MRC PDT-2250 User manual

Your purchase of this RPM CONTROLLER/MONITOR marks
a step forward for you into the field of precision
measurement. Although this CONTROLLER is a complex
and delicate instrument, its durable structure developed.
Please read the following instructions carefully and
always keep this manual within easy reach.
OPERATION MANUAL
RPM CONTROLLER
MONITOR
Model : PDT-2250
PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
MRC.11.20

Caution Symbol
Caution :
* Risk of electric shock !
Caution :
* Do not use fingers or any tool
to touch the Wire Terminals.
* Do not apply the relay contact load
current > 0.5 Amp.
* The instrument contains no user
serviceable parts and should not
be opened by the user.
* Repair or after service should be
done by a qualified technician only.
* Power supply should apply the
correct ACV power voltage
*Cleaning - Only use the dry cloth to
clean the plastic case !
*Equipment protected throughout
by Double Insulation or Reinforced
Insulation.
Environmental Condition
* Comply with EN61010.
Transient overvoltage at Mains Supply 2500V.
* Pollution Degree 2.
* Altitude up to 2000 meters.
* Indoor use.
* Relative humidity 80% max.

TABLE OF CONTENTS
1.FEATURES.................................................................1
2.SPECIFICATIONS.......................................................1
2-1 General Specifications..........................................
.
1
2-2 Electrical Specifications........................................
.
3
3.FRONT PANEL DESCRIPTION.....................................
.
4
3-1 Display................................................................4
3-2 PV ( process value ) indicator...............................
.
4
3-3 SV ( set value ) indicator......................................4
3-4 Set Button...........................................................4
3-5 Button............................................................
.
▼4
3-6 Button............................................................
.
▲4
3-7 Function Button................................................... 4
3-8 Control relay indicator..........................................4
3-9 Alarm relay indicator............................................4
3-10 RPM indicator....................................................
.
4
3-11 ft/min indicator..................................................4
3-12 m/min indicator................................................. 4
3-13 Wire terminals...................................................
.
4
3-14 Case holder ......................................................
.
4
3-15 RS232 terminal.................................................. 4
4.MEASURING PROCEDURE..........................................
.
5
4-1 Terminal connection.............................................5
4-2 RPM measurement...............................................6
4-3 Surface speed ( m/min., ft/min. )
measurement......................................................6
4-4 1st layer setting procedures.................................
.
8
4-5 2nd layer setting procedures................................
.
10
5. RS232 PC SERIAL INTERFACE....................................14
6. SYSTEM RESET.........................................................15
7. OPTIONAL PROXIMITY SENSOR, PX-01 .....................16
8.
T
HE ADDRESS OF AFTER SERVICE CENTER................
.
17

1. FEATURES
* Professional Tachometer ( RPM, m/min., ft/min. )
monitor and controller.
* Build in control relay and the alarm relay.
* Alarm Relay will make action when the reading value
reach to high/low alarm value.
* Control Relay will make action when the reading value
reach to control value.
* Hysteresis value setting for control and alarm function.
* Can set input pulse no. of each round.
* Can set the roller's diameter for the surface speed
( m/min., ft/min. ) function.
*Can cooperate the optional proximity sensor or photo
sensor.
*Large red LED display, high brightness and easy to read.
* RS232 computer interface.
* Microprocessor circuit ensures high accuracy and
provides special functions and features.
* Standard 96 X 48 mm DIN case.
* Optional data acquisition software.
* Optional GSM controller.
2. SPECIFICATIONS
2-1 General Specifications
Display 4 digits red LED, digit size : 14 mm.
Unit RPM RPM
Surface m/min ( meter per min. )
speed ft/min ( feet per min. )
Circuit Custom chip of microprocessor LSI
circuit.
1

Sampling Time Approx. 1 second.
* Input : one pulse per round, 60 RPM≧
Relay Output Number 2 relays
Function
Relay 1 :
Control relay.
Relay 2 :
High/Low alarm relay.
Max load 0.5 ACA/250 ACV
0.5 DCA/24 DCV
*
Do not apply the rela
y
contact load current
> 0.5 A, other wise th
e
relay may be dama
g
ed
permanently without
warranty.
Setting Value Control value setting.
Alarm high limit value setting.
Alarm low limit value setting.
Pulse no. of each round.
Roller diameter value setting.
Hysteresis value setting.
* Setting for RPM, m/min., ft/min.
Input Signal Pulse
Time base Quartz crystal
External DC 12 V, 50 mA max.
Power Supply
Data Output RS 232 PC serial interface.
Operating 0 to 50 .℃
Temperature
Operating Less than 80% R.H.
Humidity
Power Supply 90 to 260 ACV, 50/60 Hz.
Power Approx. 4.7 VA/AC 110V.
Consumption Approx. 5.3 VA/AC 220V.
Weight 384 g/ 0.84 LB.
Dimension DIN size : 96 x 48 mm.
Depth : 110 mm.
2

