Nikon L-ke User manual

Nikon
MICROSCOPE
MODELL-Ke
INSTRUCTIONS
NIPPONKOGAKUK.K.
Copy offered by courtesy of Dr. Steven Horii, USA

Caution
inHandlingandMaintenance
•Avoidtouchinglenssurfaceswiththefingersorwithanyroughmaterial.
Fordusting,usea featherora softlensbrushandthena washed-out
softcottonclothonlytowipeofffingermarksorgrease,wetthecloth
withxylor,butneverwithalcoholorether.
Themicroscopebodysurfacesshouldbedustedinthesamewayand
mayslightlybeoiled,
•Tightnessofthecoarsefocusingknobsshouldbeadjusted,inthistype
(L-Ke)ofmicroscope,bymeansoftheadjusting
ring,
notbytwistingthe
knobs.
•Dismantlingoftheinternalopticalpartsandthemicroscopebodyshould
notbeattempted,becauseitmayinterferewiththeperformanceof
theinstrument.Itcanbeaccomplishedonlybyanexpertortheoriginal
manufacturer.
•Donotapplyanygreaseofnotspecifiedtypetotheslidingsurfacesof
thecoarsefocusingadjustmentandofthefloatingstage.Ifnecessary,
askforyourdealerorthemanufacturer.
•Themicroscopeistobeusedata dust-freeplace.Avoidanyforcible
manipulationofthemovingparts.Atalltimestheinstrumentistobe
handledcarefullye.g.forcarryingthemicroscope,holditsbasewithone
handanditsarmwiththeother.
•Whennotinuse,itshouldbecoveredwiththevinylcoverorkeptin
thewoodencabinetwhichisavailableonorder.Storeitata dryplace.
Itisrecommendedtokeeptheobjectivesandeyepiecesina desiccator.
NOMENCLATURE3
VARIOUSCOMBINATIONS5
CONTENTS
1-InterchangeableEyepieceTubes6
"U"
Trinocular6
"T"
Trinocular6
"B"
Binocular6
"1"
InclinedMonocular6
2.
InterchangeableStages6
"R"
RectangularMechanical6
"G"
Graduated,CircularRotatable6
"C"
CircularFloating6
"P"
SquarePlain6
OBJECTIVES,EYEPIECES,CONDENSERS7
1.
Objectives7
2.
Eyepieces7
3.CombinationsofObjectivesandEyepieces8
4.
Condensers9
ATTACHINGTHELENSES10
1.
MountingtheObjectives10
2.
MountingtheEyepieces11
3.MountingtheCondenser11
ILLUMINATION12
1.
CondenserIrisDiaphragmandImages12
2.
LightSource13
3.CondenserFocusingKnob13
4.
BrightnessAdjustment14
5.PreparationandAdjustment14
6.Observation15
7.Photomicrography15
8.OtherLightSources15
9.FilterCase15
FOCUSING16
1.
FocusingAdjustment16
2.
EyepieceAdjustment16
3.CoarseFocusing17
4.
PresetDevice17
5.FineFocusing•••18
6.OilImmersion18
7.LongWorkingRangeoftheStage19
8.ExchangingtheStage19
MOVINGTHESPECIMENONTHESTAGE19
1.
RectangularMechanicalStage"R"19
2.
CircularFloatingStage"C"20
EXCHANGINGTHEOBJECTIVENOSEPIECE••••21
PHOTOMICROGRAPHY••22

—3
NOMENCLATURE(MODELLBR-Ke)
Eyepiece
—4 —

VARIOUS
COMBINATIONS
TheNikonMicroscopeModel
L-Keis
available
in
variouscombinations
with
objectives,eyepieces,condensers,eyepiecetubes,
and
stages.
Forex-
ample.
Model
LBR-Ke
consists
of
Model
L
microscopestandwithKoehler
illuminator
built-inbase,Binoculareyepiecetube
and
Rectangularmechani-
calstage.
Rectangular
Mechanical Circular
Floating Graduated,
Circular
Rotatable SquarePlain
—
5
—
1.
Interchangeable
Eyepiece
Tubes
U=Trinocular
Magnificationfactor
IX.Has
provision
for
dioptercompensation
and
interpupillarydistanceadjustmentfrom
54mm
to
74mm.
Observation
binocularsinclined45°,photo-tubeupright,360°rotatable.Withbuilt-in
sliding
prismsystem,lighttransmission
canbe
switched
over
three
ways
to
permitphotomicrographythroughverticaltubewhileviewing
throughbinoculartube;
100%
lightdirected
to
observationbinoculars
byswitching
over
lightpath;
or
wholelighttransmissiondirected
to
verticalphototube
for
photomicrography,micro-projection
or
closed-
circuit
T.V.
pickup.
T—Trinocular
Magnificationfactor
IX.Has
fixedbeam-splitratio,
50%to
observa-
tion
binoculartubes
and50%to
verticalphotomicrographictube.
In-
clined45°
and
rotatable
360°.
Has
provision
for
dioptercompensation.
Interpupillarydistanceadjustmentfrom
54mm
to
74mm.
B
=
Binocular
Magnificationfactor
IX.
Inclined45°
and
rotatable
360°.
Has
provision
fordioptercompensation.Interpupillarydistanceadjustmentfrom
54mm
to
74mm.
I
=
IndinedMonocular
Interchangeablewithothertypeeyepiecetubes.Inclined
45°and
rotatable
360°
2.
Interchangeable
Stages
R^RectangularMechanical
Stagesurface
130mm
X
140mm.
Has
low-positionedcoaxial
X andY
motion
controlswhichprovideexceptionallyfine,smoothcrosstravel
within
range
of
50mmx75mm.
Scales
graduated
to
0.1mm
on
vernier.
G
=
Graduated,CircularRotatable
Stagesurface
140mm
in
diameter.360°rotatable.Goniometerdivided
into
1°increments
and
reads
to6'
withvernier.Centerablestagepro-
videdwithclampingscrew.Suppliedwithstageclips.
C^Circuiar
Floating
Stagesurface
140mm
in
diameterProvidedwithstageclips.Accepts
attachablemechanicalstageavailable
on
order.Movessmoothly
in
anydirectionwithincirclediameter
of
18mm
in
straightand/orrotat-
ing
motion.
Canbe
clamped
inany
desiredposition.
P^Square
Plain
Stagesurface
130mm
X
130mm.
Providedwithstageclips.Accepts
attachablemechanicalstageavailable
on
order.
-
6 -

