manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Milwaukee
  6. •
  7. Measuring Instrument
  8. •
  9. Milwaukee SM801 User manual

Milwaukee SM801 User manual

Milwaukee
WARRANTY
These instruments are warranted from all defects in
materials and manufacturing for a period of two years
from the date of purchase.
The probes are warranted for a period of six months.
If during this period the repair or replacement of parts
is required, where the damage is not due to negligence
or erroneous operation by the user, please return the
parts to either distributor or our office and the repair will
be effected free of change.
Note: We reserve the right to modify the design,
construction and appearance of our products
without advance notice.
Milwaukee
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment.
If noticeable damage is found, notify your Dealer.
Each meter is supplied complete with:
•MA850 combination, amplified, double-junction,
gel pH-electrode with incorporated EC/TDS probe
and built-in temperature sensor; 1 m cable
• pH7.01, 1413 µS/cm & 1382 ppm (for SM801) or
1500 ppm (for SM802) solutions, 20 mL each
• 1x9V battery
• Instruction Manual.
OPERATION
Slide off the battery compartment cover on the back
of the meter. Install the battery while paying attention
to its polarity.
Connect the probe to the meter
securely by aligning the pins with
the socket and pushing the plug in.
Always remove the electrode pro-
tective cap before use.
Make sure that the meter has been
calibrated before taking any mea-
surements.
Turn the instrument on by pressing
the ON/OFF key.
pH MEASUREMENTS
If the probe has been left dry, soak the tip in a pH7
(M10007) buffer solution for a few minutes to reacti-
vate it.
Submerge the tip into the sample to be measured,
select the pH mode and allow the reading to stabilize
before taking any measurement.
EC/TDS MEASUREMENTS
Immerse the tip of the probe into the sample to be
tested.
Select the desired measurement range (EC or TDS)
and wait for the temperature sensor to reach thermal
equilibrium with the sample before taking any measure-
ment.
AFTER MEASUREMENTS, switch the instrument off,
clean the probe and store it with a few drops of storage
solution in the protective cap.
CALIBRATION PROCEDURE
Remove the protective cap from the probe.
Switch the meter on.
Select the desired range (pH, EC, TDS) by pressing
the appropriate key.
pH Calibration
The instrument can be manually calibrated at one
point for pH.
A) PREPARATION
If the samples to be measured are
neutral or close to neutral pH, to
obtain accurate readings use pH
7.01 buffer solution (M10007). The
pH 4.01 buffer (M10004) is suit-
able for acidic measurements,
while the pH 10.01 (M10010) buffer
should be used for alkaline
samples.
B) PROCEDURE
Rinse the tip of the probe and immerse it in the buffer
chosen for calibration. Wait a couple of minutes for
the reading to stabilize.
With an accurate thermometer
measure the temperature of the
buffer solution.
Adjust the calibration knob until the
LCD shows the pH value at the
measured temperature (see the
“pH versus temperature” chart for
reference).
The calibration is now complete and the meter is ready
for use.
USER MANUAL
PORTABLE pH/EC/TDS METER
MODEL: SM801. SM802
Smart Combined Meter
Milwaukee
SPECIFICATIONS
SM801 SM802
RANGE 0.0 to 14.0 pH 0.00 to 14.00 pH
0 to 1990 ppm 0 to 4000 ppm
0 to 1990 µS/cm 0.00 to 6.00 mS/cm
RESOLUTION 0.1 pH 0.10 pH
10 ppm 10 ppm
10 µS/cm 0.01 mS/cm
ACCURACY ± 0.2 pH ±0.20 pH
(@ 20ºC) ±2% Full scale
±2% Full scale
CAL.SOLUTIONS M10007 (pH 7.01)
M10031 (1413 µS/cm)
M10032 (1382 ppm) M10442 (1500ppm)
CONVERSION 0.5 0.68
FACTOR
CALIBRATION Manual, at 1 point
TEMPERATURECOMPENSATION
Automatic, from 0 to 50ºC
PROBE MA850 combination pH/EC/TDS probe
ENVIRONMENT 0 to 50ºC, 95%RH max
BATTERY TYPE 1x9V alkaline
BATTERY LIFE approx. 150 hours of use
DIMENSIONS 185 x 82 x 45 mm
WEIGHT 165 g
EC/TDS calibration
The instrument can be manually calibrated at one point
for conductivity range.
A) PREPARATION
For performing TDS calibration,
use 1382 ppm (M10032) solu-
tion for SM801 and 1500 ppm
(M10442) solution for SM802.
For performing EC calibration,
use 1413 µS/cm (M10031) so-
lution for both meters.
B) PROCEDURE
Immerse the tip of the probe in the calibration solu-
tion.
Wait for the reading to stabi-
lize, and then turn the EC/TDS
calibration knob until the dis-
play shows the EC or TDS
reading at 25°C.
The calibration is now complete and the meter is ready
for use.
Note: Since the conversion between EC and TDS is
made by a built-in circuit, it is request to
calibrate the meter only in EC or TDS range,
and the other range will be automatically cali-
brated.
The instrument should be recalibrated at least once a
month, or whenever the probe or battery is changed.
BATTERY REPLACEMENT
The meters are powered by a 9V battery that is
located on the rear of the instrument.
When the battery becomes weak, the instrument auto-
matically switches off.
Battery replacement must only take place in a non-
hazardous area using an alkaline 9V battery.
Slide off the battery compartment cover and replace
the rundown battery with a new one. Make sure that
the battery contacts are tight and secure before in-
serting the cover.
pH VERSUS TEMPERATURE CHART
ACCESSORIES
M10004B pH 4.01 buffer solution, 20
mL sachet (25 pcs)
M10007B pH 7.01 buffer solution, 20
mL sachet (25 pcs)
M10010B pH 10.01 buffer solution,
20 mL sachet (25 pcs)
M10031B 1413 µS/cm cal. solution,
20 mL sachet (25 pcs)
M10032B 1382 ppm cal. solution, 20
mL sachet (25 pcs)
M10442B 1500 ppm cal. solution, 20
mL sachet (25 pcs)
MA9015 Electrode storage solution,
220 mL bottle
MA850 pH/EC/TDS spare probe
with 1 m cable
ISMIL801 04/03
TEMP pH VALUES
ºC ºF
0 32 4.01 7.13 10.32
5 41 4.00 7.10 10.24
10 50 4.00 7.07 10.18
15 59 4.00 7.04 10.12
20 68 4.00 7.03 10.06
25 77 4.01 7.01 10.01
30 86 4.02 7.00 9.96
35 95 4.03 6.99 9.92
40 104 4.04 6.98 9.85
45 113 4.05 6.98 9.85
50 122 4.06 6.98 9.82
55 131 4.07 6.98 9.79
60 140 4.09 6.98 9.77
65 149 4.11 6.99 9.76
70 158 4.12 6.99 9.75
M10004B M10007B M10010B

