
VS300 Troubleshooting Manual Responding to alarms
Issue 10.0 2019-01-18 Page 1-11
+3.3V Supply
Fail
(+3.3V Fail)
PS (red) This alarm occurs if the +3.3 V supply on the exciter/control PWB
(A1) is outside of its acceptable voltage range (between +3.0 V
and +3.6 V). The VS300 takes no action on this alarm. Check for a
+5V Supply Fail alarm: If present, follow the associated
troubleshooting procedure. If not present, use a digital multimeter
to measure between TP1 (+3.3 V) and TP4 (ground) of the
exciter/control PWB. If the measured value is within the
acceptable range, suspect the sampling circuitry on the exciter/
control PWB. If not, use a digital multimeter to measure between
TP5 and TP4 (ground). The measured voltage should be between
+4.5 V and +5.5 V. If the measured voltage is within the
acceptable range, the power supply circuitry on the exciter/control
PWB has failed. Replace the exciter/control PWB (see “Exciter/
Control PWB Replacement” on page 1-41). If not, check ribbon
cable W4 for damage and that connectors W4P1 and W4P2 are
properly seated in A2J3 and A1J11 respectively. If there are no
visible problems with the ribbon cable, contact Nautel for further
support.
+5V Supply
Fail
(+5V Fail)
PS (red) This alarm occurs if the +5 V output of the low voltage power
supply (LVPS) (U3) is outside of its acceptable voltage range
(between +4.5 V and +5.5 V). Use a digital multimeter to measure
between +5V and COM of the LVPS. If the measured value is not
within the acceptable range, replace the LVPS (see “LVPS
Replacement” on page 1-38). If the measured value is within the
acceptable range, use a digital mutimeter to measure the voltage
between TP12 (+5 V) and TP7 (ground) of the PS distribution
PWB (A2). If the measured value is not within the acceptable
range, turn off ac power and use a digital multimeter to perform a
continuity check across fuse F2 of the PS distribution PWB. If the
measurement is greater than 1 , replace F2 (Nautel Part # FA57
in the ancillary kit), restore ac power and check the alarm again. If
the measurement is less than 1 , or replacing F2 does not clear
the alarm, suspect the connection between the +48 V power
supply and the PS distribution PWB, and contact Nautel for further
assistance. If the measured value is within the acceptable range,
check ribbon cable W2 for damage and that W2P1 and W2P2 are
properly seated in A1J12 and A2J2 respectively. If there are no
visible problems with the ribbon cable, suspect the sampling
circuitry on the PS distribution PWB. If necessary, replace the PS
distribution PWB (see “PS Distribution PWB Replacement” on
page 1-44) . If the alarm persists after replacing the PS
distribution PWB, suspect the exciter/control PWB (A1).
Alarm Name
AUI and
(Front Panel)
Front Panel
LED (color) Description and Troubleshooting Action