and battery, fully depress
the trigger and release one time before use.
4. Battery Release Button
Depress this button to remove a battery/power pack
from the handpiece. This button does not need to be
depressed to load a battery/power pack.
ACCESSORY ATTACHMENT
Battery Attachment
The REF 7800 Battery Sternum Saw must have a fully
charged MicroAire REF 6640-710 battery installed before
each use. The MicroAire REF 7505-710 and REF 7500-620
batteries may also be used with the REF 7800 Battery
Sternum Saw. To attach the battery to the handpiece,
place the battery under the base of the instrument with
the longer side facing the rear of the instrument. Firmly
push the battery into the base of the handpiece until the
battery latch clicks into place.

IM-7800 Rev G 2021-05
9 of 16
Sternum Guard Attachment
WARNING: DO NOT USE The REF 7800 Battery Sternum
Saw without the REF 7800-002 sternum
guard securely in place and properly
oriented with the MicroAire REF ZR-032M or
ZR-033M sternum saw blade.
INSTRUMENT SETUP
CAUTION: Disconnect battery from handpiece before loading or unloading the saw blade.
CAUTION: Battery pack should not be left attached to the handpiece when not in use. This could result in
accelerated battery drain.
To insert the REF ZR-032M or ZR-033M saw blade into the sternum saw handpiece:
1. Make sure the handpiece trigger is set to the “OFF” position and the safety is “ON” by turning the trigger to
one side.
2. If installed, remove the sternum guard from the handpiece by depressing the release button located on the
underside of the sternum guard collet.
3. Twist the blade locking collet into the open position by rotating in the direction of the arrows.
4. While holding the blade locking collet in the open position, insert a MicroAire REF ZR-032M or REF ZR-033M
saw blade until it is fully seated. The blade may be inserted in the inferior or superior position.
5. Release the blade locking collet; it should automatically spring back into the lock position.
To install the REF 7800-002 Sternum Guard on the sternum saw:
1. Make sure the handpiece trigger is set to the “OFF” position and the safety is “ON” by turning the trigger to
one side.
2. Make sure that the MicroAire REF ZR-032M or REF ZR-033M sternum saw blade is securely installed.
3. Push the sternum guard into the blade/guard collet, over the installed saw blade. The sternum guard may
be inserted more easily by depressing the sternum guard lock button as the sternum guard is installed on
the handpiece. Make sure that the sternum guard is aligned with the saw blade teeth facing away from the
sternum guard. The sternum guard must be completely locked into place prior to use. Use the alignment
arrows to make sure the sternum guard is properly aligned with the handpiece.
CAUTION: Make sure the tip of the sternum saw blade is inside the foot of the sternum guard and does not strike
any part of the sternum guard. Improper alignment of the sternum guard with the saw blade can cause
damage to the sternum saw or saw blade.
Test run the assembled handpiece for a few seconds to make sure that the blade and sternum guard are properly
aligned and secure.
1. Move the safety lock to the “OFF” position by turning the trigger back into the vertical position.
2. Run the handpiece by fully depressing the trigger.
WARNING: Always point the handpiece away from the patient and surgical team when testing.
WARNING: If the handpiece will not securely hold the saw blade or sternum guard, discontinue use and return
the instrument to the factory for service.

IM-7800 Rev G 2021-05 10 of 16
WARNING: Check Safety Lock and Trigger before operating the handpiece. If the trigger should stick in the
depressed “ON” position, do not use the handpiece. Immediately remove the battery. Return the
instrument to the factory for service. Always use the Safety Lock when the instrument is not running.
WARNING: Check that the battery is fully seated and locked.
WARNING: Disconnect battery in the event of continuous running.
TROUBLESHOOTING
1. Blade will not load into collet:
a. Make sure that the blade is the MicroAire REF ZR-032M or REF ZR-033M.
b. Mare sure the blade locking collet is fully turned in the direction indicated by the arrow on the collet.
c. Attempt to load a different REF ZR-032M or REF ZR-033M blade into the collet. If the second blade cannot be
loaded into the collet, return the handpiece for service.
2. Sternum Guard will not lock into handpiece:
a. Make sure that the sternum guard is the MicroAire REF 7800-002.
b. Make sure the arrows on the handpiece and guard are aligned properly, and the Guard Lock/Release Button is
depressed when attempting to insert the sternum guard.
c. Attempt to load a different REF 7800-002 sternum guard into the collet. If the second guard cannot be loaded
into the collet, return the handpiece for service.
3. Trigger cannot be depressed:
a. Make sure the trigger is turned to the upright “ON” position (refer to page 6).
b. Make sure the instrument was properly cleaned and sterilized.
c. If trigger cannot be depressed, return to factory for service.
d. DO NOT OIL the handpiece or trigger, as this could damage the instrument.
4. Handpiece does not run when trigger is depressed:
a. Make sure you have a fully charged MicroAire Battery Pack properly seated in the handpiece base.
b. Try a different, fully charged battery pack in the handpiece. If the handpiece still does not operate when the
trigger is depressed, return the handpiece for service.
5. Battery does not seat properly in handpiece:
a. Make sure you are using an approved MicroAire battery/power pack (part numbers listed on page 5).
b. Make sure the longer side of the battery pack is facing the rear of the instrument as shown on page 6.
c. Attempt to insert another battery pack. If second battery pack will seat properly, return the first battery for
inspection. If the second battery pack does not seat properly, return the handpiece and batteries for service.
d. Make sure the connector pins in the handpiece battery receptacle are not bent. If the pins are bent, return the
handpiece for service. DO NOT attempt to bend the pins back into place.
6. Handpiece runs slowly:
a. Make sure you have a fully charged MicroAire Battery Pack properly seated in the handpiece base.
b. Try a different, fully charged battery pack in the handpiece. If the handpiece still runs slowly when the trigger is
depressed, return the handpiece for service.

