MSE SONIPREP 150 User manual

INSTRUCTIONMANUALFORTHESONIPREP150ULTRASONICDISINTEGRATORREADBEFOREUSE!
Models:-MSS150.CX3.1MSS150.CX3.571100-1129-060206
MSE(UK)LTD
HealthandSafetyatWork
MSE(UK)Ltdisrequiredunderthe1974HealthandSafetyatWorkActandotherUKlegislationasdesigners,manufacturers,suppliersandimportersofarticlesforuseatworktoensurethat,asfarasisreasonablypracticable,articleswhichwedesign,produce,supplyorimportaresafeandwithoutrisktohealth.Wearerequiredtoprovideinformationonthesafetyandhandlingprecautionstobeobservedwheninstalling,operating,maintainingandservicingourproducts.Suchadviceiscontainedinthismanual.Wearealsoobligedtoupdatethisinformationshouldcircumstanceschangeandtooperateasystemtothisend.Weshouldalsoliketopointout,howeverthatyouasusershaveanimportantresponsibilityintheprovisionandmaintenanceofsafeworkingpracticesandconditions.Accordingly,wedrawthefollowingmatterstoyourattention:1.Thisapparatusshouldonlybeusedasintendedandwithinitsdesignparametersbysuitablyqualifiedandtrainedpersonnelwhohavereadandunderstoodtherelevantsectionsofthismanual.2.Thismanualshouldbereadilyavailabletosuchpersonnelatalltimes.3.Inadditiontothatwhichiswritteninthemanual,normalcommon-sensesafetyprecautionsmustbetakenatalltimestoavoidthepossibilityofaccidents.Particularcareisrequiredwhenworkingwithapparatusathightemperatureorpressure.4.Installation,maintenance,repairsandservicingshouldonlybecarriedoutbyanMSE(UK)Ltdapprovedengineer,andconnectiontoelectricalsuppliesshouldonlybecarriedoutbysuitablytrainedpersonnel.TECHNICALSUPPORT,WARRANTYSERVICEANDMAINTENANCEUKcustomers;ifyouareinanydoubtwhatsoeverregardingthecorrectuseofthisapparatus,orifyourequireanytechnicaldataorassistance,pleasecontacttheMSE(UK)LtdTechnicalSupportDepartmentat:
MSE(UK)Ltd
WorsleyBridgeRoadTelephone+44(0)8706094097LowerSydenhamFax+44(0)2086508408LondonE-mail:sales@mseuk.co.ukSE265AZWebPage:www.mseuk.co.ukOVERSEASCUSTOMERS:ShouldcontacttheirlocalMSE(UK)Ltddistributor.ELECTRICITYSUPPLIES:VoltageandfrequencyMSE(UK)Ltdelectricalapparatusisofferedandlabelledforone,orforachoiceoftwoormorevoltagerangesand,wherenecessary,differentfrequenciesofmainssupply.MSE(UK)Ltddoesnotacceptanyresponsibilityfortheoperationofanysuchapparatusshoulditbeconnectedtoelectricitysupplieswhicharenormallyoutside,orvaryoutside,thestatedvoltageandfrequencyvaluesforwhichitisdesigned,norforanyconsequentialloss,damageorinjury,howsoevercaused.
ReadThisBeforeUse!
THISMACHINEHASBEENFITTEDWITHANEWTHUMBTURNLOCK.
THEMACHINEWILLOPERATEONLYWHENTHEDOORISLOCKEDANDTHELATCHISFULLYENGAGEDTOENSURETHEDOORISFULLYLATCHED,TURNTHETHUMBWHEELFULLYCLOCKWISETILLTHEFLATSONTHELOCKARESITTINGHORIZONTALLYTHISHASBEENINTRODUCEDWITHTHEOPERATORINMIND.ASSOONASTHEDOORISOPENEDTHEGENERATORISTURNEDOFFWHICHINTURNSTOPSANYACCIDENTALCONTACTWITHTHEOPERATORANDTHEENERGISEDPROBE.THISALSOSTOPSANYSOUNDWAVESFROMLEAVINGTHESOUNDPROOFEDCHAMBERWHENTHEDOORISOPENED.
DECLARATIONOFCONFORMITYaccordingtoEN45014-(BS7514)
MSE (UK) Ltd Registered Office: Worsley Bridge Road, Lower Sydenha , London SE26 5AZ. Registered in England and Wales No. 5098052
Documentno:71100.1253.1MSE(UK)Ltd
WorsleyBridgeRoad,LowerSydenhamLondon,SE265AZU.K.Telephone:+44(0)8706094097Fax:+44(0)86508408E-Mail:sales@mseuk.co.ukWeb:www.mseuk.co.uk
ThankyouforbuyingaMSE(UK)LtdSoniprep150.PleasereadthisoperatingmanualbeforeusingyourSoniprep;itwillprovideyouandyourcolleagueswithusefulinformationonallaspectsoftheequipment.Asourcustomer,weshouldliketoensurethatyouaretotallysatisfiedatalltimes.Donothesitatetocontactourteam.Yourviewsareveryimportanttous
MSE(UK)LTDPRODUCTWARRANTYTermsandConditions
