Nonin 3230 User manual

111604-001-05 02/2022
© 2022 Nonin Medical, Inc.
U.S. Patents 5,490,523; 5,792,052; other patents pending.
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, MN 55441, USA
www.nonin.com
E-mail: info@nonin.com
+1 (763) 553-9968
(800) 356-8874 (US and Canada)
Fax: +1 (763) 553-7807
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tiburg, Netherlands
Email: infointl@nonin.com (Europe)
+31 (0)13 - 45 87 130 (Europe) MPS, Medical Product Service GmbH
Borngasse 20
D-35619 Braunfels, Germany
EC
REP
Nonin, NoninConnect, and CorrectCheck are trademarks of Nonin Medical, Inc.
The Bluetooth® word mark and logo are registered trademarks owned by Bluetooth SIG, Inc.
Continua™, the Continua logo and Continua Certified™ are trademarks, service marks, or certification marks of the Continua Health Alliance.
Installing AAA Batteries
WARNING: Before changing
batteries, make sure the device is
off and is not applied to a digit.
1.
Hold the 3230 so you see the
back of the device and the
arrows on the battery door
point away from you.
.
2. Place your thumbs on the ovals.
3. Slide the battery door away
from you and off the 3230.
4. If applicable, remove the
old batteries from the 3230.
Properly dispose of the
batteries.
5. Insert two new 1.5 volt AAA-
size batteries. Carefully match
the polarity markings (+ and -).
The 3230 will not work if the
batteries are inserted the
wrong way.
6. Carefully slide the battery door
back onto the device.
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
Turning On the NoninConnect Model
3230
1. Insert a digit into the Model 3230
until it touches the built-in stop.
NOTE: Make sure the finger is
centered within the finger guide and
flat (not on its side). For best results,
keep the device at heart or chest
level.
2. If the CorrectCheck screen
(see Display Symbols table)
displays, slide finger further
into device. Correct positioning
of the finger is critical for
accurate measurements.
3. The 3230 begins sensing the
pulse and displaying readings.
4. View about 4 seconds of
readings before relying on the
displayed values. Continually
verify operation. It is common
for the displayed values to vary
slightly over a period of several
seconds. If the 3230 does not
turn on or if it shuts off
unexpectedly:
• Verify batteries are correctly
inserted.
• The batteries are depleted. Replace
batteries.
If the problem persists, remove the
batteries and contact Nonin Technical
Service.
NOTE: While on the finger, do
not press the device against any
surface and do not squeeze or hold
it together. The internal spring
provides the correct pressure;
additional pressure may cause
inaccurate readings.
Instructions for Use – English
NoninConnect™Model 3230
Bluetooth®Smart Pulse Oximeter
Indications for Use
The NoninConnect Model 3230 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in measuring and
displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate of patients who are well or poorly
perfused. It is intended for spot-checking of adult and pediatric patients with digits between 0.8 – 2.5 cm (0.3 – 1.0 inch) thick.
NOTE: Use Environment—Home healthcare environments under the supervision of qualified medical professionals. Users
include current/potential users of pulse oximetry in the home and caregivers/potential caregivers of such a user.
Clinical Benefits
Nonin pulse oximeters allow for the management of patients’ medical conditions by providing fast, accurate, real-time,
noninvasive oxygen measurement in order to meet patients’ medical needs.
Warnings
Do not use the device in an MR environment, in an explosive atmosphere, or on neonatal patients.
• This device is not defibrillation proof per IEC 60601-1.
• Use the Model 3230 within its designated range (approximately 10 m/32 ft, spherical radius, line of sight when connected to
a Bluetooth Smart Ready device). Moving outside this range may cause missing, lost, and/or inaccurate data.
• Inspect the sensor application site at least every 4 hours to ensure correct sensor alignment and skin integrity. Patient sensitivity
to sensor may vary due to medical status or skin condition.
• Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
• This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
• The device must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is
hindering the pulse measurement before relying on the SpO2 measurement.
• Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
• General operation of the device may be affected by the use of an electrosurgical unit (ESU).
• Keep the oximeter away from young children. Small items such as the battery door and battery are choking hazards.
• Before changing batteries, make sure the device is off and is not applied to a digit.
• Portable RF communications equipment such as cell phones or radios (including peripherals such as antenna cables and
external antennas) should be used no closer than 30cm (12 inches) to any part of the ME system, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Cautions
• This device has no audible alarms and is intended only for spot-checking.
•
T
his device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may
degrade pulse oximeter performance or affect the accuracy of the measurement include the followi
ng:
• The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
• The device is designed to be attached only to a digit.
• This device’s display will shut off after 30 seconds of no readings or poor readings.
• In some circumstances, the device will interpret motion as good pulse quality. Minimize patient motion as much as possible.
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
• This device is a precision electronic instrument and must be repaired by Nonin Technical Service. Field repair of the device is
not possible. Do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the
warranty.
• This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or systems.
This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However,
because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in health care and other
environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the
performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be
installed and put into service according to the EMC information specified in this manual.
• When device is connected via Bluetooth, other Wi-Fi devices within 6 meters (20 feet) could interrupt the Bluetooth
• connection.
• Portable and mobile RF communications equipment including CT, diathermy, RFID, and electronic article security systems
can affect medical electrical equipment.
• Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than 30 days. Do
not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time. These
actions may cause the batteries to leak.
• Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of the device
and device components, including batteries.
• In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not
dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor
regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your
distributor’s contact information.
Symbols
NOTE: Where applicable, an additional label bearing your country radio communications license information will appear
on the side of your device. This is not a serial number or device identifier.
Display Symbols
† These indicators only display when the associated feature has been activated by an integrator.
Using the NoninConnect Model 3230
Installing AAA Batteries
Use only alkaline batteries. When batteries are low, displays. Replace low batteries as soon as possible.
See the “Installing AAA Batteries” instructions and figures at left.
Turning On the NoninConnect Model 3230
See the “Turning on the NoninConnect Model 3230” instructions and figures at left.
Connection via Bluetooth Wireless Technology
When the Model 3230 is placed on the finger and turns on, it is ready for a Bluetooth wireless connection. The 3230 stays in this
mode until it is shut off or the Bluetooth radio turns off. The symbol is white when the Bluetooth radio is on, green when
the 3230 is connected, and flashes when there is a communication error.
The Bluetooth symbol is useful for the product installer.
