CALIBRATION PROCEDURE
PXIe-4330/4331
This document contains information about verifying and adjusting National Instruments
PXIe-4330/4331 modules using NI-DAQmx 9.1 and later. For more information about
calibration, visit ni.com/calibration.
Contents
Software.................................................................................................................................... 2
Documentation.......................................................................................................................... 2
Calibration Interval................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 4
Calibration Procedure ............................................................................................................... 4
Initial Setup....................................................................................................................... 5
Connecting the Equipment ............................................................................................... 5
Install the CAL-4330 Into a Rack (Optional)........................................................... 7
Connect the Calibration and Switch Cables ............................................................. 8
Connect the DMM and Multifunction Calibrator Cables......................................... 10
Verification Procedures .................................................................................................... 11
Gain Accuracy Verification...................................................................................... 11
Input Offset Verification .......................................................................................... 14
Shunt Quarter-Bridge Calibration Accuracy Verification........................................ 15
Shunt Calibration Resistance Accuracy Verification ............................................... 17
Excitation Voltage Verification................................................................................ 18
Gain and Offset Adjustment Procedure............................................................................ 20
Test Limits................................................................................................................................ 24
Gain Accuracy Test Limits............................................................................................... 25
Input Offset Test Limits ................................................................................................... 31
Shunt Quarter-Bridge Calibration Test Limits ................................................................. 32
Shunt Calibration Resistance Test Limits ........................................................................ 32
Excitation Voltage Test Limits......................................................................................... 33
Appendix A: Connection Path Details...................................................................................... 34
Worldwide Support and Services ............................................................................................. 40