Minebea CSA-507C User manual

Transmitter
CSA-507C
Instruction Manual
EN294-1459-B


I
FOREWORD
Thank you very much for your purchasing our Transmitter CSA−507C.
This manual explains installation procedures and connecting method and also operating
method for Transmitter CSA−507C. Make use of it properly after reading through the
manual carefully.
Be sure to deliver the manual to the end user. Moreover, the end user should keep the
manual at hand after reading it over.

II
Marks and arrangements used in this manual
The following symbols are attached to the explanation on the matters that alert
“Do not do this.”, “Take care.” and “For reference”.
Be sure to read these items where these marks are attached.
Warning ●Precaution which must be followed to prevent the possibility of
accident or serious injury to the operators.
To avoid possible hazard, do not perform any procedure described
here.
●Precautions and important informations for safe operation and
service personnels.
Be sure to read the item to prevent from malfunction.
Mark during operation
●Press the switch.

III
For safe operation
Be sure to read this manual before use.
1. Installation place
●Use the instrument where the temperature/humidity specifies within
the range as follows:
Environmental temperature : −10 ℃〜50 ℃
Environmental humidity : Less than 85 %RH(Non condensing.)
(1) Location where installation is not allowed.
Warning ●Do not locate the instrument on the places such as follows:
It may cause an unexpected faulty in the instrument.
DDo not expose the instrument in direct sunlight and/or high temperature
area.
DDo not use the instrument in a high humid area.
DDo not install the instrument where there is high mechanical vibration or
shock.
DDo not use the instrument where there are excess of dusts and fine
particles.
DDo not install the instrument where there includes any corrosive gas or
any salty atmosphere.
DDo not install the instrument where there is rapid change on temperature
and humidity.
DDo not install the instrument near the devices that are magnetized or
generate an electromagnetic field.
DDo not install the instrument where there may suffer radioactivity or
radioactive rays.
DAvoid the location where chemical reaction may take place such as in a
laboratory.

IV
(2) Installation
●When installing the instrument, install as referring to the following
dimensions and secure the space around the instrument.
Each dimensions of the instrument and required dimensions for the environmental
spaces are as follows:
Dimensions for installation
Dimensions when the optional Auto zero unit(CSA507C−P99) is installed.
44 110
2−φ4.5
110
102
90
45
2−M4
Panel cut size
95
102
(9)
44 230
2−φ4.5
110
102
90
Unit:mm
(9)

V
2. Power supply
Warning ●Be sure to check that power supply is off when installing each cable.
If the work is done with the power fed, there may have the possibility
of electric shock to the operator or even the possibility of destroying
the instrument.
●Before supplying the power, check that the indication of power supply
voltage/specifications for the instrument and the power going to
supply should be the same. If they are not equal, contact us.
If you use the instrument without checking them, it may cause a
damage in the instrument or electric shock to the operator.
●Grounding wire should be connected securely. When grounding wire
is not connected, it may cause a malfunction of the instrument or
electric shock to the operator.
3. Application note
Warning ●Before using a new instrument or exchanging the strain gage applied
transducer for a new one, be sure to make calibration.
If calibration shall not be made, correct measured results may not be
obtained nor it may cause a malfunction in the instrument and there
may exist damage in peripheral equipments.
Besides, even though calibration had been made, there may occur the
similar case when the result is not correct, so make calibration again.
Warning ●In case of using the instrument, check that the connections are
executed properly. If not connected properly, correct measured result
will not be obtained, nor it may cause malfunctions of the instrument,
damage to the peripheral equipments or even worse serious accidents.
Warning ●When change of setting is made carelessly on the instrument during
measurement, it may cause malfunction in the instrument and also
possibility of damage in peripheral equipments.

VI
Warning ●Do not shock the instrument such as throwing something on it.
It may cause the damage to the case and also may cause the damage
to the instrument in resisting to environment.

VII
History of revision
Date
Instruction manual No.
Details of revised point
Sep. 2010
DRW. NO.EN294-1459
First Version
May. 2012
DRW. NO.EN294-1459-A
Duo.FN12−02078
Feb. 2018
DRW.No.EN294-1459-B
Due to ECN FN17-02017
・Delete the company name in the cover page.
・Delete the company name in the contents.


