manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Nonin
  6. •
  7. Medical Equipment
  8. •
  9. Nonin 2500 PalmSAT User manual

Nonin 2500 PalmSAT User manual

ENG – Translations of the IFU can be found using this QR code.
FRE – Les traductions de cette notice d’utilisation peuvent être retrouvées à l’aide de ce code QR.
GER – Übersetzungen dieses Handbuchs können über diesen QR-Code abgerufen werden.
ITL – Utilizzando questo codice QR, è possibile trovare le traduzioni delle Istruzioni per l’uso.
SPA – Las traducciones de este manual se pueden encontrar utilizando este código QR.
POR – Pode aceder às traduções das instruções de utilização através deste código QR.
DUT – Vertalingen van de handleiding zijn te vinden met behulp van deze QR-code.
GRK – ΜετηχρήσηαυτούτουκωδικούQRμπορείτεναβρείτεμεταφράσεις
τωνοδηγιώνχρήσης(IFU).
DAN – ScandenneQR-kodeforatndeoversættelserafdennebrugsvejledning.
SWE – Översättningar av den här guiden kan hittas med denna QR-kod.
FIN – Käyttöohjeen käännökset löytyvät tällä QR-koodilla.
POL – TłumaczeniategoprzewodnikamożnaznaleźćzapomocątegokoduQR.
NOR – OversettelseravdennebruksanvisningenkannnesvedåbrukedenneQR-koden.
Instructions for Use / Operator’s Manual
https://www.nonin.com/support/2500
P/N 114856-001-01 11/2022 13700 1st Avenue North, Plymouth, MN 55441-5443 USA nonin.com
Instructions for Use
Model 2500
©2022 Nonin Medical, Inc. All trademarks are the property of Nonin Medical, Inc. unless otherwise noted.
Cautions (continued)
• This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or systems. This
standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However, because of the
proliferation of radio-frequency transmitting equipment and other sources of electrical noise in healthcare and other environments, it is possible
that high levels of such interference due to close proximity or strength of a source might disrupt the performance of this device. Medical electri-
cal equipment needs special precautions regarding EMC, and all equipment must be installed and put into service according to the
EMCinformationspecied.
• IncompliancewiththeEuropeanDirectiveonWasteElectricalandElectronicEquipment(WEEE)2002/96/EC,donotdisposeofthisproductas
unsorted municipal waste. This device contains WEEE materials; please contact your distributor regarding take-back or recycling of the device.
If you are unsure how to reach your distributor, please call Nonin for your distributor’s contact information.
• This device’s display will go blank after 10 seconds of inadequate signals. The data update period is every 1.5 seconds.
• This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may degrade pulse
oximeter performance or affect the accuracy of the measurement include the following:
• A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor.
• AllpartsandaccessoriesconnectedtotheserialportofthisdevicemustbecertiedaccordingtoatleastIEC60950orUL1950for
data-processing equipment.
• This device is a precision electronic instrument and must be repaired by trained Nonin personnel only. Field repair of the device is not possible.
Do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the warranty.
• Anysignorevidenceofopeningthesystem,eldservicebynon-Noninpersonnel,tampering,oranykindofmisuseorabuseofthesystem,shall
void the warranty in its entirety.
• Replacebatterieswithin30secondstoavoidlosingsettings(date,time,andpatientdatastoredinmemory)orcorruptingdata.
• Radiosandcellphonesorsimilardevicescanaffecttheequipmentandmustbekeptatleast2meters(6.5feet)awayfromequipment.
• Failureofanetworkdatacoupling(serialcable/connectors/wirelessconnections)willresultinlossofdatatransfer.
Cautions – only for the Model 2500A PalmSAT Pulse Oximeter with Alarms
• Setting alarm limits to extremes can render the alarm system useless.
•Excessive ambient light •Moisture in the sensor •Cardiogreen and other intravascular dyes
•Excessive motion •Improperly applied sensor •Carboxyhemoglobin
•Electrosurgical interference •Incorrect sensor type •Methemoglobin
• Bloodowrestrictors(arterialcatheters,
blood pressure - carboxyhemoglobin cuffs,