OBJECTIVES,
EYEPIECES,
CONDENSERS
1.
Objectives
a.
OBJECTIVESformedical andbiologicaluse
Type Individual
Magnification NumericalFocal
ApertureLength Working
Distance Remarks
Achromatic
Dry
Oil-immersion
4X
"
lOX
20X
40X
S40X
NC40X
lOOX
SIOOX
lOOX
0.10
0.25
0.40
0.65
0.65
0.65
1.25
1.25
1.25
28.3mm
14.8mm
7.5mm
4.3mm
4.3mm
4.3mm
1.8mm
1.8mm
1.8mm
9.50lim
7.10mm
5.70mm
0.54mm
0.54mm
0.52mm
0.16mm
0.16mm
0.16mm
Spring-loaded
Nocoverglasstype
Spring-loaded
Withirisdiaphragm
for
dark-field
PlanAchromatic
Dry
Oil-immersion
Plan
1.2X
2X
3X
4X
lOX
40X
NC40X
lOOX
0.03
0.05
0.08
0.10
0.25
0.65
0.65
1.30
35.8mm
42.3mm
37.7mm
29.5mm
15.6mm
4.0mm
4.1mm
1.6mm
29.7mm
35.6mm
28.6mm
18.2mm
7.0mm
0.24mm
0.32mm
0.12mm
Spring-loaded
Spring-loaded,
no
coverglasstype
Spring-loaded
PlanApochromatic
Oil-immersion
Apochromatic
Dry
Oil-immersion
lOOX
40X
lOOX 0.80
„_1-40
1.6mm
4.3mm
L6mm
0.12mm
0.19
mm'
0.10mm
Spring-loaded
Spring-loaded
Spring-loaded
Long-WorkingDistanceType
Achromatic
Dry
LWD40X
4.0mm
2.0mm Fortissuecultureobservation
Phase-Contrast
Achromatic
Dry
Oil-immersion
Long-WorkingDistance
Achromatic
Dry
DLLIOX
B-MIOX
DLL20X
B-M20X
DLL40X
D-M40X
B-M40X
DLLIOOX
D-MIOOX
B-MIOOX
D-M40X
0.30
0.30
0.40
0.40
0.65
0.65
0.65
1.25
1.25
1.25
0.60
15.9mm
15.9mm
8.2mm
8.2mm
4.4mm
4.4mm
4.4mm
1.8mm
1.8mm
1.8mm
4.0mm
6.4mm
6.4mm
4.5mm
4.5mm
0.54mm
0.54mm
0.54mm
0.16mm
0.16mm
0.16mm
2.0mm
DarkContrast
BrightContrast
DarkContrast
BrightContrast
DarkContrast,Spring-loaded
DarkContrast
//
BrightContrast
DarkContrast
DarkContrast
//
BrightContrast
n
DarkContrast
if
(Fortissueculturework)
b.OBJECTIVESformetallurgicaluse
Metallurgy
Achromatic
Dry
M5X
MlOX
M20X
M40X
MIOOX
0.10
0.25
0.40
0.65
1.25
25.0mm
14.8mm
7.5mm
4.3mm
1.8mm
15mm
7.1mm
5.7mm
0.52mm
0.16mm
PlanAchromatic
Dry
PlanMlOX
PlanM40X
PlanMIOOX
0.25
0.65
1.30
15.6mm
4.1mm
1.6mm
7.0mm
0.32mm
0.12mm
Spring-loaded
Spring-loaded
2.
Eyepieces
Type Individual
Magnification Focal
Length Field
Number Remarks
Higheyepoint,compensating HK5X 50mm 21.0 Withadjustableeyepiececollar
Higheyepoint,compensating,
wide-field HKWIOX 25mm 18.0 Withadjustableeyepiececollar
Higheyepoint,compensating,
wide-field HKW15X 16.7mm 14.0 Withadjustableeyepiececollar
Compensating,wide-field WFIOX 25mm 18.0 High-powerpurpose
Compensating K20X 12.5mm 8.0 Generalpurpose
Huygenian H5X
HlOX
H15X
50mm
25mm
^17nim
21.0
12.0
8.0
Generalpurpose
Generalpurpose
Generalpurpose
Diopteradjustable,high
eye-
point,compensating,wide-field DHKWIOX 25mm 18.0 Suppliedtogetherwith
"U"
eyepiecetube
With
5X,lOX,15X
pictureframespluscross-
lines
for
framing
and
focusinsf
7
-
3.
CombinationsofObjectivesand
Eyepieces
Totalmagnifyingpowerobtainedbythecombinationistheproductof
individualobjectivepowermultipliedbyindividualeyepiecepower.A
selection
ofthecombination
will
bedecidedsoastogetthehighest
resolutionoftheimage(resolvingpower),thelargestextentofobject
area
(realfield)whichcanbeobservedwithoutmovingthestageor
slide,
orthegreatestthicknessofobject(depthoffocus)whichcan
be
distinctly
seen
withoutraisingorloweringthemicroscopestage,
depending
uponthepurposeofusingthemicroscope.Shownbelow
are
theresultscompiled
from
thedifferentcombinationsofobjectives
and
eyepieces:
Objective Eyepiece Total
Magnifying
Power
Working
Distance
Resolution
or
MinimumResolved
Distance RealField
ofView Depth
of
Focus
Objective Eyepiece Total
Magnifying
Power (mm) inobject
(^) inimage
(mm) (mm) (^)
4X 5X
lOX
15X
20X
40X
60X 9.5
2.7-5.5
0.05-0.11
0.11-0.22
0.16-0.32
5.25
4.5
3.5
100
64
52
lOX 5X
lOX
15X
50X
lOOX
150X 7.1
1.1-2.2
0.05-0.11
0.11-0.22
0.17-0.33
2.1
1.8
1.4
16
10
8
20X 5X
lOX
15X
lOOX
200X
300X 5.7
0.69-1.38 0.07-0.14
0.14-0.28
0.21-0.42
1.05
0.9
0.7
6
4
3
40X 5X
lOX
15X
200X
400X
600X 0.54
0.42-0.84 0.08-0.17
0.17-0.34
0.25-0.50
0.52
0.45
0.35
1.8
1.2
1.0
lOOX 5X
lOX
15X
500X
lOOOX
1500X 0.16
0.22-0.44
0.11-0.22
0.22-0.44
0.33-0.66
0.21
0.18
0.14
0.6
0.44
0.38
•
Theworkingdistanceisthe
clearance
betweentheuppersurfaceofthe
cover
glass
andthelowestedgeoftheobjectivecriticallyfocused.Note
that,asshownintheabovetable,theworkingdistancebecomesvery
small
for
high
powerobjectives.
•
Theresolutionofminimumresolveddistance(thelimitofresolvingpower)
is
theminimumdistancebetweenobjectpointsdiscernibleasseparate
under
themicroscopeilluminatedbythe
light
ofwavelength
550myu.
Theshorterthe
light
wavelength,thehighertheresolvingpower,that
is,
thesmallresolvingdistanceresults.Inthetable,thesmallervalues
indicate
theresolutionobtainedbyobliqueandthelargervaluesby
central
illumination,(see"Illumination"onpage11)
#The
minimumresolveddistanceintheimageisthevalueintheobject
multipliedbythetotalmagnificationofthemicroscope.Iftheresolving
powerofthemicroscopeisimportant,choosetheeyepiecebytheuse
ofwhichtheimageresolutioncomes
within
thatofthenakedeye
0.15
—0.3mm
(whentheobjectis
seen
from
thedistance
25mm);
generally
accepted
criterion ontheupperlimitofthetotalmagnificationofa
-8-