WehopethatyoudonothavetheneedtousetheextensivewarrantycoverthatMSE(UK)Ltdextendstoyou.Howevershouldyouhaveaproblem,ourpromptresponseisgreatlyhelpedifyouhavefilledinandpostedthepre-paidWarrantyRegistrationCardsuppliedwithyournewequipment.MSE(UK)Ltdgivesaone-yearwarrantyfromthedateofdelivery.Duringthisperiod,componentpartsproventobedefectiveinmaterialsorworkmanshipwillberepairedorreplacedatourexpense.Installation,commissioningandcalibrationarenotcoveredbythiswarrantyagreement.TheMSE(UK)Ltdapprovedserviceagentmustbecontactedforwarrantydeterminationanddirectionpriortoanyworkbeingcarriedout.Thesewarrantiesareonlyapplicabletonewproducts,andnotsecondhandnorrefurbishedproductsevenifrepairedbyMSE(UK)Ltd.Anysuchproductsarecoveredbyseparatewarrantytermsandconditionswhichwillbemadeavailableonrequest.ReplacementorrepairofcomponentpartsorequipmentunderthiswarrantyshallnotextendthewarrantytoeithertheequipmentorthecomponentpartbeyondtheoriginaloneyearwarrantyperiodunlessagreedinwritingbyMSE(UK)Ltd.Theabovewarrantiesareextendedtotheoriginalpurchaseruponfullinvoicepayment.Apurchasereceiptorotherproofofpurchasemayberequiredbeforewarrantyservicewillbeperformed.Thesewarrantiesonlycoverfailuresduetodefectiveworkmanshipwhichoccurduringthenormaloperationoftheproductbytheoriginalpurchaser,andnotfailureswhichresultfromaccident,misuse,abuse,neglect,mishandling,misapplication,alteration,faultyinstallation,electricalpowerfluctuations,dust,orotherenvironmentalextremes,modificationorserviceotherthanbyanapprovedserviceagentorfollowingthewrittenauthorityofthemanufacturer,ordamagethatisattributabletoactsofGod.Expendableitemssuchasmotorbrushes,doorseals,lidseals,‘O’ringsorlampsareexcluded.MSE(UK)Ltd,oritsapprovedserviceagent,reservestherighttorepairdefectiveequipmentonthepremisesofthecustomer,orataservicestation,atthesolediscretionofMSE(UK)Ltdortheirapprovedagent.Intheeventofreturntoanapprovedservicecentrethecustomerisresponsibleforthesafepackagingoftheinstrumentandnotificationtotheservicecentre.NeitherMSE(UK)Ltdnoritsagentsareresponsibleforanydamageoccurringduringshipment.SpecificationandMaterialChanges:MSE(UK)Ltdreservestherighttosupplyourlatestandimprovedmodelsattimeofshipment.Taxes:ThepricesquoteddonotincludeanytaxesimposedbytheStateorCountryinwhichthepurchasewasmade.Installation:Installationofallequipmentshallbeattheexpenseofthepurchaserunlessstatedotherwise.Accesstothesite,andtheprovisionofrequiredutilitiese.g.power,wateranddrainagetosuitableconnections,willbetheresponsibilityofthepurchaser,andatthepurchaser'sexpense.Limitationofliability:Innoevent,whetherasaresultofbreachofcontractorwarranty,shallMSE(UK)Ltdbeliableforanyconsequentialorincidentaldamagesincluding,butnotlimitedto,lossofprofitorrevenues,lossofuseoftheequipmentoranyassociatedequipment,downtimecosts,costsofsubstituteequipment,costsoflabour,costsduetodelaysorclaimsofpurchaser'sowncustomersforsuchdamages.ThepurchaseragreestoindemnifyMSE(UK)Ltdandtoholdthemharmlessfromanyandallliability,claims,demands,actions,suits,expensesorcosts,includingattorney'sfeesrelatingtosuchconsequentialorincidentaldamages.Allexpressedandimpliedwarranties,includingthewarrantiesofmerchantablequalityandfitnessforaparticularpurposearelimitedtotheapplicationperiodofoneyear.Validity:Legalrightsvaryfromcountrytocountryandstateswithincountries,sosomeoralloftheexclusionsorlimitationslistedabovemaynotapply,butifanypartoftheseconditionsshallbefoundtobeunenforceableitshallnotaffectthevalidityorenforceabilityoftheremainderoftheconditions.
ELECTRICITYSUPPLY
