Trotec T660 User manual

TRT-BA-T660-TC2014-04-003-EN
T660
EN
OPERATING MANUAL
MOISTURE MEASURING
DEVICE

2 EN
moisture measuring device T660
Table of contents
Notes regarding the operating manual.................................2
Safety .....................................................................................2
Information about the device................................................3
Transport and storage...........................................................5
Operation ...............................................................................5
Measuring principle ..............................................................9
PC software..........................................................................10
Maintenance and repair ......................................................11
Errors and faults..................................................................11
Disposal ...............................................................................12
Notes regarding the operating manual
Symbols
Warning of electrical voltage
This symbol indicates dangers to the life and health of
persons due to electrical voltage.
Warning
This signal word indicates a hazard with an average
risk level which, if not avoided, can result in serious
injury or death.
Caution
This signal word indicates a hazard with a low risk
level which, if not avoided, can result in minor or
moderate injury.
Note
This signal word indicates important information (e.g.
material damage), but does not indicate hazards.
Info
Information marked with this symbol helps you to carry
out your tasks quickly and safely.
Follow the manual
Information marked with this symbol indicates that the
operating manual must be observed.
You can download the current version of the operating manual
and the EU declaration of conformity via the following link:
T660
https://hub.trotec.com/?id=44145
Safety
Read this manual carefully before starting or using the
device. Always store the manual in the immediate vicinity
of the device or its site of use.
Warning
Read all safety warnings and all instructions.
Failure to follow the warnings and instructions may
result in electric shock, fire and/or serious injury.
Save all warnings and instructions for future
reference.
This appliance can be used by children aged from
8years and above and persons with reduced physical,
sensory or mental capabilities or lack of experience
and knowledge if they have been given supervision or
instruction concerning use of the appliance in a safe
way and understand the hazards involved.
Children shall not play with the appliance. Cleaning and
user maintenance shall not be made by children
without supervision.
• Do not use the device in potentially explosive rooms or
areas and do not install it there.
• Do not use the device in aggressive atmosphere.
• Do not immerse the device in water. Do not allow liquids to
penetrate into the device.
• The device may only be used in dry surroundings and must
not be used in the rain or at a relative humidity exceeding
the operating conditions.
• Protect the device from permanent direct sunlight.
• Do not expose the device to strong vibrations.
• Do not remove any safety signs, stickers or labels from the
device. Keep all safety signs, stickers and labels in legible
condition.
• Do not open the device.
• Observe the storage and operating conditions as given in
the Technical data chapter.

EN 3
moisture measuring device T660
Intended use
Only use the device for the moisture measurement of building
materials. Observe and comply with the technical data.
To use the device for its intended use, only use accessories and
spare parts which have been approved by Trotec.
Improper use
Do not use the device in potentially explosive atmospheres, for
measurements in liquids or at live parts.
Any unauthorised changes, modifications or alterations to the
device are forbidden.
Personnel qualifications
People who use this device must:
• have read and understood the operating manual, especially
the Safety chapter.
Residual risks
Warning of electrical voltage
There is a risk of a short-circuit due to liquids
penetrating the housing!
Do not immerse the device and the accessories in
water. Make sure that no water or other liquids can
enter the housing.
Warning of electrical voltage
Work on the electrical components must only be
carried out by an authorised specialist company!
Warning
Risk of suffocation!
Do not leave the packaging lying around. Children may
use it as a dangerous toy.
Warning
The device is not a toy and does not belong in the
hands of children.
Warning
Dangers can occur at the device when it is used by
untrained people in an unprofessional or improper way!
Observe the personnel qualifications!
Caution
Keep a sufficient distance from heat sources.
Note
To prevent damages to the device, do not expose it to
extreme temperatures, extreme humidity or moisture.
Note
Do not use abrasive cleaners or solvents to clean the
device.
Information about the device
Device description
The material moisture measuring deviceT660 is a dielectric
moisture indicator, which can be used for the quick and non-
destructive localization of humidity or moisture distributions.
The device enables the detection of near-surface moisture in
walls or flooring. The measured values will be displayed in real
time.
A minimum, maximum and average function is available for the
direct analysis of the measured data. Besides, the currently
measured value can be recorded via the hold function.
The device further features an alarm function. Once the preset
limit value is exceeded, the device alerts the user through
emitting an acoustic signal!
The device is suited for the preliminary check of the building
materials' readiness for covering for subsequent measurements
according to the calcium carbide method (CM measurement).
The moisture indication allows the localization of the most
significant measuring spots to extract material for the CM
measurement.
You can connect the device to a computer by using the USB
cable included in the scope of delivery. Then you can extract
and analyse your measured results with the optional
MultiMeasure Studio software.
Device depiction
1
4
3
2
5
6
No. Designation
1 Display
2 Cross control
3 Battery compartment with cover
4 Power button
5 USB interface
6 Measuring head

