Morehouse 4215-HS Guide

ISO/IEC 17025 / ANSI/NCSLI Z540.3 Accredited
Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
P: (717) 843-0081
F: (717) 846-4193
info@mhforce.com
www.mhforce.com
PM-4101
Rev. 3/2022
Operation and Instruction Manual
4215-HS Indicator

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 2
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Table of Contents
Introducon .................................................................................3
Quick Start...................................................................................4
Run Mode ...................................................................................6
Setup Mode Summary .........................................................................9
Remote Operaon ...........................................................................19
RS232 Run Mode Command Summary...........................................................20
Rs232 Setup Mode Command Summary .........................................................25
System Calibraon Mode ......................................................................43
Force Averaging (Oponal) ....................................................................43
Appendix A - 4215-HS Specicaons ............................................................45
Appendix B - Cables and Connectors ............................................................47

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 3
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Introducon
The Model 4215-HS is a versale precision instrument intended for the digital readout of strain gage sensors
such as load cells. Opcal encoders are also supported.
Features:
* Full bi-polar 6 digit display (±999,999) using two line, 20 character, vacuum uorescent display for clear, pre-
cise indicaon of measured quanes and limit status.
* No knobs or dials. No internal jumpers. All adjustment and calibraon is done using user-friendly front panel
menus. Keylock opon to prevent accidental or unauthorized changes to the unit’s setup.
* Measure load in pounds, kilograms, Newtons, PSI, MegaPascals, kilo-pounds, kiloNewtons, metric tons,
mV/V, or grams. User dened base area for PSI and MPa. Tare at any me.
* Capture the load peak and valley, display in pounds, kilograms, Newtons, PSI, MegaPascals, kilo-pounds,
kilo-Newtons, metric tons, or mV/V. Reset at any me.
* Measure posion and velocity with oponal encoder input. Display in English units, metric units, or in per-
cent of a user dened base length. Reset at any me.
* Four contact closure Limit switches. You can set which signal drives the limit (load, peak, valley, posion,
or velocity), normally open / normally closed, set point, and manual reset (latched) or automac reset (not
latched).

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 4
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
* Print current readings of load, peak, valley, posion, and velocity using the currently selected units at the
touch of one buon.
* Automac idencaon of calibrated load cells with auto-id opon.
* Calibraon for load cells using manufacturer’s milli-volt per volt calibraon factor, 5 point milli-volt per volt,
2 or 5 point known mass, or internal precision shunt calibraon. A back panel switch allows you to choose be-
tween two values used for the shunt calibraon so there is no need to open the unit.
* Two Channels of Analog output of any signal (load, peak, valley, posion, velocity) with user dened scale
factor and oset. Large, buered ± 10.00 volt output.
* Remote operaon using standard RS232 ASCII command set.
* Back panel digital inputs for tare, peak and valley reset, posion reset, and print. Other funcons available as
custom programmed opons.
* Oponally measure the average force using a contact closure to begin and end the averaging period.
This secon will help you get your Model 4215-HS set up and operang in just a couple of minutes. The back
panel is shown below. You will need to make at least two connecons here. First be sure that the power switch
is o (the lower part of the rocker should be pressed in). Connect the AC power line and plug it in.
Next connect a load cell to the 9 pin female connector labeled “Sensor A”. You should use the cabling supplied
with your Model 4215-HS and load cell. If you do not have a factory supplied cable, see Appendix B for the
recommended load cell connecon.
Model 4215-HS Back Panel
Quick Start