microscopesetupisabout
500—lOOOX
ofthenumericalapertureofthe
objectivetobeused.Notethatinphotomicrographyitisuselessto
raisethemagnificationbeyondtheresolvingpoweroftheemulsion
(usuallyabout
0.05mm).
However,
sincetheresolutionoftheemulsion
is
higherthanthatofthenakedeye,photographsareusuallytakenat
alowermagnificationandthereaftertheyareenlarged.
•
Realfieldofview(inmm)representstheextentoftheobjectthatcomes
underobservation.Inhighermagnificationitbecomesextremely
small.
Consequently,itisadvisabletotakeaimattheobjectpointtobeexa-
mined
firstunderlowermagnificationandthenrevolvethenosepieceto
highermagnification.
•
Whenfocusisonthecenterofthe
field,
itscircumferencewillusuallybe
blurred,
becausea curvatureoftheimageplaneisunavoidableinthe
microscope,exceptwhenusinga plantypeobjectives.Inordertoget
sharpcornerimage,itisnecessarytoadjustthefinefocusingknoband
switchthefocusfromthecenterontothecorners.
•
Depthoffocusrepresentsthethicknessorheightoftheobjectinfi sharply
seenwhenobservationismadethroughthemicroscope.Inphotomicrogra-
phythedepthoffocusbecomessmallerthanthefigureshowninthe
previoustable.Therefore,carefulattentionmustbemadeinfocusing
when
taking
microscopepictures.
Byclosingthecondenserdiaphragm,thedepthoffieldcanbemade
greaterthanthevalueshowninthetable.
4.
Condensers
Type
NumericalAperture Remarks
Abbe
1.30
Twolenses For centralillumination(withoutobliqueilluminationslider)
Abbe
1.30
Twolenses For centralandobliqueillumination(withoblique
illumi-
nation
slider)
Aplanatic
1.40
Three
lenses For high-classmicroscopy
Achromatic
1.25
Five
lenses For high-classmicroscopy
Achromatic
(longfocal-lengthtype)
0.7
Four
lenses Longfocal-lengthtype.Workingdistance12mm.With
turret-mounted
annulardiaphragmsforphase-contrastand
phase-interferencemicroscopy.
UniversalDark-Field
Condenser
1.20-1.40
for
dark-field
Suppliedincenterablemount.Withouterdiameter36.8
mm.
ObjectivestobeusedlOXupto
lOOX.
Ideallysuited
for
fluorescencework.
lOOX
objectiveusedshouldhave
built-in
adjustableirisdiaphragm.Thicknessofslideglass
to
beused,mustbelessthan
1.2mm.
Low-Power
Condenser
Singlelens For low-powermacro-objectives,e.g.,1.2X,2X,& 3X
PlanAchromatic
These
condensersarenotonlycapableofconcentratinglight-beamforbetter
illumination
oftheimage
field,
butalsogivesa greatinfluenceontheresolu-
tion
ofmicroscopeimage,imagecontrastanddepthoffocus.Inprecise
observationandphotomicrography,theuseofanachromaticoraplanatic
condenserwiththemaximumnumericalaperture1.25providedwithan
obliqueilluminationdeviceanda filtercaseisspeciallyrecommended.
—
9 —
ATTACHING
THELENSES
Beforeattachingtheobjectiveandtheeyepiecetothemicroscope,first
cleantheirouterlenssurfaces.
Even
a slightlystainedfingermarkmay
often
interferewiththeimagecontrast.
1.
Mounting
the
Objectives
Take
specialcareinhandlingtheobjectives.
Beforeattachingtheobjectivesonthenosepiecerevolver,lowerthe
microscopestagesufficiently.Securingeachobjectivewiththefingers
ofonehand,screwit
into
eachnosepiece
holewiththoseofthe
otherhand(Fig.4).
It
isrecommendedto
mount
theobjectives
onthenosepieceor-
derlyfromlowtohigh
powerssothatthe
magnification
ofeach
objectiveaugmentsas
the
revolverisrotated
clockwiseorcounter-
clockwise.
MicroscopeModelL-Kehasonitsuppersurfaceofthenosepiecerevolver
five
blankspots(Fig.5)oneachofwhichtheobjectivemagnificationcan
beInkedso
that,
by
looking
theseIndica-
tions
fromabove,the
objectivelocatedbelow
canbeidentified
with-
out
havingtoturn
away
your
eyes
from
the
eyepiece.
Whenrotatingthere-
volver,holditsouter
milled
rimwith
your
thumb
andfirstfinger,
but
donotpushthe
objectivebarrels,other-
wisealignmentofthe
objectivesmaybe
trou-
bled.
-10-