IM-7800 Rev G 2021-05
11 of 16
ROUTINE CLEANING/DECONTAMINATION
7800 INSTRUMENT CLEANING INSTRUCTIONS
1.Remove the battery, sternum guard, and sternum saw blade from the handpiece.
2.Scrub the handpiece, sternum guard, and batteries using warm water, a soft brush, and mild detergent. Scrub the
handpiece with the brush, paying close attention to the instrument crevices. Take care to keep the instrument as
upright as possible to restrict water from flowing into the battery cavity and contacts of the battery.
3. Rinse all items thoroughly under running water, again paying close attention to the battery cavity by keeping the
instrument as upright as possible. Dry the instrument and accessories with a lint-free towel.
INSTRUMENT STERILIZATION INSTRUCTIONS
[see separate Battery Sterilization Instructions in IM-PWRBATT]
MicroAire surgical instruments are normally sterilized using steam autoclaves by either gravity discharge, or pre-
vacuum sterilization. The following sterilization processes have been validated based on AAMI Protocols. The
exposure times and temperatures are the minimum validated requirements to ensure sterility.
CAUTION: Disconnect sternum guard, battery, and saw blade prior to sterilization. Batteries must be sterilized
separate from the handpiece and sternum guard.
HANDPIECE AND STERNUM GUARD WRAPPED OR UNWRAPPED
1. Steam Sterilization
Steam sterilize using one of the following cycles:
Sterilization Cycle Instrument Minimum Time & Temp Min Heated Dry time
Dynamic Air Removal
(Pre-vacuum)
Single Instrument
3 minute full cycle @
134 – 137°C (273 – 279°F) 8 minutes
4 minute full cycle @
132 – 135°C (270 – 275°F) 8 minutes
Sterilization Tray
3 minute full cycle @
134 – 137°C (273 – 279°F) 10 minutes
4 minute full cycle @
132 – 135°C (270 – 275°F) 10 minutes
Gravity Displacement
Single Instrument 30 minute full cycle @
132 – 135°C (270 – 275°F) 8 minutes
Sterilization Tray 45 minute full cycle @
132 – 135°C (270 – 275°F) 20 minutes
CAUTION: DO NOT run instruments while warm. Cool by exposure to room temperature.
CAUTION: DO NOT immerse in liquid to cool.
2.Flash sterilization
Flash sterilization is not recommended for powered surgical instruments. The REF 7800 Sternum Saw has not been
validated for ash sterilization.
3.Ethylene Oxide Sterilization.
Ethylene oxide is NOT recommended for powered surgical instruments because lengthy aeration time is needed to
assure that no ethylene oxide is left in the internal mechanisms or on the surface of the instrument.
4.Peracetic Acid
DO NOT process powered surgical instruments in equipment that uses peracetic acid as a liquid sterilant.
Disposal
Follow federal, state and local ordinances governing the proper disposal of medical equipment.

