Nikon Eclipse Ci-E User manual

Turn on the power.
Turn on the power.
Turn on the power switch (press to the "|" position)
for the microscope. (The power LED on
the front of the main body will light up.)
M569 E
11.8.NF.1 (2/2)
Bright-field Microscopy Quick Guide
Bright-field Microscopy Quick Guide
1
2
3
4
5
Press the nosepiece forward/reverse rotation buttons, or
the nosepiece address button on the remote control pad
to bring the 10x objective into the optical path.
6
Set the setting mode switch to the left side
Set the setting mode switch to the left side
(operation mode).
(operation mode).
Set the setting mode switch on the remote control pad to the left side.
Turn on the dia-illumination lamp.
Turn on the dia-illumination lamp.
Press the dia-illumination ON/OFF switch
to turn on the lamp.
Select the setup target with the toggle pattern switch,
Select the setup target with the toggle pattern switch,
and then switch to the setting mode to configure
and then switch to the setting mode to configure
the motorized functions.
the motorized functions.
Before using a motorized swing-out condenser for the first time,
configure the swing-out of the top lens. (

See Chapter 2, Section 13.1 in the
instruction manual for details.) To use the toggle function, configure the toggle
pattern. (

See Chapter 2, Section 12.2 in the instruction manual for details)
(1)
(2)
P
O
WER
10x
Lower the condenser slightly from the uppermost position.
Lower the condenser slightly from the uppermost position.
Turn the condenser focus knob until the condenser is positioned at
the upper limit (where it clicks to a stop), and then lower it a little.
Fully open the field diaphragm and aperture diaphragm.
Fully open the field diaphragm and aperture diaphragm.
Turn the field diaphragm dial and the aperture diaphragm lever
clockwise to open them completely.
Bring the 10x objective into the optical path.
Bring the 10x objective into the optical path.
Focus on the specimen.
Focus on the specimen.
(
See Chapter 2, Section 2 in the instruction manual for details)
See Chapter 2, Section 2 in the instruction manual for details)
When using the trinocular tube or ergonomic tube, push in the
optical path switching lever to distribute 100% light to the
binocular section.
Look into the eyepiece and turn the coarse focus knob away to
raise the stage to the upper limit. From there, focus on the
specimen by lowering the stage.
CLAMP
TORQUE
0.1
0.3
0.4
0.5
0.2
0.9
0.8
CLAMP
TORQUE
0.1
0.3
0.4
0.5
0.2
0.9
0.8
CLAMP
TORQUE
0.1
0.3
0.4
0.5
0.2
0.9
0.8
OFF12
TOGGLE
PATTERN
MEMORY
7
9
(2)
(1)
Rotate the stage knob to move the stage and bring the target into
the optical path.
(So that the sample sealed under the cover glass will be lighted.)
Place a specimen on the stage, and move the stage
Place a specimen on the stage, and move the stage
to bring the target into view.
to bring the target into view.
Open the claw of the specimen holder's moving part and place
the specimen onto the stage, gently stowing the claw back
to fix the specimen.
0.9
0.7
0.6
0.5
0.8
0.1
0.2
POWER
0.1
0.2
0.3
0.40.6
0.7
0.8
0.9
8
11
17
8
5
1
18
7
13
159
16
14
4713
6
12
2
3
10
TORQUE
P
O
WER
0
90
80
70
60
50
40
30
20
10
OFF
TOGGLE
PATTERN
MEMORY
12
OFF
TOGGLE
PATTERN
MEMORY
12
9
Opening the field diaphragm too much
results in stray light entering the field
of view, generating flare and reducing
the image contrast.
Be sure to adjust the field diaphragm
each time you change the objective.
Adjust the field diaphragm
Adjust the field diaphragm
Turn the field diaphragm dial to adjust the field diaphragm so that
it almost circumscribes the field of view.
16
13
View the specimen.
View the specimen.
Rotate the stage knob to move the target. If the target is not in focus,
use the focus knob to adjust the focus.
17
14
Turn off the power.
Turn off the power.