infusionlines,etc.)
•Inadequate signal •Dysfunctional hemoglobin
•Venous pulsations •Articialnailsorngernailpolish
•Anemia or low hemoglobin concentrations •A sensor not at heart level
Nonin Model 2500C Charger Stand
Indications for Use/Intended Use/Intended Purpose
The Nonin Model 2500C Charger Stand is intended for use with the PalmSAT Models 2500 and 2500A Pulse Oximeters and the Model 2500B
RechargeableNiMH(NickelMetalHydride)BatteryPack.
Warnings
• Do not use this product in an MR environment.
• Do not use this product in an explosive atmosphere.
• ThisproductisnotdebrillationproofperIEC60601-1.
• As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement, strangulation, or injury to
the patient.
• This product should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the product should be
observed carefully to verify normal operation.
• The use of accessories, sensors, cables, and power supplies other than those listed in the Parts and Accessories List may result in increased
electromagnetic emission and/or decreased immunity of this product.
• Topreventimproperperformanceand/orpatientinjury,verifycompatibilityofthemonitor,sensor(s),andaccessoriesbeforeuse.
• Nomodicationstothisdeviceareallowedasitmayaffectdeviceperformance.
• PortableRFcommunicationsequipmentsuchascellphonesorradios(includingperipheralssuchasantennacablesandexternalantennas)
shouldbeusednocloserthan30cm(12inches)toanypartoftheMEsystem,includingcablesspeciedbythemanufacturer.Otherwise,
degradation of the performance of this equipment could result.
Cautions
• This equipment complies with International Standard 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or
systems. This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However,
because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in healthcare and other
environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the performance
of this product. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be installed and put into
serviceaccordingtotheEMCinformationspeciedinthismanual.
• Do not connect this product to an AC outlet controlled by a wall switch. If the switch is accidentally turned off before the battery pack is
recharged, the pulse oximeter may not function.
• Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked “Hospital Only” or “Hospital Grade”.
• This product contains sensitive electronic components and must be repaired by trained Nonin personnel only.
• Do not immerse this product in liquid.
• Do not place liquids on top of this product.
• Do not use caustic or abrasive cleaning agents on this product.
• Do not remove any covers from the product. There are no user-serviceable parts inside the unit.
• Do not attempt to charge disposable batteries. Disposable batteries may leak or explode if used improperly.
• Follow local, state and national governing ordinances and recycling instructions regarding disposal or recycling of the product and product
components, including batteries. Use only Nonin-approved battery packs.
• IncompliancewiththeEuropeanDirectiveonWasteElectricalandElectronicEquipment(WEEE)2002/96/EC,donotdisposeofthisproduct
as unsorted municipal waste. This product contains WEEE materials; please contact your distributor regarding take-back or recycling of the
product. If you are unsure how to reach your distributor, please call Nonin for your distributor’s contact information.
Adverse Event Statement
Users and/or patients should report adverse events involve their Nonin device to Nonin Medical, Inc. and the competent authority of the EU Member
State in which the user and/or patient is established, if applicable.
Nonin Model 2500 PalmSAT®Pulse Oximeter and Model 2500A PalmSAT®Pulse Oximeter with Alarms
Indications for Use/Intended Use/Intended Purpose
The Nonin®Model 2500 PalmSAT®Pulse Oximeter and Model 2500A PalmSAT®Pulse Oximeter with Alarms are indicated for use in measuring
anddisplayingfunctionaloxygensaturationofarterialhemoglobin(SpO2)andpulserateforadult,pediatric,andneonatalpatients.Thesedevices
