NEULOG AMMONIUM LOGGER SENSOR GUIDE
NeuLog ammonium logger sensor
NUL-240
The NeuLog ammonium ion sensor can be used for any science experiment which
utilizes ammonium concentration readings such as in the fields of Environmental
Science, Ecology, Chemistry, Biology, etc.
Among hundreds of possible experimental subjects that can be studied with the NUL-
240 sensor are: chemical reactions, water quality, environmental health, small animal
behavior, ecological studies, and many more.
The ammonium sensor’s measurement units are:
Milligrams per liter (mg/L): SI measurement unit for density of a solid
dissolved in a liquid.
Parts per million (ppm): The total amount of a specific molecule (ammonium
for example) per million molecules in a sample.
Ammonium sensor usage:
The ammonium sensor’s electrode uses a PermaFil (non-refillable), ion-selective
electrode. The reference chamber is gel filled and sealed; therefore, no reference
filling solution is required.
Usage guide:
Concentration range
1M to 5x10-6M
0.01 to 18,000 ppm
pH range
4 to 10 pH
Temperature range
0 to 50°C
Minimum sample size
3mL in a 50 mL beaker
Reproducibility
±4%
Electrode resistance
1 to 4 megaohms
Equipment required:
Wash bottle with distilled or deionized water
Several clean beakers
1 mL, 10 mL and 50 mL pipettes (a 50 ml graduated cylinder can also be used).
Technical notes:
All standard solutions should be prepared fresh.
The standard solutions should be at the same temperature as the sample.
Use ISA in all solutions.
Note: ISA is added to all solutions to ensure the samples and standard solutions
have the same ionic strength; the ratio is 1:50
Electrode preparation:
1. Remove the protective cap encasing the ammonium probe’s electrode.
CAUTION: Do not touch the PVC membrane at the end of the probe with
your fingers.
2. Rinse the electrode with deionized water and blot dry. Be sure not to rub as this
will harm the sensor.
3. Soak the electrode in deionized water for 10 minutes, then in a diluted
ammonium standard solution (either 0.0001 M or 1 ppm) for two hours until
ready for use.
4. For preparing the 1 ppm standard solution, mix one drop of the 1000 ppm
standard solution with 20 drops of ISA and add deionized water up to a total of
40 ml solution (assuming one drop equals 0.04 ml). It is also possible to use a
micropipette if available.
Offsetting the sensor:
Be sure to offset the ammonium sensor once daily during periods of regular usage.
1. Prepare the electrode as described in the “Electrode preparation” guide above. If
you have already prepared the electrode in the last three days and stored it as
described in the "Electrode storage" section then you do not have to repeat it
before offsetting the sensor.
2. Prepare the offsetting solution: Mix 1 ml of the ISA solution with 0.5 ml of the
1000 ppm standard solution. Add deionized water up to a total of 50 ml. You will
get a 10 ppm solution.
3. Rinse the electrode with deionized water, blot dry and place in the beaker.
4. After a stable reading has been reached, press the “Start/Stop” button located
on the sensor’s faceplate for 4 seconds.
5. Rinse the electrode once more with deionized water and blot dry.
6. The ammonium sensor is now ready for use.
NEULOG AMMONIUM LOGGER SENSOR GUIDE
Electrode storage:
Short term (overnight or weekend):
Rinse the electrode thoroughly with deionized water and place the tip in a diluted
standard solution (around 10 ppm) between measurements.
Long term:
Rinse the electrode thoroughly with deionized water and blot dry. Attach the hard
plastic protective electrode cap again to protect the sensing element. Follow
procedures in the "Electrode preparation" section when using the electrode again.
Quick start procedure:
PC or Mac Computer
Materials needed:
NUL-240 Ammonium Sensor
USB-200 USB Module
A USB to mini USB cable (included with the USB-200 module)
Before use, please review the “Ammonium sensor usage”, “Electrode preparation”,
“Electrode storage”, and “Offsetting the sensor” sections of this document to ensure
the proper use and longevity of the NeuLog ammonium ion sensor.
Your ammonium sensor needs to be connected to a USB-200 module. The USB-200
module then connects to a computer via a USB to mini-USB cable. Please note that
you cannot plug the ammonium sensor directly into the computer.
The browser based application can be downloaded for free at www.NeuLog.com as
well as a full application user guide.
Choose "Downloads" on the main menu and then choose "Software and Application".
Procedure:
1. Install the NeuLog application.
2. Connect the USB-200 module to the PC or Mac.
3. Connect the ammonium sensor to the USB-200 module (they directly plug
together).
4. Open the NeuLog application by clicking on the shortcut on the screen.
5. Once an ammonium sensor module box appears on the left side of the screen
the probe has been automatically identified and you can begin experimentation.
6. If the ammonium sensor is not automatically identified, click the “Search for