Due to the wide variety of wireless environments, the Bluetooth connection between the 3230 and the host device must be tested
before using the 3230’s Bluetooth capabilities.
Turning Off the NoninConnect Model 3230
The Model 3230 will automatically turn off approximately 10 seconds after the digit is removed, or after a 2-minute period of poor
signals.
Cleaning the NoninConnect Model 3230
CAUTIONS:
!
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
1. To clean, wipe the device’s surfaces with a soft cloth dampened with one of the following:
• A 10% bleach solution (household bleach [5.25% sodium hypochlorite]).
• Warm, soapy water (hand dishwashing detergent – see note below), and then rinse the cleaned surfaces with a soft cloth
dampened with water (home use only).
2. Dry with a soft cloth, or allow to air dry. Ensure that all surfaces are completely dry.
NOTE: The hand dishwashing detergent that was tested includes these ingredients: Sodium Lauryl Sulfate, Sodium Laureth
Sulfate, Lauramine Oxide, Sodium Chloride, PPG-26, PEG-8 Propylheptyl Ether, and Phenoxyethanol.
Warranty
For warranty information please refer to: http://www.nonin.com/warranty.
All non-warranty work shall be done at Nonin's standard rates and charges in effect at the time
of delivery to Nonin. All repairs include a complete restest of the Nonin 3230 using factory test
fixtures.
Users and/or patients should report adverse events involving their Nonin device to Nonin Medical, Inc.
Specifications
Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 321 beats per minute (BPM)
Declared Accuracy*: The table below shows Arms values measured using the Model 3230 in a clinical study.
NOTE: If your national regulatory authority recognizes accuracy in motion, please contact
regulatory@nonin.com for accuracy data.
SpO2 Low Perfusion Accuracy (Arms)*: 70 to 100% ±2 digits
Pulse Rate Declared Accuracy Range (Arms)*: 20 to 250 BPM ±3 digits
Low Perfusion Pulse Rate Declared Accuracy Range (Arms)*: 40 to 240 BPM ±3 digits
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW max. average
Infrared: 910 nanometers @ 1.2 mW max. average
Temperature:
Operating: -5 °C to 40 °C / 23 °F to 104 °F
Storage/Transportation: -40 °C to 70 °C / -40 °F to 158 °F
Time (from storage) for device to be ready for its intended use:
3 minutes to warm from -40 °C to -5 °C
8 minutes to cool from 70 °C to 40 °C
Humidity:
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing
Altitude:
Operating: Up to 4,000 meters / 13,123 feet
Hyperbaric Pressure: Up to 4 atmospheres
Battery Life:
Operating: Approximately 2,200 spot checks (25 sec. per spot-check), within 10 meters/32 feet of
collector with streaming data
Storage: 1 month, with batteries installed. CAUTION: Remove batteries if the device will be
stored for more than 30 days.
* ±1 Arms represents approximately 68% of measurements at zero bias.
** This information is especially useful for clinicians performing photodynamic therapy.
Bluetooth Wireless Technology Information
Bluetooth Compliance: Version 4.0 single mode low energy
Operating Frequency: 2.4 to 2.4835 GHz
Output Power: TX: +3 dBM
Operating Range: 10 meter radius (line of sight)
Network Topology: Star - bus
Operation: Slave
Model 3230
Antenna Type: Integrated chip type antenna
Modulation Type: Frequency Hopping Spread Spectrum
Data Rate: 1 Mbit/second
Data Latency: 6 ms
Data Integrity: Adaptive Frequency Hopping
24-bit CRC (cyclic redundancy check)
32-bit message integrity check
Data Format: Nonin Proprietary: Sends data packets once per second. Includes a second counter
that allows the host to detect if packets are missing and the device to retransmit.
Bluetooth SIG Standard: Compliant with Bluetooth SIG Pulse Oximeter Profile
specifications adopted by Continua.
Quality of Service: This device uses Bluetooth Smart technology for wireless communications, which
allows for reliable communications in electrically noisy environments, and transmits
physiological data. If the connection is lost, the device will become available for a
connection in a few seconds.
Bluetooth Profiles Supported: GATT-based Nonin Proprietary Oximeter Profile; GATT-based Bluetooth SIG Pulse
Oximeter Profile
Authentication and Encryption: Supported
Encryption Key Size: 128 bits AES (advanced encryption standard)
Bluetooth Security
The Bluetooth radio contained in the 3230 is a Bluetooth Smart single-mode, low-energy radio. The 3230 supports an encryption
key size of 128 bits. While the 3230 is in a Bluetooth connection, it will be unavailable for other connections. Apart from the
standard Bluetooth security measures, Nonin has two non-standard security measures that are available.
For additional technical information, please see the insert, “NoninConnect Model 3230 Technical Description.”
WARNING: Portable RF communications equipment such as cell phones or radios
(including peripherals such as antenna cables and external antennas) should be used no
closer than 30cm (12 inches) to any part of the ME system, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Manufacturer’s DeclarationRefer to the following tables for specific information regarding this device’s compliance to 60601-1-2. Component compliance
level is determined by system compliance level.
Essential Performance
Essential performance of the 3230 is defined as SpO2 accuracy and pulse rate accuracy or an indication of abnormal
operation. Accuracies may be affected as a result of exposure to electromagnetic disturbances that are outside of the
environments listed in the Indications For Use. If issues are experienced, move the Nonin system away from the source of
electromagnetic disturbances.
Symbol Definition Symbol Definition
!
Caution! Not for continuous monitoring (no alarm for
SpO2)
Follow Instructions for Use. Battery orientation
Medical Device
Importer
MR
MR
MR unsafe
IP32
Protected against vertically falling water drops
when enclosure is tilted up to 15 degrees and
ingress of solid foreign objects greater than or
equal to 2.5 mm (0.1 in.) in diameter per
IEC 60529.
Type BF Applied Part (patient isolation
from electrical shock)
UL Mark for Canada and the United
States with respect to electric shock, fire,
and mechanical hazards only in
accordance with IEC 60601-1,
UL 60601-1 and CAN/CSA-C22.2
No. 601.1.
SN Serial Number
BDA Bluetooth Device Address
CE Marking indicating conformance to
EC Directive No. 93/42/EEC
concerning medical devices. Storage/shipping temperature range
Non-ionizing electromagnetic radiation.
Equipment includes RF transmitters.
Interference may occur in the vicinity of
equipment marked with this symbol.
