BKI 12 ATEX 0012X/1 ▪ txp1710a0600h_02 ▪ 2 / 2
2.1 EX TEMPERATURE CLASSIFICATION
Temperature class T6 T5 T4
Max. ambient temperature +65 °C +70 °C +70 °C
Max. medium temperature +85 °C +100 °C +135 °C
2.2 ACCESSORIES
• User’s Manual
• Warranty Card
• Declaration of Conformity
2.3 ORDER CODES
THERMOCONT TXP– – 1
PROCESS CONNECTION CODE SENSOR TYPE /PT-100 CODE PROBE LENGTH CODE EX CERTIFICATE CODE
1” NPT 1 Class A, single, 2-wire 1 120 mm 0 None 0
M33x2 V Class A, dual, 3-wire 4 160 mm 1
Class A, single, 4-wire 7 200 mm 2
Ex d ia / ½” NPT
without cable glands 62
250 mm 3 Ex ia 7
1The order code of an Ex version should end in „Ex” 300 mm 4 Ex d ia 8
2Only allowed to use with approved cable glands with “d” protection type 350 mm 5 Ex d 9
400 mm 6
450 mm 7
500 mm 8
2.4 DIMENSIONS
TXP-1-TXP-V-
Ø14
Ø28
1"NPT
SW41
Ø21.4
Ø53
E
C
151
281
75
Ø28
M33x2
L (max. 500 mm)
88
86
22
(2)
SW41
Label of the incorporated sensor
99.1.4.
A
B
The sensor has a unique
manufacturing identification number.
LEGENDS:
A: Drawing number,
B: Identification number
C: Dimension is depending on the
applied cable glands
D, E: As per order codes
5. WIRING
A
B
A
A: No.1 Pt100
B: Marking
C: No.2 Pt100
B
Pt 100
Wiring of 2-wire single
temperature sensor
Wiring of 4-wire single
temperature sensor
5
2
4
6
3
1
Pt 100 Pt 100
Wiring of 3-wire dual temperature sensor
3. MOUNTING
The temperature sensor instruments have threaded process
connection. Sealing of the wiring cables are provided by the
cable glands. The cable glands can be normal or Ex approved
type.
4. SPECIAL CONDITIONS FOR SAFE USEAGE
Do not remove the housing cover when energized!
The housing of the device must be connected to the
equipotential network.
The device must be operated only with using properly sealed
Ex d IIC or Ex d IIB certified protective cable gland.
The unused cable gland connection should be closed with
certified closing bolt.
When the wiring is finished and the housing cover is closed
the security chain should be mounted.
Take into consideration that process connection of the
instrument could be the same as the temperature of the
medium to be measured. So the temperature class of the
applied instrument depends on the highest medium
temperature.
Tightening torque for the NPT 1” process connection types
should be 45 Nm.
The II 1/2 G Ex d ia IIB T__ Gb approved temperature sensor
devices must be operated from certified [Ex ia] IIC
or [Ex ia] IIB approved intrinsically safe circuits complying with
the following technical specifications:
Uomax <= 30V, Iomax <= 100mA, Pomax <= 750mW.
6. MAINTENANCE, REPAIR
The devices do not require maintenance on a regular basis. In some
instances, the sensor probe may need occasional cleaning to remove
surface deposits. This must be carried out gently, without harming the
sensor. Repairs during or after the warranty period are effected at the
Manufacturers.
7. STORAGE CONDITIONS
Ambient temperature: -25 °C ... +55 °C.
txp1710a0600h_02
2016. April
NIVELCO reserves the right to change
technical specifications without notice.