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 5
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
You are now ready to turn on the power. Each me power is applied to the Model 4215-HS you will see a se-
quence of 6 messages appear on the front display. For example,
Each of these messages will be displayed for about 3 seconds. Aer that the front display will start showing the
current readings from the load cell and sensor.
To change what item is being displayed press the lemost buon. This will cycle through the list of items (Load
A, Peak A, Valley A, Load B, Peak B, Valley B, Posion, Velocity, Limits, oponally: Avg A, Avg B).
The rightmost buon will change the units that are being used. Repeated presses of this buon will cycle
through the list of units available for the currently displayed item. For example, with Load A showing you
would see Lb, kg, N, PSI, MPa, Klb, kN, t, mV/V, g and then back to Lb.
# Message Meaning
1 ESL .X.Y.Z This is the version number.
2 Serial # 12345
Option # 114236
This is the serial number.
This is the option number.
3Last ID A 90437
Last ID B 12030
The serial number of the load cell on Ch A.
The serial number of the sensor on Ch B
4Ch A Max 4.00Lb Ch B
Max 10.00In
Maximum rated load of the sensor on channel A.
Maximum rating of the sensor on channel B.
5
Ch A Cal 4.00000 mVv
Ch B Cal 4.00000 mVv
The mV/V constant used with the load cell on channel A.
The mV/V constant used with the sensor on channel B.
Note: If either of these sensors was calibrated by 5 –Point Mass, then 4 mV/V constants will be displayed here.
6Ch A Exc 5.00 Volt
Ch B Exc 2.50 Volt
Excitation voltage used with the load cell on channel A.
Excitation voltage used with the sensor on channel B.

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 6
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
The Model 4215-HS has three modes of operaon, Run mode, Setup mode, and System Calibraon mode. At
power-on it will be in Run mode.
The rst line of the display show an item label (Load A), a value (0.0000), and a unit label (Lb).
The second line of the display can show the status of the 4 contact closure limit switches, show another dis-
play, show text, or be blank.
To change what item is being displayed press the lemost buon. This will cycle through the list of items (Load
A, Peak A, Vall A, Load B, Peak B, Vall B, Pos, Vel, and Limits, oponally Avg A, Avg B).
The rightmost buon will change the units that are being used. Repeated presses of this buon will cycle
through the list of units available for the currently displayed item. For example, with Load showing you would
see Lb, kg, N, PSI, MPa, Klb, kN, t, mV/V, g and then back to Lb.
The table below summarizes the opons.
All of these measurements are acquired by the Model 4215-HS at all mes. The display just selects which par-
cular item you wish to see and what units are used to display or print this data.
The row of buons have the funcons indicated by the upper set of labels (Item, Tare, Reset, View, Print, Unit).
Each of these is described below.
Message Meaning
Load A Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Peak A Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Valley A Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Load B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Peak B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Valley B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Position Inches, Centimeters, Percent of Base Length.
Velocity Inches per Minute, Centimeters per Minute.
Limits Shows state of each switch, 1 for on, 0 for o, dash for disabled, asterisk for source error.
Avg A Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Avg B Pounds, Kilograms, Newtons, PSI, MegaPascals, Kilo-Pounds, Kilo-Newtons, Metric Tons, mV/V, grams.
Run Mode

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 7
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Item Buon
The Item buon cycles through the list of dierent items that can be displayed (Load A, Peak A, Vall A, Load B,
Peak B, Vall B, Pos, Vel, Limits, and oponal Avg A and Avg B). Use the Unit buon to change the units that are
used for the currently displayed item.
Tare Buon
The Tare buon sets the current load to zero for channel A, channel B, or channels A and B. A submenu will
allow you to select what channel you want to tare. It does not maer what is being displayed when you press
this buon. The load is always reset to zero. A single channel unit will automacally set Load A to zero.
Reset Buon
The Reset buon brings up a list of things that you might want to set to zero:
If this is a 2-channel unit, the following menu will appear:
> Reset Ch A + Ch B will reset the peak, valley, max, min for channels A and B.
> Reset Ch A Only will reset the peak and valley for channel A.
> Reset Ch B Only will reset the max and min for channel B.
> Reset Posion will reset the posion to zero.
> Reset Limits will reset any limits that require manual reseng.
For a single channel unit, the following menu will appear:
> Reset Peak+Vall will reset the peak and valley measurements.
> Reset Peak Only will reset the peak measurement.
> Reset Vall Only will reset the valley measurement.
> Reset Posion will reset the posion to zero.
> Reset Limits will reset any limits that require manual reseng.
Pressing either the “+” buon or the “>“ buon will scroll forward through this list. Pressing either the “-” but-
ton or the “<“ buon will scroll backward through the list. Pressing “Enter” will reset the currently displayed
item.
Pressing “Esc” will return to Run Mode without reseng anything. Note that all menu lists on the Model
4215-HS start with the “>“ symbol and that the Esc buon will always back out of a menu without making any
changes.

