Milwaukee MI412 User manual

USER MANUAL
MI412
PRO Phosphate
Low Range Photometer
milwaukeeinstruments.com (USA & CAN)
milwaukeeinst.com
yUnder range. A blinking “0.00”
indicates that the sample
absorbs less light than the zero
reference. Check the procedure
and make sure you use the same
cuvet for reference (zero) and
measurement.
yA flashing value of the maximum
concentration indicates an over
range condition. The concentration
of the sample is beyond the
programmed range: dilute the
sample and re-run the test.
BATTERY REPLACEMENT
Battery replacement must only take place in a non-
hazardous environment.
Simply rotate the battery cover on the back of the meter.
Detach the battery from the terminals and attach a fresh
9V battery while paying attention to the correct polarity.
Insert the battery and replace the cover.
ACCESSORIES
MI512-100 Phosphate low range reagent (100 tests)
MI0001 Glass cuvets (2 pcs)
MI0002 Caps for cuvets (2 pcs)
MI0003 Stoppers for cuvets (2 pcs)
MI0004 Tissue for wipping cuvets (4 pcs)
MI0005 9V battery (1 pc)
CERTIFICATION
Milwaukee Instruments conform to the CE
European Directives.
Disposal of Electrical & Electronic
Equipment. Do not treat this product as
household waste. Hand it over to the
appropriate collection point for the recycling of electrical
and electronic equipment.
Disposal of waste batteries. This product
contains batteries. Do not dispose of them
with other household waste. Hand them
over to the appropriate collection point for recycling.
Please note: proper product and battery disposal
prevents potential negative consequences for human
health and the environment. For detailed information,
contact your local household waste disposal service or
go to www.milwaukeeinstruments.com (USA & CAN)
or www.milwaukeeinst.com.
RECOMMENDATION
Before using this product, make sure it is entirely suitable
for your specific application and for the environment in
which it is used. Any modification introduced by the user
to the supplied equipment may compromise the meter’s
performance. For your and the meter’s safety do not use
or store the meter in hazardous environment. To avoid
damage or burn, do not perform any measurement in
microwave ovens.
WARRANTY
These instruments are warranted against defects in
materials and manufacturing for a period of 2years from
the date of purchase. This warranty is limited to repair or
free of charge replacement if the instrument cannot be
repaired. Damage due to accidents, misuse, tampering
or lack of prescribed maintenance is not covered by
warranty. If service is required, contact your local
Milwaukee Instruments Technical Service. If the repair is
not covered by the warranty, you will be notified of the
charges incurred. When shipping any meter, make sure it
is properly packaged for complete protection.
ISTMI412 07/20
SATISFACTION
GUARANTEED
MI412 PRO Phosphate
Low Range Photometer
Dear Customer,
Thank you for choosing a Milwaukee Instruments
product. This manual will provide you with the necessary
information for the correct use of the instrument. Please
read it carefully before using the meter.
SPECIFICATIONS
Range 0.00 to 2.50 mg/L PO4
Resolution 0.01 mg/L
Precision ±0.07 mg/L @ 1.00 mg/L
Light Source Tungsten lamp
Light Detector Silicon Photocell and 610 nm narrow
band interference filter
Method Adaptation of ascorbic acid method
Environment 0 to 50 °C (32 to 122 °F)
100% RH max
Battery Type 9 volt (1 pc)
Auto-Shut o After 10’ of non-use
Dimensions 192 x 104 x 52 mm
(7.5 x 4.1 x 2”)
Weight 380 g
MEASUREMENT PROCEDURE
1. Turn the meter on by pressing ON/OFF.
2. When the LCD displays “- - -”, the
meter is ready.
3. Fill the cuvet with 10 mL of sample, up
to the mark, and replace the cap.
4. Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
5. Press ZERO and “SIP” will blink on the
display.
6. After a few seconds the display will
show “-0.0-”. The meter is now zeroed
and ready for measurement.
▶
7. Remove the cuvet. Open it.
8. Add the content of one packet of MI512-0 reagent.
9. Reinsert the cuvet into the holder
and ensure that the notch on the
cap is positioned securely into the
groove.
10. Press and hold READ for 3 seconds
and the display will show the
countdown prior to the measurement.
Alternatively, wait for 3 minutes and
just press READ. In both cases “SIP”
will blink during measurement.
11. The instrument directly displays concentration in
mg/L of phosphate.
Interferences
yIron above 50 mg/L
ySilica above 50 mg/L
ySilicate above 10 mg/L
yCopper above 10 mg/L
yHydrogen sulfide, arsenate, turbid sample and highly
buered samples also interfere.
GUIDE TO DISPLAY CODES
This prompt appears for 1 second
each time the instrument is turned
on.
The dashes “- - -” indicates that the
meter is in a ready state and zeroing
can be performed.
Sampling In Progress. Flashing “SIP”
prompt appears each time the meter
is performing a measurement.
“-0.0-”, the meter is in a zeroed state
and measurement can be performed.
3’
The blinking “BAT” indicates that the
battery voltage is getting low and the
battery needs to be replaced.
“-bA-”, the battery is dead and must
be replaced. Once this indication
is displayed, the meter will lock up.
Change the battery and restart the
meter.
“Conf”, the meter has lost its
configuration. Contact your dealer
or the nearest Milwaukee Customer
Service Center.
ERROR MESSAGES
On zero reading
yBlinking “-0.0-” indicates that the
zeroing procedure failed due to a
low signal-to-noise ratio. In this
case press ZERO again.
y“no L”, the instrument can not
adjust the light level. Please check
that the sample does not contain
any debris.
y“L Lo”, there is not enough light to
perform a measurement. Please
check the preparation of the zero
cuvet.
y“L Hi”, there is too much light to
perform a measurement. Please
check the preparation of the zero
cuvet.
On sample reading
y“-SA-”, there is too much light for
the sample measurement. Please
check if the right sample cuvet is
inserted.
y“Inv”, the sample and the zero
cuvet are inverted.
y“ZErO”, a zero reading was not
taken. Follow the instruction in
the measurement procedure for
zeroing the meter.
10 ml
▶