Beforeconnectingthisapparatustotheelectricitysupply,checktheinformationgivenontheapparatusratingplateandensurethat:A)YoursupplyissinglephaseAC(alternatingcurrent)ofthestatedfrequencywiththeneutralnominallyatearthpotential.B)Yoursupplyvoltageiswithinthestatedrange.C)Thecurrentratingiswithinthecapacityofyoursupplyoutlet.D)Yourplugorelectricitysupplycircuitisfittedwithasuitablefuse.FuseRating220v-240v110v-120v5ampSeenote100V-120VWARNING!ThisapparatusmustbeearthedThewiresinthemainsleadarecolouredinaccordancewiththefollowingcode:220v-240v110v-120vLiveBrownBlackNeutralBlueWhiteEarthGreen&YellowGreenConnectthewirestoanon-reversible3pinplugasfollows:GreenandYelloworGreenwiretoterminalmarkedE(Earth),G(GroundcolouredGreenandYelloworGreen)ormarkedwiththeEarthsymbol.BlueorWhitetoterminalmarkedN(Neutral)orcommonorcolouredBlue.BrownorBlacktoterminalmarkedL(Live4)orPhaseorcolouredBrown.NOTE:100v-120vinstallationstocomplywithNationalandStateWiringCodesIMPORTANT:Consultanelectricianifindoubtorifyoursupplysystemhasanyofthefollowing:Noearth.Acolourcodedifferentfromabove.Reversibleplugs.Supplyandreturnleadsthatarebothaboveearthpotential.NOTE:Donottouchplugpinswithintwosecondsofplugremovalfromsocket.
CONTENTS8CONTENTS
1.INTRODUCTION....................................................................................................91.1GeneralDescription................................................................................91.2ControlPanelDescription....................................................................112.INSTALLATION...................................................................................................132.1PowerSupplyandFuses......................................................................133.PROBES..............................................................................................................143.1DescriptionandFunction.....................................................................143.2ExponentialMicroprobe.......................................................................143.3FittingaprobetotheTransducer........................................................143.4Positionoftheprobeintheliquidfortreatment................................143.5ProbeMaintenance...............................................................................153.6TestingProbePerformance.................................................................164.PROCESSINGSAMPLES...................................................................................174.1VesselsforSamples.............................................................................174.2BasicProcessing..................................................................................174.3BatchTreatmentVessel........................................................................184.4MeterFlowVessel.................................................................................184.5ContinuousFlowVessel.......................................................................185.OPERATINGINSTRUCTIONS............................................................................195.1ManualControl......................................................................................195.2TimerControl.........................................................................................195.3OptionalProcessTimerUnit(230Vunitsonly)..................................195.4OperatingInstructions..........................................................................196.CLEANING...........................................................................................................217.FAULTFINDING..................................................................................................217.1General...................................................................................................217.2Faultconditionscouldbe:-..................................................................