Turn off the power switch (press to the " "
position) for the microscope. (The power
LED on the front of the main body will turn off.)
18
15
10
11
Focus on the specimen.
Focus on the specimen.
12
(1)
(2)
(3)
(4)
(5)
(1)
(2)
(3)
(3)
(4)
POWER
0.1
0.2
0.3
0.40.6
0.7
0.8
0.9
TORQU E
CL
A
P
O
WER
0
90
80
30
20
10
POWER
0.1
0.2
0.3
0.40.6
0.7
0.8
0.9
0.1
0.2
0.3
0.40.6
0.7
0.8
0.9
(Continuation)
(Continuation)
Focus on the specimen.
Focus on the specimen.
(
See Chapter 2, Section 2 in the instruction manual for details)
See Chapter 2, Section 2 in the instruction manual for details)
When the focus was roughly adjusted using coarse focus knob,
turn the fine focus knob to accurately adjust the focus.
Adjust the brightness of the field of view by turning
the dia-illumination lamp brightness control knob.
Avoid the following actions, which can
cause equipment malfunction.
Notes on controlling the focus knobs
Notes on controlling the focus knobs
Rotating the right and left focus
knobs in opposite directions.
Turning the coarse focus knob past
its limit point.
Adjust the diopter.
Adjust the diopter.
(
See
See
Chapter 2, Section 4 in the instruction manual for details)
Chapter 2, Section 4 in the instruction manual for details)
Turn the diopter adjustment ring on the right and
left eyepieces to align the end face of the diopter
adjustment ring with the line.
(This is the diopter adjustment reference position.)
Focus on the specimen using the 40x objective.
Bring the 10x (or 4x) objective into the optical path.
Look into the right eyepiece with your right eye and the left
eyepiece with your left eye. Turn the diopter adjustment ring
of each eyepiece to focus on the specimen.
At this point no focus knobs are used.
Line
Repeat steps (2) through (4) to make sure the focus has been
adjusted properly.
For easy adjustment, look into
the eyepiece as if you were looking
at a distant object.
Adjust the interpupillary distance.
Adjust the interpupillary distance.
Look into both eyepieces and rotate the binocular part to adjust
the binocular part's opening until the fields of view for the right
and left eyes coincide.
Focus and center the condenser.
Focus and center the condenser.
(
See Chapter 2, Section 5 in the instruction manual for details)
See Chapter 2, Section 5 in the instruction manual for details)
Look into the eyepiece with the field diaphragm stopped down
to the minimum. Focus on the field diaphragm image using
the condenser focus knob, then adjust the condenser centering
screws to center the diaphragm image within the field of view.
Bring the 40x objective into the optical path to check the focus
and centering of the field diaphragm image. Make adjustments
in the same way as step (1) as necessary.
Turn the field diaphragm dial and adjust the field diaphragm image
so that its size is almost the same as the field of view.
Select the desired objective.
Select the desired objective.
Press the nosepiece forward/reverse rotation buttons, or
the nosepiece address button on the remote control pad to move
the desired objective into the optical path.
Field diaphragm Field of view
OFF
Be sure to adjust the aperture
diaphragm each time you change
the objective.
(You can see the aperture diaphragm
image with the centering telescope.)
Adjust the aperture diaphragm.
Adjust the aperture diaphragm.
(
See
See
Chapter
Chapter
2, Section 6 in the instruction manual for details)
2, Section 6 in the instruction manual for details)
Turn the aperture diaphragm lever on the condenser to adjust
the aperture diaphragm so that it is set to 70 to 80% of
the numerical aperture of the objective used.
Pupil of
objective
Appropriate size of
the aperture diaphragm
70 to 80
100
Aperture diaphragm
image
Circumscribe around
the field of view
Field diaphragm image
Eyepiece field of view
Field diaphragm image
Eyepiece field of view

