CALIBRATION
Calibration may be required when the scale is rst set up for use, or if the scale is moved to a dierent altitude or new location. This is necessary because the weight
of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur.
Standard Calibration
For this calibration method you must have an accurate 200kg weight or combination of weights in order to calibrate.
1. From the o setting press & hold the [ZERO ] & [TARE] buttons and turn on the scale. Release both keys when the display shows“CAL”and an AD value.
2. Press the [TARE] key, the display will show the calibration weight“200kg”.
3. Place the correct calibration weight on the center of the platform and press [TARE].
4. The display will show the A/D value and return to weighing mode.
5. Remove the weight. Verify the scale is calibrated correctly. Repeat the process, if necessary.
Any Weight Calibration
For this calibration method you can select the weight and unit in which to calibrate the scale. For best accuracy we recommend a minimum weight of 20kg.
1. From the o setting press & hold [ZERO] and turn on the scale. The display will show“SET”(Do not release zero).
2. While holding [ZERO] press [UNIT] 3 times. Release [ZERO]when the display will shows “CAPu”.
3. Press [ZERO] twice and when the display shows“CALu”, press [UNIT] to enter the weight unit menu.
4. Press [TARE] to select the calibration unit of lb or kg. Press [ZERO] to conrm, the display will show“CALu”.
5. Press [ZERO] again the display will show“CAL”. Press [UNIT] to enter the calibration weight set menu.
6. The display will show“x0000”, press [HOLD] to move the ashing digit to the right and [TARE] to increase. Press [ZERO] to conrm the selection. (The calibration
weight should be a minimum of 20kg). The display will show“CAL”.
7. Press [TARE] the display will show the AD value (a series of random numbers). Press [TARE] again when the stable indicator shows.
8. The display will ash the selected calibration weight. Place the correct calibration weight on the center of the platform and press [TARE] when the stable indicator
shows.
9. The display will show“-----”and the AD value. Calibration is complete.
10. Remove the weight. Turn the scale o and on. Verify the scale is calibrated correctly. Repeat the process, if necessary.
AUTO OFF / BACKLIGHT SETTINGS
TheVHD3 allows you adjust the default settings of the scale.
Programming Auto-O Time
1. Press and hold [HOLD] (do not release). Turn on the scale, release when the display shows the auto-o time setting
2. Press [ZERO] to toggle auto-o settings:
i) A-120=120 seconds.
ii) A-180=180 seconds.
iii) A-300= 300 seconds.
iv) A-OFF = Auto-o is disabled.
3. Press [HOLD] to enter backlight setting mode or turn the scale OFF and ON to return to normal weighing mode.
Adjusting Backlight Setting
4. The display will show settings “L-Auto, L-ON, L-OFF or L-STB”.
5. Press [ZERO] to toggle the settings:
i) AUTO= backlight enabled when weight is placed on platform.
ii) ON= backlight is on constantly.
iii) OFF= backlight is disabled.
iv) STB= backlight turns o if scale is not in use for more than 60 seconds.
6. Turn the scale OFF and ON to return to normal weighing mode.