4 EN
moisture measuring device T660
Cross control
7
10
9
8
11
No. Designation
7 Up button
8 Right/back button
9 Down button
10 OK button
11 Left/menu button
Display
12
16
15
14
13
17
No. Designation
12 CFG symbol (configuration mode display)
13 Measuring mode
14 Measurement value display top
15 Indication of the unit (moisture in % or CM)
16 Measurement value display bottom
17 Configuration mode
Technical data
Parameter Value
Model T660
Material moisture
Measuring range 1 to 200 digits or
0% to 7.6%
Accuracy 0.1digits or
±0.1%
Resolution 0.1digits
Penetration depth (non-
destructive)
max. 40mm
General technical data
Measuring method Capacitive
Display LCD
Interface USB
Operating conditions 0°C to 50°C with <95 %RH
(non-condensing)
Storage conditions -10°C to 60°C with <95%RH
(non-condensing)
Power supply 4x 1.5 V, AAbatteries
Weight approx. 285 g
Dimensions
(lengthx widthx height)
209mmx 63mmx 35mm
Scope of delivery
• 1x Device T660
• 4x 1.5V batteries, typeAA
• 1 x Factory test certificate
• 1 x Quick guide
• 1 x Display protection film
• 1 x USB cable

EN 5
moisture measuring device T660
Transport and storage
Note
If you store or transport the device improperly, the
device may be damaged.
Note the information regarding transport and storage of
the device.
Transport
When transporting the device, ensure dry conditions and and
protect the device from external influences e.g. by using a
suitable bag.
Storage
When the device is not being used, observe the following
storage conditions:
• dry and protected from frost and heat
• protected from dust and direct sunlight
• The storage temperature is the same as the range given in
the Technical data chapter.
• Remove the batteries from the device.
Operation
Inserting the batteries
Insert the supplied batteries before first use.
Caution
Make sure that the surface of the device is dry and the
device is switched off.
3
1. Remove the battery compartment cover(3).
2. Insert the batteries with correct polarity.
3. Insert the battery compartment cover(3).
ðThe device can now be switched on.
Switch-on
Note:
The cross control is very sensitive. Therefore, avoid dirt on the
control panel, because it could be misinterpreted by the device
as keystroke.
Before use make sure that the touchscreen control panel is dirt-
free.
If required clean the touchscreen control panel according to
chapter Cleaning the device.
1. Hold the device pointing into clear space.
– The device should point away from the body and not at
or near material surfaces. When failing to observe the
above instructions, the following calibration will be corrupt.
2. Press the On/Off button(4) until a beep is emitted.
ðThe device performs a short self-test.
ðThe device name and firmware version is shown on the
display.
ðThe battery charge is indicated on the display.
3. The device carries out an automatic calibration. During
calibration the display indicates the letteringCAL. These
flashing characters are accompanied by brief acoustic
signals. An extended acoustic signal indicates that the
calibration has been completed.
ðThe device is ready for operation.