Accessories Instruction manual......................1 PC
Included Case holder with screw...............
.
2 PCs
Optional * Proximity sensor, PX-01.
Accessories * Photo sensor, PI-06.
* Data Acquisition software,
SW-U801-WIN.
*RS232 cable, UPCB-02.
* USB cable, USB-01.
* GSM controller, GSM-889.
* Interface cable ( cable between meter
to GSM-889 ), GMCB-89.
2-2 Electrical Specifications (23±5 )℃
Range
Photo Tachometer :
10 to 99,999 RPM
Surface Speed ( m/min. ) :
0.05 to 9,999 m/min.
Surface Speed ( ft/min. ) :
0.2 to 32,805 ft/min.
Accuracy ± ( 0.1 % + 1 digit ).
Resolution RPM 0.1 RPM < 1,000 RPM
1 RPM 1,000 to 9,999 RPM
10 RPM 10,000 to 99,999 RPM
m/min. 0.01 m/min
.
< 100 m/min.
0.1 m/min. 100 to 999 m/min.
1 m/min. 1,000 to 9,999 m/min.
ft/min. 0.1 ft/min. < 1000 ft/min.
1 ft/min. 1,000 to 9,999 ft/min.
10 ft/min. 10,000 to 32,805 ft/min.
*
Above specification tests under the environment RF Field Stren
g
th less than
3
V
/M & frequency less than 30 MHz only.
3

3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display
3-2 PV ( process value ) indicator
3-3 SV ( set value ) indicator
3-4 Set Button
3-5 Button▼
3-6 Button▲
3-7 Function Button
3-8 Control relay indicator
3-9 Alarm relay indicator
3-10 RPM indicator
3-11 ft/min indicator
3-12 m/min indicator
3-13 Wire terminals
3-14 Case holder
3-15 RS232 terminal
4

4. MEASURING PROCEDURE
Terminal layout Fig. 2
4-1 Terminal connection
1)Input the ACV power ( 90 to 260 ACV ) to T1, T2.
Do not input the
over voltage to the
AC input terminals.
2)Connect the " Control Relay " output from T3, T4.
Connect the " Alarm Relay " output from T5, T6.
3)Connect the following optional sensor ( or similar sensor ).
* Proximity sensor, PX-01.
* Photo sensor, PI-06.
to the input terminal T8, T9, T10.
T8 is DC 12 V power supply terminal.
T9 is the ground ( - ) terminal.
T10 is the pulse input terminal.
5

4-2 RPM measurement
1)Power on the meter, the " Display " ( 3-1, Fig. 1 ).
2)Press the " Function Button " ( 3-7, Fig. 1 ) once until
the " RPM indicator " ( 3-10, Fig. 1 ) and the " PV
indicator " ( 3-2, Fig. 1 ) is lit.
The " Display " ( 3-1 , Fig. 1 ) will show the RPM value.
3) The RPM value equal :
Total input signal ( pulse ) per min.
÷ ( divide )
setting pulse no. per round
For example :
If the total signal ( pulse ) input per minute
is 5, 124, the setting pulse no. per round is 1,
the RPM value will be 5,124/1=5,124.
The procedures that to set the
*" Pulse no. per round "
refer to the page 10 " 4-5 2nd layer setting procedures ".
4-3 Surface speed ( m/min., ft/min. )
measurement
1)Power on the meter, the " Display " ( 3-1, Fig. 1 ).
2)* Press the " Function Button " ( 3-7, Fig. 1 ) once until
the " m/min indicator " ( 3-12, Fig. 1 ) is lit. The "
Display " ( 3-1 , Fig. 1 ) will show the m/min. value.
* Press the " Function Button " ( 3-7, Fig. 1 ) once until
the " ft/min indicator " ( 3-11, Fig. 1 ) is lit. The "
Display " ( 3-1 , Fig. 1 ) will show the ft/min. value.
* Under the surface speed measurement, the
" PV indicator " ( 3-2, Fig. 1 ) will light too.
6