Handle with care
Keep dry
UDIUDI
Unique Identifier
Date of Manufacture Federal law (USA) restricts this device to sale
by or on the order of a licensed practitioner.
Continua Certified™ signifies this device
meets Continua certification testing
requirements, which support
interoperability in personal health
devices (continuaalliance.org).
US
US
Country of Manufacture
Manufacturer
Indicates separate collection for electrical
and electronic equipment (WEEE)
EC
REP
Authorized Representative in the European
Community
REFREFREFREF
Catalogue number
QTYQTY
Quantity
Symbol Description
Nonin’s CorrectCheck™ senses that the finger has not been correctly inserted. If you see this symbol,
slide finger further into device.
The number next to this symbol is the amount of oxygen in your blood (functional oxygen saturation
of arterial hemoglobin).
MR
MR
• applying the pulse oximeter on the
same arm as a blood pressure cuff,
arterial catheter or infusion line(s)
(IVs)
• excessive light, such as sunlight or
direct home lighting
• excessive motion
• moisture in the device
• improperly applied device
• finger is outside recommended
size range
• poor pulse quality
• venous pulsations
• cardiogreen and other
intravascular dyes
• anemia or low hemoglobin
concentrations
• carboxyhemoglobin
• methemoglobin
• dysfunctional hemoglobin
• artificial nails or fingernail polish
• residue (e.g., dried blood, dirt,
grease, oil) in the light path
The number next to this animated symbol is your pulse rate. Pulse rate is the number of times your
heart beats per minute.
– – – Dashes replace the readings when the 3230 is unable to detect a usable signal.
White symbol – Radio is on.
Green symbol – 3230 is connected.
Flashing symbol – Connection error. The radio will reset.
No symbol – Radio is off.
Poor signal. Steady your hand, reposition finger, warm finger by rubbing, or select a different finger.
Low battery. Replace batteries.
Critical battery. Flashing indicator on full screen. The device will not work until the batteries are
replaced.
† Spot-check complete. While Spot-check is in progress, a clockwise spinning circular icon displays.
† Measurement complete (full screen).
Symbol Description
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, Minnesota 55441, USA
(800) 356-8874 (USA/Canada)
+1 (763) 553-9968 (outside USA and Canada)
E-mail: technicalservice@nonin.com
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tilburg, Netherlands
+31 (0)13 - 45 87 130 (Europe)
E-mail: technicalserviceintl@nonin.com
nonin.com
70 75 80 85 90 95 100
−6−4−20246
SaO2(%)
SpO2−SaO2(%)
Bias=0
Limits of Agreement
Linear Regression
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
22
2
2
22 2
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
44
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
6
6
6
6
66
6
6
6
66
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
7
7
7
7
7
7
7
7
7
7
7
7
77
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
9
9
9
9
9
9
9
9
9
9
99
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
10
10
10
1010
10
10
10
10
10
10
10
10
10
10
10
11
11
11
11
11
11
11
11
11
11
11
11
11 11
11
11
11
11
11
1111
11
11 11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11 12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
13
13
13
13
13
13
13
13 13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
14
14
1414
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14 14
14
14
SpO2−SaO2(%)
Finger
Thumb
This graph shows plots of the error
(SpO2 – SaO2) by SaO2 using the
3230 with a linear regression fit and
upper 95% and lower 95% limits of
agreement. Each sample data point is
identified by subject from a clinical
study in non-motion conditions.
Accuracy Summary – Finger and Thumb
Range Specified Oxygen
Saturation (Arms)Thumb Oxygen
Saturation (Arms)
70 – 100% ± 2 ± 1.56
70 – 80% ± 2 ± 1.91
80 – 90% ± 2 ± 1.21
90 – 100% ± 2 ± 1.49
Finger Oxygen
Saturation (Arms)
± 1.31
± 1.65
± 1.05
± 1.18
Table 1: Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified in the Indications for Use section. The user of this
device should ensure that it is used in such an environment.
Emissions Test Compliance
RF Emissions
CISPR 11 Group 2, Class B
Table 2: Electromagnetic Immunity
Immunity Test Compliance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Power Frequency (50/60 Hz) Magnetic Field
IEC 61000-4-8 30 A/m
Radiated RF
IEC 61000-4-3
80 MHz – 2.7 GHz 10 V/m
380 – 390 MHz 27 V/m
430 – 470 MHz 28 V/m
704 – 787 MHz 9 V/m
800 – 960 MHz 28 V/m
1.7 – 1.99 GHz 28 V/m
2.4 – 2.57 GHz 28 V/m
5.1 – 5.8 GHz 9 V/m
Table 3: Not Applicable
Harmonic Emissions (IEC 61000-3-2), Voltage Flicker Emissions (IEC 61000-3-3), Electrical Fast Transients (IEC 61000-4-4),
Surge (IEC 61000-4-5), Voltage dips (IEC 61000-4-11), Conducted Immunity (IEC 61000-4-6)
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
111604-001-05 02/2022
© 2022 Nonin Medical, Inc.
U.S. Patents 5,490,523; 5,792,052; other patents pending.
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, MN 55441, USA
www.nonin.com
E-mail: info@nonin.com
+1 (763) 553-9968
(800) 356-8874 (US and Canada)
Fax: +1 (763) 553-7807
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tiburg, Netherlands
Email: infointl@nonin.com (Europe)
+31 (0)13 - 45 87 130 (Europe) MPS, Medical Product Service GmbH
Borngasse 20
D-35619 Braunfels, Germany
EC
REP
Nonin, NoninConnect, and CorrectCheck are trademarks of Nonin Medical, Inc.
The Bluetooth® word mark and logo are registered trademarks owned by Bluetooth SIG, Inc.
Continua™, the Continua logo and Continua Certified™ are trademarks, service marks, or certification marks of the Continua Health Alliance.
Installing AAA Batteries
WARNING: Before changing
batteries, make sure the device is
off and is not applied to a digit.
1.
Hold the 3230 so you see the
back of the device and the
arrows on the battery door
point away from you.
.
2. Place your thumbs on the ovals.
3. Slide the battery door away
from you and off the 3230.
4. If applicable, remove the
old batteries from the 3230.
Properly dispose of the
batteries.
5. Insert two new 1.5 volt AAA-
size batteries. Carefully match
the polarity markings (+ and -).
The 3230 will not work if the
batteries are inserted the
wrong way.
6. Carefully slide the battery door
back onto the device.
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
Turning On the NoninConnect Model
3230
1. Insert a digit into the Model 3230
until it touches the built-in stop.