6 EN
moisture measuring device T660
Carrying out a measurement
Note:
Note that moving from a cold area to a warm area can lead to
condensation forming on the device's circuit board. This
physical and unavoidable effect can falsify the measurement. In
this case, the display shows either no measured values or they
are incorrect. Wait a few minutes until the device has become
adjusted to the changed conditions before carrying out a
measurement.
Please observe the information regarding the measurement
principle.
TH
HOLD
%WT
g/m
3
g/kg
Pabs[hPa]
SENS MAT TEMP Alarm
Alt.[ft] CAL12
gr/lb dp °C °F % rH
CM m/s ppm °F °C
IR
MAX
DP
MIN AVG
CFG
ACT
1. Hold the measuring device at the black rubber surfaces
and maintain this position, for otherwise the measurement
results might be falsified.
2. Firmly place the measuring head in a vertical position on
the good to be measured.
3. Observe a minimum distance of 8 to 10 cm to corners.
4. Read the measured value from the display.
Carrying out a comparative measurement
1. Look for a preferably dry spot at the component.
2. Perform a measurement as previously described.
ðThe thus determined measured value will be the
reference value for dry.
3. Look for a preferably damp or wet spot at the component,
if any.
4. Perform a measurement as previously described.
ðWith regard to the present conditions, the thus
determined measured value will be the reference value
for damp/wet.
5. Carry out further measurements at the component.
Higher measured values usually signify a higher moisture
content in the near-surface area of the material to be measured.
Key lock
1. Briefly press the On/Off key(4) during operation.
ðThe device emits a short beep.
ðLoC on is indicated on the display.
ðKey lock is activated.
2. Press the On/Off key(4) again.
ðThe device emits a short beep.
ðLoC oFF is indicated on the display.
ðKey lock is no longer activated.
Configuration mode
1. Press the Left/menu key(11) for approx. 2seconds.
ðThe device emits a short beep.
ðThe CFG symbol(12) is displayed in the upper right
corner.
2. Use the buttons on the cross control(2) to select the
desired option.
3. Confirm the selection by pressing the OK button(10).
ðThe selected symbol will be illuminated.
12
17
Configuration
mode
Description
ALARM Setting the threshold value for the alarm
Lamp Adjusting the display illumination
CAL Setting the offset value.
MAT Material settings

EN 7
moisture measuring device T660
Setting the alarm limit value
Here you can determine the limit value for the alarm function.
Upon exceeding this value, the device emits an acoustic signal
and the indication ALARM flashes in the configuration
mode(17). The alarm function draws on the current measured
value.
The limit value can be determined within a range of 0 to 200.
1. Select ALARM whilst in configuration mode(17).
2. Press the OK button(10) to confirm.
ðThe upper measurement value display(14) flashes.
3. Press the Up(7) or Down(9) button to activate or
deactivate the alarm.
ðOn or Off will be indicated in the upper measurement
value display(14).
4. Press the Right/back button (8).
ðDepending on the selection, the alarm is either activated
or deactivated.
ðThe lower measurement value display(16) flashes.
5. Press the buttons Right/back(8) or Left/menu(11) to select
a digit.
ðThe selected digit flashes.
6. Press the Up(7) or Down(9) button to change the value of
the selected digit.
7. Repeat the steps 5. and 6. until the value is set as desired.
8. Press the OK button(10) for approx. 2 seconds.
ðThe alarm was set based on your choice.
ðThe device switches to measuring mode.
ðWith activated alarm function the indicationALARM
continues to be displayed in the configuration
mode(17).
Adjusting the display illumination
The display illumination can be adjusted within a range of20
to100%. Another available setting is AL.on (always on). AL.on
features a brightness of 100 % and deactivates the automatic
switch-off function.
1. Select the lamp(18) whilst in configuration mode.
2. Press the OK button(10) to confirm.
3. Select the desired value by use of the Up(7) or Down(9)
button.
4. Press the OK button(10) for approx. 2 seconds.
ðThe set value will be adopted.
ðThe device switches to measuring mode.
Material settings
Here you can select the construction material for the
measurement. The following options are available:
•0 no: No material selected.
•1 An: The selected material is anhydrite screed.
•2 CE: The selected material is cementitious screed.
Please note that in case of the settings1 and 2 an additional
value will be indicated in the upper measurement value
display(14).
It is a combination of the material(1or2) and the calculated
percentage.
Example: 2_7.5 (7.5 % with cementitious screed).
1. Select MAT whilst in configuration mode(17).
2. Press the OK button(10).
3. Select the desired value by use of the Up(7) or Down(9)
button.
4. Press the OK button(10) for approx. 2 seconds.
ðThe device switches to measuring mode.
ðWith either setting 1 or 2 the indication MAT continues
to be displayed in configuration mode(17).
Setting the offset
By use of CAL a single-point calibration can be carried out for
the selected sensor indications. All sensors are already factory-
calibrated and have a corresponding characteristic calibration
curve. By stating a calibration value (offset) a global shift of the
calibration curve, which has an effect on the entire measuring
range, is performed for the single-point calibration! The offset
value to be entered is that value by which the calibration curve
will be shifted.
Example:
The displayed value is always "5" too high => change the offset
value for this measurement channel to "-5".
The offset value's default setting is 0.0.
Note
Please note that changing the offset value brings about
an automatic reset of the measured values.
1. Select CAL whilst in configuration mode(17).
2. Press the OK button(10).
3. Use the buttons Left or Right to select a digit.
ðThe selected digit flashes.
4. Press the Up(7) or Down(9) button to change the value of
the selected digit.
5. Repeat the steps 3. and 4. until the value is set as desired.
6. Press the OK button(10) for approx. 2 seconds.
ðThe offset is set.
ðThe device switches to measuring mode.
ðWith set offset value the indication CAL continues to be
displayed whilst in configuration mode(17).