217.3LooseProbe...........................................................................................217.4FatiguedProbeOrCouplingStud........................................................218.SPECIFICATIONS...............................................................................................228.1SpecificationForSoniprep...................................................................228.2SpecificationForProcessTimer.........................................................229.OPTIONALACCESSORIESFORTHESONIPREP150...................................2310.PRODUCTDISPOSAL-ISO14001COMPLIANCE..........................................23
INTRODUCTION91.INTRODUCTIONTheSoniprep150UltrasonicDisintegratorisaself-containedunitdesignedforconvenienttabletopoperation.Itincorporatesanacousticcabinet,whichreducesexternallyaudiblenoiseandultrasoundtoasafelevel.Atransparentdoorallowsobservationofthesampleduringsonication.Theapparatusisfactoryadjustedtotheuser'selectricalsupplyvoltageandfrequency.1.1GeneralDescriptionThegeneratorisasolid-statetransistoramplifier,whichprovideselectricalenergy,atanoutputfrequencyof23KHz,toahighefficiencypiezo-electrictransducer,whichconvertstheelectricalenergytomechanicalenergy.Thetransducer,mountedinthetopofthedisintegratorchamber,transmitsultrasoundbyhighfrequencyvibrationsviaatitaniumalloyprobe.Thereisachoiceofthreeprobes,whichscrewintothetransducerheadandtransmitenergyofapproximateamplitudeforanyspecificapplication.Eachprobeisdesignedtooperateatanominalfrequencyof23KHzconcentratingtheresultantultrasonicenergy,asrequired.Table1ProbeDetails
DescriptionTipDiameter
TransformationRatio
LargeProbeAssy.(solid)19mm(3/4’)3.8:1SmallProbeAssy.(solid)9.5mm(3/8’)5.5:1ExponentialMicroprobe3mm(1/8’)7:1
Thedisintegratorisdesignedtocopewithdifferentrequirementsandwillensurethatthereisthemaximumpossibleacousticenergyatthetipfortheprobe.Oncetheoutputlevelhasbeenselecteditiscontinuouslymaintainedbytheautomaticfeedbackfromthetransducer.Ameter,calibratedinmicrons,allowstheamplitudetobemonitoredcontinuously.Anadjustabletableandaretortclamparemountedonarodwithinthechamber.Thespecimentobetreatediscontainedinasuitabletubeclampedinpositionorinavesselplacedontheadjustabletable.
INTRODUCTION10
INTRODUCTION111.2ControlPanelDescriptionThecontrolpanelissituatedontherighthandsideoftheunit.Abriefdescriptionoftheirindividualfunctionsfollows:AmplitudeMeter-Themeteriscalibratedinmicronsandmonitorstheprobeamplitude,thatis,theactualmovementpeak-to-peakthatistakingplaceatthetransducer/probeinterface.Multiplyingthemeterindicationbythetransformationratiooftheparticularprobeinuse(SeeTable1)allowsactualmovementoftheprobetiptobedetermined.PowerControl-Clockwiserotationofthepowercontrolincreasestheamplitudeleveldisplayedontheamplitudemeter.Thiscontrolpermitssteplessselectionofpoweroutputtothetransducer,providinganaccurateandreproducibleamplitudesettingfortheparticularprobeinuse.Timer-Thisisa15minutetimingunitwithswitchingcontactsandaresettingorcancellationfacility.Itoperatesindependentlyofthestartswitchaftertheoutputlevelhasbeendetermined,thetimerequiredisselected.Theoperatinglampwillbeilluminated,ultrasonicationwillcommenceandthetimerwillstartrun-downtozero.Whenthetimeselectedhasruntheunitwillswitchoffautomaticallyandtheoperatinglampwillgooff.OverrideSwitch-Whenpressed‘ON’,allowstheunittooperatecontinuouslyregardlessofthetimeset.OperatingLamp-Whenilluminated,indicatesthatthepoweristhroughtothetransducer.
INTRODUCTION12
ControlPanel
INSTALLATION132.INSTALLATION2.1PowerSupplyandFusesWhentheequipmentisfirstreceived,makesurethevoltageselectorswitchiscorrectlypositionedtosuityourpowersupply.Thisswitchislocatedattherearoftheunit,nexttotheprocesstimerconnection.Table2-VoltageSelectorSwitchPositions