are intended for continuous monitoring and/or spot- checking of patients during both motion and no-motion conditions, and for patients who are
well or poorly perfused.
Warnings
• Do not use this device in an MR environment.
• Explosion Hazard:Donotuseinanexplosiveatmosphereorinthepresenceofammableanestheticsorgasses.
• ThisdeviceisnotdebrillationproofperIEC60601-1.
• This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of assessing clinical signs
and symptoms.
• Oximeterreadingsofthisdevicemaybeaffectedbytheuseofanelectrosurgicalunit(ESU).
• Inspect the sensor application site at least every 6 to 8 hours to ensure correct sensor alignment and skin integrity. Patient sensitivity to sensors
and/or double-backed adhesive strips may vary due to medical status or skin condition.
• To avoid patient injury, use only with Nonin-branded PureLight®pulse oximeter sensors. These sensors are manufactured to meet the accuracy
specicationsforNoninPulseOximeters.Usingothermanufacturers’sensorscanresultinimproperpulseoximeterperformance.
• Topreventimproperperformanceand/orpatientinjury,verifycompatibilityofthemonitor,sensor(s),andaccessoriesbeforeuse.
• Nomodicationstothisdeviceareallowedasitmayaffectdeviceperformance.
• Do not use a damaged sensor. If the sensor is damaged in any way, discontinue use immediately and replace the sensor.
• As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement, strangulation, or injury
to the patient.
• This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device should be
observed carefully to verify normal operation.
• Theuseofaccessories,sensors,cables,andpowersuppliesotherthanthosespeciedinthePartsandAccessoriesListmayresultinincreased
electromagnetic emission and/or decreased immunity of this device.
• This device must be able to measure the pulse properly to obtain an accurate SpO2measurement. Verify that nothing is hindering the pulse
measurement before relying on the SpO2measurement.
• Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
• Discontinue use of adhesive tape strips if the patient exhibits an allergic reaction to the adhesive material.
• Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
• When a system fault occurs, the patient will no longer be monitored.
• To comply with relevant product safety standards, ensure that all alarm volumes are set appropriately and are audible in all situations.
Do not cover or otherwise hinder any speaker openings.
• The device turns off after approximately 10 minutes when at low battery capacity.
• Before changing the batteries, make sure the device is off and the sensor is not attached to a digit.
• PortableRFcommunicationsequipmentsuchascellphonesorradios(includingperipheralssuchasantennacablesandexternalantennas)
shouldbeusednocloserthan30cm(12inches)toanypartoftheMEsystem,includingcablesspeciedbythemanufacturer.Otherwise,
degradation of the performance of this equipment could result.
Warnings – only for the Model 2500A PalmSAT Pulse Oximeter with Alarms
• Verify all alarm settings and limits during system startup to ensure that they are set as intended.
• A hazard can exist if different presets are used on multiple 2500A monitors in one care area.
• Because operating environments vary, use caution to ensure that all audible alarms and indicators can be heard. Users must determine the
acceptable audible distance of all alarms.
• Donotplacethisdeviceinanenvironmentwhereitsspeakeropeningmaybecomeblocked;alarmsmaybecomemufedorinaudible.
• Turning off the alarm volume creates a situation that is not compliant with relevant safety standards. The alarm silence indicator is lit solid
when the alarm volume is turned off or set below 45 dBA.
Cautions
• Before use, carefully read the package insert provided with the sensors.
• This device is not an apnea monitor.
• Verifythatallvisibleindicatorsilluminateandthatanaudibleindicatorsoundsduringthestartup(initialization)sequence.Ifanyindicatoris
not lit or the audible indicator does not sound, do not use the device. Contact Nonin Technical Service for assistance.
• Review all limits to ensure they are appropriate for the patient.