2.
Mountingthe
Eyepieces
Formounting,simplydrop
the
eyepieceinto
the
eyepiecetube.
Itis
recommended
to
leave
an
eyepiece
in
placeevenwhen
itisnotinuse,in
order
to
prevent
the
entrance
of
dustinto
the
eyepiecetube.
Or
put
onthe
eyepiece
cap
in
place
ofthe
eyepieceremoved.The
inclinedeyepiecebody
tube,
trinocular,bino-
cular
or
monocular,can
berotatedafterloosen-
ing
the
clampscrew
(Fig.
6)for
conveni-
ence
in
viewingfrom
anydesireddirection
without
moving
the
microscopestand.
By
furtherreleasing
the
clampscrew
the
bodytube
can
be
removed
and
replacedwithanothertypeeyepiecebodytube.
3.
IVIounting the
Condenser
To
mount
the
condenser,unlock
the
lockscrew,
and
insert
the
condenser
beneaththe condenser
holder
asfarasit
will
go.Then,tighten
the
lockscrew.
In
this
caselocate
the
dia-
phragmeccentering
lever
andits
screw
at
aproperplace
soas
to
facilitatetheir
manipulation
with
one
hand
(Fig.7).The
correctdistance
tore-
tain
immersion
oilbe-
tween
the
lowersur-
face
ofthe
slide
and
the
topof
condenser
issecuredwhen
the
condenser
is
raised
tothe
upper
limit
by
turning
the
condenserfocusing
knob.
-11—
ILLUMINATION
Resolution
and
contrast
of
image
are
greatlyaffected
bythe
method
of
illumination.
1.
Condenser
IrisDiaphragmand
Images
Stopdown
the
condenseririsdiaphragm
and
slide
itinthe
radialdirec-
tion
fromcenter
to
edge.
The
farther
the
irisdiaphragm
is
off-centered,
the
higherbecomes
the
contrast
and
resolution,whichdistinguishdetails
of
object
by
increased
and
unsymmetricalshadow
atthe
boundary
of
object
(Fig.
8,a,b).
(a)
Centralillumination(b)Obliqueillumination(c)Darkfieldillumination
When
the
irisdiaphragm
is
positioned
soastoletthe
lightbundleenterinto
the
object
atan
angle
o'f
incidence
the
same
asthe
apertureangle
ofthe
objective
the
resolutionreachesmaximum
and
twice
as
much
asthe
resolution
by
centralillumination.
If
the
diaphragm
is
furthereccentered
to
such
an
extent
asnotto
introduce
the
lightbundledirectlyinto
the
objective,darkfield
illumi-
nationwill
be
obtained.
Ifthe
irisdiaphragm
is
widelyopened,images
byvariousangleilluminations
are
integrated.
An
illuminationangle
not
favorable
forthe
object
maybe
included.
In
the
centralillumination
the
maximumresolution
is
obtainedwhen
the
opening
ofthe
irisdiaphragmjustcorresponds
tothe
apertureangle
of
the
objective.
In
thiscase
excessive
outer
rays
tobe
used
as
dark
field
illumination
arecutoffand
flare
is
minimized.
Ifthe
opening
is
made
smaller,
the
contrast
is
enhanced,though
the
resolution
is
lowered.
But
if
the
irisdiaphragm
is
largeenough
to
cover
60-70%
ofthe
objective
aperture,
the
decrease
of
resolutionwill
notbe
remarkable.
If
the
diaphragm
is
stoppeddown
to
minimum
for
admittingonly
very
—12—