VoltagePositionIndication220/240Downward240120Upward120
Twoanti-surgefusesarefactoryfittedintheholderslocatedbeneaththemainsinletconnectionsocket.LocationoftheInstrumentandElectricalConnectionsPlacetheSoniprep150onasuitabletableorbenchandensurethatattherearoftheinstrumentisatleast50mm(2’)awayfromthewalltoallowadequateaircirculation.See‘ELECTRICITYSUPPLYSHEET’(frontofmanual)forplug/mainscabledetails.
PROCESSINGSAMPLES143.PROBES3.1DescriptionandFunctionTheprobeisaresonantelementusedforthetransmissionofvibratingmechanicalenergyfromthetransducertothesample.Theprimepurposeoftheprobeistoincreasethetransduceramplitudetoausablelevelor,toselecttheacousticpowertransfer.Ideally,materialusedforsuchprobesshouldhavegoodacousticproperties,highfatiguestrength,lowacousticlossanddensity.Itisessentialthatitshouldalsobecorrosionanderosionresistantandrelativelyinerttothematerialsbeingsonicated.Thematerialthatpossessesallthesequalitiesistitaniumalloyanditisfromthismaterialthatprobesavailableforusewiththedisintegrator,aremanufactured.Note:ProbesarespeciallydesignedtosuittheSoniprep150modelsandaretunedto23KHz.3.2ExponentialMicroprobeThisprobehasbeenspecificallydesignedforthetreatmentofsmallquantitiesofsample,upto5ml.Itshould,preferably,beusedwiththe20mltubeandsealingcapbutanyothersuitablysizedtubemaybeemployed.3.3FittingaprobetotheTransducerCleanthematingsurfacesoftheprobeandtransducerwithasoftcloth.Beforefittingtheprobe,screwthecouplingstudintothetransducerapproximately10mmleavingapproximately10mmprotruding.Screwtheprobeontothestudinthetransducerhead,byhand.Thentightenfirmlyusingashortscrewdriver'sshaftlocatedinthethroughholeintheprobe.Itisimportantthattheprobeisclampedsecurelytothetransducer,asfailuretodosowillresultinalargepowerlossandanirritatinghigh-pitchednoise.NOTE:Thecouplingstudisarelativelyloosefitinboththetransducerandtheprobetoensurethatthecontactsurfacesareallowedtomateperfectly.Anextremelysmallsmearofsiliconegreaseontheprobeheadwillfacilitatesurfacemating.3.4PositionoftheprobeintheliquidfortreatmentWhenmaximumpowerisnecessaryforadifficultapplication,theprobetipshouldbeimmersedintothesolutiontoanoptimumdepth.Asaguide,for50mlsampledownto10mlsample,thebestpositionfortheprobetipis1cmbelowthesurfaceofthesample.Ifthesampleislessthan10ml,thebestpositionfortheprobetipshouldbefoundbytrialanderror.Excessivedepthandalsotheviscosityofthesolutionwillcausethepowertobedamped,whichwillbeevidentbyalowerindicationontheamplitudemeter.Conversely,iftheprobetipisinsufficientlybelowthesurface,therewillbeatendencytogeneratefoaminthesolutionortocausescatteringofdropletsoutofthevessel.Thisisespeciallysowhenusinghighamplification.Ifthesolutionfoams,therewillbeadistinctchangeinsound,i.e.virtuallyunloadedandlittleenergyisbeingtransferredtothesolution.
PROCESSINGSAMPLES15Sometimes,whenthesolutionisthoroughlyaerated,theusualprocessnoisefromthebeakerortesttubewillstop,i.e.nobuzzorvibrationwillbeheard.Again,thisisbecausethereislittletransferofenergy.Whentheseconditionsoccur,turnthepoweroffforafewsecondsandsettheprobedeeper.Graduallyincreasethepowerandthesolutionwillde-gas.Thelowerthepowersetting,thesmallerwillbethequantitythatmaybeprocessedwithoutfoamingthesolution.Donotallowtheprobetocometoocloseto,ormakecontactwith,thesidesorbottomofglasscontainersaspowderedglassmaybeformedinthesolution,orbreakagemayoccur.Plastictubesaremoreresistanttofailureduringsonication,buttheinsulationpropertiesofplasticmaymakeitdifficulttokeepthesamplecool.CAUTIONItmustbeborneinmindthatsonicationcanweakenglasstubes.Itistherefore,advisablethatglasstubesusedinsonicationarenotsubsequentlycentrifugedathighspeedasfailureofthetubemayresult.3.5ProbeMaintenanceInusetheprobewillerodebytheactionofcavitationintheliquid.Thisdissolutionofthemetalfromthetipoftheprobe,whichoccurs,withanymetalduringuseathighsoundintensitiesis,however,reducedtoaminimumbytheuseoftitanium.Theerosionshowsitselfinconcentricringsontheradiatingface.Anerodedsurfacetransfersenergylessefficientlyintotheliquidthanasmoothonebecauseoftheformationofazoneofcavitationadjacenttothesurface.Energynottransferredintotheliquidmediumwilltendtoheattheendoftheprobe.Furthermore,erosioncontinuesatanacceleratedrateontheprobetipsurface,whichisalreadypitted,thusreducingtheusefullifeoftheprobe.Forthesereasonsthetipoftheprobeshouldbepolishedsmoothbymeansofaveryfineemeryclothonaflatsurface,butnotethefollowingtwocautions.CAUTION:a)ProbesWEAROUT,althoughprobesmaybepolished/reground,thelengthoftheprobemustnotbeshortenedbymorethan2mm.PROBESSHOULDBEREPLACEDANNUALLYORASREQUIRED.b)Thetransducer/probematingsurfacesmustnotbecleanedwithanabrasivecleaner.Shouldtheendrequirepolishing,ONLYALIQUIDNONABRASIVEPOLISHshouldbeused.
PROCESSINGSAMPLES163.6TestingProbePerformanceThefollowingtestshouldbecarriedoutsothatusersmaymonitortheperformanceofeachprobe.Measurethetemperatureriseobtainedatmaximumpowerafter5minuteswiththefollowingvolumesofchilledwater(i.e.below10°C).ProbeVolumeofChilledWaterExponential15ml9.5mm50ml19mm100mlAtemperatureriseofapproximately35°Cshouldbeobtained.Iftheriseisnotobtainable,theprobeorthetransducerisatfault.Topinpointthepartofthesystem,whichisfaulty,repeatthetestwithanUNUSEDprobe.Ifyoustilldonotobtainatemperatureriseofapproximately35°C,thencontacttheservicecentreattheaddressinsidethefrontcover.Theservicecentrewillgiveyoumoreadvice.
PROCESSINGSAMPLES174.PROCESSINGSAMPLES4.1VesselsforSamplesNOTE:MinimumVolume-Volumesoflessthan1mlmaybetreatedwiththeexponentialmicroprobe.Thesampleshouldbeheldinamicrotube.MaximumVolume-Amaximumof100mlmaybetreatedatanyonetime.However,largervolumesmaybeprocessedbyusingthetreatmentvessels.Itispreferabletouseanarrowvesselratherthanawideone.Twicethevolumeinanarrowvesseltakesashortertimetoprocessthanthesamevolumeinawidervessel.Twicethevolumeorconcentrationtakestwiceaslongtoprocess.Materialsthatareparticularlyresistanttoultrasonicenergy,however,shouldbeprocessedinthesmallestareapracticable.Sincetheenergyradiatesfromtheprobetipintheshapeofacone,themoreconfinedthesampleunderthetipthebettertheresults.Anarrow,relativelyshorttesttubeisalsocalledforwhenprocessingsmallquantities,sayoftheorderof4ml.Theprobetipshoulddisplaceasmuchliquidaspossibletoforcealongliquidheightor‘head’abovethevibratingtip.Thispreventsfoamingorwhippingthesolutionoutofthetesttube.Awidebeakerisnotnormallysuitable.Thereisconsiderableevidencethattheshapeofthecontainerhasgreatinfluenceontherateofreleaseofsolubleproteinortheefficiencyofthecellularbreakage.Thisisprobablymorethefunctionofmixingthananythingelse,sincetheindividualcomponentsarenotdirectlyexposedtotheactionoftheprobetipformorethanasecondortwo.Ifthevolumeisverylarge,thenindividualcellsarenotintheareaofcavitationactivityforasufficientperiodoftime.Theprobetipshouldbebelowthesurfaceoftheliquidbyapproximately13mm(1/2’)andshouldnotbetoofarfromthebottomofthevessell(againapproximately13mm).Ifthetipisnotsatisfactorilyimmersed,aconsiderablequantityofairmaybetrappedinthesolutionthusdissipatingthepowerdeliveredandnotallowingittobetransmittedtothesolution.AsthereissufficientforcegeneratedbytheSoniprep150tofractureorshatterglass,theprobetipmustnotbeallowedtocomeincontactwith,orcomecloseto,thebottomofthevessel.4.2BasicProcessingTesttubesof20mland100mlnominalcapacityareavailableforcertainapplications.Inadditiontothese,cooledtreatmentvesselsareavailableforusewhenitisimportantthatthemediabeingtreatedshouldbesubjectedtominimaltemperaturerise.Arangeofrubbersealingcapsareavailableforalltubesandcooledtreatmentvessels.Theprobeisinsertedthroughthesealingcapwhichfitsoverthemouthofthevessel,ortube,andprovidesanairtightseal.Foreaseofmanipulationandtoeffectthebestpossibleseal,theedgeoftheholeinthesealingcapshouldbelightlysmearedwithsiliconegrease.Whenusingsealingcaps,ensurethetubeissecurelyclampedtopreventdownwardmovementofthetubeandconsequentdisengagementfromtheprobe.