• Thepresenceofadebrillatormayinterferewiththeperformanceofthisdevice.
• This device may not work on all patients. If you are unable to achieve stable readings, discontinue use.
• This device has motion tolerant software that minimizes the likelihood of motion artifact being misinterpreted as good pulse quality. In some
circumstances, however, the device may still interpret motion as good pulse quality. Minimize patient motion as much as possible.
• EarClipandReectancesensorsarenotrecommendedforpediatricorneonataluse.Theaccuracyofthesesensorshasnotbeenestablished
for pediatric or neonatal use.
• Do not autoclave or immerse the device or sensors in liquid. Do not expose the device or components to excessive moisture or liquids.
• Do not use caustic or abrasive cleaning agents on the device or the sensors.
• Theoximetersensormightnotworkoncoldextremitiesduetoreducedcirculation.Warmorrubthengertoincreasecirculation,orreposition
the sensor.
• Replace the batteries as soon as possible after a low-battery indication. Always replace the batteries with fully charged batteries.
• UseonlyNonin-speciedbatterytypeswiththisdevice.
• Do not use fully charged and partially charged batteries at the same time. This may cause the batteries to leak.
• Do not remove any covers other than the battery cover when replacing batteries. There are no user-serviceable parts inside other than the
replaceable batteries.
• Follow local, state and national governing ordinances and recycling instructions regarding disposal or recycling of the device and device
components, including batteries.
• Batteries may leak or explode if used or disposed of improperly.
• Remove the batteries if the device will be stored for more than 1 month.
REP
EC
Warranty
The device warranty
is 3 years.
nonin.com/warranty
Symbol
Glossary
nonin.com/symbols
Compliance
This product complies
withISO10993.
Not made from
natural rubber latex.
MPS, Medical Product Service GmbH
Borngasse 20
D-35619Braunfels,Germany
MedEnvoy Switzerland
Gotthardstrasse 28, 6302 Zug
Switzerland
REP
CH
For summary of safety and clinical data see above QR code.
To use the Nonin Model 2500 (Model 2500A with alarms) PalmSAT® pulse
oximeter, please follow the simple instructions below.
Install Batteries or Use Rechargeable Battery Pack
Instructions for Use
Model 2500
P/N 114856-001-01 11/2022 13700 1st Avenue North, Plymouth, MN 55441-5443 USA nonin.com
Ensure device is turned off and sensor is not applied to a digit.
1. Press the battery cover latch and remove the battery cover on the bottom of the unit
2. Insert four new AA-size alkaline batteries or a rechargeable NiMH battery pack*, as
indicatedbythepolaritymarkings(+and-)insidethebatterycompartment
3. Replace the battery cover and turn on the device
P/N 114856-001-01 11/2022 13700 1st Avenue North, Plymouth, MN 55441-5443 USA nonin.com
Instructions for Use
Model 2500
Taking Measurements
Verify the pulse oximeter sensor is properly positioned
on the patient. If the Pulse Quality display is blinking red
or amber or is blinking erratically, reposition, or replace
the sensor.
If the sensor is not properly positioned, or no sensor
isattachedtothepulseoximeterafterstartup(afew
secondsafterpoweringon),boththeSpO2and Pulse
Rate displays will display a single dash until a valid
pulse signal is detected.
Please refer to the Operator’s Manual on nonin.com/support/2500
for troubleshooting guidance and more detailed operating instructions.
* Model 2500B Rechargable NiMH Battery Pack included with Model 2500C Charger Stand.
2500C Charger Stand will only work with the NiMH Battery Pack not with alkaline batteries.
Care and Maintenance
Cleaning the pulse oximeter
• Clean the device separately from the sensors. For instructions on cleaning sensors, refer to the
respective sensor Instructions for Use.
• Clean the device with a soft cloth dampened with isopropyl alcohol. Do not pour or spray any liquids
onto the device, and do not allow any liquid to enter any openings in the device. Allow the device to
dry thoroughly before reusing.
Alarm Functions – Only Applicable to 2500A