small
lightbundles,theeffectsbydiffraction,reflection,refraction,etc.,
maybeexaggeratedsothatfringesmaybeseenattheimageedgeswhich
maylikelyinducemisinterpretationoftheimage,butitmaybeeffective
forspecialoccassions(e.g.definitionofgeneralstructureofnonstained
specimens).(Fig.8,c).
2.
Light
Source
Asalreadycited,theirisdiaphragmplaysanimportantroleInthe
illumi-
nation
onmicroscopy.Asa principlethediaphragmshouldbesoadjusted
that
thenumericalapertureofthecondenserisequaltothatofthe
objectivebeingused,inordertoobtainthemaximumresolution.In
practice,however,tokeep
outstraylightwhichwould
reduceimagecontrast,
closingtheapertureofthe
condenserdownto60-
70%
ofthatoftheobjec-
tivelenswillbringabout
agoodresultinmost
cases.Thecoincidence
ofcondenserdiaphragm
aperturewiththeopening
(exitpupil)oftheobjec-
tivecanbeascertained
bylookingthroughthe
microscopetubeafterre-
moving
theeyepieceand
closingthediaphragm
slowly.Anexperienced
user,however,maydispensewithsucha procedure,andwillobtainthe
sameresultbyadjustingthediaphragmopeninguntilsatisfactorydistinct-
nessoftheimageisobtained.
Ifa highresolutionand,atthesame
time,
a highcontrastaredesirable,
obliqueilluminationwillbeeffective.
This
issuitedespeciallyforlowdyed
specimens,transparentphase-contrastspecimens,etc.
However,
inthis
illumination,
a remarkablepolarityincontrastandresolutionmayappear;
it
isnecessaryforobservationtochangethedirectionofillumination,by
turning
theirisdiaphragm.
Thecondenserapertureisoff-centeredinanydirectionbyrotatingandat
the
sametimebyradiallyslidingthediaphragm.
This
manipulationcanbe
doneonlybyusingonehand,thethumbandthefirstfingerforoff-centering
andthemiddlefingerforopeningorclosingthediaphragm.(SeeFig.9).
3.
Condenser
Focusing
Knob
Thecondenserfocusingismadebyturningthecondenserfocusingknob.
This
manipulationisnecessaryonlyincaseofKoehlerilluminationordark
-13-
field
observation.Thecondenserisusuallytostayattheupper
limit
and
neednottobelowered,exceptwhena strayimageofanoutsideobject
superimposesthespecimenimagetoosharplyinsucha caseasindirect
sunlightilluminationorassomeunevenbrightnessinterfereswiththe
observation.
4.
Brightness
Adjustment
Forthispurposeadjustthevoltageofcurrentsupplybyregulatingthe
transformeroftheilluminatororbyusinga properneutraldensityfilter.
Adjustmentbymeansofthecondenserdiaphragmisnotatallpossible.
5.
Preparation
and
Adjustment
(1)LampandSocket
AsshowninFig.10,lining
upthereddotonthesocket
retaining
ringtothaton
the
lamphousing,fitthe
sockettothehousingby
bayonetfashion.
(2)CenteringandFocusingof
the
Lamp
Bringtheillumination
change-overleverinthe
position"M".Stopdown
the
condenserdiaphragm
beneaththemicroscope
stage.Movethelampback
andforthtobringtheimage
ofthefilamentintosharp
focusonthebottomsur-
faceofthediaphragm.For
this
purpose,itisconve-
nienttoutilizethereflection
onthe
accessory
reflecting
mirror,
asshowninFig.11.Then,turnrightorleftonlythesocket
ring
tofastenthelampsocketinposition.
Bringtheilluminationchange-overlevertotheposition"H".Manipu-
lating
thelampcenteringscrews,makecenteringoftheimageonthe
condenserdiaphragm.
(3)CenteringoftheIlluminationFieldDiaphragm
Stopdowntheopeningoftheilluminationfielddiaphragmtoabout
2mm
indiameter.
Viewingthroughthemicroscope,focustheobjective40Xtotheimage
ofspecimen.
Movingupanddownthecondenser,makea sharpfocusedimageof
the
illuminationfielddiaphragmonthespecimen.
—
14—