PROCESSINGSAMPLES18Whenfittingsealingcaps,theairpressurewhichmayhavebuiltupcanbereleasedbyliftingtheedgeofthecapslightly.Thecapsaremanufacturedfromamaterialwhichissuitableforautoclaving.Whentreatingsamplesinanyofthetesttubesthetemperatureofthesuspensionwillrise,especiallyifthequantitiesofthematerialbeingtreatedaresmall.Toavoidthistemperaturerise,immersethetubecontainingthemediuminabeakercontainingcrushediceorasuitablecoolingmixture.Thebeakercanbeplacedontheadjustableplatformandthevesselcontainingmediumheldinthelaboratoryclamp.However,ifthemediumbeingtreatedisparticularlysensitivetotemperature,thenoneofthecooledtreatmentvesselsdescribedinsections4.3,4.3and4.4shouldbeused.4.3BatchTreatmentVesselThisconsistsofaninnerchambertocontainupto50mlofsampletobetreated,surroundedbyacoolingjacketthroughwhichasuitablecoolingliquid(e.g.icedwateroralcohol)canbecirculated.4.4ContinuousFlowVesselThisisavailableforthetreatmentofevenlargervolumeswheretherequireddisintegrationtimeisrelativelyshort.Thisvesselhasbothinletandoutlettubestotheinnerchamberwhichallowsthesampletobepassedcontinuouslythroughatameasuredflowrateasdisintegrationtakesplace.Aperturesareprovidedinthebackoftheultrasoniccabinetsothattheliquidfeed-linescanbepassedthroughtothetreatmentvesselsandpermitsonicationwiththecabinetdoorclosed.
OPERATINGINSTRUCTIONS195.OPERATINGINSTRUCTIONSMisusesinclude:a)Impedingairflow.b)Incorrectpositioningoftheprobeswithrespecttothetransducer(seesection3.4).c)Lackofprobemaintenance(seesection3.5).d)Probeinliquid(foaming).5.1ManualControla)Plugthe3pinplugintothepoweroutlet.Switchthegreensupplyrockerswitchtoits‘ON’position(itsinnerLEDwillilluminate).b)Openthedoor,placethevesselcontainingthesampletobetreatedontheplatform,orifatesttube,clampinpositionandsetupforsonicationasrequired.Closethedoor.c)Presstheoverrideswitch,theamberlampwillilluminateandsonicationcancommenceonceamplificationisset(ifnotsetcorrectlyfromapreviousoperation).d)Ifnotset,thenselectamplitudelevelrequired.e)Aftersonicationiscomplete,switchthesupplyrockerswitchtoits’OFF’position.5.2TimerControla)Carryoutinstructionsinsection5.1a)andb).b)Settime,1to10minutes,byturningknobclockwise.ImmediatelytheTimerbuttonisturnedclockwisetheStartbuttonlampwilllightup,andafterthetimeselectedonthetimerhasrunout,sonicationwillceaseandthelampwillgooff.5.3OptionalProcessTimerUnit(230Vunitsonly)a)PlacetheProcesstimerUnitontopoftheSoniprep.b)Connectthetwocablesfromtheprocesstimer-oneterminatinginathree-pinplug(power)andtheotherterminatinginafive-pinplug(control)-tothecorrespondingsocketsontherearoftheSoniprep.5.4OperatingInstructionsa)Setupthedesiredoperatingconditionsonthedigitalswitchesoftheprocesstimerunit.b)EnsuretheSonipreppowerisONandthattheamplitudeissetorsetwhenthesonicationstarts.TheoverrideswitchshouldbeturnedOFF.c)Switchontheprocesstimermainsswitch.Thegreenmainslampwillilluminate.d)Pressthestartswitchpush-buttonontheprocesstimerunit.TheSoniprepwillnowoperateunderthecontroloftheprocesstimerunit.Ifitisrequiredtostoptheprocessduringsonication,pressthestopbuttonontheprocesstimerunit.Sonicationwillceaseandthetimerwillresetforthenextcycle.
OPERATINGINSTRUCTIONS20
ProcessTimer