High-priorityalarmsarepatient-specicandareindicatedbyaashingredalarmbarandahigh
priorityaudiblealarmsignal.Medium-priorityalarmsareequipment-specicandareindicatedbya
ashingamberalarmbarandamediumpriorityaudiblealarmsignal.Thesealarmcyclesrepeatuntil
silenced. Silence alarms by pressing the On/Off button twice.
Users may adjust alarm settings when the device is in Setup mode.
For additional information on adjusting alarm settings, please reference the Operator’s Manual.
Cleaning the charger stand – only applicable to 2500C (sold separately)
Unplug the power supply from the AC power outlet. Clean the product with a soft cloth dampened
with isopropyl alcohol. Do not pour or spray any liquids onto the product, and do not allow any
liquid to enter any openings in the product. Allow the product to dry thoroughly before reusing.
P/N 114856-001-01 11/2022 13700 1st Avenue North, Plymouth, MN 55441-5443 USA nonin.com
Instructions for Use
Model 2500
Thank you for trusting Nonin with your patients’ healthcare needs. We sincerely
appreciate your business. Please read your Operator’s Manual carefully and direct any
further questions to a Nonin Technical Service representative. If you would like to order
products or check the status of a current order, please contact Customer Service.
Contact Customer Service or Technical Service
Toll Free: 800.356.8874 (U.S.andCanadaonly)
Phone:+1763.553.9968
Email: [email protected]
Nonin Medical Inc.
13700 1st Avenue North
Plymouth, MN 55441-5443, USA
Nonin B.V.
Doctor Paul Janssenweg 150
5026 RH Tilburg, Netherlands
Have other questions or want to learn more?
Visit nonin.com to read more about our history, product offerings, and more.
Nonin is committed to sustainable practices. This paper is made with
FSC-certiedmaterials,printedwithsoyink,andis100%recyclable.
Battery Pack
Battery
Orientation
AA-size
Alkaline Batteries
→
→
Speaker
Advance Button
On/Off Button
Alarm Bar
(Bicolor)
Low Battery
Indicator
Alarm Silence
Indicator
Pulse Rate
Display
SpO2Display
Pulse Oximeter
Sensor
Connector
Pulse Quality
Display
(Tricolor)
Power On and Off
• Turn on the device by pressing and releasing the button on the front of the unit
• Turn off the device by pressing and holding the button for approximately two seconds
The device automatically powers off after 10 minutes of inactivity. Change battery when low
battery indicator light appears. AA-sized batteries should provide 60 hours of continuous use while the
rechargeable NiMH pack will provide 40 hours of continuous use.
Charging the Battery Pack Using the Charger Stand–Only Applicable to 2500C
(sold separately)
To use, place a PalmSAT Pulse Oximeter containing a rechargeable battery pack into the charger
stand. Next, connect the charger power supply to the charger, then plug the power supply into an
appropriateACpowersource(walloutlet).Thebatterypackwillbefullychargedin
approximately 180 minutes.
Connect the Sensor Connect the pulse oximeter sensor with the Nonin logo facing
up to the top of the device as shown. Ensure the sensor is
rmlypluggedin.
RefertotheOperator’sManualorthespecicsensorpackageinsert
for pulse oximeter sensor positioning information.
Memory Functions
EachtimetheModel2500isturnedon(exceptduringSetupmode),dataareautomaticallycollectedin
memory. The device can collect and store up to 72 hours of SpO2and pulse rate information. Refer to the
Operator’s Manual.
Clinical Benets
Nonin pulse oximeters allow for the management of patients’ medical conditions by providing fast,
accurate, real-time, noninvasive oxygen measurement in order to meet patients’ medical needs.
2500/2500A Specications
*AdditionalspecicationsareavailableintheOperator’sManualthatcanbelocatedusingtheQRCode.
• Oxygen saturation measured from 0%-100% SpO2
• Pulse rate measured from 18bpm-321bpm
• ThedevicemustmaintainaccuracyinaccordancewithISO80601-2-61andISO9919forpulserate
and SpO2atthelowestobtainablepulseamplitude(0.3%modulation).
Measurement Wavelengths and Output Power:
• Red: 660 nanometers @ 0.8 mW maximum average
• Infrared:910nanometers@1.2mWmaximumaverage

