MTP MTP-1325 User manual

MTP
Operating Manual
Model MTP-1325
Non-Contact Infrared Thermometers
58
7p
912
11p
14
3p
6p
5p 9p 10p
MTP Instruments
www. .com information@itm.com1.800.561.8187

2
TABLE OF CONTENTS
INTRODUCTION……………………………………….. 3
FEATURES…………………………………………….. 3
WIDERANGEAPPLICATION …………………………... 3
SAFETY………………………………………………... 4
DISTANCE&SPOTSIZE………………………………… 4
SPECIFICATIONS……………………………………… 5
FRONTPANELDESCRIPTION………………………….. 6
INDICATOR…………………………………………….. 6
MEASURMENTOPERATION…………………………… 7
BATTERYREPLACEMENT…………………………….. 8
NOTES………………………………………………… 8
MAINTENANCE ……………………………………….11
www. .com information@itm.com1.800.561.8187

3
INTRODUCTION
Thank you for purchase of the IR Thermometer. This is
capable of non-contact (infrared) temperature
measurements at the touch of a button. The built-in laser
pointer increases target accuracy while the backlight LCD
and handy push-buttons combine for convenient, ergonomic
operation.
The Non-contact Infrared Thermometers can be used to
measure the temperature of objects’ surface that is improper
to be measured by traditional (contact) thermometer (such
as moving object, the surface with electricity current or the
objects which are uneasy to be touched.)
Proper use and care of this meter will provide years of
reliable service.
FEATURES:
Rapid detection function
Precise non-contact measurements
Unique flat surface, modern housing design
Automatic Data Hold
°C/°F switch
MAX temperature displays
Backlight LCD display
Automatic selection range and Display Resolution 0.1º
C(0.1ºF)
WIDE RANGE APPLICATION:
Food preparation, Safety and Fire inspectors, Plastic
molding, Asphalt, Marine and Screen printing, measure ink
and Dryer temperature, HVAC/R, Diesel and Fleet
maintenance.
www. .com information@itm.com1.800.561.8187

4
SAFETY
Use extreme caution when the laser beam is turned on.
Do not let the beam enter your eye, another person’s
eye or the eye of an animal.
Be careful no to let the beam on a reflective surface
strike your eye.
Do not allow the laser light beam impinge on any gas
which can explode.
Distance & Spot Size
As the distance (D) from the object increases, the spot size
(S) of the area measured by the unit becomes larger. The
relationship between distance and spot size for each unit is
listed below. The focal point for each unit is 914mm (36”).
The spot sizes indicate 90% encircled energy.
300
100
75
25
900 1200
D:S= 12:1
www. .com information@itm.com1.800.561.8187

5
1. SPECIFICATIONS
Temperature range D: S
-50 to 450 °C (-58°F ~ 842°F) 12:1
Display resolution 0.1 °C (0.1°F)
Accuracy for targets:
Assumes ambient operating temperature of 23 to 25 °C (73
to 77°F)
-50 ~ 20°C (-58°F ~ 68°F) ±2.5°C (4.5°F)
20°C ~300°C (68°F ~572°F) ±1.0% ±1.0°C (1.8°F)
300°C ~450°C (572°F ~ 842°F) ±1.5%
Repeatability
-50~20°C (-58~68°F) : ±1.3°C (2.3°F)
20~450°C (68~842°F) : ±0.5% or ±0.5°C (0.9°F)
Response time 150ms
Spectral response 8~14um
Emissivity Fixed at 0.95
Over range indication LCD will show “----”
Polarity Automatic (no indication for positive polarity);
Minus (-) sign for negative polarity
Diode laser output <1mW,Wavelength 630~670nm,
Class 2 laser product
Operating temp. 0 to 50°C (32 to 122°F)
Storage temp. –10 to 60°C (14 to 140°F)
Relative humidity 10%~90%RH operating,
<80%RH storage
Power supply 9V battery, NEDA 1604A or IEC 6LR61,
or equivalent
Safety “ CE ” Comply with EMC
www. .com information@itm.com1.800.561.8187

6
Note:
Field of View: Make sure that the target is larger than
the unit’s spot size. The smaller the target, the closer
you should be to it. When accuracy is critical, make sure
the target is at least twice as large as the spot size.
2. FRONT PANEL DESCRIPTION
①IR sensor
②LCD Display
Laser pointer beam
③Measurement Trigger
④Battery Cover
⑤Handle Grip
3. INDICATOR
①Data hold
②Laser ” on” symbols
③°C /°F symbol
④Low power symbols
⑤Emissivity symbol and value
⑥Temperature values for the MAX
⑦Symbols for MAX
⑧Current temperature value
www. .com information@itm.com1.800.561.8187