sensors” icon to find the sensor.
7. Click on the “On-line experiment” icon; this will open a graph below.
8. Click on the “Module setup” button located in the ammonium sensor module box
to change the sensor’s settings if need be.
9. Click on the "Experiment set up" icon to change the experiment settings if need
be (experiment duration for example).
10. The ammonium sensor will give a live reading in its module box while plugged in.
11. To run an experiment and collect data, click the “Run experiment” icon.
12. To end data collection early, click the “Stop experiment” icon.
Tablet, smart phone device
Materials needed:
NUL-240 Ammonium Sensor
WIFI-201 WiFi module
BAT-200 Battery
Before use, please review the “Ammonium sensor usage”, “Electrode preparation”,
“Electrode storage”, and “Offsetting the sensor”sections of this document to ensure
the proper use and longevity of the NeuLog ammonium ion sensor.
Your ammonium sensor needs to be connected to a WIFI-201 module. The WIFI-201
module will create a closed NeuLog WiFi network which will stream the NeuLog data
to a device of your choosing. Once your device is wirelessly connected to the NeuLog
network, you can run experiments and collect data through a browser of your
choosing.
Procedure:
1. Connect the ammonium sensor directly to the left side of a WIFI-201 (no wires
required).
2. Connect a BAT-200 module to the right side of the WIFI-201 module.
3. Although not required, we recommend plugging the BAT-200 to an outlet using a
USB to mini USB charger (such as a typical cell phone charger). The WIFI-201
module will run for 60-90 minutes (depending on the sensor) without being
plugged in.
NEULOG AMMONIUM LOGGER SENSOR GUIDE
4. The WIFI-201 module can be powered directly using a mini to USB cord and
plugging it into your computer or a wall charger. Please note this function is only
available on WIFI-201 modules with the USB icon on the cover.
5. For further WIFI-201 instructions or the WIFI-201 quick start guide please visit:
www.NeuLog.com; Choose "Products" on the main menu,
then choose "WiFi Communication module".
6. The WIFI-201 indicator lights will flash; take no action until the LED to the far left
turns blue; this can take up to one minute.
7. Take your tablet or smart phone, go to the WiFi settings and select the NeuLog
network (NeuLog0184 for example) which matches the ID found on the back of
the WIFI-201 module (0184 for example).
8. Give your device about 20 seconds to connect to the WIFI-201 module.
9. Once the device is connected, go to your browser and type the website
wifi201.com into the URL bar, then wait for 30-60 seconds.
10. You will see a “Control mode” icon in the application, click on it.
11. The application will then load a new screen and begin to auto detect the
sensor(s); this can take a minute. (You can stop the search when the sensor is
found.)
12. If the application does not auto detect the sensor(s), select “Search for sensors”.
13. Once the sensor is found you will see a module box on the left side of the screen
for the ammonium sensor. In this box, real time data will be displayed.
14. Click on the “Module setup” button located in the ammonium sensor module box
to change the ammonium probe settings if need be.
15. Click on the “Experiment setup” icon to change the experiment settings if need
be (Experiment duration for example).
16. To run an experiment and collect data, click on the “On-Line experiment” icon,
then click on the "Run experiment" icon.
17. If you have a newer generation WIFI-201 module (with the USB icon on the
cover) this can be used as a USB-200, meaning you can tether the probes
directly to the computer using the WIFI-201. To engage operation in this fashion,
press 3 times on the panel button. Repeating this operation will return the
module to WiFi mode.
Operation with NeuLog Graphic Display Module (viewer)
Materials needed:
NUL-240 Ammonium Sensor
VIEW-101 Graphic display Module
BAT-200 Battery Module
Before use, please review the “Ammonium sensor usage”, “Electrode preparation”,
“Electrode storage”, and “Offsetting the sensor” sections of this document to ensure
the proper use and longevity of the NeuLog ammonium ion sensor.
Procedure:
1. Connect the ammonium sensor to the left side of the viewer.
2. Connect the BAT-200 module to the right side of the viewer.
3. The VIEW-101 will auto detect the sensor and you will see it appear on the left
side of the screen.
4. Once the sensor appears it will be monitoring data in real time.
5. To run an experiment and collect data press the run icon (little green person).
Off-line experiments
(Off-line experiments are when you do not have a sensor connected directly to a
computer, tablet, smartphone, or NeuLog viewer).
Materials needed:
NUL-240 Ammonium Sensor
BAT-200 Battery Module
Before use, please review the “Ammonium sensor usage”, “Electrode preparation”,
“Electrode storage”, and “Offsetting the sensor” sections of this document to ensure