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 8
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
View Buon
The View buon switches between two virtual displays stored in the memory. Each of these displays will re-
member the units you have selected for each item, and which item is currently being shown. This may save you
from scrolling through the list of items if there are two parcular results that you oen use. For instance you
might set the display to show Load in Lb, press View, and then set the display for Peak in PSI. Repeated press-
es of the View buon will now switch back and forth between Load in Lb and Peak in PSI with a single buon
push. If you choose to have the second line show a display, then the View buon toggles between the display
currently being acvated. The acvated display will have a “>” symbol to the le of it.
Print Buon
T The Print buon will send a set of data to the printer. This will contain the current readings of load, peak, val-
ley, posion, velocity, and limit status. The units used for prinng will be the units last selected on the numeri-
cal display.
The printer must be a serial printer that supports ESC/P protocol. Many Epson printers with serial ports are
known to be compable.
Appendix B gives the pin assignments on the serial connector.
Unit Buon
The Unit buon cycles through the list of units that are available for the item that is currently being displayed.

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 9
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
The table below summarizes the setup mode menus. Detailed information about each individual item is
presented in the following sections.
Setup Mode Summary

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 10
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Setup Mode
Setup mode is used to change the setup of the Model 4215-HS. To enter setup mode press the two end but-
tons at the same me ( “<“ and “>“ ). The row of four buons will now funcon as indicated by the lower set of
labels (Plus, Minus, Enter, and Escape).
If the oponal keylock is in the locked posion you will see “>Setup is Locked” when you try to enter setup
mode. This prevents accidental or unauthorized changes to the setup of the unit.
To exit from setup mode and return to run mode press the escape buon (Esc). When you are entering numer-
ical data or selecng opons from a sub-menu the escape buon will back you out without making any chang-
es.
The main menu in setup mode contains the following 6 entries which will be described in detail in the follow-
ing secons. Note that all main menu items begin with the “>“ symbol and that all submenu items begin with
the “>>“ symbol.
• > User Data Entry -- allows you to set any common math opons such as the base area used to display load
in PSI or the base length used to display posion in percent.
• > Analog Output -- is used to select the analog channel, source, and scaling for the analog output channels.
• > Sensor Select -- lets you manually set which load cell calibraon data is used. If you use auto-id load cells
you will never need this menu.
• > Calibraon -- provides access to calibraon for load cells, extension sensor, or opcal encoders.
• > Limits -- allows you to set the source and opons for the 4 contact closure switches.
• > System Opons -- contains entries to set the printer baud rate, enable or disable automac sensor iden-
caon, enable or disable automac load zeroing, and to set the serial communicaons address, baud rate, and
line feed opons.
• > Display Opons -- lets you choose the ltering that is applied to the load measurements, the number of
decimal digits that are shown on the display and on the print, the count by values for channels A and B (1, 2,
5, 10, or 20), and whether the second line of the display shows the limit switch status, a blank line, another
display, or text.
The Plus or Right buon ( “+” or “>“ ) will scroll forward through this menu. The Minus or Le buon ( “-” or
“<“ ) will scroll backward through this menu. The Enter buon selects the currently displayed menu item. The
Esc buon will return you to Run Mode.
Entering Numerical Data
At mes you will need to enter numerical data into the Model 4215-HS. The method is consistent in all cases. A
number is presented on the le side of the display with an underline cursor.

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 11
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Entering Numerical Data
At mes you will need to enter numerical data into the Model 4215-HS. The method is consistent in all cases. A
number is presented on the le side of the display with an underline cursor.
• Le ( “<“ ) moves the cursor le one posion. If you are on the lemost digit a new zero will be added. Thus
10.000 will become 010.00 where the cursor has moved from the tens place to the hundreds place.
• Right ( “>“ ) moves the cursor right one posion. If you are at the far right a new decimal will be added if
there is room. Thus 12.34 will become 12.340 where the cursor has moved from the tenths to the hundredths
place.
• Plus ( “+” ) increments the digit above the cursor. Thus 12.34 becomes 13.34. Note that 9 will “wrap around”
to 0.
• Minus ( “-” ) decrements the digit above the cursor. Thus 12.34 becomes 11.34. Note that 0 will “wrap
around” to 9.
• Plus and Minus (both “+” and “-” at the same me). This will change the sign of the number. Thus 12.34 be-
comes -12.34. Note that cursor posion does not maer when you change the sign.
• Enter accepts the currently displayed number.
• Esc escapes from numerical entry and discards any changes that have been made.
>User Data Entry
This menu is used to enter numerical constants used by the math opons. These constants are listed in a sub-
menu that works exactly like the main setup mode menu. Typical items in this submenu are:
>>Base Area Ch A
This is the area in square inches that is used to calculate load in pounds per square inch (PSI) or Mega Pascals
(MPa).
>>Base Length
This is the length in inches that is used to calculate posion in percent.