Other manuals for 2500 PalmSAT

2

Other Nonin Medical Equipment manuals

Nonin WristOx2 3150 User manual

Nonin

Nonin WristOx2 3150 User manual

Nonin Onyx Vantage 9590 User manual

Nonin

Nonin Onyx Vantage 9590 User manual

Nonin 9570 User manual

Nonin

Nonin 9570 User manual

Nonin Onyx 9500 User manual

Nonin

Nonin Onyx 9500 User manual

Nonin G02 ARCHIEVE User manual

Nonin

Nonin G02 ARCHIEVE User manual

Nonin 2120 User manual

Nonin

Nonin 2120 User manual

Nonin 7500 User manual

Nonin

Nonin 7500 User manual

Nonin Sensmart H500 User manual

Nonin

Nonin Sensmart H500 User manual

Nonin 2500 PalmSAT User manual

Nonin

Nonin 2500 PalmSAT User manual

Nonin 8500 User manual

Nonin

Nonin 8500 User manual

Nonin 2120 User manual

Nonin

Nonin 2120 User manual

Nonin 8600V User manual

Nonin

Nonin 8600V User manual

Nonin NoninConnect 3245 User manual

Nonin

Nonin NoninConnect 3245 User manual

Nonin 3150 User manual

Nonin

Nonin 3150 User manual

Nonin CO-Pilot H500 User manual

Nonin

Nonin CO-Pilot H500 User manual

Nonin 7500 Troubleshooting guide

Nonin

Nonin 7500 Troubleshooting guide

Nonin 9843 User manual

Nonin

Nonin 9843 User manual

Nonin RespSense User manual

Nonin

Nonin RespSense User manual

Nonin 3100 WristOx User manual

Nonin

Nonin 3100 WristOx User manual

Nonin Onyx Vantage 9590 User manual

Nonin

Nonin Onyx Vantage 9590 User manual

Nonin 2500 PalmSAT User manual

Nonin

Nonin 2500 PalmSAT User manual

Nonin 7500 User manual

Nonin

Nonin 7500 User manual

Nonin WristOx2 3150 User manual

Nonin

Nonin WristOx2 3150 User manual

Nonin GO2 Achieve User manual

Nonin

Nonin GO2 Achieve User manual

Popular Medical Equipment manuals by other brands

Delton Fingertip Oximeter user manual

Delton

Delton Fingertip Oximeter user manual

Stryker S3 MEDSURG BED quick start guide

Stryker

Stryker S3 MEDSURG BED quick start guide

Nidek Medical Mark 5 Nuvo OCSI Lite 3 user guide

Nidek Medical

Nidek Medical Mark 5 Nuvo OCSI Lite 3 user guide

MEDIPIA Endoclean Series user manual

MEDIPIA

MEDIPIA Endoclean Series user manual

DARAY MAG700 Range installation manual

DARAY

DARAY MAG700 Range installation manual

KLARITY Supine Breast System Instructions for use

KLARITY

KLARITY Supine Breast System Instructions for use

College Park Espire Classic Plus Technical manual

College Park

College Park Espire Classic Plus Technical manual

Moretti RespiAdvance ST587 instruction manual

Moretti

Moretti RespiAdvance ST587 instruction manual

Kyoto Kagaku PICC instruction manual

Kyoto Kagaku

Kyoto Kagaku PICC instruction manual

Beurer IL 11 Instructions for use

Beurer

Beurer IL 11 Instructions for use

Saebo MAS mini product manual

Saebo

Saebo MAS mini product manual

Gima SUPERVEGA EVO manual

Gima

Gima SUPERVEGA EVO manual

Bard Peripheral Vascular LifeStent 5F Instructions for use

Bard Peripheral Vascular

Bard Peripheral Vascular LifeStent 5F Instructions for use

Siemens Healthineers UniPOC RAPIDPoint 500 Configuration guide

Siemens

Siemens Healthineers UniPOC RAPIDPoint 500 Configuration guide

Ossur Form Fit Night Splint Instructions for use

Ossur

Ossur Form Fit Night Splint Instructions for use

Medela Invia Liberty Instructions for use

Medela

Medela Invia Liberty Instructions for use

Verathon GlideScope GVL user manual

Verathon

Verathon GlideScope GVL user manual

OSI Systems Spacelabs Ultraview 1700 Service manual

OSI Systems

OSI Systems Spacelabs Ultraview 1700 Service manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.