8 EN
moisture measuring device T660
Exiting configuration mode
Configuration mode will automatically be terminated after 8
seconds without entry.
You can quit configuration mode yourself anytime. Please note
that in such an event no changes made to the settings will be
adopted.
1. Press the Right/back key(8) for approx. 2 seconds.
ðConfiguration mode will be terminated.
Measuring mode
1. Press the keys Right/back(8) or Left/menu(11) until the
desired measuring mode is displayed.
2. The selected measuring mode(13) will be indicated on the
display(1)
13
15
14
16
The device comes with the following measuring modes:
Measuring
mode
Description
ACT Measured value in real time
AVG Average value of measurements since switch-on
Min Minimum measured value
Max Maximum measured value
HOLD Measured value will be held
Holding the measured value
1. Set the measuring mode to HOLD.
ðThe current measured value will be held and displayed.
ðThe device will hold this value until the measured values
are set back or the device is switched off.
Resetting the measured values
1. Press the OK button(10) for approx. 2 seconds.
ðAll previously stored measured values of the measuring
modes AVG, MIN, MAX and HOLD will be set back.
ðAll previously stored measured values of the measuring
modes AVG, MIN, MAX and HOLD will be set back.
Measured value storage
Please note that measured values cannot be saved on the
device itself. In order to save measured values, the device has
to be connected to a PC via a USB cable using the MultiMeasure
Studio software.
1. Briefly press the OK button(10).
ðThe displayed measured value will be saved in the
software.
Checking current settings
1. Press the Right/back key(8) for approx. 2 seconds.
ðThe selected settings are indicated in the two
measurement value displays(14, 16).
USB interface
The device can be connected to a PC via the USB interface(5).
See chapter PC software.
Switch-off
1. Press and hold the On/Off button(4) for approx.3seconds
until a beep is emitted.
ðThe device is switched off.