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 12
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
>Analog Output Channels
This menu is used to set up the analog outputs. Instead of a submenu there is a series of selecons. Pressing
escape at any me will leave the analog output unchanged and return you to the setup mode main menu.
1. Channel: -- This entry allows you to select the analog output channel A or B. (use either the le/right or
plus/minus buons) for the selected source data that will drive the analog output. Press enter to select the
displayed Analog channel.
2. Source: -- This entry allows you to scroll through a list (using the le/right or plus/minus buons) to select
the source of the data that will drive the analog output. The standard list will include load, peak, valley, posi-
on, and velocity. Press enter to select the displayed item.
3. Units: -- This entry allows you to scroll through the list of available units for the data source you have chosen
in the previous step.
4. Full -- This entry species the value of source data (in the selected units) that will cause the Model 4215-HS
to output posive full scale analog voltage (+10.0 volts).
5. Zero -- This entry species the value of source data (in the selected units) that will cause the Model 4215-HS
to output zero analog voltage (0.0 volts).
Once you have entered these 4 selecons the analog output is setup.
>Sensor Select
This menu is used to manually select load cell calibraon data from a list stored in the Model 4215-HS. When
you select this menu on a 2 channel unit, you are asked to select Channel A or B. The load cell serial number of
the selected channel is displayed. Use either the le/right or plus/minus buons to page through the list. The
escape buon will return to the setup mode main menu without changing the load cell that is selected. The
enter buon will select the currently displayed load cell and briey display the rated load, last calibraon date,
and excitaon voltage.
To add new load cells to this list use the Calibraon entry on the setup mode main menu described below.
If you are using auto-id load cells this menu will allow you to see which load cell is selected but will not allow
you to change the selecon. This ensures that the correct calibraon informaonis always used when you use
Auto-ID load cells.

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 13
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Sensor Delete
The Model 4215-HS can store calibraon data for 4 dierent load cells / sensors (25 slots oponal). To delete
a load cell from this list choose “>Sensor Select” from the setup mode main menu. Press the le/right or plus/
minus keys unl the serial number of the cell you wish to delete is displayed. (You may need to shut o auto-id
if it is on, see the System Opons menu for details). Then press both the plus and minus keys at the same me.
The Model 4215-HS will ask “Delete Sensor?” to be sure that you intend to delete this calibraon data. Press-
ing Enter will delete the cell data from the list. Pressing Escape will return to the load cell select without chang-
ing the list.
>Calibraon
This menu is used to access all methods of calibraon that are supported by the Model 4215-HS. Selecng cal-
ibraon from the setup mode main menu will bring up a submenu. This submenu is used exactly like the main
menu. Press le/right or plus/minus to change the menu item. Press escape to return to the setup mode main
menu. Press enter to select the displayed menu item.
If you are re-calibrang a load cell, the Model 4215-HS will recognize it using either the autoid or a manually
entered serial number. Any me you overwrite exisng load cell calibraon data you will rst be warned with
the “Overwrite Data?” message. Pressing enter will overwrite; pressing escape will cancel the calibraon. An-
other method of overwring exisng data is to delete the sensor rst and then calibrate.
If a load cell with a new serial number (or new auto-id) is calibrated the data is added as a new entry in the
load cell list.
>>Cal Check
Selecng this item will run a calibraon check on the load cell / sensor aached to Channel A or B. The Model
4215-HS will measure the shunt value of the load cell and compare it to the shunt value that was recorded
when the load cell was last calibrated. The rated load and last calibraon date for the load cell are displayed
followed by the currently measured shunt value and the shunt value that was recorded when the load cell was
calibrated.
Note that the Model 4215-HS has a switch on the back panel that allows you to select either a 30K shunt or a
60K shunt. Be sure that this switch is set correctly for the load cell that you are using.
Since this shunt check electronically applies approximately half the rated load, it is important that the load cell
/ sensor is not loaded when this check is run. If you see the “Cell Overloaded” message, remove any load on
the cell and try the check again.