Other Milwaukee Measuring Instrument manuals

Milwaukee MW11 User manual

Milwaukee

Milwaukee MW11 User manual

Milwaukee LDM 30 User manual

Milwaukee

Milwaukee LDM 30 User manual

Milwaukee MW14 User manual

Milwaukee

Milwaukee MW14 User manual

Milwaukee Mi453 User manual

Milwaukee

Milwaukee Mi453 User manual

Milwaukee MC310 User manual

Milwaukee

Milwaukee MC310 User manual

Milwaukee Mi 180 User manual

Milwaukee

Milwaukee Mi 180 User manual

Milwaukee PH55 User manual

Milwaukee

Milwaukee PH55 User manual

Milwaukee MC310 User manual

Milwaukee

Milwaukee MC310 User manual

Milwaukee 48-22-5112 User manual

Milwaukee

Milwaukee 48-22-5112 User manual

Milwaukee MC125 User manual

Milwaukee

Milwaukee MC125 User manual

Milwaukee MW11 User manual

Milwaukee

Milwaukee MW11 User manual

Milwaukee PRO Lux MW700 User manual

Milwaukee

Milwaukee PRO Lux MW700 User manual

Milwaukee MW301 User manual

Milwaukee

Milwaukee MW301 User manual

Milwaukee M12 3PL User manual

Milwaukee

Milwaukee M12 3PL User manual

Milwaukee C 12 IC AVD User manual

Milwaukee

Milwaukee C 12 IC AVD User manual

Milwaukee 3712 User manual

Milwaukee

Milwaukee 3712 User manual

Milwaukee MW100 User manual

Milwaukee

Milwaukee MW100 User manual

Milwaukee MW802 User manual

Milwaukee

Milwaukee MW802 User manual

Milwaukee MR100ATC User manual

Milwaukee

Milwaukee MR100ATC User manual

Milwaukee M12 3632-20 User manual

Milwaukee

Milwaukee M12 3632-20 User manual

Milwaukee M12 CLLP User manual

Milwaukee

Milwaukee M12 CLLP User manual

Milwaukee MW306 MAX User manual

Milwaukee

Milwaukee MW306 MAX User manual

Milwaukee MR32ATC User manual

Milwaukee

Milwaukee MR32ATC User manual

Milwaukee SM100 User manual

Milwaukee

Milwaukee SM100 User manual

Popular Measuring Instrument manuals by other brands

Tektronix TCP312 Service manual

Tektronix

Tektronix TCP312 Service manual

Thermo Scientific DensityPRO-T Series user manual

Thermo Scientific

Thermo Scientific DensityPRO-T Series user manual

ABB 6553 Series instruction manual

ABB

ABB 6553 Series instruction manual

Major tech MT940 instruction manual

Major tech

Major tech MT940 instruction manual

Kobold MIM Series operating instructions

Kobold

Kobold MIM Series operating instructions

TOOLCRAFT CL12 operating instructions

TOOLCRAFT

TOOLCRAFT CL12 operating instructions

TSI Instruments 3936 SMPS instruction manual

TSI Instruments

TSI Instruments 3936 SMPS instruction manual

WIKA 632.50.100 operating instructions

WIKA

WIKA 632.50.100 operating instructions

Anritsu IQ Fiber Master MT2780A user guide

Anritsu

Anritsu IQ Fiber Master MT2780A user guide

Dräger Flow Check Instructions for use

Dräger

Dräger Flow Check Instructions for use

Keysight Technologies E2655C user guide

Keysight Technologies

Keysight Technologies E2655C user guide

Salter Brecknell S750 Series Operation/setup manual

Salter Brecknell

Salter Brecknell S750 Series Operation/setup manual

TPI 717R instruction manual

TPI

TPI 717R instruction manual

Elster 2 Series Commissioning and Maintenance Instructions

Elster

Elster 2 Series Commissioning and Maintenance Instructions

Mannix DCFM8906 operating instructions

Mannix

Mannix DCFM8906 operating instructions

SolaX Power X1-Hybrid/Fit G4 manual

SolaX Power

SolaX Power X1-Hybrid/Fit G4 manual

Shimpo DT-2100 instruction manual

Shimpo

Shimpo DT-2100 instruction manual

i4Home i4H-TX-3AC user manual

i4Home

i4Home i4H-TX-3AC user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.