NOTE: Make sure the finger is
centered within the finger guide and
flat (not on its side). For best results,
keep the device at heart or chest
level.
2. If the CorrectCheck screen
(see Display Symbols table)
displays, slide finger further
into device. Correct positioning
of the finger is critical for
accurate measurements.
3. The 3230 begins sensing the
pulse and displaying readings.
4. View about 4 seconds of
readings before relying on the
displayed values. Continually
verify operation. It is common
for the displayed values to vary
slightly over a period of several
seconds. If the 3230 does not
turn on or if it shuts off
unexpectedly:
• Verify batteries are correctly
inserted.
• The batteries are depleted. Replace
batteries.
If the problem persists, remove the
batteries and contact Nonin Technical
Service.
NOTE: While on the finger, do
not press the device against any
surface and do not squeeze or hold
it together. The internal spring
provides the correct pressure;
additional pressure may cause
inaccurate readings.
Instructions for Use – English
NoninConnect™Model 3230
Bluetooth®Smart Pulse Oximeter
Indications for Use
The NoninConnect Model 3230 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in measuring and
displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate of patients who are well or poorly
perfused. It is intended for spot-checking of adult and pediatric patients with digits between 0.8 – 2.5 cm (0.3 – 1.0 inch) thick.
NOTE: Use Environment—Home healthcare environments under the supervision of qualified medical professionals. Users
include current/potential users of pulse oximetry in the home and caregivers/potential caregivers of such a user.
Clinical Benefits
Nonin pulse oximeters allow for the management of patients’ medical conditions by providing fast, accurate, real-time,
noninvasive oxygen measurement in order to meet patients’ medical needs.
Warnings
Do not use the device in an MR environment, in an explosive atmosphere, or on neonatal patients.
• This device is not defibrillation proof per IEC 60601-1.
• Use the Model 3230 within its designated range (approximately 10 m/32 ft, spherical radius, line of sight when connected to
a Bluetooth Smart Ready device). Moving outside this range may cause missing, lost, and/or inaccurate data.
• Inspect the sensor application site at least every 4 hours to ensure correct sensor alignment and skin integrity. Patient sensitivity
to sensor may vary due to medical status or skin condition.
• Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
• This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
• The device must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is
hindering the pulse measurement before relying on the SpO2 measurement.
• Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
• General operation of the device may be affected by the use of an electrosurgical unit (ESU).
• Keep the oximeter away from young children. Small items such as the battery door and battery are choking hazards.
• Before changing batteries, make sure the device is off and is not applied to a digit.
• Portable RF communications equipment such as cell phones or radios (including peripherals such as antenna cables and
external antennas) should be used no closer than 30cm (12 inches) to any part of the ME system, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Cautions
• This device has no audible alarms and is intended only for spot-checking.
•
T
his device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may
degrade pulse oximeter performance or affect the accuracy of the measurement include the followi
ng:
• The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
• The device is designed to be attached only to a digit.
• This device’s display will shut off after 30 seconds of no readings or poor readings.
• In some circumstances, the device will interpret motion as good pulse quality. Minimize patient motion as much as possible.
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
• This device is a precision electronic instrument and must be repaired by Nonin Technical Service. Field repair of the device is
not possible. Do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the
warranty.
• This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or systems.
This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However,
because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in health care and other
environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the
performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be
installed and put into service according to the EMC information specified in this manual.
• When device is connected via Bluetooth, other Wi-Fi devices within 6 meters (20 feet) could interrupt the Bluetooth
• connection.
• Portable and mobile RF communications equipment including CT, diathermy, RFID, and electronic article security systems
can affect medical electrical equipment.
• Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than 30 days. Do
not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time. These
actions may cause the batteries to leak.
• Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of the device
and device components, including batteries.
• In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not
dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor
regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your
distributor’s contact information.
Symbols
NOTE: Where applicable, an additional label bearing your country radio communications license information will appear
on the side of your device. This is not a serial number or device identifier.
Display Symbols
† These indicators only display when the associated feature has been activated by an integrator.
Using the NoninConnect Model 3230
Installing AAA Batteries
Use only alkaline batteries. When batteries are low, displays. Replace low batteries as soon as possible.
See the “Installing AAA Batteries” instructions and figures at left.
Turning On the NoninConnect Model 3230
See the “Turning on the NoninConnect Model 3230” instructions and figures at left.
Connection via Bluetooth Wireless Technology
When the Model 3230 is placed on the finger and turns on, it is ready for a Bluetooth wireless connection. The 3230 stays in this
mode until it is shut off or the Bluetooth radio turns off. The symbol is white when the Bluetooth radio is on, green when
the 3230 is connected, and flashes when there is a communication error.
The Bluetooth symbol is useful for the product installer.
Due to the wide variety of wireless environments, the Bluetooth connection between the 3230 and the host device must be tested
before using the 3230’s Bluetooth capabilities.
Turning Off the NoninConnect Model 3230
The Model 3230 will automatically turn off approximately 10 seconds after the digit is removed, or after a 2-minute period of poor
signals.
Cleaning the NoninConnect Model 3230
CAUTIONS:
!
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
1. To clean, wipe the device’s surfaces with a soft cloth dampened with one of the following:
• A 10% bleach solution (household bleach [5.25% sodium hypochlorite]).
• Warm, soapy water (hand dishwashing detergent – see note below), and then rinse the cleaned surfaces with a soft cloth
dampened with water (home use only).
2. Dry with a soft cloth, or allow to air dry. Ensure that all surfaces are completely dry.
NOTE: The hand dishwashing detergent that was tested includes these ingredients: Sodium Lauryl Sulfate, Sodium Laureth
Sulfate, Lauramine Oxide, Sodium Chloride, PPG-26, PEG-8 Propylheptyl Ether, and Phenoxyethanol.
Warranty
For warranty information please refer to: http://www.nonin.com/warranty.
All non-warranty work shall be done at Nonin's standard rates and charges in effect at the time
of delivery to Nonin. All repairs include a complete restest of the Nonin 3230 using factory test
fixtures.
Users and/or patients should report adverse events involving their Nonin device to Nonin Medical, Inc.
Specifications
Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 321 beats per minute (BPM)
Declared Accuracy*: The table below shows Arms values measured using the Model 3230 in a clinical study.
NOTE: If your national regulatory authority recognizes accuracy in motion, please contact
regulatory@nonin.com for accuracy data.