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 14
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
>>Cal by mV/volt
This item is used to calibrate a cell using the manufacturer’s milli-volt per volt calibraon factor or a series
of milli-volt per volt points with corresponding masses (usually listed on the load cell calibraon cercate).
Before you enter in the desired informaon, another menu will ask if you want to calibrate by milli-volt per
volt or 5 Point milli-volt per volt. You will be asked for the channel, the cell serial number, the current date,
the calibraon unit, the cell’s maximum rated load, and the desired excitaon voltage (5.0 or 10.0 volts). If
milli-volt per volt calibraon was selected, you are then prompted to enter the manufacturer’s milli-volt value.
If 5 point milli-volt per volt calibraon was selected, you are prompted for each mass and milli-volt per volt
value. One of these points should be zero mass and zero milli-volts per volt. Enter each of these selecons or
press the escape buon at any me to return to the setup mode main menu without changing the calibraon.
The cell should not be under load when this calibraon method is used since a shunt value will be measured
and recorded for later use with the “Cal Check” feature described above. The “No Masses Ready” message is
intended to remind you of this fact.
>> Cal by Masses
This item is used to calibrate a load cell, channel A, using two known masses. Before you enter in the desired
informaon, another menu will ask if you want to calibrate by 2 Point, 5 Point, or 10 Point (Non standard
oponal conguraon) You will then be asked for the channel, the load cell serial number, the date, the cali-
braon unit, the cell’s maximum rated load, and the desired excitaon voltage (5.0 or 10.0 volts). You are then
prompted to enter the value of the rst mass. The “Mass 1 Ready?” message will wait while you actually hang
the mass. When the Model 4215-HS is reading the load cell it will display the “Reading...” message for about 10
seconds. During this me, it is important that the masses are not disturbed since all the readings taken are av-
eraged together to obtain the calibraon data. The second mass is then entered and read. If 5 Point calibraon
was selected, the process will connue unl all masses have been read. A “No Masses” reading is also required
during which me a shunt check value is recorded.
>>Cal by Shunt Val
This item is used to calibrate a load cell using the manufacturer’s shunt calibraon factor expressed in pounds.
You will be asked for the channel, the load cell serial number, the date, the calibraon unit, the cell’s maximum
rated load, the desired excitaon voltage (5.0 or 10.0 volts), and the shunt calibraon constant. The Model
4215-HS will then automacally perform the shunt calibraon of the load cell using its internal precision shunt
resistor. Note that there is a back panel switch that allows you to select either a 30K or 60K resistor for shunt
measurements (see Appendix B for a circuit diagram). Be sure to select the appropriate value for your load cell.
During this calibraon the “Reading...” message will be displayed. Do not disturb the load cell while it is being
read. Like the “Cal by mV/volt” opon, the load cell should be free of load during this calibraon.