Other MSE Laboratory Equipment manuals

MSE Kestrel User manual

MSE

MSE Kestrel User manual

MSE CENTAUR 2 User manual

MSE

MSE CENTAUR 2 User manual

MSE HARRIER User manual

MSE

MSE HARRIER User manual

MSE MicroCentaur Plus User manual

MSE

MSE MicroCentaur Plus User manual

Popular Laboratory Equipment manuals by other brands

Drawell DW-FD Sereries user manual

Drawell

Drawell DW-FD Sereries user manual

Glass Expansion D-Torch installation guide

Glass Expansion

Glass Expansion D-Torch installation guide

Pasco Scientific TD-8578 Instruction manual and experiment guide

Pasco Scientific

Pasco Scientific TD-8578 Instruction manual and experiment guide

Thermo Electron 3911 Operating and maintenance manual

Thermo Electron

Thermo Electron 3911 Operating and maintenance manual

Thermo Scientific Omni FTIR Multi Gas CEMS instruction manual

Thermo Scientific

Thermo Scientific Omni FTIR Multi Gas CEMS instruction manual

Fritsch PULVERISETTE 7 operating instructions

Fritsch

Fritsch PULVERISETTE 7 operating instructions

Merck Millipore CellASIC ONIX2 user guide

Merck

Merck Millipore CellASIC ONIX2 user guide

VWR avantor LED Digital Rotary Evaporator V17 instruction manual

VWR

VWR avantor LED Digital Rotary Evaporator V17 instruction manual

Tuttnauer Valueklave 1730 Operation & maintenance manual

Tuttnauer

Tuttnauer Valueklave 1730 Operation & maintenance manual

Gilson Pipetman Fixed user guide

Gilson

Gilson Pipetman Fixed user guide

Emerson Dosaodor-D V

Emerson

Emerson Dosaodor-D V

Argos FireStar ST instruction manual

Argos

Argos FireStar ST instruction manual

CIANFLONE 2701 operating instructions

CIANFLONE

CIANFLONE 2701 operating instructions

Precisionary CRYOSTAT CF-6100 user manual

Precisionary

Precisionary CRYOSTAT CF-6100 user manual

LW Scientific MiniVID USB instruction manual

LW Scientific

LW Scientific MiniVID USB instruction manual

Selecta MACROTRONIC BLT instruction manual

Selecta

Selecta MACROTRONIC BLT instruction manual

3B SCIENTIFIC PHYSICS ELWE U8400830 instruction sheet

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS ELWE U8400830 instruction sheet

Thermo Scientific Sorvall BIOS A instruction manual

Thermo Scientific

Thermo Scientific Sorvall BIOS A instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.