The centrifuge shall not be located near source of heat and shall not be
subjected to direct sunlight.
Centrifuge should be located on a stable and flat-levelled table top.
Centrifuge should be set horizontally on a rigid base.
It is necessary to ensure a ventilation zone of a minimum of 30cm round the
centrifuge from every direction. Do not obstruct ventilation holes !
Benching/Table for centrifuge should have a safety zone of a minimum of
30cm round the centrifuge from every direction (this is needed in case of
malfunction according to EN 61010-020).
Benching/Table for centrifuge should be free of restraints.
Parameters of the centrifuge refer to the above temperatures (see 2.Technical
specification).
When changing from cold to warm, condensation of water will occur inside the
centrifuge. It is therefore important that sufficient time be provided for drying
out of the centrifuge prior to further use of the centrifuge (min. 4 hours).
Do not position the centrifuge so that it is difficult to operate the power
switch.
Supply voltage given on the name plate has to be consistent with local supply
voltage. MSE CENTRIFUGES LTD laboratory centrifuges are 1st class safety
devices and they are provided with the three-core cable with the plug resistant
to dynamic loadings. Mains socket shall be provided with the safety pin
(protective earth (PE)).