EN 9
moisture measuring device T660
Measuring principle
The measurement is effected according to the dielectric
measuring principle.
Notes regarding the measurement principle
• The measurement results are only to be used as reference
for a rough orientation.
• An important variable influencing the measured value is
the bulk density of the good to be measured. The higher
the bulk density, the higher the measured value.
• Before measuring, the measuring point is to be cleared of
any impurities (e.g. paint residues, dust).
• If the material to be measured contains metal (e.g. nails,
screws, lines, pipes, etc.) and is situated within the
sensor's measuring field, the measured value skyrockets.
In that event the measurement is not conclusive.
• When the ball head is held to corners (e.g. window
frames), the measured value is generally higher, because
there is more matter in the measuring head's stray field.
Toward a corner a distance of more than 8 to 10cm has to
be observed.
• During the measurement always hold the measuring
head(6) in a vertical position to the measured material,
press it to the surface and do not tilt it.
• Rough surfaces will always result in a too low measured
value.
• Depending on the material's bulk density, the device's
impact depth amounts to 20 to 40mm. Conclusions
regarding zones located at a lower level cannot be drawn.
• With material thicknesses of less than 20 mm there is a
danger of humidity values from adjacent material layers
affecting the measured value.
• The dielectric measurement method's main area of use
consists in comparative measurements at the same
construction material or similar components. Therefore,
this method is well suited for the examination of water
damage, leak detection and the localization of moisture
zones for the purpose of marking the measuring spot for
CM measurements.
Measured value assessment for wood
The measurement result is either in the dry or the saturated
area.
Saturated means that in addition to the water bound within the
wood there is further liquid. If there is, hence, a measured value
of more than 80, the wood is most like soaked.
Display [digit] Wood moisture zone
< 50 digits dry
> 80 digits limit to the saturated range
Since the measurement value displays of the dielectric
measurement method – depending on marginal conditions – are
subject to great fluctuations, a resistance measurement is
always preferable, in particular for the determination of wood
moisture.

10 EN
moisture measuring device T660
Measured value assessment for building material
For building materials the measurement results of the dielectric
measuring method can only be used as reference for a rough
orientation. Conclusions with respect to absolute humidity in
mass % (M%) can only be drawn for measurements, performed
with the exact same marginal conditions and compositions of
the building material as indicated for the test set-up in the chart
below.
This graph was created in collaboration with the Institute of
Building Materials Research of the RWTH Aachen (IBAC) and
illustrates the correlation between the measured value and the
mass-related moisture content of the examined building
materials.
Key
a Moisture content (M%)
b Measured value (digits)
Anhydrite floating screed (conversion: M% = CM%)
Cementitious screed (conversion: CM% = M% -1.5 to 2)
C 30/37 concrete (conversion not possible)
The spread of the achieved measurement results for the
dielectric measuring method is clearly higher than for the
resistance measuring method. The dielectric measuring method
is not suited for the quantitative, but only for the qualitative
moisture measurement in near-surface areas.
The indicated measured values are not the result of a qualified
moisture measurement as per VOB (German Construction
Contract Procedures) or relevant specialist regulations.
The measured values are only to be interpreted as indicators
(dry, damp, wet).
For building materials the measurement results of the dielectric
measuring method can only be used as reference for a rough
orientation.
Display [digit] Building moisture zone
< 40 digits dry
40 – 80 digits damp
> 80 digits wet
PC software
Use the MultiMeasure Studio Standard PC software (free
standard version) or MultiMeasure Studio Professional (paid
professional version, dongle required) to carry out a detailed
analysis and visualisation of your measured results. You can
only use all configuration, visualisation and functional options of
the device when using this PC software and a TROTEC® USB
dongle (professional).
Installation requirements
Ensure that the following minimum requirements for installing
the MultiMeasure Studio Standard or MultiMeasure Studio
Professional PC software are fulfilled:
• Supported operating systems (32 or 64bit version):
– Windows XP from service pack 3
– Windows Vista
– Windows 7
– Windows 8
– Windows 10
• Software requirements:
– Microsoft Excel (to display stored Excel files)
– Microsoft .NET Framework 3.5SP1 (is automatically
installed during software installation, where applicable)
• Hardware requirements:
– Processor speed: 1.0GHz, minimum
– USB connection
– Internet connection
– 512MB RAM, minimum
1GB hard disk space, minimum
optional: TROTEC® USB dongle (Professional) for using the
professional version of the PC software
Installing the PC software
1. Download the current PC software from the Internet. To do
so, visit the website www.trotec.de. First click on products
and services, then on services and downloads. Then select
Software in the category menu. Select the MultiMeasure
Studio Standard software from the list. If you want to use
the optionally available professional version of the PC
software "MultiMeasure Studio Professional" (dongle),
then get in touch with your TROTEC® customer service.
2. Double-click on the downloaded file to start the
installation.
3. Follow the instructions of the installation wizard.