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 15
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
>>Counts per Inch
This entry allows you to set the number of counts per inch that the quadrature encoder input will use.
>Limits
This menu is used to set up the 4 contact closure switches. Instead of a submenu there are a series of selec-
ons. Pressing Enter will accept the displayed entry. Pressing Escape will return to the setup mode main menu
without changing the limit setup.
1. Select Limit X -- The rst selecon chooses which of the limits you want to set up (1-4). Use the le/right or
plus/minus buons to change the number. Press enter to accept the displayed number. Press escape to return
to the setup mode main menu.
2. Normally Open / Closed -- This selecon determines if the switch contacts are open or closed when the limit
is de-acvated.
3. Enabled / Disabled -- Choosing enable allows the limit to operate as specied by the following selecons.
Choosing disable places the limit in its “normal” state and completes the limit setup.
4. Source: -- This selecon allows you to scroll through a list (use either the le/right or plus/minus buons) to
select the source of the data that will drive the limit. The standard list will include load, peak, valley, posion,
velocity and, oponally, average force. Press enter to select the displayed item.
5. Units: -- This entry allows you to scroll through the list of available units for the data source you have chosen
in the previous step.
6. Set -- This is the numerical value of the source at which the limit will be acvated (set point). It is specied
in the units you chose in the previous step.
7. Trip if source < / > Set -- Choosing source>set means that the limit will be acvated when the source be-
comes larger than the set point. Choosing source<set means the limit will be acvated when the source be-
comes less than the set point.
8. Latching on/o -- Selecng latching o means that the limit will be de-acvated automacally when the
source passes the reset point (see next selecon). Selecng latching on means that once the limit is acvated
you must manually reset it using the Reset buon. If you select latching on then the limit setup is done.
9. Rst -- This selecon determines the numerical value of the source at which the limit will be deacvated (re-
set point). If you selected “source>set” in step 7 then the limit is deacvated when the source is less than the
reset point. If you selected “source<set” in step 7 then the limit is deacvated when the source is greater than
the reset point

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 16
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Notes:
The Set and Reset condions for each limit are evaluated independently. If both condions are met then Reset
will take priority over Set.
Limit status may show an asterisk if the limit is enabled but there is an error in the signal source. For example,
if the limit source is load but there are no load cells currently calibrated, or the load cell has a rang of 0.0 Lb,
or a millivolt per volt constant of zero. Another example would be if the limit source is posion and zero has
been entered for the encoder counts-per-inch constant. Limits that are in this error condion will be forced to
the o posion as if they were disabled.
>System Opons
This menu contains some system wide parameters that are seldom changed. These are listed in a submenu
that works exactly like the setup mode main menu.
>>Printer Baud
The front panel print buon and the RS232 print commands will use this baud rate for communicaon with
your printer. This seng has no eect on units that are not equipped with the prinng opon. 57.6 K baud is
commonly used on older serial printers 230.4 K baud is used on more recent serial printers.
>>Auto Idenfy / TEDS Capability
This allows you to turn automac load cell idencaon on or o. Usually you can leave auto-id on even if
you never use automac load cell idencaon. In order to use this funcon you must either have an au-
to-id equipped load cell or use an in-line auto-id module aached to your load cell cable (TEDS). Meters can
be equiped with TEDS Template 40 and 41 read & write capability. Template 40 and 41 are avaialble in the
“4215-HS-Enahanced Model”.
>>Auto Zero Ch A/B
This menu lets you enable or disable auto zeroing for load measurements on Channel A or B. When enabled
this funcon will tare the load to zero any me that it has remained within 10 counts of zero on the display for
10 seconds..
>>Com Address
This item allows you to set the address of this unit used for the RS232 ASCII command set. Valid values are 1
through 254. If the RS232 mul-drop opon is installed you must be sure that all the aached units have dier-
ent addresses to prevent data collisions.

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 17
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
>>Com Baudrate
This sets the baud rate for the RS232 ASCII command set. Available values are: 300, 600, 1200, 2400, 4800,
9600, and 19.2k. Note that the “Special” entry is reserved for future use.
>>Com Line Feed
This controls whether a line feed is added to each carriage return that is sent out by the Model 4215-HS when
the RS232 ASCII command set is being used.
>>Retain Tare
This controls whether the most recent tare value is retained through power down of the unit and used again
immediately on power up. Choosing this opon will retain tare on channel A in a single channel unit and both
channels A and B in a twochannel unit.
>>Display Opons
This menu contains opons that eect how load, peak, and valley are displayed and printed. These are listed in
a submenu that works exactly like the setup mode main menu.
>> Filter Select
There are two types of lters available on the Model 4215-HS: Type I, and Type II. Type I lters are good for
removing most kinds of noise but may leave some jier on the end digits. Type II are more advanced lters
that are opmized for the typical industrial environment. Both lter types allow you to select from 4 levels of
ltering with Level 1 being the least ltering and Level 4 being the most ltering.
The seling me for each of the 4 levels of ltering is shown in the table below.
Level 1 2 2 4
Setting Time (seconds) <1 2 10 30
Read Rate (per second) 60 60 30 10

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 18
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
>>Decimals Ch A/B
This lets you set the maximum number of digits to the right of the decimal that are displayed or printed for
load, peak, and valley. When large load cells are selected the number of decimal digits shown may be limited,
for example a 100 Lb cell will allow 3 digits, while a 1000 Lb cell will only allow 2 digits. Also if a number is too
large to t on the display some decimal digits may be dropped (the display will “auto-range”). The number of
digits shown for posion data depends only on the number of counts per inch of the encoder and is not eect-
ed by this seng.
>>Second Line
This opon allows you to choose what you want to have displayed on the second line. You can select limits,
select a display, select text, or leave the line blank.