the proper use and longevity of the NeuLog ammonium ion sensor.
Materials needed to configure your offline experiment:
USB-200 Module, WIFI-201 module or VIEW-101 Module
USB to mini USB cable (included with the USB-200 module).
Procedure:
1. To determine the experiment settings, first connect to your chosen device (PC,
Mac, tablet, smart device) and click on the “Off-line experiment” icon after the
sensor search has ended. (For more information on how to change the
experiment settings, view the quick start procedure section for your chosen
device in this document.)
2. Connect the ammonium sensor directly to the left side of a charged BAT-200
module.
3. When ready to collect data, press the “Start/Stop” button on the ammonium
sensor, a red light will turn on for the duration of the experiment.
NEULOG AMMONIUM LOGGER SENSOR GUIDE
4. After the experiment has concluded (the “Start/Stop” button is pressed again or
the data collection period ends) connect the sensor to your chosen device (you
can find full instructions in each device’s quick start procedure section).
5. Open the NeuLog application.
6. Click the “Off-line experiment” icon.
7. Click the “Load data from sensors” icon.
8. Select which experimental data to upload (5 experimental runs can be stored at
one time on the NUL-240 ammonium sensor).
Restoring sensor’s factory default settings:
Procedure:
1. Connect the USB-200 to a PC, Mac, tablet, or smart device.
2. Connect the ammonium sensor to a USB-200 module (they directly plug
together).
3. Open the NeuLog application.
4. Click the “Tools” icon.
5. Click the “Restore sensor’s factory defaults” on the menu.
Included with the sensor:
NeuLog NUL-240 Ammonium Sensor Guide (this document)
NeuLog ammonium (NH4+) sensor electrode
Ammonium (NH4+) Ionic Strength Adjuster (ISA)
Ammonium (NH4+) 1000 ppm standard
Sensor’s specifications
mg/L
ppm
Range and
operation modes
0.02 to 18,000
0.02 to 18,000
ADC resolution
15 bit
Resolution
0.03 at 0.02 to 100
mg/L
0.1 at 100 to 1,000
mg/L
130 at 1,000 to
18,000 mg/L
0.03 at 0.02 to 100
ppm
0.1 at 100 to 1,000
ppm
130 at 1,000 to
18,000 ppm
Max sample rate
(S/sec)
100
Experiment Duration: 1 second to 31 days.
Sensor’s features:
Fully digital data
Rugged plastic ergonomic case
PermaFil ion selective electrode
Ammonium probe connected to the sensor’s body by a flexible rubber-coated wire
Protective hard-plastic storage cap
Push button switch for Start/Stop experiments in off line mode
LED indicator of experiment status (blinks while collecting data)
Note: NeuLog products are intended for educational use.
Videos and experiment examples:
Videos, literature and other probes can be found at www.NeuLog.com.
In order to access the ammonium sensor's page, choose "Products" on the main
menu and then "Ammonium logger sensor".
In order to find the science experiments, choose "Experiments" on the main
menu of the NeuLog website, then choose "Physics", "Chemistry", "Biology" or
"Environmental Science".
Biology Experiments:
oAmmonium and Nitrate in an Aquarium (B-45)
NEULOG AMMONIUM LOGGER SENSOR GUIDE
Technical background:
The philosophy behind NeuLog’s plug and play technology is based on each sensor’s
ability to store its own data due to an internal flash memory chip and micro-controller
in each plastic NeuLog body. This technology allows the sensor to collect and then
store the digital data in the correct scientific units (oC, oF, Lux, %, ppm, for example).
The sensor is pre-calibrated at the factory. The built-in software in the logger can be
upgraded for free at any time using the provided firmware update.
The ammonium sensor uses a permanently filled ion selective electrode to accurately
measure the amount of ammonium ion in a solution.
The Ionic Strength Adjustor (ISA) is used to standardize the strength of the ions in
various solutions to give the most accurate readings.
Maintenance and storage:
Before use, please review the “Ammonium sensor usage”, “Electrode preparation”,
“Electrode storage”, and “Offsetting the sensor” sections of this document above to
ensure the proper storage and longevity of the NeuLog ammonium ion sensor.
Never submerge the NeuLog plastic body in any liquid
Do not allow liquid into the ammonium sensor’s body
After use, gently wipe away any foreign material from the ammonium sensor
Store in a box at room temperature out of direct sunlight.
Warranty:
We promise to deliver our sensor free of defects in materials and workmanship. The
warranty is for a period of 3 years from the date of purchase and does not cover
damage of the product caused by improper use, abuse, or incorrect storage. Sensors
with a shelf life such as ion selective probes have a warranty of 1 year. Should you
need to act upon the warranty, please contact your distributor. Your sensor will be
repaired or replaced.
Thank you for using NeuLog!
Flexible, simple, fast, forward thinking.
W: www.neulog.com
E: info@neulog.com
A: 850 St Paul Street, Suite 15, Rochester, NY 14605
P: 1.866.553.8536
V11042014