3) The m/min. value equal :
Total input signal ( pulse ) per min.
÷ ( divide )
setting pulse no. per round
x ( multiple )
setting roller diameter ( cm )
x ( multiple )
3.14
÷ ( divide )
100 1 meter = 100 cm
For example :
If the total input signal ( pulse ) per minute
is 501, the setting pulse no. per round is 1,
the roller diameter is 10 cm
the m/min. value will be 501/1x10x3.14/100=157.3
The ft/min. value equal :
Total input signal ( pulse ) per min.
÷ ( divide )
setting pulse no. per round
x ( multiple )
setting roller diameter ( cm )
x ( multiple )
3.14
÷ ( divide )
30.48 1 ft = 30.48 cm
7

For example :
If the total input signal ( pulse ) per minute
is 501, the setting pulse no. per round is 1,
the roller diameter is 10 cm
the ft/min. value will be 501/1x10x3.14/30.48=516.1
* The procedures that to set the
" Pulse no. per round "
" roller diameter ( cm ) "
refer to the page 10, 11, " 4-5 2nd layer setting
procedures ".
4-4 1st layer setting procedures
CtSP Control value setting
LoLt Low limit value setting
HILt High limit value setting
Control Value Setting
1)Press the " Set Button " ( 3-4, Fig. 1 ) once, the
" Display " will show " CtSP ", now the meter is ready
for the " Control value " setting.
2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring " Control
value ".
Remark :
* When adjust the value, the " SV indicator " ( 3-3, Fig. 1 )
will light.
* The function of " Control value " setting, refer to
page 12, Fig. 3.
8

Low Limit Value Setting
1)Press the " Set Button " ( 3-4, Fig. 1 ) twice, the
" Display " will show " LoLt ", now the meter is ready
" Low Limit value " setting.
2)Use the " " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring " Low
limit value ".
Remark :
* When adjust the value, the " SV indicator " ( 3-3, Fig. 1 )
will light.
* The function of " Low limit value " setting, refer to
page 13, Fig. 4.
High Limit Value Setting
1)After set the " Low Limit " value, press the " Set
Button " ( 3-4, Fig. 1 ) twice, the " Display " will show
" HILt ", now the meter is ready for the humidity
" High Limit " value setting.
2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring " High
limit value ".
Remark :
* When adjust the value, the " SV indicator " ( 3-3, Fig. 1 )
will light.
* The function of " High limit value " setting, refer to
page 13, Fig. 4.
9

Consideration of 1st layer setting
1)Under the normal " Display ", if the " Function Button
" ( 3-7, Fig. 1 ) select to " RPM function ", the
" RPM indicator " ( 3-10, Fig. 1 ) light, the " 1st
layer setting " will ready for " RPM " function .
2)Under the normal " Display ", if the " Function Button "
( 3-7, Fig. 1 ) select to " m/min. function ", the
" m/min indicator " ( 3-12, Fig. 1 ) light, the " 1st layer
setting " procedures will ready for " m/min. " function.
3)Under the normal " Display ", if the " Function Button "
( 3-7, Fig. 1 ) select to " ft/min. function ", the
" ft/min indicator " ( 3-11, Fig. 1 ) light, the " 1st layer
setting " procedures will ready for " ft/min. " function.
4-5 2nd layer setting procedures
PULS Pulse no. per round setting
dIA Roller diameter value setting
CtHy Control Hysteresis value setting
ALHy Alarm Hysteresis value setting
Pulse no. per round setting
1)Press the " Set Button " ( 3-4, Fig. 1 ) continuously at
least two seconds, the " Display " will show " PULS ",
now the meter is ready for the " Pulse no. per round "
setting.
2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring
" Pulse no. per round ".
10

Remark :
* When adjust the value, the " SV indicator " ( 3-3, Fig. 1 )
will light.
* The function of " Pulse no. per round setting ", refer to page 6, 7.
Roller diameter value setting
1)After set the " Pulse no. per round ", press the
" Set Button " ( 3-4, Fig. 1 ) twice, the " Display "
will show dIA ", now the meter is ready for the
the " Roller diameter value " setting.
2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring
" Roller diameter value " in the unit of cm.
* When adjust the value, the " SV indicator " ( 3-3, Fig. 1 )
will light.
* The function of " Roller diameter value setting ", refer to page 7.
Control Hysteresis value setting
1)After set the " Roller diameter value ", press the
" Set Button " ( 3-4, Fig. 1 ) twice, the " Display "
will show CtHy ", now the meter is ready for the
" Control Hysteresis value " setting.
2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring
" Control Hysteresis value ".
Remark :
* When adjust the " Control Hysteresis value " , the " SV indicator "
( 3-3, Fig. 1 ) will light.
* The function of " Control Hysteresis value " setting, refer to
page 12, Fig. 3.
11