7
Functional Design
the switches of C/F sat in a cell switching
Switching C/F
Select the temperature units (°C or °F) using
the oC/oFswitch ( ①)
Max indicate the max record that displays between the
pressing and releasing the "ON/OFF" button each time
MEASUREMENT OPERATION
①Hold the meter by its Handle Grip and point it toward the
surface to be measured.
②Pull and hold the Trigger to turn the meter on and begin
testing. The display will light if the battery is good.
Replace the battery if the display does not light.
③Release the Trigger and the HOLD display icon will
appear on the LCD indicating that the reading is being
held.
④The meter will automatically power down after
approximately 7 seconds after the trigger is
released.(Unless the unit is locked on)
Note: Measurement considerations
Holding the meter by its handle, point the IR Sensor
toward the object whose temperature is to be measured.
The meter automatically compensates for temperature
deviations from ambient temperature. Keep in mind that
it will take up to 30 minutes to adjust to wide ambient
temperatures are to be measured followed by high
www. .com information@itm.com1.800.561.8187

8
temperature measurements, some time (several
minutes) is required after the low (and before the high)
temperature measurements are made.
This is a result of the cooling process, which must take
place for the IR sensor.
4. BATTERY REPLACEMENT
①As battery power is
not sufficient, LCD
will display “ ”
replacement with one
new battery type 9V is required.
②Open battery cover, then take
out the battery from instrument
and replace with a new 9-Volt
battery and place the battery cover back.
6. NOTES:
How it Works
Infrared thermometers measure the surface
temperature of an object. The unit’s optics sense
emitted, reflected, and transmitted energy, which is
collected and focused onto a detector. The unit’s
electronics translate the information into a temperature
reading, which is display on the unit. In units with a
laser, the laser is used for aiming purposes only.
Open the battery over
9v battery
Low battery indicator
www. .com information@itm.com1.800.561.8187

9
Field of View
Make sure that the target is larger than the unit’s spot
size. The smaller the target, the closer you should be to
it. When accuracy is critical, make sure the target is at
least twice as large as the spot size.
Distance & Spot Size
As the distance (D) from the object increases, the spot
size (S) of the area measured by the unit becomes
larger. See: Fig: 1.
Locating a hot Spot
To find a hot spot aim the thermometer outside the area
of interest, then scan across with an up and down
motion until you locate hot spot.
Reminders
①Not recommended for use in measuring shiny or
polished metal surfaces ( stainless steel, aluminum,
etc.).See Emissivity
②The unit cannot measure through transparent surfaces
such as glass. It will measure the surface temperature
of the glass instead.
③Steam, dust, smoke, etc., Can prevent accurate
measurement by obstructing the unit’s optics.
Emissivity
Emissivity is a term used to describe the
energy-emitting characteristics of materials.
Most (90% of typical applications) organic materials
and painted or oxidized surfaces have an emissivity of
0.95 (pre-set in the unit). Inaccurate readings will result
from measuring shiny or polished metal surfaces. To
www. .com information@itm.com1.800.561.8187

10
compensate, cove the surface to be measured with
masking tape or flat black paint. Allow time for the tape
to reach the same temperature as the material
underneath it. Measure the temperature of the tape or
painted surface.
Emissivity Values
Substance
Thermal
emissivity
Substance
Thermal
emissivity
Asphalt
0.90 to 0.98
Cloth (black)
0.98
Concrete
0.94
Human skin
0.98
Cement
0.96
Lather
0.75 to 0.80
Sand
0.90
Charcoal (powder)
0.96
Earth
0.92 to 0.96
Lacquer
0.80 to 0.95
Water
0.92 to 0.96
Lacquer (matt)
0.97
Ice
0.96 to 0.98
Rubber (black)
0.94
Snow
0.83
Plastic
0.85 to 0.95
Glass
0.90 to 0.95
Timber
0.90
Ceramic
0.90 to 0.94
Paper
0.70 to 0.94
Marble
0.94
Chromium oxides
0.81
Plaster
0.80 to 0.90
Copper oxides
0.78
Mortar
0.89 to 0.91
Iron oxides
0.78 to 0.82
Brick
0.93 to 0.96
Textiles
0.90
www. .com information@itm.com1.800.561.8187

11
7. MAINTENANCE
Repairs or service are not covered in this manual and
should only be carried out by qualified trained
technician.
Periodically, wipe the body with a dry cloth. Do not use
abrasives or solvents on this instrument.
For service, use only manufacturer’s specified parts.
www. .com information@itm.com1.800.561.8187

12
Les Instruments
MTP
Instruments
Siège Social
4409, rue Charleroi
Montréal-Nord, Québec
H1H 1T6
Téléphone : (514) 326-7167 Télécopieur : (514)
326-7835
Ligne directe : 1-888-326-7167
Site Web : www.mtpinc.com Courrier
www. .com information@itm.com1.800.561.8187
Table of contents
Other MTP Thermometer manuals