SpO2 Low Perfusion Accuracy (Arms)*: 70 to 100% ±2 digits
Pulse Rate Declared Accuracy Range (Arms)*: 20 to 250 BPM ±3 digits
Low Perfusion Pulse Rate Declared Accuracy Range (Arms)*: 40 to 240 BPM ±3 digits
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW max. average
Infrared: 910 nanometers @ 1.2 mW max. average
Temperature:
Operating: -5 °C to 40 °C / 23 °F to 104 °F
Storage/Transportation: -40 °C to 70 °C / -40 °F to 158 °F
Time (from storage) for device to be ready for its intended use:
3 minutes to warm from -40 °C to -5 °C
8 minutes to cool from 70 °C to 40 °C
Humidity:
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing
Altitude:
Operating: Up to 4,000 meters / 13,123 feet
Hyperbaric Pressure: Up to 4 atmospheres
Battery Life:
Operating: Approximately 2,200 spot checks (25 sec. per spot-check), within 10 meters/32 feet of
collector with streaming data
Storage: 1 month, with batteries installed. CAUTION: Remove batteries if the device will be
stored for more than 30 days.
* ±1 Arms represents approximately 68% of measurements at zero bias.
** This information is especially useful for clinicians performing photodynamic therapy.
Bluetooth Wireless Technology Information
Bluetooth Compliance: Version 4.0 single mode low energy
Operating Frequency: 2.4 to 2.4835 GHz
Output Power: TX: +3 dBM
Operating Range: 10 meter radius (line of sight)
Network Topology: Star - bus
Operation: Slave
Model 3230
Antenna Type: Integrated chip type antenna
Modulation Type: Frequency Hopping Spread Spectrum
Data Rate: 1 Mbit/second
Data Latency: 6 ms
Data Integrity: Adaptive Frequency Hopping
24-bit CRC (cyclic redundancy check)
32-bit message integrity check
Data Format: Nonin Proprietary: Sends data packets once per second. Includes a second counter
that allows the host to detect if packets are missing and the device to retransmit.
Bluetooth SIG Standard: Compliant with Bluetooth SIG Pulse Oximeter Profile
specifications adopted by Continua.
Quality of Service: This device uses Bluetooth Smart technology for wireless communications, which
allows for reliable communications in electrically noisy environments, and transmits
physiological data. If the connection is lost, the device will become available for a
connection in a few seconds.
Bluetooth Profiles Supported: GATT-based Nonin Proprietary Oximeter Profile; GATT-based Bluetooth SIG Pulse
Oximeter Profile
Authentication and Encryption: Supported
Encryption Key Size: 128 bits AES (advanced encryption standard)
Bluetooth Security
The Bluetooth radio contained in the 3230 is a Bluetooth Smart single-mode, low-energy radio. The 3230 supports an encryption
key size of 128 bits. While the 3230 is in a Bluetooth connection, it will be unavailable for other connections. Apart from the
standard Bluetooth security measures, Nonin has two non-standard security measures that are available.
For additional technical information, please see the insert, “NoninConnect Model 3230 Technical Description.”
WARNING: Portable RF communications equipment such as cell phones or radios
(including peripherals such as antenna cables and external antennas) should be used no
closer than 30cm (12 inches) to any part of the ME system, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Manufacturer’s DeclarationRefer to the following tables for specific information regarding this device’s compliance to 60601-1-2. Component compliance
level is determined by system compliance level.
Essential Performance
Essential performance of the 3230 is defined as SpO2 accuracy and pulse rate accuracy or an indication of abnormal
operation. Accuracies may be affected as a result of exposure to electromagnetic disturbances that are outside of the
environments listed in the Indications For Use. If issues are experienced, move the Nonin system away from the source of
electromagnetic disturbances.
Symbol Definition Symbol Definition
!
Caution! Not for continuous monitoring (no alarm for
SpO2)
Follow Instructions for Use. Battery orientation
Medical Device
Importer
MR
MR
MR unsafe
IP32
Protected against vertically falling water drops
when enclosure is tilted up to 15 degrees and
ingress of solid foreign objects greater than or
equal to 2.5 mm (0.1 in.) in diameter per
IEC 60529.
Type BF Applied Part (patient isolation
from electrical shock)
UL Mark for Canada and the United
States with respect to electric shock, fire,
and mechanical hazards only in
accordance with IEC 60601-1,
UL 60601-1 and CAN/CSA-C22.2
No. 601.1.
SN Serial Number
BDA Bluetooth Device Address
CE Marking indicating conformance to
EC Directive No. 93/42/EEC
concerning medical devices. Storage/shipping temperature range
Non-ionizing electromagnetic radiation.
Equipment includes RF transmitters.
Interference may occur in the vicinity of
equipment marked with this symbol.
Handle with care
Keep dry
UDIUDI
Unique Identifier
Date of Manufacture Federal law (USA) restricts this device to sale
by or on the order of a licensed practitioner.
Continua Certified™ signifies this device
meets Continua certification testing
requirements, which support
interoperability in personal health
devices (continuaalliance.org).
US
US
Country of Manufacture
Manufacturer
Indicates separate collection for electrical
and electronic equipment (WEEE)
EC
REP
Authorized Representative in the European
Community
REFREFREFREF
Catalogue number
QTYQTY
Quantity
Symbol Description
Nonin’s CorrectCheck™ senses that the finger has not been correctly inserted. If you see this symbol,
slide finger further into device.
The number next to this symbol is the amount of oxygen in your blood (functional oxygen saturation
of arterial hemoglobin).
MR
MR
• applying the pulse oximeter on the
same arm as a blood pressure cuff,
arterial catheter or infusion line(s)
(IVs)
• excessive light, such as sunlight or
direct home lighting
• excessive motion
• moisture in the device
• improperly applied device
• finger is outside recommended
size range
• poor pulse quality
• venous pulsations
• cardiogreen and other
intravascular dyes
• anemia or low hemoglobin
concentrations
• carboxyhemoglobin
• methemoglobin
• dysfunctional hemoglobin
• artificial nails or fingernail polish
• residue (e.g., dried blood, dirt,
grease, oil) in the light path
The number next to this animated symbol is your pulse rate. Pulse rate is the number of times your
heart beats per minute.
– – – Dashes replace the readings when the 3230 is unable to detect a usable signal.
White symbol – Radio is on.
Green symbol – 3230 is connected.
