NoiseKen LSS-6330 Series User manual

INSTRUCTION MANUAL
INJECTIONUNIT
ForLIGHTNINGSURGESIMULATOR
LSS-6330/LSS-F03SERIES
MODEL:LSS-INJ6401SIG
NOISELABORATORYCO.,LTD. 1.02 Edition
AEE00585-00E-0

1
NOTICE
The contents of this booklet are subject to change without prior notice.
No part of this booklet may be reproduced or transferred, in any form, for any purpose,
without the permission of Noise Laboratory Co., Ltd.
The contents of this booklet have been thoroughly checked. However, if a doubtful point,
an error in writing or a missing is found, please contact us.
Noise Laboratory Co., Ltd. shall have no liability for any trouble resulting from the misuse
or improper handling of this product regardless of the contents of this booklet or arising
from the repair or remodeling of this product by a third party other than Noise Laboratory
Co., Ltd. or its authorized person.
Noise Laboratory Co., Ltd. shall have no liability for any trouble resulting from the
remodeling or modification of this product.
In no event shall Noise Laboratory Co., Ltd. be liable for any results arising from the use
of this product


1
1. IMPORTANT SAFETY PRECAUTIONS
The following is very important matters in order to safely handle this product (called the unit
hereafter). Read carefully and strictly observe them.
1. As the unit outputs high voltage (6600V maximum) and great current
(157A maximum) by surge from LSS-6330 / LSS-F03 series, carefully
handle it. Mishandling orcarelessoperationmayresultina fatalwound.
2. Use of the unit and the lightning surge simulator (LSS-6330 / LSS-F03
series) unit in an explosive area such as "No fire" area etc. is prohibited.
If used in such an area, it is liable to cause combustion or ignition due to
discharge.
3. Any person who has an implanted pacemaker in the body should not
operate this unit. Furthermore, such a person should not enter the test
area while this unit is operating.
4. Testequipmenttobeusedwiththisunitshouldbecapableofinsulatingagainstat
leastavoltageof6600V.The equipment under test (called EUT hereafter)using
this unitshould be performed ina protective enclosure or cover againstscattering
broken pieces,fireelectricshock,etc.
5. When connecting this unit to accessories for test waveforms and testconditions,
optional equipment and other equipment, press STOP button of the lightening
surge simulator (LSS-6330 series) and check that lamp of STOP button is lighting
up beforehand. Otherwise, you may receive an electric shock. There may be
residual voltage even if the test is stopped. Therefore, start operation after a lapse
ofmorethan3seconds.

2
2. ACCESSORIES
A:Accessorybag...................................................................................... 1pc
B:Surge InCable(1m) ........................................................................... 2pcs
C:LineOutCable(1m)............................................................................ 5pcs
D:ConnectionPlug................................................................................... 7pcs
E:Capacitor/GasArrestorUnit.............................................................. 1pc
F:GasArrestorUnit.................................................................................. 1pc
G:InterlockConnectionCable................................................................. 1pc
H:PECable(2m)...................................................................................... 1pc
I:InstructionManual.................................................................................. 1 pc
When it is connected with LSS-F03 series, cable 05-T1578 is required additionally.
ABC
DEF
GHI

3
3. APPLICATION FORM FOR INSTRUCTION MANUAL
Weplaceanorderforan instruction manual.
Model: LSS-INJ6401SIG
Serial No.:
Applicant:
Companyname:
Address:
Department:
Personincharge:
TelNo.:
FaxNo.
Cut off this page "APPLICATION FORM FOR INSTRUCTION
MANUAL" from this volume andkeep it for future use with care.
WhenanINSTRUCTIONMANUALisrequired,fillintheaboveApplicationFormandmail
orfaxittothefollowingsalesdepartment ofourcompany.
To:Noise LaboratoryCo.,Ltd.
1-4-4Chiyoda Chuo-kuSagamihara City,
KanagawaPref.,252-0237 Japan
Tel:+81-(0)42-712-2051 Fax:+81-(0)42-712-2050
Cut
line