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 19
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
Digital Inputs
There are 4 digital inputs available on the Model 4215-HS. They are accessible on the Digital I/O connector on
the back panel (pins 9 - 16). These inputs are individually optoisolated and include current liming resistors. An
input voltage anywhere from +4 to +22 volts DC may be used to obtain the “on” state. See Appendix B for a full
descripon of the Digital I/O connector. If you do not require isolaon for your applicaon you can use the +5
or +12 volt and GND available on the Digital I/O connector (pins 24-26).
These 4 inputs are called IN1 through IN4 and have the following standard funcons:
• IN1 will Tare the Load to zero while it is in the On state.
• IN2 will Reset the Peak and Valley measurements while it is in the On state.
• IN3 will Reset the Posion measurement while it is in the On state.
• IN4 will send one set of measurements (load, peak, valley, posion, velocity, and limits) to the serial printer
when it transions from O to On.
Other funcons may be assigned to these inputs by custom programming at the factory.
RS232 Operaon
All the features and sengs of the Model 4215-HS can be accessed remotely using the RS232 ASCII command
set. These commands fall into two groups: commands that are like Run Mode, and commands are like Setup
Mode.
The Model 4215-HS supports remote operaon using a standard RS232 Interface consisng of ASCII characters.
The baud rate is set using the Com Baudrate item on the System Opons menu. Supported rates are 300, 600,
1200, 2400, 4800, 9600, and 19.2k baud. There must be 8 data bits, no parity, and 1 stop bit. To support the
mul-drop RS232 opon each unit is assigned a communicaons address which is set using the Com Address
item on the System Opons menu. All the commands have the following format:
@123XYZ<CR> The @ symbol iniates the command, all commands must begin with this symbol. The number
123 is the address of the Model 4215-HS which should respond to this command. This MUST be 3 digits so
type 001 for unit one, and 026 for unit twenty-six. The command itself will be a sequence of leers such as the
XYZ shown above. If the command requires a number as one of its parameters the number must be followed
by the pound sign “#”. Numbers that are less than one should include a leading zero so you should type 0.123#
to enter one-hundred twenty three thousandths. Commands are not processed unl a carriage return <CR> is
sent.
No Model 4215-HS will respond to address 000. All Model 4215-HS’s will respond to address 255. Throughout
the rest of this secon we will assume that you are using address 123. Whenever an Model 4215-HS executes a
remote command it will issue an acknowledge that includes the address and any required data.
Remote Operaon

Morehouse Instrument Company, Inc.
1742 Sixth Ave., York, PA 17403-2675 USA
Phone: (717) 843-0081
www.mhforce.com
Page 20
Rev. 3/2022
4215-HS Indicator
Manual
(PM-4101)
RS232 Run Mode Commands
The commands listed in this secon mirror (and in some cases extend) the funcons that are available from
the front panel of the Model 4215-HS when it is in Run Mode.
Each entry will include the name of the command, followed by a short descripon that includes the command
format, an example that shows what you would type (shown in plain font) and what the Model 4215-HS would
return (shown in italics).
Hello Command (H)
This command fetches the model, version, and serial number of any unit at the specied address.
Example:
@123H
Acknowledge:
@123 ESL Model 4215-HS Version 4.4.0 Serial #-980500
Queson Mark Command (?)
This command returns a list of item numbers and unit numbers that are available on the Model 4215-HS that is
addressed. These numbers are used as arguments for the many of the commands shown below.
Example:
@123?
RS232 Run Mode Command Summary
This manual suits for next models
1
Table of contents
Other Morehouse Measuring Instrument manuals