Flashing symbol – Connection error. The radio will reset.
No symbol – Radio is off.
Poor signal. Steady your hand, reposition finger, warm finger by rubbing, or select a different finger.
Low battery. Replace batteries.
Critical battery. Flashing indicator on full screen. The device will not work until the batteries are
replaced.
† Spot-check complete. While Spot-check is in progress, a clockwise spinning circular icon displays.
† Measurement complete (full screen).
Symbol Description
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, Minnesota 55441, USA
(800) 356-8874 (USA/Canada)
+1 (763) 553-9968 (outside USA and Canada)
E-mail: technicalservice@nonin.com
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tilburg, Netherlands
+31 (0)13 - 45 87 130 (Europe)
E-mail: technicalserviceintl@nonin.com
nonin.com
70 75 80 85 90 95 100
−6−4−20246
SaO2(%)
SpO2−SaO2(%)
Bias=0
Limits of Agreement
Linear Regression
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
22
2
2
22 2
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
44
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
6
6
6
6
66
6
6
6
66
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
7
7
7
7
7
7
7
7
7
7
7
7
77
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
9
9
9
9
9
9
9
9
9
9
99
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
10
10
10
1010
10
10
10
10
10
10
10
10
10
10
10
11
11
11
11
11
11
11
11
11
11
11
11
11 11
11
11
11
11
11
1111
11
11 11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11 12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
13
13
13
13
13
13
13
13 13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
14
14
1414
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14 14
14
14
SpO2−SaO2(%)
Finger
Thumb
This graph shows plots of the error
(SpO2 – SaO2) by SaO2 using the
3230 with a linear regression fit and
upper 95% and lower 95% limits of
agreement. Each sample data point is
identified by subject from a clinical
study in non-motion conditions.
Accuracy Summary – Finger and Thumb
Range Specified Oxygen
Saturation (Arms)Thumb Oxygen
Saturation (Arms)
70 – 100% ± 2 ± 1.56
70 – 80% ± 2 ± 1.91
80 – 90% ± 2 ± 1.21
90 – 100% ± 2 ± 1.49
Finger Oxygen
Saturation (Arms)
± 1.31
± 1.65
± 1.05
± 1.18
Table 1: Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified in the Indications for Use section. The user of this
device should ensure that it is used in such an environment.
Emissions Test Compliance
RF Emissions
CISPR 11 Group 2, Class B
Table 2: Electromagnetic Immunity
Immunity Test Compliance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Power Frequency (50/60 Hz) Magnetic Field
IEC 61000-4-8 30 A/m
Radiated RF
IEC 61000-4-3
80 MHz – 2.7 GHz 10 V/m
380 – 390 MHz 27 V/m
430 – 470 MHz 28 V/m
704 – 787 MHz 9 V/m
800 – 960 MHz 28 V/m
1.7 – 1.99 GHz 28 V/m
2.4 – 2.57 GHz 28 V/m
5.1 – 5.8 GHz 9 V/m
Table 3: Not Applicable
Harmonic Emissions (IEC 61000-3-2), Voltage Flicker Emissions (IEC 61000-3-3), Electrical Fast Transients (IEC 61000-4-4),
Surge (IEC 61000-4-5), Voltage dips (IEC 61000-4-11), Conducted Immunity (IEC 61000-4-6)
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
111604-001-05 02/2022
© 2022 Nonin Medical, Inc.
U.S. Patents 5,490,523; 5,792,052; other patents pending.
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, MN 55441, USA
www.nonin.com
E-mail: info@nonin.com
+1 (763) 553-9968
(800) 356-8874 (US and Canada)
Fax: +1 (763) 553-7807
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tiburg, Netherlands
Email: infointl@nonin.com (Europe)
+31 (0)13 - 45 87 130 (Europe) MPS, Medical Product Service GmbH
Borngasse 20
D-35619 Braunfels, Germany
EC
REP
Nonin, NoninConnect, and CorrectCheck are trademarks of Nonin Medical, Inc.
The Bluetooth® word mark and logo are registered trademarks owned by Bluetooth SIG, Inc.
Continua™, the Continua logo and Continua Certified™ are trademarks, service marks, or certification marks of the Continua Health Alliance.
Installing AAA Batteries
WARNING: Before changing
batteries, make sure the device is
off and is not applied to a digit.
1.
Hold the 3230 so you see the
back of the device and the
arrows on the battery door
point away from you.
.
2. Place your thumbs on the ovals.
3. Slide the battery door away
from you and off the 3230.
4. If applicable, remove the
old batteries from the 3230.
Properly dispose of the
batteries.
5. Insert two new 1.5 volt AAA-
size batteries. Carefully match
the polarity markings (+ and -).
The 3230 will not work if the
batteries are inserted the
wrong way.
6. Carefully slide the battery door
back onto the device.
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
Turning On the NoninConnect Model
3230
1. Insert a digit into the Model 3230
until it touches the built-in stop.
NOTE: Make sure the finger is
centered within the finger guide and
flat (not on its side). For best results,
keep the device at heart or chest
level.
2. If the CorrectCheck screen
(see Display Symbols table)
displays, slide finger further
into device. Correct positioning
of the finger is critical for
accurate measurements.
3. The 3230 begins sensing the
pulse and displaying readings.
4. View about 4 seconds of
readings before relying on the
displayed values. Continually
verify operation. It is common
for the displayed values to vary
slightly over a period of several
seconds. If the 3230 does not
turn on or if it shuts off
unexpectedly:
• Verify batteries are correctly
inserted.
• The batteries are depleted. Replace
batteries.
If the problem persists, remove the
batteries and contact Nonin Technical
Service.
NOTE: While on the finger, do
not press the device against any
surface and do not squeeze or hold
it together. The internal spring
provides the correct pressure;
additional pressure may cause
inaccurate readings.
Instructions for Use – English
NoninConnect™Model 3230
Bluetooth®Smart Pulse Oximeter
Indications for Use
The NoninConnect Model 3230 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in measuring and
displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate of patients who are well or poorly
perfused. It is intended for spot-checking of adult and pediatric patients with digits between 0.8 – 2.5 cm (0.3 – 1.0 inch) thick.
NOTE: Use Environment—Home healthcare environments under the supervision of qualified medical professionals. Users
include current/potential users of pulse oximetry in the home and caregivers/potential caregivers of such a user.
Clinical Benefits
Nonin pulse oximeters allow for the management of patients’ medical conditions by providing fast, accurate, real-time,
noninvasive oxygen measurement in order to meet patients’ medical needs.