EN 11
moisture measuring device T660
Starting the PC software
1. Connect the device to your PC via the USB connection
cable provided in the scope of delivery.
Note
Step 2 only needs to be performed, when using the
Professional software functions.
If you only use the Standard software functions, please
proceed to step 3.
2. In order to enable the Professional functions, connect the
TROTEC® USB dongle to a free USB port on your PC.
The TROTEC® USB dongle (Professional) is automatically
detected by the operating system.
If you only connect the TROTEC® USB dongle (Professional)
to your PC after starting the PC software, click the
"Parameters" menu item in the PC software. Afterwards,
click the USB symbol (dongle check) to read the connected
TROTEC® USB dongle (Professional).
3. Switch the device on (see chapter Switch-on and
measurements).
4. Start the MultiMeasure Studio software. Depending on the
activation process you will be asked to insert the access
code that has been previously assigned to you. Only then
the dongle for releasing the according Professional tools of
the software will be activated.
Note
Information regarding the use of the MultiMeasure
Studio software is provided in the help text of the
software.
Maintenance and repair
Battery change
Change the batteries when the message Batt lo is displayed
upon switch-on or the device can no longer be switched on.
See chapter Operation Inserting the batteries.
Cleaning
Clean the device with a soft, damp and lint-free cloth. Make
sure that no moisture enters the housing. Do not use any
sprays, solvents, alcohol-based cleaning agents or abrasive
cleaners, but only clean water to moisten the cloth.
Repair
Do not modify the device or install any spare parts. For repairs
or device testing, contact the manufacturer.
Errors and faults
The device has been checked for proper functioning several
times during production. If malfunctions occur nonetheless,
check the device according to the following list.
The device does not switch on:
• Check the charging status of the batteries. Change the
batteries when the message Batt lo is displayed upon
switch-on.
• Check that the batteries are properly positioned. Check the
polarity is correct.
• Never carry out an electrical check yourself; instead,
contact your TROTEC® customer service.

12 EN
moisture measuring device T660
Disposal
The icon with the crossed-out waste bin on waste
electrical or electronic equipment stipulates that this equipment
must not be disposed of with the household waste at the end of
its life. You will find collection points for free return of waste
electrical and electronic equipment in your vicinity. The
addresses can be obtained from your municipality or local
administration. For further return options provided by us please
refer to our website https://de.trotec.com/shop/.
The separate collection of waste electrical and electronic
equipment aims to enable the re-use, recycling and other forms
of recovery of waste equipment as well as to prevent negative
effects for the environment and human health caused by the
disposal of hazardous substances potentially contained in the
equipment.
In the European Union, batteries and accumulators must not be
treated as domestic waste, but must be disposed of
professionally in accordance with Directive 2006/66/EC of the
European Parliament and of the Council of 6September 2006
on batteries and accumulators. Please dispose of batteries and
accumulators according to the relevant legal requirements.

Trotec GmbH
Grebbener Str. 7
D-52525 Heinsberg
+49 2452 962-400
+49 2452 962-200
www.trotec.com
Other manuals for T660
1
Table of contents
Other Trotec Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

TEXA
TEXA GAS MOBILE user manual

AESSEAL
AESSEAL FLOWTRUE FTP-50-145 Installation, operation & maintenance manual

Andor Technology
Andor Technology Zyla sCMOS 4.2 PLUS Hardware guide

Medica
Medica EasyLyte Na/K Service manual

LI-COR
LI-COR LI-250A instruction manual

Spinner
Spinner BN 53 70 84 Handling & Cleaning Instructions