To use the SFP instruments, you must set up the DAQ device to
communicate with the benchtop workstation. To do this, complete the
following steps:
1. Ensure that the DAQ device is properly installed and configured in
MAX and that the benchtop workstation is powered on.
2. Launch the NI ELVIS Traditional Instrument Launcher by selecting
Start»Programs»National Instruments»NI ELVIS Traditional»
NI ELVIS Traditional.
3. Click the Configure button to launch the Configure Hardware dialog
box.
Note If there is an error, the Configure button is the only button enabled.
4. Select the DAQ device that is cabled to the NI ELVIS hardware from
the DAQ Device control. If only one DAQ device is detected inside the
computer, this device is selected by default.
5. Click the Check button in the Communications portion of the window
to verify communication with the NI ELVIS Benchtop Workstation.
6. If the configuration attempt is successful, a message appears in the
Status window, indicating that the NI ELVIS Benchtop Workstation
has been detected and configured. If the attempt fails, an error message
appears, and the message in the Status window indicates that the
configuration attempt has failed.
7. If the computer initializes the NI ELVIS hardware through the selected
DAQ device, the NI ELVIS Traditional software is now configured.
You can now begin to use NI ELVIS. If the configuration attempt fails,
follow the recommendation given in the error dialog or return to the
DAQ Device control and select a different device.
Tip You can click the Reset button available on this dialog box even after you have
completed configuring the DAQ device. Click this button when you want to reset the
NI ELVIS Benchtop Workstation using software. The DAQ device does not reset.
Refer to the NI ELVIS Online Traditional Help and NI Educational Virtual
Instrumentation Suite (NI ELVIS) Hardware User Manual for more
information about using the NI ELVIS Traditional instruments within
LabVIEW or SignalExpress, or for more information about NI ELVIS
features.