Manipulatetwocenteringscrewsuntiltheimageoftheillumination
field
diaphragmcomestothecenter.
Youneednotmindabouta slightdeviationfromthiscenteringcaused
byswitching
over
tootherobjectivesbecausethiswillnotproduce
anyobjectionableresult.
6.
Observation
Forusualcaseofobservationwithobjectivesfrom4Xto
lOOX
(oil-immer-
sion),itsufficestoleavethechange-overleveratthepositionL.Whenan
exactilluminationorextremebrightnessisrequired,asisthecasewith
photomicrography,itisnecessarytomakeadjustmentforeachmagnifi-
cationbyswitching
over
thelever.
7.
Photomicrography
Makeexactadjustment,dependinguponthemagnificationoftheobjec-
tivebeingused,byswitchingthechange-overleverasbelow:
Objectivemagnification4X
10-20X
40-lOOX
Positionofthechange-overleverL Mi H
i
Whenusinganobjective3Xorlower,setthechange-overlevertoH
andreplacethecondenserwiththeonespecificallydesignedforlower
magnification.
TheilluminationfortheModelL-Kewillnotchangeitscolortemper-
aturebythechange-overabovementioned.
8.
Other
Light
Sources
Whenusinganyother
light
sourceoutsideof
the
microscope,replace
the
illuminationfield
lenswiththe
accessory
reflecting
mirror,as
showninFig.12.
9.
Filter
Case
Afilterofthediameter
45mm
canbemountedontheilluminationfield
—15—
lens.A filterofthediameter
33mm
canbeplacedinthecondenseror
in
thefiltercaseofthemicroscope.
This
filtercasetobeattachedonto
the
microscopebase isavailableonorder.
FOCUSING
1.
Focusing
Adjustment
ThemicroscopemodelL-Keisprovidedwithcoaxial,coarseandfine
focusingknobs,bothofwhicharelocatednearthemicroscopebase(Fig.
13).
Forwardrotationofeitherofthefocusingknobsbytheoperatorraises
the
microscopestageandvice
versa
(Fig.13,14).
2.
Eyepiece
Adjustment
Whenusinga binocularortrinoculareyepiecetubeforobservationthe
adjustment
oftheuser'seye-sight(diopter)discrepancybetweenthe
right
andleft
eyes
isnecessarywhichismadebyrotatingtheadjusting
ring
onthelefthandeyepiece.
Afterfocusingwith
your
righteyebyraisingorloweringthemicroscope
stage,turntheadjustingringleftorrighttoobtainthesharpimagewith
your
lefteye,too.Then,regulatetheinterpupillarydistanceofthebino-
cularortrinoculartubebyslidingtheeyepiecesleftorrightbymeansof
the
knob(Fig.15),untiltheviewfieldsofbotheyepiecesmergetogether.
It
willbeadvantageoustomemorizetheattaineddiopterandinter-
pupillarydistancereadingsforfutureuse.
Thereddotengravedontheinterpupillarydistancescaleindicatesthe
positionwherethemechanicaltubelengthbecomesexactly
160mm.
The
HK(higheyepointtype)eyepieceshaveaneyecupontop,therotationof
—
16—