Warnings
Do not use the device in an MR environment, in an explosive atmosphere, or on neonatal patients.
• This device is not defibrillation proof per IEC 60601-1.
• Use the Model 3230 within its designated range (approximately 10 m/32 ft, spherical radius, line of sight when connected to
a Bluetooth Smart Ready device). Moving outside this range may cause missing, lost, and/or inaccurate data.
• Inspect the sensor application site at least every 4 hours to ensure correct sensor alignment and skin integrity. Patient sensitivity
to sensor may vary due to medical status or skin condition.
• Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
• This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
• The device must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is
hindering the pulse measurement before relying on the SpO2 measurement.
• Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
• General operation of the device may be affected by the use of an electrosurgical unit (ESU).
• Keep the oximeter away from young children. Small items such as the battery door and battery are choking hazards.
• Before changing batteries, make sure the device is off and is not applied to a digit.
• Portable RF communications equipment such as cell phones or radios (including peripherals such as antenna cables and
external antennas) should be used no closer than 30cm (12 inches) to any part of the ME system, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Cautions
• This device has no audible alarms and is intended only for spot-checking.
•
T
his device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may
degrade pulse oximeter performance or affect the accuracy of the measurement include the followi
ng:
• The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
• The device is designed to be attached only to a digit.
• This device’s display will shut off after 30 seconds of no readings or poor readings.
• In some circumstances, the device will interpret motion as good pulse quality. Minimize patient motion as much as possible.
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
• This device is a precision electronic instrument and must be repaired by Nonin Technical Service. Field repair of the device is
not possible. Do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the
warranty.
• This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or systems.
This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However,
because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in health care and other
environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the
performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be
installed and put into service according to the EMC information specified in this manual.
• When device is connected via Bluetooth, other Wi-Fi devices within 6 meters (20 feet) could interrupt the Bluetooth
• connection.
• Portable and mobile RF communications equipment including CT, diathermy, RFID, and electronic article security systems
can affect medical electrical equipment.
• Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than 30 days. Do
not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time. These
actions may cause the batteries to leak.
• Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of the device
and device components, including batteries.
• In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not
dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor
regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your
distributor’s contact information.
Symbols
NOTE: Where applicable, an additional label bearing your country radio communications license information will appear
on the side of your device. This is not a serial number or device identifier.
Display Symbols
† These indicators only display when the associated feature has been activated by an integrator.
Using the NoninConnect Model 3230
Installing AAA Batteries
Use only alkaline batteries. When batteries are low, displays. Replace low batteries as soon as possible.
See the “Installing AAA Batteries” instructions and figures at left.
Turning On the NoninConnect Model 3230
See the “Turning on the NoninConnect Model 3230” instructions and figures at left.
Connection via Bluetooth Wireless Technology
When the Model 3230 is placed on the finger and turns on, it is ready for a Bluetooth wireless connection. The 3230 stays in this
mode until it is shut off or the Bluetooth radio turns off. The symbol is white when the Bluetooth radio is on, green when
the 3230 is connected, and flashes when there is a communication error.
The Bluetooth symbol is useful for the product installer.
Due to the wide variety of wireless environments, the Bluetooth connection between the 3230 and the host device must be tested
before using the 3230’s Bluetooth capabilities.
Turning Off the NoninConnect Model 3230
The Model 3230 will automatically turn off approximately 10 seconds after the digit is removed, or after a 2-minute period of poor
signals.
Cleaning the NoninConnect Model 3230
CAUTIONS:
!
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
1. To clean, wipe the device’s surfaces with a soft cloth dampened with one of the following:
• A 10% bleach solution (household bleach [5.25% sodium hypochlorite]).
• Warm, soapy water (hand dishwashing detergent – see note below), and then rinse the cleaned surfaces with a soft cloth
dampened with water (home use only).
2. Dry with a soft cloth, or allow to air dry. Ensure that all surfaces are completely dry.
NOTE: The hand dishwashing detergent that was tested includes these ingredients: Sodium Lauryl Sulfate, Sodium Laureth
Sulfate, Lauramine Oxide, Sodium Chloride, PPG-26, PEG-8 Propylheptyl Ether, and Phenoxyethanol.
Warranty
For warranty information please refer to: http://www.nonin.com/warranty.
All non-warranty work shall be done at Nonin's standard rates and charges in effect at the time
of delivery to Nonin. All repairs include a complete restest of the Nonin 3230 using factory test
fixtures.
Users and/or patients should report adverse events involving their Nonin device to Nonin Medical, Inc.
Specifications
Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 321 beats per minute (BPM)
Declared Accuracy*: The table below shows Arms values measured using the Model 3230 in a clinical study.
NOTE: If your national regulatory authority recognizes accuracy in motion, please contact
regulatory@nonin.com for accuracy data.
SpO2 Low Perfusion Accuracy (Arms)*: 70 to 100% ±2 digits
Pulse Rate Declared Accuracy Range (Arms)*: 20 to 250 BPM ±3 digits
Low Perfusion Pulse Rate Declared Accuracy Range (Arms)*: 40 to 240 BPM ±3 digits
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW max. average
Infrared: 910 nanometers @ 1.2 mW max. average
Temperature:
Operating: -5 °C to 40 °C / 23 °F to 104 °F
Storage/Transportation: -40 °C to 70 °C / -40 °F to 158 °F
Time (from storage) for device to be ready for its intended use:
3 minutes to warm from -40 °C to -5 °C
8 minutes to cool from 70 °C to 40 °C
Humidity:
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing
Altitude:
Operating: Up to 4,000 meters / 13,123 feet
Hyperbaric Pressure: Up to 4 atmospheres
Battery Life:
Operating: Approximately 2,200 spot checks (25 sec. per spot-check), within 10 meters/32 feet of
collector with streaming data
Storage: 1 month, with batteries installed. CAUTION: Remove batteries if the device will be
stored for more than 30 days.
* ±1 Arms represents approximately 68% of measurements at zero bias.