CONTENTS
FORWARD Ⅰ.......................................................
Marks and arrangements used in this manual Ⅱ......................
For safe operation Ⅲ................................................
1. Installation place Ⅲ.........................................................
2. Power supply Ⅴ.............................................................
3. Application note Ⅴ..........................................................
Record of revision Ⅶ.............................................................
1. General 1.................................................................
2. Each function and name 1..................................................
3. Installation procedure 2....................................................
3−1. Installation place 2....................................................
3−2. Location where installation is not allowed. 2..............................
3−3. Installation 3.........................................................
4. Connecting method 4......................................................
4−1. Layout of terminals 4..................................................
4−2. Note on connections 5..................................................
4−3. Connections 6.........................................................
4−3−1. Connection with strain gage applied transducers 6...................
4−3−2. Connection with external control input 8...........................
4−3−3. Connection with power supply and ground 9........................
5. Calibration procedures 10...................................................
5−1. Preparation 10.........................................................
5−2. Calibration method 10..................................................
5−2−1. Calibration with actual load 11.....................................
5−2−2. Electrical calibration with CALIB input. 12..........................
5−3. Zero adjustment by external resistance. 14................................
5−3−1. Installation of external resistance 14................................
5−3−2. Confirmation of initial value 16.....................................
6. Function and Operation 17..................................................
6−1. Setting of sensitivity 18.................................................
6−2. Setting of frequency response 18.........................................
6−3. Setting of bridge power supply voltage 19.................................
7. Option 20..................................................................
7−1. Current output (CSA507C−P07) 20...................................
7−2. Auto zero (CSA507C−P99)20.................................
7−2−1. Equivalent circuit of input section 20................................
7−2−2. Function 20......................................................
7−3. Power supply voltage AC200 V (CSA507C−P63)21.....................
7−3−1. Layout of the terminals 21.........................................
7−3−2. Connection with power supply and ground 22........................
8. Trouble shooting 23........................................................
8−1. Execute trouble shooting 24.............................................

9. Specifications 29...........................................................
9−1. Specifications 29........................................................
9−2. General specifications 29................................................
9−3. Standard specifications at the shipment 29................................
9−4. Accessories 30..........................................................
9−5. Options 30.............................................................
9−5−1. Current output 30................................................
9−5−2. Auto zero 30......................................................
9−5−3. Power supply voltage 30...........................................
9−6. Outlines 31............................................................
10. Warranty 32................................................................
10−1. Warranty 32...........................................................
10−2. Repair 32..............................................................
11. Appendix 33...............................................................
11−1. Replacement of fuse 33..................................................

1
1. General
The instrument is a transmitter for the application of strain gage applied transducers.
It obtains analog output corresponding to load, pressure, torque, and so on by amplifying output from
various kinds of strain gage applied transducers.
2. Each function and name
①
②
③
④
⑥
⑤
1“POWER” LED
Lights on when the power is turned on.
2“CAL” switch
Used when it will turn ON/OFF of the CALIB value.
3“ZERO” trimmer
Zero adjustment trimmer.
4“SPAN” trimmer
Amplitude adjustment trimmer.
5Terminals No.1
Terminals to connect with various kinds of strain gage applied transducers such as load cell,
and so on, and cable for voltage output signals.
6Terminals No.2
Terminals to connect with the cables for current output signal(Option), external control input
for “A/Z ON” and “A/Z OFF”, grounding wire and AC power cable.

2
3. Installation procedure
3−1. Installation place
●Use the instrument where the temperature/humidity specifies within
the range as follows:
Environmental temperature :−10 ℃〜50 ℃
Environmental humidity :Less than 85 %R.H.(Non condensing.)
3−2. Location where installation is not allowed.
Warning ●Do not locate the instrument on the places such as follows:
It may cause an unexpected faulty in the instrument.
DDo not expose the instrument in direct sunlight and/or high temperature area.
DDo not use the instrument in a high humid area.
DDo not install the instrument where there is high mechanical vibration or shock.
DDo not use the instrument where there are excess of dusts and fine particles.
DDo not install the instrument where there include any corrosive gas or any salty
atmosphere.
DDo not install the instrument where there is rapid change of temperature and humidity.
DDo not install the instrument near the devices that are magnetized or generate an
electromagnetic field.
DDo not install the instrument where there may suffer radioactivity or radioactive rays.
DAvoid the location where chemical reaction may take place such as in a laboratory.

3
3−3. Installation
●When installing the instrument, install as referring to the following
dimensions and secure the space around the instrument.
Each dimensions of the instrument and required dimensions for the environmental spaces are as
follows:
Dimensions for installation
Panel cut size
Unit:mm
44
2−φ4.5
110
102
110
90
45
2-M4
95
102
Dimensions when the optional Auto zero unit(CSA507C−P99) is installed.
230
90
Unit:mm
(9)
44
2−φ4.5
110
102