whichwillgivepropereye-to-lensdistance.
For
thoseweareyeglasses,
eyecupshould
be
screwed
in.
3.
Coarse
Focusing
Thecoarseadjustment
maybe
eased
or
tightened
by
means
ofthe
adjusting
ring:
If
the
revolution
ofthe
coarsefocusingknob
istoo
loose,turn
thead-
justing
ringcounterclockwise.
Too
muchtightness
maybe
adjusted
bythe
clockwiseturning.
Never
twist
the
focusingknobs
for
thisadjustment
asinthe
traditional
microscopewhosefocusingknobs,coarse
and
fine,
are
locatedseparate
(notcoaxial).Focusing
maybe
performed
as
follows:First,raise
the
microscopestageuntil
the
distancebetween
the
specimen
andthe
objectivebecomesless longerthan
the
workingdistance
ofthe
objective
to
be
used
(see
table
onp.8),
thenlookingthrough
the
eyepiece,lower
the
stageuntil
the
specimen
tobe
examined
is
plainlyvisible.
4X,
lOX,20X,40XandlOOX
objectives
are
parfocal
, andare
approxi-
mately
in
focuswhenrevolvedintoposition
one
afteranother,
theuse
of
the
finefocusingknobonlybeingrequired
for
criticalfocusing.
4.
Preset
Device
Theright-handfocusingknob
hasa
presetlever
onits
drum
(Fig.16).
When
the
lever
is
fastened
by
turningclockwise
(as
indicated
bythe
arrow
engravedbesides)until
it
stops,
the
coarsefocusingknobscannot
be
turned
in
the
direction
to
drive
the
stagecloser
tothe
objective.
This
presetting
is
—
17--
utilized
for
quickrefocusing
after
thestage
has
beenlowered
andoutfocused
for
changing
the
specimen
or
applying
im-
mersionoil.
The
presetdevice,
whenfastened,prevents
atthe
same
time
thedanger
of
damag-
ingthe
objectivefront
and
slide
glass.
5.
Fine
Focusing
Manipulation
ofthe
finefocus-
ing
knob
is
necessary:
a.
To
obtainthesharpestimage.
b.
To
transfer
the
focusfrom
center
toa
corner
ofthe
viewfield.
c.
To
focusupon
the
differentlayers
ofa
thickspecimen.
d.
To
correct
a
slightblurringwhich
may
takeplacewhenshifting
the
slide.
e.Tomeasure
the
thickness
of
objectunderexamination.
Themicroscope
isso
designedthat
one
revolution
ofthe
finefocusing
knob
raises
or
lowers
the
microscopestage
0.2mm.
This
permitsdirect
reading
onthe
right-handknobscale
upto
0.002mm
(2«).The
whole
range
of
finemovement
is
38mm;
the
same
asof
coarsefocusing.
6.
Oil
Immersion
Whenusing
lOOX
objective,
the
application
of
immersion
oilinthe
minute
space(0.1mm)between
the
objective
topandthe
cover
glass
is
necessary
to
attain
the
specifiednumericalaperture.
For
criticalwork
the
immersion
oilistobe
filledbetween
thetop
lens
ofthe
condenser
and
the
slide
as
well
as
between
the
objective
andthe
cover
glass.
Oilim-
mersionobservation
is
performed
as
follows:First,using
lOXor40Xob-
jective
(dry
system),bring
the
specimen
in
focus
andinthe
center
ofthe
viewfield.
Setthe
presetlever
by
turningclockwise.
Lower
the
micro-
scopestage
and
revolve
the
nosepiecerevolver
tolOOX
objective.After
applying
a
drop
of
immersion
oil
onto
the
cover
glass,raise
the
stage
up
to
the
preset
limit.
Then,focus
upby
lookingthrough
the
eyepiece
and
raising
carefully
the
stage
by
manipulating
the
finefocusingknob.
The
oil
immersion
lOOX
objective
is
designed
to
attain
its
criticalfocusing
by
about
1/3
forwardrotation
ofthe
finefocusingknob,that
is,
bringing
the
stageabout
0.08mm
closer
tothe
objectivefrom
the
parfocalposition.
Airbubbles
inthe
immersion
oil,
which
may
sometimesintervene
the
microscopeimage
andare
visiblewhenlookinginto
the
microscope
tube
without
the
eyepiece,
canbe
removed
by
repeatingslightmovement
of
the
nosepiecerevolver,
by
adding
a
certainquantity
of
immersion
oil
or
by
means
ofa
needle.
Remaining
stiffened
oilmay
oftenimpair
the
image.Therefore,
im-
18

