If this product is to be unused for an extended period of time, remove the
batteries from the unit. Failure to do so may cause them to leak or
discharge, leading to product failure.
Even when this product is to be unused for an extended period of time,
turn on the product once every 6 months and check that the pump
normally works (by running the pump for approx. 3 minutes). If it is unused
for an extended period of time, the grease inside the pump motor may
become hardened, resulting in a motor malfunction.
This product is explosion-proof. Do not disassemble, modify, or alter the
structure of this unit or its electrical circuits. Doing so may impair the
performance of the explosion-proof characteristics.
Do not modify or use the detector for any purposes other than those
intended/described in this manual. Doing so may impair the performance
of the product.
Do not leave the product in high temperature/humidity conditions for a
long period of time. Doing so may impair the performance of the product.
Avoid using the product outside the specified operating
temperature/humidity/pressure range. Also, avoid exposing the product
to abrupt temperature/humidity/pressure changes. Failure to do so may
impair the performance of the product.
Avoid rapid changes in pressure. Failure to do so may impair the sensor
performance or may damage the sensors.
Avoid strong mechanical shock, impact or vibration to the product by
dropping or bumping. Failure to do so may impair the performance of the
product.
If you drop or bump the product by mistake then the reading is fluctuated;
allow enough time for the reading to stabilize before use, or perform a
zero adjustment/restart the product in clean air.
If condensation is present on the product, remove it and make sure the
unit is completely dried and checked for abnormalities before use.
This detector may detect gases or solvent gases that are not target gases.
Take the usage environment into consideration.
Do not use the product in a place or near a place where silicone
sealant/vapor may be present. Doing so may compromise the
performance of the product.
Detecting high concentration of sulfur dioxide or chlorine may shorten the
sensor life or increase errors.
As the oxygen sensor has pressure dependence, make necessary
pressure adjustments when using the detector at a place other than sea-
level, such as at high altitude.
See “Detection Principle” on page 55.
The gas sensors may contain harmful substances. For disposal, return
them to New Cosmos or treat them as industrial waste.
When used at low temperature, the battery life will be shorter than when
used at room temperature due to the battery’s characteristics.
Keep the detector away from wireless devices, while in use. Failure to do
so may cause a fault alarm or fluctuations in the reading due to radio
wave interference.
Wiring and installation should only be performed by a qualified electrician
with knowledge of wiring/installation procedures.