4
Memorandum

5
4. TABLE OF CONTENTS
1.IMPORTANTSAFETYPRECAUTIONS...................................................................................................................1
2.ACCESSORIES.............................................................................................................................................................2
3.APPLICATIONFORMFORINSTRUCTIONMANUAL..........................................................................................3
4.TABLEOFCONTENTS...............................................................................................................................................5
5.PREFACE.......................................................................................................................................................................6
5-1. PREFACE............................................................................................................................................................................. 6
5-2. FEATURE ............................................................................................................................................................................. 6
6.BASICSAFETYPRECAUTIONS.................................................................................................................................7
7.APPEARANCEOFTHEUNIT...................................................................................................................................9
8.HOWTOCONNECTEQUIPMENT......................................................................................................................... 11
8-1 CONNECTIONOFREARPANEL.....................................................................................................................................11
8-2 CONNECTIONOFFRONTPANEL...................................................................................................................................11
8-3.CONNECTIONOFTHEFRONTPANEL(INCASEOFLSS-F03SERIES)..........................................................................12
8-4SETTINGWHENARRESTORSCOUPLING.......................................................................................................................13
8-5SETTINGWHENCAPACITORSCOUPLING......................................................................................................................15
9.BLOCKCHART.......................................................................................................................................................... 17
10.SPECIFICATIONS................................................................................................................................................... 18
11.WARRANTY............................................................................................................................................................. 19
12.MAINTENANCE....................................................................................................................................................... 21
13.NOISELABORATORYSUPPORTNETWORK................................................................................................ 22

6
5. PREFACE
5-1. Preface
We thank you very much for your purchase of our injection unit for the Lightening Surge
Simulator for LSS-6330 / LSS-F03 SERIES (LSS-INJ6401SIG).
The unit is the injection unit for tests for interconnection lines described in IEC61000-4-5.
Please fully read and utilize this manual before using this unit.
5-2. Feature
1. The unit is used for the surge immunity tests (1.2/50μs combination waveform)for
interconnection lines described in IEC61000-4-5 Edition.3.
2.EUT’s power capacity is DC50V/1A. The unit can be used to inject high voltage surge
(maximum 6600V) to interconnection lines.
3.Bypassing 20mH inductor is available by connecting the connection plug in the accessory to
the inductor bypass terminal.
4.The protective arrestor in the accessory is attachable between each line and ground.
⚫The instruction manual was prepared so that any person who can
observe the prescribed instruction method and operating
precautions may safely handle and fully utilize the Unit.
⚫Keep the Manual by your side or other proper location so that it
may be readily available when using the Unit.
⚫This instruction manual doesn’t have details about Lightening
Surge Simulator (LSS-6330/LSS-F03 series).For details about
how to use Lightening Surge Simulator, refer to the
instruction manual of each model.
⚫

7
6. BASIC SAFETY PRECAUTIONS
1. Any person who has an implanted pacemaker in the body should not operate this unit.
Furthermore, such a person should not enter the test area while this unit is operating.
Otherwise, the electronic medical device can malfunction so that can result fatal.
2. Use of the unit and the lightning surge simulator (LSS-6330/LSS-F03 series) unit in an
explosive area such as "No fire" area etc. is prohibited. If used in such an area, it is liable
to cause combustion or ignition due to discharge.
3. Firstly, it must be confirmed that the ‘STOP’button of lightning surge simulator is pushed
and test is paused. Then turn off the ‘EUT LINE switch’of lightning surge simulator. Please
connect or set up lightning surge simulator (LSS-6330/LSS-F03 series), options, other
equipment and accessories after EUT power supply is cut off.
If this is not obeyed, it may result in fatal wound by electric shock from high voltage. Also,
it may result in burning damage on this unit by EUT power supply.
4. As the unit outputs high voltage (6600V maximum) and great current (157A maximum) by
surge from LSS-6330 series, carefully handle it. Mishandling or careless operation may
result in a fatal wound. Do not touch the surge in connector, DC line out connector, DC line
in terminal in conducting tests because high voltage is generated.
5. Mishandling or careless operation may result in fatal wounds.
6. Test equipment to be used with this unit should be capable of insulating against at least a
voltage of 6600V. The equipment under test (called EUT hereafter) using this unit should
be performed in a protective enclosure or cover against scattering broken pieces, fire
electric shock, etc.
7. Do not touch surge in cables, line out cables etc. Otherwise, a fatal accident may happen.
8. In case of touching the surge in connector, the socket of DC line out (the hole for inserting
the connection plug), and DC line in terminals just after conducting tests, as it takes about
3 seconds for automatic elimination of electricity of the lightening surge simulator, do not
touch them for this interval.
9. Never fail to connect PE terminal to the FG terminal of the lightening surge simulator (LSS-
6330/LSS-F03 series)by PE cable (accessory).
10. EUT can be electrified by high voltage after injecting surge. Eliminate electricity of EUT
before touching it.
11. In case of touching DC line out connector, never fail to make high voltage circuit “OFF”
status by pressing STOP button or EMERGENCY button.
12. Connect the connection plug, the capacitor / arrestor unit, the arrestor unit, and the surge
protection arrestor correctly and firmly depending on the test application. Misconnection
may cause very dangerous situation because of residual high voltage.
13. Never fail to watch the unit, the lightening surge simulator, and EUT in operation.