mediatelyafterfinishingthework,cleanofftheremainingoilfromthelens
usinga softcottonclothwettedwithxylor.Neverusealcoholorimmerse
thetopoftheobjectiveinxylor.
Becarefulnottouseimmersionoilthathasbeenthickenedbyage.
Therefractiveindexoftheimmersionoilshouldbe
1.515.
7.LongWorkingRangeoftheStage
ThemicroscopeL-Keis
designedtopermitlarge-
rangeloweringofitsstage
(71mm)forconvenience
inobservinga thickspeci-
mensuchasformetallurgy
orculture.Unlockthelock
screwbeneaththestage
(Fig.
17),andthestage
canbelowered33mm^^^^^^^
andbymanipulationof
^^W^^^^^gnm^^j^^^^
'[J
thecoarsefocusingknob
farther38mm.
Thestageloweredtosucha greatextentmaybeusefulinmacrophoto-
graphyorwhena specialstageisused.
8.ExchangingtheStage
Lowerthestagebymeansofthecoarsefocusingknobandthenunlock
thestagelockscrew.Thestage,then,canberemoved.
MOVINGTHESPECIMENONTHESTAGE
Clip
1.
RectangularMechanicalStage'*R"
Thisstageenablesfinecrosswisetravellingoftheslideintherange
of50X 75mm,allowingreading
ofthemovementdownto0.1
mmbytheuseofthevernier
provided.
Forsecuringtheslideonthe
stageinposition,openthe
clip.
Eachtravellingisperformed
byrotationoftwocoaxialknobs
iDcated
oneabovetheotheron
theverticalrodprotruded
belowontheleftsidefrom
thefront,theupperknobbeing
--19—
forlongitudinalandthelower
oneforlateraltravelofthe
slideonthestage.
Influorescencemicroscopy
orinusingoilimmersionob-
jectives,wheretheclearance
betweenthecondenserand
theslidealsoshouldbeoil-
immersed,
stiffenedoilmay
causeunsmoothtravelofthe
slide.
Inthiscase,removing
19ofthecircularopeningplate
atthecenterofthestageor
fasteningoftheclampscrew
willbehelpfultoa positive
traveloftheslide(Fig.18).
Byunlockingthelockknobon
theedgeofthestagethestage
canberotatedhorizontallyfor
convenienceintheobservation
fromtheoppositesideofthe
microscope(Fig.19),wherethe
eyepiecetubeistoberotated
180°.
Thisrotationofthestage
mayoftenbeofuseinphoto-
micrography,whenthepicture
formatischangedfrom
verti-
caltohorizontalorviceversa.
Itisrecommendedtousethe
slideadapteronthestage
(Fig.
20)fora sufficientlongi-
tudinaltraveloftheslidein
suchreversedposition.
2.
CircularFloatingStage
Thecircularfloatingstage
(Fig.
21)glidesandrotates
smoothlyandpreciselyinany
desireddirectionsimplyby
pushingtherimofthestage
bythefingers,wichina circle
of18mmindiameter.
Tofastenthefloatingstage
inposition,pressingthestage

22
downward,
turn
the
rinn
ofthe
stagecounterclockwise.
Fastening
ofthe
floatingstage
isnecessary,whenusing
an
attachablemechanicalstage
(Fig.
22),
whichwill
be
avail-
able
on
order.
Thecircularrotatingstage
type
G is
alsoavailable,which
permitsmeasurement
ofthe
rotatingangle
of
specimen
with
its
circularscale.(Fig.
23).
EXCHANGING
THE
OBJECTIVENOSEPIECE
It
may
becomenecessary
toexchange
the
objective
nosepiece
in
such
a
casewhen
aseries
of
phase-contrast
or
metallurgicalobjectives
are
mounted
asa
unit
onthe
other
objectivenosepiece.
First,unlock
the
revolver
lockscrew,
and
draw
outthe
nosepiece
inthe
direction
ofthe
arrow
(Fig.24).In
place,
in-
sert
the
othernosepiecegently
until
it
stops,making
the
guid-
ingsurfaces
ofthe
dove-tail
fitparallel
to
eachother.
Fasten
the
revolver
in
position
bymeans
ofthe
lockscrew
(Fig.
25).
-21—
PHOTOMICROGRAPHY
TheMicroscopeModel
L-Ke,
incorporating
the
Koehlerilluminationdevice
withlightsourcebuilt
inthe
microscopebase,enablesconvenient
and
perfectmicrophotographonly
byadditionally
mounting
a
cameraconnected
to
the
microscopeeyepiecewith
any
photomicrographicadapter.
Therefore,whentakingphotographs
ofthe
microscopicimage
on
35mm
film,
itis
recommended
tousethe
NikonMicroflexModel
EFM
(withbuilt-in
exposuremeter)
orAFM
(withbuilt-inexposuremeterpermittingautomatic
exposuresetting)
andthe
Nikon
F or
Nikkormatcamera
or
NikonDark
Box
for
usingwith
the
aboveattachments.
For
details
of
photomicrographic
methods,refer
tothe
"Instructions
of
UsingNikonMicroflex
EFM"or
"AFM"
andany
generalworks
on
photomicrography.
-22—

7r
O
Q
•D
Q
o
o
Table of contents
Other Nikon Microscope manuals

Nikon
Nikon 50i User manual

Nikon
Nikon Eclipse Ti-E User manual

Nikon
Nikon N-STORM User manual

Nikon
Nikon ECLIPSE Si User manual

Nikon
Nikon eclipse Ci-S User manual

Nikon
Nikon A1 User manual

Nikon
Nikon T1-FM User manual

Nikon
Nikon SMZ-10A User manual

Nikon
Nikon Eclipse 800 User manual

Nikon
Nikon ECLIPSE E200 POL User manual

Nikon
Nikon LABOPHOT2-POL User manual

Nikon
Nikon ECLIPSE 80i User manual

Nikon
Nikon Eclipse L150 User manual

Nikon
Nikon Optiphot User manual

Nikon
Nikon Elcipse TE2000 User manual

Nikon
Nikon E800 User manual

Nikon
Nikon Ti-E User manual

Nikon
Nikon SMZ1500 User manual

Nikon
Nikon Eclipse Ci-E User manual

Nikon
Nikon Eclipse E400 POL User manual