Other NeuLog Accessories manuals

NeuLog NUL-202 User manual

NeuLog

NeuLog NUL-202 User manual

NeuLog NUL-227 User manual

NeuLog

NeuLog NUL-227 User manual

NeuLog NUL-229 User manual

NeuLog

NeuLog NUL-229 User manual

NeuLog NUL-202 User manual

NeuLog

NeuLog NUL-202 User manual

NeuLog NUL-216 User manual

NeuLog

NeuLog NUL-216 User manual

NeuLog NUL-216 User manual

NeuLog

NeuLog NUL-216 User manual

NeuLog NUL-215 User manual

NeuLog

NeuLog NUL-215 User manual

NeuLog NUL-216 User manual

NeuLog

NeuLog NUL-216 User manual

NeuLog NUL-215 User manual

NeuLog

NeuLog NUL-215 User manual

NeuLog NUL-212 User manual

NeuLog

NeuLog NUL-212 User manual

NeuLog NUL-219 User manual

NeuLog

NeuLog NUL-219 User manual

NeuLog NUL-246 User manual

NeuLog

NeuLog NUL-246 User manual

NeuLog NUL-227 User manual

NeuLog

NeuLog NUL-227 User manual

NeuLog NUL-205 User manual

NeuLog

NeuLog NUL-205 User manual

NeuLog NUL-243 User manual

NeuLog

NeuLog NUL-243 User manual

NeuLog NUL-236 User manual

NeuLog

NeuLog NUL-236 User manual

NeuLog NUL-235 User manual

NeuLog

NeuLog NUL-235 User manual

NeuLog NUL-206 User manual

NeuLog

NeuLog NUL-206 User manual

NeuLog NUL-218 User manual

NeuLog

NeuLog NUL-218 User manual

NeuLog NUL-210 User manual

NeuLog

NeuLog NUL-210 User manual

NeuLog NUL-237 User manual

NeuLog

NeuLog NUL-237 User manual

NeuLog NUL-210 User manual

NeuLog

NeuLog NUL-210 User manual

NeuLog NUL-245 User manual

NeuLog

NeuLog NUL-245 User manual

NeuLog NUL-225 User manual

NeuLog

NeuLog NUL-225 User manual

Popular Accessories manuals by other brands

SBOX Knife Box Instructions and installation guide

SBOX

SBOX Knife Box Instructions and installation guide

Silvercrest RC DP3 4001-A GB 3726 Operation and safety notes

Silvercrest

Silvercrest RC DP3 4001-A GB 3726 Operation and safety notes

Cooper Surgical K-Systems G185 SensorTech user manual

Cooper Surgical

Cooper Surgical K-Systems G185 SensorTech user manual

EQ3 homematic IP HmIP-SCTH230 Installation instructions and operating manual

EQ3

EQ3 homematic IP HmIP-SCTH230 Installation instructions and operating manual

Pro-Aqua JL027130NPF instruction manual

Pro-Aqua

Pro-Aqua JL027130NPF instruction manual

Bin Master FD-2000 manual

Bin Master

Bin Master FD-2000 manual

Silvercrest SPBH 5.200 A1 Operating instructions and safety instructions

Silvercrest

Silvercrest SPBH 5.200 A1 Operating instructions and safety instructions

ABQINDUSTRIAL Plug&Test MR01-50 user guide

ABQINDUSTRIAL

ABQINDUSTRIAL Plug&Test MR01-50 user guide

Silvercrest SPBD 5000 A1 operating instructions

Silvercrest

Silvercrest SPBD 5000 A1 operating instructions

Air King BS48 Specifications

Air King

Air King BS48 Specifications

STEEL-PLAY Powerbank user manual

STEEL-PLAY

STEEL-PLAY Powerbank user manual

HAMPTON BAY HB-7951-02 Use and care guide

HAMPTON BAY

HAMPTON BAY HB-7951-02 Use and care guide

Sewhacnm SI 4300 Operation manual

Sewhacnm

Sewhacnm SI 4300 Operation manual

HYDACELECTRONIC HAT 1400 CAN operating manual

HYDACELECTRONIC

HYDACELECTRONIC HAT 1400 CAN operating manual

Lenord, Bauer GEL 2477 operating instructions

Lenord, Bauer

Lenord, Bauer GEL 2477 operating instructions

Lenord, Bauer GEL 2471 operating instructions

Lenord, Bauer

Lenord, Bauer GEL 2471 operating instructions

Rocket Fish RF-PSV1303 Quick setup guide

Rocket Fish

Rocket Fish RF-PSV1303 Quick setup guide

Magicfx MFX0370 manual

Magicfx

Magicfx MFX0370 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.