8
14. Do not open the unit because high voltage is generated internally.
15. In case of bypassing the inductor of each line’s decoupling network, as the surge can be
back to the input side with some EUT impedance condition, connecting protection
instruments is recommendable. The surge protection arrestor in the accessory can connect
between each line and ground. The discharge starting voltage of the arrestor is
approximately 350V under 1kV/μs condition. Even if the arrestor is connected, the voltage
less than the discharge starting voltage may return to line input side and damage the power
supply. In such a case, connect a protective instrument between the power supply and line
in.
16. Avoid using the unit in an environment exposed to water drop like rain. Also, avoid storing
the unit in high or low temperature environment.(Operating temperature range:15~35℃)
17. Avoid using the unit in an extremely humid or dusty place.(Operating humidity range:25~
75%RH)
18. If dewing or condensation occurs, thoroughly dry it before operating the unit.
19. Our company and sales agents shall have no responsibility for any accident resulting in
injury or death, any breakage or resultant damages due to irresponsible handling.
20. Only a qualified engineer can perform repair, maintenance and internal adjustment of the
unit.
21. Use the accessories and optional equipment supplied by our company for safe handling.
22. Avoid giving a strong shock to the unit
23. Do not wipe the unit with thinner, alcohol or similar solvent. When the body is dirty, soak a
cloth in detergent, squeeze the cloth and wipe the body with it.

9
7. APPEARANCE OF THE UNIT
FRONT PANEL
1 2 3 4 GND
INTERCONECTION LINE UNIT LSS-INJ6400SIG
COM HOT
6.6kV MAX
C
L L L L
1 2 3 4 GND
INDUCTORS BY-PASS
SURGE PROTECTION
R
COM
COM
COM
COM
COM
OUT1
OUT2
OUT3
OUT4
GND
HOT
HOT
HOT
HOT
SURGE RETURN SURGE INJ
LINE SELECT
ARRESTORS
COUPLING
CAPACITORS
COUPLING
LINE SELECTLINE IN
LINE OUT1 LINE O UT2 LINE O UT3 LINE OUT4 LINE O UT GND SURGE IN CO M
LINE OUT
SURGE IN
SURGE
IN HOT
①⑦
③
②
④
⑤
⑥
①SURGEINCONNECTOR[SURGEIN:HOT,COM]:Usedforinputtingsurgefromthe
lightningsurgesimulator.
②ARRESTORS COUPLING PART [ARRESTORS COUPLING]: Connects the
capacitor/arrestorunitorthearrestorunit(accessory) incaseofarrestorcouplingtest.
③CAPACITORS COUPLING PART [CAPACITORS COUPLING]: Connects the
connectionplug(accessory)incaseofcapacitorcouplingtest.
④SURGE INJECTION SETTING PART [SURGE INJ LINE SELECT] [SURGE
RETURNLINESELECT]:Selectssurgeinjectionlineandreturnline.
⑤SURGE PROTECTION INSTRUMENT CONNECTOR [SURGE PROTECTION]:
Used for connecting surge protection instruments to avoid surge return. A gas arrestor
(90V) is connected when shipped. Remove the gas arrestor or connect other surge
protectioninstrumentsdependingoncases.
⑥INDUCTOR BY-PASS CONNECTOR [INDUCTORS BY-PASS]: Connects the
connection plug in case of bypassing inductors in the decoupling network. In case of
bypassinginductors,neverfailtobypassalllines.
⑦DCLINEOUTCONNECTOR[LINEOUT:1,2,3,4,GND]:SuppliespowertoEUTandinjects
surgetolinessetbysurgeinjectionsetpart.

