Thank you for purchasing the My Weigh® iBalance® 3100™ digital scale. Please read all operating instructions carefully before use.
This electronic scale is a precision instrument. With normal care and proper treatment, it will provide years of reliable service. For
more information please visit www.myweigh.com
Never load the scale with more than the maximum capacity. Although the iBalance® 3100™ is designed to be extremely durable with
extra overload protection built into the case, overloading will permanently damage it! Avoid any exposure to extreme heat or cold,
your scale works better when operated at normal room temperature. Keep your scale in a clean environment. Dust, dirt, moisture,
vibration, air currents and/or a close proximity to other electronic equipment can all cause an adverse eect on the reliability and
accuracy of your scale. Handle with care. Gently apply all items to be weighed onto tray top. Avoid shaking, dropping or otherwise
shocking the scale. Scales are delicate instruments and unlike cellular phones, scales have delicate sensors that determine how much
an item weighs. If you drop or shock your scale, these sensors“feel” the shock and are sometimes destroyed. This happens with all
digital scales. We design our scales to be as resistant to shock or drops as possible, however there is no way for us to protect 100%
against load cell or sensor damage.
Failure to follow these instructions will void your warranty.
Always allow the scale to acclimate to normal room temperature for at least one hour before use. Give your scale sucient warm up
time. Usually 30-60 seconds before calibration to give the internal components a chance to stabilize.
PRECAUTIONS BEFORE USING THE BALANCE
1. Matter charged with static electricity can aect accuracy. Discharge all static electricity. For example, one method is to use
Static-Guard spray, and spray it on both sides of the weighing platform.
2. The balance must be in an exactly horizontal position in order to achieve accurate measurement results. In order to bring the
balance into a horizontal position, the adjustable feet are turned either clockwise or counter-clockwise until the air bubble on the
front panel is in the center of the marked circle.
3. Please use an independent power outlet to avoid interference from other electrical appliances.
4. Don’t place any objects on the platform before powering on.
5. When possible please allow the scale to warm up for several minutes before operation.
6. Items should always be placed on the center of the platform when being weighed.
COUNTING FUNCTION
1. Press to cycle through the weighing modes until the display will show“ CA 10 PCS”(means sample size is 10 pcs)
2. Press to toggle sample size,“10”,“20”,“50”,“100”,“200”,“500”,“1000”pcs will appear in. Stop at the one you want to use.
3. Put the exact quantity of samples desired on the platform and press , the set sample size will appear.
4. Keep adding objects to be counted on the pan, the total number of the objects will be displayed. If the unit weight is too small for
the counting resolution, the display will show“Err”
5. To return to normal weighing press
WEIGHT RESPONSE SPEED
The i3100 allows you to adjust the scale reaction time and division selction.
1. Press and hold (do not release) and power the scale on with . Wait until the display shows“nb0 - nb4”and release
2. Press to select your response speed. (nb0: fastest, nb4 slowest) Press to conrm.
COMMUNICATION AND BACKLIGHT
1. Press and hold and press to power the scale on. The scale will power on and go through the self test.
COMMUNICATION
2. The scale should now read “bAUd” followed by“12”, “24”, “48”or“96”. These values represent the baud rate settings: 1200, 2400,
4800, 9600.
3. Use to scroll through the settings and press to conrm.
4. The display will now read “Pr”(Press Sending),“Co”(Stable Sending”or“St”(Continuous Sending)
5. Use to scroll through the settings and then press to conrm
BACKLIGHT
6. The display should now read“bL On”(Backlight ON),“bL OFF”(Backlight OFF) or“bL AUTO”(Automatic Backlight).
7. Use to scroll through the settings and press to conrm
8. Once is pressed the scale will return to normal weighing.
SELECTING DESIRED WEIGHING MODES
The 3100™comes with a preset 9 units with the option to add another 9 units if required. To enable or disable each unit weighing
modes follow these steps:
1. Press and hold and together, then press and release , the scale will power on. Release all the keys when the
display shows“PASS”.
2.Press 4 times and the display will show“YES”and the corresponding unit. Press again to enable (yes) or disable (no) the
selected unit.
3.Press to move to the next unit.The scale will return to normal weighing mode when all the unit settings are nished.