IM-7800 Rev G 2021-05 12 of 16
WARRANTY, SERVICE AND REPAIR
Periodic inspection and service is essential to keep MicroAire instruments running properly.
IN HOSPITAL SERVICE
All MicroAire equipment should be inspected and tested periodically in accordance with the facility’s
bioengineering policy.
CAUTION: Repairs or alterations to MicroAire products made by anyone other than MicroAire or an Authorized
MicroAire Repair Facility will void that product’s warranty, and the customer will be responsible for
any costs related to returning the product to working condition
MICROAIRE REPAIR SERVICE
If a problem with your equipment should arise, contact MicroAire Customer Service Department:
Telephone: Fax: Email:
USA: 800-722-0822 800-438-6309 [email protected]
Outside USA: 434-975-8000 434-975-4134 intlsvc@microaire.com
NOTE: Mailing address is located on back cover.
PERIODIC INSPECTION
Because of the stressful nature of surgical use, decontamination, and sterilization, we recommend that all
instruments be returned for routine inspection and service at least once a year. There is no charge for service during
the warranty period.
7800 BATTERY STERNUM SAW & ACCESSORY WARRANTY
MicroAire Surgical Instruments, LLC warrants its REF 7800 Battery Sternum Saw and the REF 7800-002 Sternum
Guard to be free from defects in material and workmanship for a period of 1 year from the original purchase date by
the end customer. The warranty is limited to the repair or replacement of the product without charge.
This warranty is void in the event of abuse, misuse, or use in other than normal surgical environment, or in the event
disassembly, alteration, or repair of the product not authorized by the manufacturer, or in the event that the product
has not been used in compliance with the written instructions furnished by the Manufacturer.
All other expressed or implied warranties of fitness and merchantability are excluded here from, and manufacturer
shall have no liability of any kind for incidental or consequential damages.
EXTENDED WARRANTY / SERVICE AGREEMENT
Extended warranties and service agreements are available on MicroAire power equipment. Extended warranties may
be purchased while the equipment is covered by the original warranty. If the equipment is out of warranty, it must
first be restored, if necessary, to full serviceable condition before being eligible for a service agreement.

IM-7800 Rev G 2021-05
13 of 16
BATTERY INSTRUCTIONS
Battery Charging
See separate operating instructions provided with the MicroAire UL Classied Battery Charger (REF 7500-700).
Battery Installation
The REF 7800 Battery Sternum Saw handpiece must have a fully charged MicroAire battery pack (REF 6640-710, REF
7505-710 or Aseptic Battery) installed before use. To attach the battery to the handpiece, place the battery under the
base of the instrument with the longer side facing the rear of the instrument. Firmly push the battery into the base of
the handpiece until the battery latch clicks into place.
Aseptic Battery
The REF 7800 Battery Sternum Saw is designed to operate using the MicroAire Aseptic Battery. The Aseptic Battery
uses a non-sterile battery pack REF 7500-620 that is placed in a sterilized System housing REF 7500-615 with
locking door, and removes the need to steam sterilize battery packs. See separate instructions provided with
the REF 7500-6XX Aseptic Battery System.
CAUTION: Leaving a battery pack attached to a handpiece when not in use may accelerate battery drain.
MICROAIRE BATTERY WARRANTY
MicroAire Surgical Instruments warrants the REF 7505-710 Large Battery Pack and REF 6640-710 Small Battery Pack to
be free from defects in material and workmanship for a period of one (1) year from the original purchase date by the
end customer. MicroAire warrants the REF 7500-620 Aesptic Battery to be free from defects and workmanship in their
manufacture for a period of ninety (90) days from the original purchase date by the end customer. The warranty is
limited to the repair or replacement of the product without charge.
This warranty is void in the event of abuse, misuse, or use in other than normal surgical environment, or in the event
disassembly, alteration, or repair of the product not authorized by the manufacturer, or in the event that the product
has not been used in compliance with the written instructions furnished by the Manufacturer.
All other expressed or implied warranties of fitness and merchantability are excluded here from, and manufacturer
shall have no liability of any kind for incidental or consequential damages.
Extended warranties and/or service agreements are not available on MicroAire batteries.
Part Number
7505
Drill Reamer
Battery\Electric
7506
Oscillating Saw
Battery\Electric
7507
Reciprocating Saw
Battery\Electric
Power Output kW-KiloWatts 0.0135 0.13 0.13
Vibration Exposure
ahv(m/s2) 4.1 4.1 4.1
Uncertainty
K (m/s2)1.5 1.5 1.5
Noise Emission
Value
LPA (db(A)) 84 84 84
LC,peak (db(C)) - - -
LWA (dbA)) 100 100 100
Mass Weight (kg) 1.04 1.19 1.16