4
4. Connecting method
4−1. Layout of terminals
There are 2 pieces of terminals connector, one has 7 points of terminals connector(Terminals No.1)
and the other has 8 points of terminals connector(Terminals No.2).
Layout of the terminals are shown in the following figure.
DTerminals
Terminal name Description Application
A Bridge power supply(+)
1
B Amplifier input(−)
N
o.1
C Bridge power supply(−)Strain gage applied transducer
n
als
N
D Amplifier input(+)
r
min
a
E Shield
Ter
m
[V]OUT+Voltage output terminal(+)
V
oltage
output
T
[V]OUT−Voltage output terminal(−)
V
o
l
tage output
[I]OUT+Current output terminal(+)Current out
p
ut
[I]OUT−Current output terminal(−)
Current
output
(Option)
N
o.2
A/Z ON Auto zero control terminal
Et l
tl
it
a
ls N
o
A/Z OFF Auto zero control terminal External control input
(Option)
m
ina
l
A/Z COM Auto zero control terminal
(Option)
T
er
m
GND Grounding terminal
T
AC100 Power supply input terminal Power supply
AC100 Power supply input terminal
pp y
●“[V]OUT−” for voltage output terminal and “[I]OUT−” for current
output terminal are connected internally.

5
4−2. Note on connections
Warning ●Keep strictly to the following items in the connection with the
instrument. If neglected, it may cause an unexpected faulty or a
damage in the instrument.
DTurn the power supply OFF when connection is made.
DSince the terminals at the front side of the instrument is made of resin, take care not to
drop it down nor apply strong impact.
DRecommended torque to tighten the terminal screws for terminals is 0.8 N・m at maximum.
DThe crimp−type terminal lugs for the terminals of the instrument are to be applied as
follows:
Width of crimp−type terminal lugs Suitable crimp−type terminal lugs
6.0 mm or less 1.25−3 or Y type 1.25−3.5
DKeep the connecting cable with the instrument away from the noise source like power
supply line and I/O line for control as far as possible.
DConduit wiring should be the type of exclusive one, and avoid using with another line
together.
DAll of the connections should be executed securely according to the Instruction manual.

6
4−3. Connections
4−3−1. Connection with strain gage applied transducers
The instrument can connect with strain gage applied transducers, such as load cell, pressure
transducer and so on. Here, we will describe the example of connections with load cell, so
connect with another type of strain gage applied transducers in the same way.
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and output of “+”
direction is required, connect “Green” with B terminal, and “Blue”
with D terminal individually. As there is a case which standard
wiring color is different, please confirm the inspection data sheet of
the load cell being used.
※2 When the total length of CAB−502 specifies more than 30 m, the
accuracy may be out of warranty because the resistance of cable
makes the input voltage of the instrument decreased.
※3 When the length of CAB−502 cable is applied more than 10 m, or
when the system is using the zener barrier, the CALIB is not
applicable.
1Connection with one set of load cell and CSA−507C
3m RED
※1 GRN
WHT
(Shield)
※1BLU
YEL

7
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and output of “+”
direction is required, connect “Green” with B terminal, and “Blue”
with D terminal individually. As there is a case which standard
wiring color is different, please confirm the inspection data sheet of
the load cell being used.
※2 When the total length of CAB−502 specifies more than 30 m, the
accuracy may be out of warranty because the resistance of cable
makes the input voltage of the instrument decreased.
※3 When the length of CAB−502 cable is applied more than 10 m, or
when the system is using the zener barrier, the CALIB is not
applicable.
2Connection with one set of load cell and junction box for extension use B−304 and CSA−507C
Junction Box
B−304
CAB−502
※2(Length of CAB−502 should be
within 30 m totally.)
3m
RED
WHT
(Shield)
YEL
※1 GRN
※1BLU
B−304 internal terminal connection
to CSA−507C
to load cell
RED WHTBLU GRN
YEL(Shield)
Terminal pitch 9.5 mm
Suitable crimp−type
terminal lugs:1.25−4or2−4
RED WHTBLU GRN
YEL(Shield)

8
4−3−2. Connection with external control input
Connection of external control input of “A/Z ON”, “A/Z OFF” should be made by using contact
or open collector as below figure.(Effective only when optional Auto zero unit(CSA507C−P99)
is installed.)
Refer to the paragraph 7−2 for each input function.
or
Shield
Warning ●Connection with external control input should be made securely as the
figure indicated. If neglected, it may cause an unexpected trouble or
malfunction on the instrument.
●For the connection with external control input, be sure to apply
shielded cable, and that shield should be connected with GND
terminal. If not connected, it may cause malfunction due to the effect
from external noises and so on.
Table of contents
Other Minebea Transmitter manuals

Minebea
Minebea CSA-524 User manual

Minebea
Minebea CSA-528-80-4 User manual

Minebea
Minebea CSA-522B User manual

Minebea
Minebea OPT-564 User manual

Minebea
Minebea CSA-521 User manual

Minebea
Minebea NS100A User manual

Minebea
Minebea OPT-564-71 User manual

Minebea
Minebea OPT-563B User manual

Minebea
Minebea OPT-563B-71 User manual

Minebea
Minebea CSA-593 User manual