10
REARPANEL
1 2 3 4
PE
INTERLOCK
1 2
5
LINE IN
⑧
⑩ ⑨
⑧DCLINEINTERMINAL [LINEIN]:InputspowerforEUT.
⑨INTERLOCKCONNECTOR[INTERLOCK:1,2]:Connect[1]tothelighteningsurgesimulator’s
interlockconnectorbytheinterlockconnectioncable(accessory).Connectinterlockconnector
(ofthesurgesimulator’s accessory)to[2].
⑩PE TERMINAL [PE]: Protective Earth terminal of the unit. Never fail to connect [PE] to FG
terminalofthe lighteningsurgesimulator(LSS-6000series)byPEcable(accessory).
1 2 3 4 G

11
8. HOW TO CONNECT EQUIPMENT
8-1 Connection of Rear Panel
Connect PE terminal to FG terminal of LSS series by PE cable (accessory) as indicating in the figure
below. Connect the interlock connector [1]to the LSS seriesinterlock connector by the interlock cable
(accessory).Asnotattached,preparecabletoLineInterminal(material:AGW10-24).
1 2 3 4
PE
INTER LOCK
1 2
5
LINE IN
EUT供給電源へ
(添付品)
インターロックケーブル
LSS-6000シリーズ
インターロックコネクタへ
FG端子へ
PEケーブル
(添付品)
LSS-6000シリーズ
8-2 Connection of Front Panel (in case of LSS-6330)
Connect the surge in connector [SURGE IN: HOT, COM] to the surge out terminal
[HOT,COM] by the surge in cable (accessory). Connect line out connecter [LINE OUT:
1,2,3,4,GND] to EUT by DC line cable.
TotheFGofLSSseries
PEcable(Accessory)
TothepowersupplyforEUT
Interlockcable(Accessory)
To the Interlock connector of
LSSseries
LSS-6330
ToEUT
Lineoutcable
(Accessory)
Surgeincable
(Accessory)
1 2 3 4 G
1 2 3 4 GND
INTERCONECTION LINE UNIT LSS-INJ6400SIG
COM HOT
6kV MAX
C
L L L L
1 2 3 4 GND
INDUCTORS BY-PASS
SURGE PROTECTION
R
COM
COM
COM
COM
COM
OUT1
OUT2
OUT3
OUT4
GND
HOT
HOT
HOT
HOT
SURGE RETURN SURGE INJ
LINE SELECT
ARRESTORS
COUPLING
CAPACITORS
COUPLING
LINE SELECTLI N E I N
LI N E O U T1 LI N E O U T2 LI N E O U T3 LI N E O U T4
LI N E O U T G N DSU RG E I N C O M
LI N E O U T
SU RG E I N
SURGE
I N H O T

12
8-3. Connection of the front panel (in case of LSS-F03 series)
Connection of the LSS-INJ6401SIG to the LSS-F03 is done by using the
Surge In Cable(05-T1578) sold separately. Connect the safety socket side of the cables to the
SURGE IN terminals (HOT,COM) on the LSS-INJ6401SIG and High Voltage terminal side to the
SURGE OUTPUT (HOT,COM) terminals on the LSS-F03.
Next, connect the safety socket side of the supplied Line Out Cables to the
LINE OUT terminals (LINE OUT1, 2, 3, 4, GND). Connect the EUT to the other
end of the cables (clip side)
EUT(供試体)へ
1 2 3 4 GND
INTERCONECTION LINE UNIT LSS-INJ6400SIG
COM HOT
6kV MAX
C
L L L L
1 2 3 4 GND
INDUCTORS BY-PASS
SURGE PROTECTION
R
COM
COM
COM
COM
COM
OUT1
OUT2
OUT3
OUT4
GND
HOT
HOT
HOT
HOT
SURGE RETURN SURGE INJ
LINE SELECT
ARRESTORS
COUPLING
CAPACITORS
COUPLING
LINE SELECTLI N E I N
LI N E O U T1 LIN E O U T2 LI N E O U T3 LI N E O U T4 LIN E O U T G N DSU RG E I N C O M
LI N E O U T
SU RG E I N
SURGE
I N H O T
ライン出力ケーブル
(添付品)
For surge-in cable 05-T1578
(sold separately)
ToEUT
Lineoutcable
(Accessory)