IM-7800 Rev G 2021-05 14 of 16
ELECTROMAGNETIC COMPATIBILITY
NOTE: The 7800 Sternum Saw is suitable for use in hospitals and surgery centers. These systems should not
be used near Magnetic Resonance Imaging equipment.
EMC Test & Standard Test Levels / Limits Compliance
Mains terminal disturbance voltage
(conducted emissions): CISPR 11
Not Applicable for Battery Powered
Equipment Not Applicable
Electromagnetic radiation disturbance
(radiated emissions): CISPR 11 CISPR 11 Class A Complies with Class A Limits
Harmonic distortion: IEC 61000-3-2 Not Applicable for Battery Powered
Equipment Not Applicable
Voltage fluctuations and flicker: IEC
61000-3-3
Not Applicable for Battery Powered
Equipment Not Applicable
Electrostatic discharge immunity:
IEC 61000-4-2
Discharge Level (kV)
Contact – Direct: + /- 8
Contact – Indirect: +/- 8
Air +/- 2, 4, 8, 15
Complies Per Note*
Complies Per Note*
Complies Per Note*
Radiated RF electromagnetic field
immunity: IEC 61000-4-3
3V/m
80MHz to 2700 MHz
80% AM at 1 KHz
Complies Per Note*
Immunity to proximity fields from RF
wireless communications equipment:
IEC 61000-4-3
IEC 60601-1-2 Table 9 Complies Per Note*
Immunity to High Frequency Surgical
Equipment:
IEC 60601-2-2
Tested for immunity to modern
(non-spark gap) HF Surgical
Equipment per the guidelines of
IEC60601-2-2:2017 Annex BB.
Cut Mode at 150 Watts:
Coagulate Mode at 70 Watts:
The HF Surgical generator was
in operation as part of the EM
environment for all other immunity
tests.
Complies Per Note*
Complies Per Note*
See Other Immunity Tests
Electrical fast transient/burst immunity
– IEC 61000-4-4
AC mains:
I/O SIP/SOP Ports:
Not Applicable for Battery Powered
Equipment
Series 7000 Instruments
have no I/O SIP/SOP Ports
Not Applicable
Not Applicable
Surge Immunity: IEC 61000-4-5
Input Power Ports
(Line to Line)
(Line to Earth)
Not Applicable for Battery Powered
Equipment
Not Applicable

IM-7800 Rev G 2021-05
15 of 16
ELECTROMAGNETIC COMPATIBILITY
EMC Test & Standard Test Levels / Limits Compliance
Immunity to conducted disturbances
induced by RF fields (conducted RF
disturbance immunity) –
AC mains: IEC 61000-4-6
Patient Connected Ports: IEC 61000-4-6
Not Applicable for Battery Powered
Equipment
Not Applicable
Power Frequency Magnetic Field
Immunity: IEC 61000-4-8
Frequency: 50 Hz or 60 Hz
Test Level: 30 A/m Complies Per Note*
Voltage Dips: IEC 61000-4-11 Not Applicable for Battery Powered
Equipment Not Applicable
Voltage Interruptions Immunity: IEC
61000-4-11
Not Applicable for Battery Powered
Equipment Not Applicable
NOTE *: The 7800 Sternum Saw has no Essential Performance. Compliance with Immunity Testing is defined
as a) the instrument will remain functional after the test, and b) the instrument in a standby mode
will not run for more than 1 second as a result of the test.
WARNING: Use of the 7800 Sternum Saw adjacent to or stacked with other equipment should be avoided be
cause it could result in improper operation. If such use is necessary, the 7800 Sternum Saw and the
adjacent equipment should be observed to verify that they are operating normally.
WARNING: Portable RF communications equipment should be used no closer than 30 cm (12 inches) away
from any part of the 7800 Sternum Saw. Otherwise degradation of the performance of the 7800
could result.
WARNING: Use of Battery Packs other than those specied in the List of Compatible Accessories could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and
result in improper operation.
WARNING: 7800 Sternum Saw has been tested for use in the vicinity of High Frequency Electrosurgical / Electro
coagulation equipment. Use of such equipment in close proximity to the 7800 Sternum Saw may
cause unintended interruptions or unintended motion of durations up to 1 second. Place the
Instrument on a safe surface when not in use. Do not place powered Instruments on the patient
when not in use. Avoid entwining HF surgical cables with powered instruments.
WARNING: The following items should be periodically inspected for signs of damage and repaired or replaced as
needed to ensure continued safety with regard to electromagnetic disturbances over the life of the
system:
• Check that the Battery Connection base, the rear end cap and the front of the instrument
remain securely connected to the main instrument body.
• Check for damage to the battery pack housings.
NOTE: The emissions characteristics of this product make it suitable for use in industrial areas and hospitals
(CISPR 11 Class A). If it is used in a residential environment (for which CISPR 11 Class B is normally
required) this equipment might not offer adequate protection to radio frequency communication
services. The user might be required to take mitigation measures, such as relocating or re-orienting
the equipment.

©2021 MicroAire Surgical Instruments LLC | IM-7800 Rev. G, Printed in USA 05/2021
MicroAire Surgical Instruments, LLC
3590 Grand Forks Boulevard
Charlottesville, Virginia 22911 USA
Phone: (800) 722-0822 (434) 975-8000
Order Fax: (800) 648-4309 or (434) 975-4131
www.microaire.com
MediMark Europe
11, rue Emile Zola - BP 2332
F-38033 Grenoble Cedex 2
France 2797
Other manuals for 7800
1
Table of contents
Other MicroAire Medical Equipment manuals