Other manuals for Eclipse Ci-E

1

Other Nikon Microscope manuals

Nikon Eclipse E400 User manual

Nikon

Nikon Eclipse E400 User manual

Nikon Eclipse L200 User manual

Nikon

Nikon Eclipse L200 User manual

Nikon L-ke User manual

Nikon

Nikon L-ke User manual

Nikon E800 User manual

Nikon

Nikon E800 User manual

Nikon OPTIPHOT-2 User manual

Nikon

Nikon OPTIPHOT-2 User manual

Nikon Eclipse TS100 User manual

Nikon

Nikon Eclipse TS100 User manual

Nikon ECLIPSE Ni-E User manual

Nikon

Nikon ECLIPSE Ni-E User manual

Nikon Eclipse Ti TIFT 1454 User manual

Nikon

Nikon Eclipse Ti TIFT 1454 User manual

Nikon eclipse e100 User manual

Nikon

Nikon eclipse e100 User manual

Nikon ECLIPSE TE2000-E User manual

Nikon

Nikon ECLIPSE TE2000-E User manual

Nikon Eclipse Ti-E User manual

Nikon

Nikon Eclipse Ti-E User manual

Nikon 50i User manual

Nikon

Nikon 50i User manual

Nikon Eclipse TE300 User manual

Nikon

Nikon Eclipse TE300 User manual

Nikon ECLIPSE MA200 User manual

Nikon

Nikon ECLIPSE MA200 User manual

Nikon ECLIPSE Ni-E User manual

Nikon

Nikon ECLIPSE Ni-E User manual

Nikon TSM User manual

Nikon

Nikon TSM User manual

Nikon SMZ1000 User manual

Nikon

Nikon SMZ1000 User manual

Nikon Eclipse E600W User manual

Nikon

Nikon Eclipse E600W User manual

Nikon A1 User manual

Nikon

Nikon A1 User manual

Nikon Eclipse LV100DA-U User manual

Nikon

Nikon Eclipse LV100DA-U User manual

Nikon Ee1078 User manual

Nikon

Nikon Ee1078 User manual

Nikon YS100 User manual

Nikon

Nikon YS100 User manual

Nikon Eclipse L150 User manual

Nikon

Nikon Eclipse L150 User manual

Nikon ECLIPSE E200 MV Series User manual

Nikon

Nikon ECLIPSE E200 MV Series User manual

Popular Microscope manuals by other brands

Zeiss IV FL instructions

Zeiss

Zeiss IV FL instructions

KERN Optics OZM-98 Series User instructions

KERN Optics

KERN Optics OZM-98 Series User instructions

Jeulin DELio 601 871 manual

Jeulin

Jeulin DELio 601 871 manual

Celestron 44308 quick start guide

Celestron

Celestron 44308 quick start guide

Supereyes B006 user guide

Supereyes

Supereyes B006 user guide

KERN Optics OZM 912UK User instructions

KERN Optics

KERN Optics OZM 912UK User instructions

NetPeppers WFM-100 quick start guide

NetPeppers

NetPeppers WFM-100 quick start guide

Carson zPix 300 user guide

Carson

Carson zPix 300 user guide

Cole Parmer MSS-400 Series Installation and operating instructions

Cole Parmer

Cole Parmer MSS-400 Series Installation and operating instructions

THORLABS FS200 user guide

THORLABS

THORLABS FS200 user guide

3B SCIENTIFIC PHYSICS 400 1003276 instruction manual

3B SCIENTIFIC PHYSICS

3B SCIENTIFIC PHYSICS 400 1003276 instruction manual

Ken A Vision T-1955 brochure

Ken A Vision

Ken A Vision T-1955 brochure

Keeler KSL-H Instructions for use

Keeler

Keeler KSL-H Instructions for use

Park Systems XE-100 user manual

Park Systems

Park Systems XE-100 user manual

LEUCHTTURM DM 5 operating instructions

LEUCHTTURM

LEUCHTTURM DM 5 operating instructions

Optika B-290 Operation manual

Optika

Optika B-290 Operation manual

Leica SP8 manual

Leica

Leica SP8 manual

LW Scientific Revelation III LED Specifications

LW Scientific

LW Scientific Revelation III LED Specifications

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.