13
8-4 Setting When Arrestors Coupling
(1)ToinjectsurgetoLine-Line
Example: In case of injection line: [OUT1] and return line: [OUT2]
1 2 3 4 GND
COM HOT
C
L L L L
1 2 3 4 GND
INDUCTORS BY-PASS
SURGE PROTECTION
R
COM
COM
COM
COM
COM
OUT1
OUT2
OUT3
OUT4
GND
HOT
HOT
HOT
HOT
SURGE RETURN SURGE INJ
LINE SELECT
ARRESTORS
COUPLING
CAPACITORS
COUPLING
LINE SELECTLINE IN
LINE OUT1 LINE OUT2 LINE OUT3 LINE OUT4 LINE OUT GND SURGE IN COM
LINE OUT
SURGE IN
SURGE
IN HOT
6.6kV MAX
①Connect the capacitor / arrestor unit (accessory) or the arrestor unit (accessory) to
[ARRESTORSCOUPLING].
*Usethecapacitor/arrestorunitiftransmissionsignalofEUTislessthan5kHz,andusethe
arrestorunitifmorethan5kHz.
②Connect the connection plug to [SURGE INJ LINE SELECT] [SURGE RETURN LINE
SELECT].Theabovefigureshowstheexampleofsetting[OUT1]asthesurgeinjectionline
and[OUT2]asthesurgereturnline.
③Thegasarrestor(90V)isconnectedto[SURGEPROTECTION].Removethegasarrestor
orconnectinganothersurgeprotectioninstrumentdependingonnecessity.
*Thegasarrestorisconnectedwhenshipped.Thegasarrestorshouldbechangedifitdoes
notemitlightwhensurgeinjected.Incaseofusingsurgeabsorbersetc.otherthanthegas
arrestor,connectasurgeprotectioninstrumentdependingontestconditions.
④Incaseofbypassingtheinductor(20mH)inthedecouplingnetwork,connecttheconnection
plug(accessory)to[INDUTORSBY-PASS].Neverfailtobypassalllines.
*In case of bypassing the inductor (20mH), the surge may return to the line input side and
damageEUT.Besides,thesurgemaynot injecttoEUT.

14
(2) To inject surge to Line-Ground
Example: In case of injection line: [OUT1] and return line: [GND].
1 2 3 4 GND
COM HOT
C
L L L L
1 2 3 4 GND
INDUCTORS BY-PASS
SURGE PROTECTION
R
COM
COM
COM
COM
COM
OUT1
OUT2
OUT3
OUT4
GND
HOT
HOT
HOT
HOT
SURGE RETURN SURGE INJ
LINE SELECT
ARRESTORS
COUPLING
CAPACITORS
COUPLING
LINE SELECTLINE IN
LINE OUT1 LINE OUT2 LINE OUT3 LINE OUT4 LINE OUT GND SURGE IN COM
LINE OUT
SURGE IN
SURGE
IN HOT
6.6kV MAX
①Connect the capacitor / arrestor unit (accessory) or the arrestor unit (accessory) to
[ARRESTORSCOUPLING].
*Usethecapacitor/arrestorunitiftransmissionsignalofEUTislessthan5kHz,andusethe
arrestorunitifmorethan5kHz.
②Connect the connection plug to [SURGE INJ LINE SELECT] [SURGE RETURN LINE
SELECT].Theabovefigureshowstheexampleofsetting[OUT1]asthesurgeinjectionline
and[GND]asthesurgereturnline.
③Thegasarrestor(90V)isconnectedto[SURGEPROTECTION].Removethegasarrestor
orconnectinganothersurgeprotectioninstrumentdependingonnecessity.
*Thegasarrestorisconnectedwhenshipped.Thegasarrestorshouldbechangedifitdoes
notemitlightwhensurgeinjected.Incaseofusingsurgeabsorbersetc.otherthanthegas
arrestor,connectasurgeprotectioninstrumentdependingontestconditions.
④Incaseofbypassingtheinductor(20mH)inthedecouplingnetwork,connecttheconnection
plug(accessory)to[INDUTORSBY-PASS].Neverfailtobypassalllines.
*Incase of bypassing theinductor (20mH), the surge mayreturn tothe line input side and
damageEUT.Besides,thesurgemaynotinjecttoEUT.