This manual suits for next models

1

Other Milwaukee Measuring Instrument manuals

Milwaukee PRO MI415 User manual

Milwaukee

Milwaukee PRO MI415 User manual

Milwaukee MW150 User manual

Milwaukee

Milwaukee MW150 User manual

Milwaukee MA881 User manual

Milwaukee

Milwaukee MA881 User manual

Milwaukee M12 CPL User manual

Milwaukee

Milwaukee M12 CPL User manual

Milwaukee MI 406 User manual

Milwaukee

Milwaukee MI 406 User manual

Milwaukee 3421 User manual

Milwaukee

Milwaukee 3421 User manual

Milwaukee 4933478101 User manual

Milwaukee

Milwaukee 4933478101 User manual

Milwaukee MI806 User manual

Milwaukee

Milwaukee MI806 User manual

Milwaukee Smart PH Meter User manual

Milwaukee

Milwaukee Smart PH Meter User manual

Milwaukee 2205-20 User manual

Milwaukee

Milwaukee 2205-20 User manual

Milwaukee MW13 User manual

Milwaukee

Milwaukee MW13 User manual

Milwaukee CALIBRATION 2238-20 User manual

Milwaukee

Milwaukee CALIBRATION 2238-20 User manual

Milwaukee MI 806 User manual

Milwaukee

Milwaukee MI 806 User manual

Milwaukee pH51 User manual

Milwaukee

Milwaukee pH51 User manual

Milwaukee MW12 User manual

Milwaukee

Milwaukee MW12 User manual

Milwaukee EC40 User manual

Milwaukee

Milwaukee EC40 User manual

Milwaukee MC310 User manual

Milwaukee

Milwaukee MC310 User manual

Milwaukee 3712 User manual

Milwaukee

Milwaukee 3712 User manual

Milwaukee pH52 User manual

Milwaukee

Milwaukee pH52 User manual

Milwaukee Mi451 User manual

Milwaukee

Milwaukee Mi451 User manual

Milwaukee MW11 User manual

Milwaukee

Milwaukee MW11 User manual

Milwaukee MW100 User manual

Milwaukee

Milwaukee MW100 User manual

Milwaukee 2237-20 User manual

Milwaukee

Milwaukee 2237-20 User manual

Milwaukee T75 User manual

Milwaukee

Milwaukee T75 User manual

Popular Measuring Instrument manuals by other brands

Fluke 405 Safety information

Fluke

Fluke 405 Safety information

METREL EurotestAT MI 3101 Short instructions

METREL

METREL EurotestAT MI 3101 Short instructions

Sensitech TempTale 4 Instructions for use

Sensitech

Sensitech TempTale 4 Instructions for use

Siemens SITRANS F FS200 Utility Programming manual

Siemens

Siemens SITRANS F FS200 Utility Programming manual

FineTek EPD30 operation instruction

FineTek

FineTek EPD30 operation instruction

Organon TOF-Watch Operator's manual

Organon

Organon TOF-Watch Operator's manual

Elster A100C Operating & maintenance instructions

Elster

Elster A100C Operating & maintenance instructions

LR DM 80-S operating instructions

LR

LR DM 80-S operating instructions

Hanna Instruments Chromium VI quick start guide

Hanna Instruments

Hanna Instruments Chromium VI quick start guide

EUTECH INSTRUMENTS Ion 700 instruction manual

EUTECH INSTRUMENTS

EUTECH INSTRUMENTS Ion 700 instruction manual

Axis FA Series Operation manual

Axis

Axis FA Series Operation manual

Keysight M9391A quick start guide

Keysight

Keysight M9391A quick start guide

Endress+Hauser Levelflex FMP55 operating instructions

Endress+Hauser

Endress+Hauser Levelflex FMP55 operating instructions

horiba HEMAPREP RAB316AEN user manual

horiba

horiba HEMAPREP RAB316AEN user manual

Tegam 1316A quick start guide

Tegam

Tegam 1316A quick start guide

TESTO 340 instruction manual

TESTO

TESTO 340 instruction manual

janitza UMG 96 RM-E installation manual

janitza

janitza UMG 96 RM-E installation manual

Reed Instruments R5020 instruction manual

Reed Instruments

Reed Instruments R5020 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.