
ULTRAFLO 2000 INSTALLATION & OPERATING INSTRUCTIONS
17
W6:Pulse at Max
This message occurs when the pulses have
exceeded the maximum setting. It is possible to
re-scale the pulse output to cope with the higher
flow rate.
APPLICATION NOTES
The ULTRAFLO 2000 is a “Transit Time”
Ultrasonic flowmeter which has been designed to
work with “Clamp On” transducers, thus enabling
liquid flowing within a closed pipe to be measured
accurately without the need for any mechanical
parts to be inserted either through the pipe wall or
protrude into the flow system.
The meter is controlled by a micro-processor
containing a wide range of data which enables the
instrument to measure flow in any pipe diameter
from 13mm bore up to 5000mm, made of any
material and over a wide range of operating
temperatures.
The system operates as follows:
When ultrasound is transmitted from transducer
“X” to transducer “Y” the speed at which the sound
travels through the liquid is accelerated slightly by
the velocity of the liquid. If sound is transmitted in
the opposite direction from “Y” to “X” it is
decelerated because it is travelling against the
flow of the liquid. The difference in time taken to
travel the same distance but in opposite directions
is directly proportional to the flow velocity of the
liquid.
Having measured the flow velocity and knowing
the pipe cross-sectional area, the volumetric flow
can be easily calculated. All of the calculations
required to first determine the correct siting of the
transducers and subsequently compute the actual
flow are carried out by the microprocessor.
To measure flow, it is first necessary to obtain
detailed information about each application, which
is then programmed into the processor via the
keypad. This information must be accurate
otherwise flow measurement errors will occur.
Further having calculated the precise position at
which the transducers must be clamped onto the
pipe wall, it is equally important to align and
separate the transducers accurately with respect
to one another, as failing to do so will again cause
errors in measurement.
Finally to ensure accurate flow measurement it is
imperative that the liquid is flowing uniformly within
the pipe and that the flow profile has not been
distorted by any upstream or downstream
obstructions.
To obtain the best results from the Ultraflo 2000 it
is absolutely necessary that the following rules for
positioning the transducers are adhered to and
that the condition of the liquid and the pipe wall
are suitable to allow transmission of the sound
along its predetermined path.
Selecting A Meter Position
As the transducers with the Ultraflo 2000 are
clamped to the outside surface of the pipe, the
meter has no way of determining exactly what is
happening to the liquid. The assumption therefore
has to be made that the liquid is flowing uniformly
along the pipe, either under fully turbulent
conditions or under laminar flow conditions.
It is further assumed that the flow velocity profile is
uniform for 360° around the pipe axis. The Ultraflo
2000 is normally supplied calibrated for use on
turbulent flows but by adjustment of the CAL
FACTOR the instrument can be used for laminar
flow applications.
Figure 11
Figure 11 shows a uniform profile as compared to
a distorted profile.
The difference between (a) and (b) is that the
“Mean Velocity” of the flow across the pipe is
different and the Ultraflo 2000 expects a uniform
flow as in (a). The distorted flow in (b) will give
measurement errors which cannot be predicted or
compensated for.
Flow profile distortions result from upstream
disturbances such as bends, tees, valves, pumps
and other similar obstructions. To ensure a
uniform profile the transducers must be mounted
far enough away from any cause of distortion such
that it no longer has an effect.
“X” “Y”
FLOW