15
8-5 Setting When Capacitors Coupling
(1)ToinjectsurgetoLine-Line
Example: In case of injection line: [OUT1] and return line: [OUT2]
1 2 3 4 GND
COM HOT
6.6kV MAX
C
L L L L
1 2 3 4 GND
INDUCTORS BY-PASS
SURGE PROTECTION
R
COM
COM
COM
COM
COM
OUT1
OUT2
OUT3
OUT4
GND
HOT
HOT
HOT
HOT
SURGE RETURN SURGE INJ
LINE SELECT
ARRESTORS
COUPLING
CAPACITORS
COUPLING
LINE SELECTLINE IN
LINE OUT1 LINE OUT2 LINE OUT3 LINE OUT4 LINE OUT GND SURGE IN COM
LINE OUT
SURGE IN
SURGE
IN HOT
①Connect the connection plug (accessory) to [CAPACITORSCOUPLING].
②Connect the connection plug to [SURGE INJ LINE SELECT] [SURGE RETURN LINE
SELECT].Theabovefigureshowstheexampleofsetting[OUT1]asthesurgeinjectionline
and[OUT2]asthesurgereturnline.
③Thegasarrestor(90V)isconnectedto[SURGEPROTECTION].Removethegasarrestor
orconnectinganothersurgeprotectioninstrumentdependingonnecessity.
*Thegasarrestorisconnectedwhenshipped.Thegasarrestorshouldbechangedifitdoes
notemitlightwhensurgeinjected.Incaseofusingsurgeabsorbersetc.otherthanthegas
arrestor,connectasurgeprotectioninstrumentdependingontestconditions.
④Incaseofbypassingtheinductor(20mH)inthedecouplingnetwork,connecttheconnection
plug(accessory)to[INDUTORSBY-PASS].Neverfailtobypassalllines.
*In case of bypassing the inductor (20mH), the surge may return to the line input side and
damageEUT.Besides,thesurgemaynotinjecttoEUT.

16
(2) To inject surge to Line-Ground
Example: In case of injection line: [OUT1] and return line: [GND].
1 2 3 4 GND
COM HOT
C
L L L L
1 2 3 4 GND
INDUCTORS BY-PASS
SURGE PROTECTION
R
COM
COM
COM
COM
COM
OUT1
OUT2
OUT3
OUT4
GND
HOT
HOT
HOT
HOT
SURGE RETURN SURGE INJ
LINE SELECT
ARRESTORS
COUPLING
CAPACITORS
COUPLING
LINE SELECTLINE IN
LINE OUT1 LINE OUT2 LINE OUT3 LINE OUT4 LINE OUT GND SURGE IN COM
LINE OUT
SURGE IN
SURGE
IN HOT
6.6kV MAX
①Connect the connection plug (accessory) to [CAPACITORSCOUPLING].
②Connect the connection plug to [SURGE INJ LINE SELECT] [SURGE RETURN LINE
SELECT].Theabovefigureshowstheexampleofsetting[OUT1]asthesurgeinjectionline
and[GND]asthesurgereturnline.
③Thegasarrestor(90V)isconnectedto[SURGEPROTECTION].Removethegasarrestor
orconnectinganothersurgeprotectioninstrumentdependingonnecessity.
*Thegasarrestorisconnectedwhenshipped.Thegasarrestorshouldbechangedifitdoes
notemitlightwhensurgeinjected.Incaseofusingsurgeabsorbersetc.otherthanthegas
arrestor,connectasurgeprotectioninstrumentdependingontestconditions.
④In case of bypassing the inductor (20mH) in the decoupling network, connect the
connectionplug(accessory)to[INDUTORSBY-PASS].Neverfailtobypassalllines.
*In case of bypassing the inductor (20mH), the surge may return to the line input side and
damageEUT.Besides,thesurgemaynotinjecttoEUT.

17
9. BLOCK CHART
HOT
COM
INTER LOCK1
LINE IN (DC50V/1A)
1
2
3
4
G
LINE OUT
OUT1
OUT2
OUT3
OUT4
GND
1
2
3
4
5
1
2
3
4
5
アクリル扉部インターロックスイッチ
40Ω
0.5μF
C
SURGE INJ.
LINE SELECT
SURGE RETURN
LINE SELECT
HOT
PE
20mH
20mH
20mH
INTER LOCK2
BY-PASS
INDUCTORS
SURGE IN
PROTECTION
SURGE
PROTECTION
SURGE
20mH
HOT HOT HOT
OUT1OUT2OUT3OUT4GND
COM COM COM COM COM
R
ARRESTORS
COUPLINGCOUPLING
CAPACITORS
INTERLOCKofthedoor
This manual suits for next models
2
Table of contents
Other NoiseKen Laboratory Equipment manuals

NoiseKen
NoiseKen IJ-02 User manual

NoiseKen
NoiseKen LSS-6330-A20A User manual

NoiseKen
NoiseKen SWCS-931SD User manual

NoiseKen
NoiseKen ESS-S3011A User manual

NoiseKen
NoiseKen ESS-801GL User manual

NoiseKen
NoiseKen FNS-AX4-A20 User manual

NoiseKen
NoiseKen JSS-003 User manual

NoiseKen
NoiseKen ESS-2000AX User manual

NoiseKen
NoiseKen LSS-6330-A20 User manual

NoiseKen
NoiseKen LSS-6330-B63A User manual