6
Power supplies
The Kombi-LWUS indicates the powerup by blinking the led next to the micro USB socket.
-RHT-DP, -RHT, -RHT-IP42 variants
These variants are powered with 3 pcs LR6 batteries (AA 1.5 V alkaline). High quality batteries should be
used, for example Energizer EN91.
The device is supplied with batteries already installed, so it is ready to be used.
The Kombi-LWUS can alternatively be powered with an external supply. When an external power supply is
used, the batteries can be omitted or used as a backup supply. Use the micro USB socket to connect an
external supply into the device.
-Dust13, -Dust40, -RHT-CO2-TVOC-Dust40-DP
These variants with PM sensor must be powered with an external supply. Use the micro USB socket to
connect an external supply into the device.
Settings
The Kombi-LWUS works with its default settings and no parameter configuration is normally required.
However, if it is necessary to change the settings of the device, it can be done with Nokeval’s MekuWin
software (available for free at www.nokeval.com). Following procedure should be followed in order to get
access to the settings:
•Connect a micro-USB cable to the connector and the other end of the cable to a computer.
•If Windows requests for a driver, download it at www.nokeval.com > Support, unzip it to a
temporary folder, and show that directory as the location for the driver. If there are problems
during the installation, try pushing the button every 5-15 seconds or keeping it pressed so that
the Kombi-LWUS will not shut down its USB port.
•Launch the MekuWin program.
•In MekuWin, choose the right COM-port form the Ports-menu. If the port is not visible, try
pushing the button on the Kombi-LWUS to wake up the port.
•From the Port settings -menu, choose Protocol = ModbusRTU, Address = 1.
•Click Direct.
•A new window will open for the settings.
The configuration menu includes the following settings:
•Protection
oDescribes password protection status “None”, “Locked”, “Unlocked”.
•Password input/setup
oUnlock protection or change password. See “Protecting the settings” section below for
details.
•Period
oTime between measurements with options from ranging 5 minutes to 6 hours. Default is
15 min.
•LoRaWAN
oIf the Kombi-LWUS is used with the Nokeval cloud platform, the default settings are ok. If
you wish to manually set the LoRaWAN credentials and/or integrate it with your cloud
platform, see “LoRaWAN settings” and “Uplink payload structure” chapters.