Fig. 3
* For example :
Control value : 500
Control Hysteresis value : 5
The control relay will On when measuring
value up to 500. The control relay will Off
again when measuring value down to 495.
Alarm Hysteresis value setting
1)After select the " Control Hysteresis value " , press the
" Set Button " ( 3-4, Fig. 1 ) twice, the " Display "
will show " ALHy ", now the meter is ready for the
the " Alarm Hysteresis value " setting.
2)Use the " Button " ( 3-5, Fig. 1 ) and the "▼▲
Button " ( 3-6, Fig. 1 ) to adjust the desiring
" Alarm Hysteresis value " .
* When adjust the " Alarm Hysteresis value " , the " SV indicator "
( 3-3, Fig. 1 ) will light.
Remark :
* When adjust the " Alarm Hysteresis value " , the " SV indicator "
( 3-3, Fig. 1 ) will light.
* The function of " Alarm Hysteresis value " setting, refer to
page 13, Fig. 4.
12

Fig. 4
* For example :
High limit value : 100
Low limit value : 20
Alarm Hysteresis value : 5
a.The alarm relay will On when measuring
value up to 100. The alarm relay will Off
again when measuring value down to 95.
b.The alarm relay will On when measuring
value down to 20. The alarm relay will
Off when measuring value up to 25.
13

5. RS232 PC SERIAL INTERFACE
The instrument has RS232 PC serial interface via a 3.5
mm terminal ( 3-15, Fig. 1 ).
The data output is a 16 digit stream which can be
utilized for user's specific application.
A RS232 lead with the following connection will be
required to link the instrument with the PC serial port.
Meter PC
(3.5 mm jack plug) (9W 'D" Connector)
Center Pin...............................Pin 4
Ground/shield........................
.
Pin 2
Pin 5
The 16 digits data stream will be displayed in the
following format :
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
14

Each digit indicates the following status :
D15 Start Word
D14 4
D13 When send the upper display data = 1
When send the lower display data = 2
D12 & D11 Annunciator for Display
RPM = 27 ft/min = 11 m/min = 60
D10 Polarity
0 = Positive 1 = Negative
D9 Decimal Point(DP), position from right to the
left
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D8 to D1 Display reading, D8 = MSD, D1 = LSD.
For example :
If the display reading is 1234, then D8 to
D1 is : 00001234
D0 End Word
RS232 setting
Baud rate 9600
Parity No parity
Data bit no. 8 Data bits
Stop bit 1 Stop bit
6. SYSTEM RESET
Power on the meter, use the two fingers to press " Set
Button " ( 3-4, Fig. 1 ) and " Function Button " ( 3-7,
Fig. 1 ) continuously more than 5 seconds until the
Display show the text " rSt ", release the buttons. After
" rSt " text flashing 2 times will return to the normal
screen. The meter system will be reset, all the
calibration data will be cleared, the meter's internal
function will return the default value.
15

7. OPTIONAL PROXIMITY SENSOR
Model : PX-01
Application * Sensor for PDT-2250
Panel RPM Controller,
Monitor
* Sensor for DT-2240D
Panel Tachometer
* Position Detector.
Output method PNP, NO(normal open)
Sensing direction Vertical.
Sensing distance 4 mm.
Operating voltage 10-30 VDC.
Standard Sensing object Iron 18 x 18 x 1 mm.
Current consumption 8 mA, no load.
Response frequency 600 Hz.
Output current 250 mA max.
Leakage current 1.1 mA.
Residual voltage 0.9 VDC.
Operating Temperature -20 to 70 ℃
Operating Humidity 35 to 95% RH.
Wire Dimension 4.7 mm Dia. x 2 m x 3 cores.
Application * Sensor for PDT-2250
DT-224OD.
* Position Detector.
16
Table of contents
Other MRC Controllers manuals

MRC
MRC THROTTLEPACK 9900 User manual

MRC
MRC IOW-924 User manual

MRC
MRC PRODIGY EXPLORER User manual

MRC
MRC PURE POWER DUAL User manual

MRC
MRC TECH 6 0001203 User manual

MRC
MRC VC-100 User manual

MRC
MRC 0001325 User manual

MRC
MRC TECH 6 User manual

MRC
MRC TECH 6 2.0 User manual

MRC
MRC Prodigy Express2 User manual