** This information is especially useful for clinicians performing photodynamic therapy.
Bluetooth Wireless Technology Information
Bluetooth Compliance: Version 4.0 single mode low energy
Operating Frequency: 2.4 to 2.4835 GHz
Output Power: TX: +3 dBM
Operating Range: 10 meter radius (line of sight)
Network Topology: Star - bus
Operation: Slave
Model 3230
Antenna Type: Integrated chip type antenna
Modulation Type: Frequency Hopping Spread Spectrum
Data Rate: 1 Mbit/second
Data Latency: 6 ms
Data Integrity: Adaptive Frequency Hopping
24-bit CRC (cyclic redundancy check)
32-bit message integrity check
Data Format: Nonin Proprietary: Sends data packets once per second. Includes a second counter
that allows the host to detect if packets are missing and the device to retransmit.
Bluetooth SIG Standard: Compliant with Bluetooth SIG Pulse Oximeter Profile
specifications adopted by Continua.
Quality of Service: This device uses Bluetooth Smart technology for wireless communications, which
allows for reliable communications in electrically noisy environments, and transmits
physiological data. If the connection is lost, the device will become available for a
connection in a few seconds.
Bluetooth Profiles Supported: GATT-based Nonin Proprietary Oximeter Profile; GATT-based Bluetooth SIG Pulse
Oximeter Profile
Authentication and Encryption: Supported
Encryption Key Size: 128 bits AES (advanced encryption standard)
Bluetooth Security
The Bluetooth radio contained in the 3230 is a Bluetooth Smart single-mode, low-energy radio. The 3230 supports an encryption
key size of 128 bits. While the 3230 is in a Bluetooth connection, it will be unavailable for other connections. Apart from the
standard Bluetooth security measures, Nonin has two non-standard security measures that are available.
For additional technical information, please see the insert, “NoninConnect Model 3230 Technical Description.”
WARNING: Portable RF communications equipment such as cell phones or radios
(including peripherals such as antenna cables and external antennas) should be used no
closer than 30cm (12 inches) to any part of the ME system, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Manufacturer’s DeclarationRefer to the following tables for specific information regarding this device’s compliance to 60601-1-2. Component compliance
level is determined by system compliance level.
Essential Performance
Essential performance of the 3230 is defined as SpO2 accuracy and pulse rate accuracy or an indication of abnormal
operation. Accuracies may be affected as a result of exposure to electromagnetic disturbances that are outside of the
environments listed in the Indications For Use. If issues are experienced, move the Nonin system away from the source of
electromagnetic disturbances.
Symbol Definition Symbol Definition
!
Caution! Not for continuous monitoring (no alarm for
SpO2)
Follow Instructions for Use. Battery orientation
Medical Device
Importer
MR
MR
MR unsafe
IP32
Protected against vertically falling water drops
when enclosure is tilted up to 15 degrees and
ingress of solid foreign objects greater than or
equal to 2.5 mm (0.1 in.) in diameter per
IEC 60529.
Type BF Applied Part (patient isolation
from electrical shock)
UL Mark for Canada and the United
States with respect to electric shock, fire,
and mechanical hazards only in
accordance with IEC 60601-1,
UL 60601-1 and CAN/CSA-C22.2
No. 601.1.
SN Serial Number
BDA Bluetooth Device Address
CE Marking indicating conformance to
EC Directive No. 93/42/EEC
concerning medical devices. Storage/shipping temperature range
Non-ionizing electromagnetic radiation.
Equipment includes RF transmitters.
Interference may occur in the vicinity of
equipment marked with this symbol.
Handle with care
Keep dry
UDIUDI
Unique Identifier
Date of Manufacture Federal law (USA) restricts this device to sale
by or on the order of a licensed practitioner.
Continua Certified™ signifies this device
meets Continua certification testing
requirements, which support
interoperability in personal health
devices (continuaalliance.org).
US
US
Country of Manufacture
Manufacturer
Indicates separate collection for electrical
and electronic equipment (WEEE)
EC
REP
Authorized Representative in the European
Community
REFREFREFREF
Catalogue number
QTYQTY
Quantity
Symbol Description
Nonin’s CorrectCheck™ senses that the finger has not been correctly inserted. If you see this symbol,
slide finger further into device.
The number next to this symbol is the amount of oxygen in your blood (functional oxygen saturation
of arterial hemoglobin).
MR
MR
• applying the pulse oximeter on the
same arm as a blood pressure cuff,
arterial catheter or infusion line(s)
(IVs)
• excessive light, such as sunlight or
direct home lighting
• excessive motion
• moisture in the device
• improperly applied device
• finger is outside recommended
size range
• poor pulse quality
• venous pulsations
• cardiogreen and other
intravascular dyes
• anemia or low hemoglobin
concentrations
• carboxyhemoglobin
• methemoglobin
• dysfunctional hemoglobin
• artificial nails or fingernail polish
• residue (e.g., dried blood, dirt,
grease, oil) in the light path
The number next to this animated symbol is your pulse rate. Pulse rate is the number of times your
heart beats per minute.
– – – Dashes replace the readings when the 3230 is unable to detect a usable signal.
White symbol – Radio is on.
Green symbol – 3230 is connected.
Flashing symbol – Connection error. The radio will reset.
No symbol – Radio is off.
Poor signal. Steady your hand, reposition finger, warm finger by rubbing, or select a different finger.
Low battery. Replace batteries.
Critical battery. Flashing indicator on full screen. The device will not work until the batteries are
replaced.
† Spot-check complete. While Spot-check is in progress, a clockwise spinning circular icon displays.
† Measurement complete (full screen).
Symbol Description
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, Minnesota 55441, USA
(800) 356-8874 (USA/Canada)
+1 (763) 553-9968 (outside USA and Canada)
E-mail: technicalservice@nonin.com
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tilburg, Netherlands
+31 (0)13 - 45 87 130 (Europe)
E-mail: technicalserviceintl@nonin.com
nonin.com
70 75 80 85 90 95 100
−6−4−20246
SaO2(%)
SpO2−SaO2(%)
Bias=0
Limits of Agreement
Linear Regression
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
22
2
2
22 2
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
44
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
6
6
6
6
66
6
6
6
66
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
7
7
7
7
7
7
7
7
7
7
7
7
77
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
9
9
9
9
9
9
9
9
9
9
99
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
10
10
10
1010
10
10
10
10
10
10
10
10
10
10
10
11
11
11
11
11
11
11
11
11
11
11
11
11 11
11
11
11
11
11
1111
11
11 11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11 12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
13
13
13
13
13
13
13
13 13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
14
14
1414
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14 14
14
14
SpO2−SaO2(%)
Finger
Thumb
This graph shows plots of the error
(SpO2 – SaO2) by SaO2 using the
3230 with a linear regression fit and
upper 95% and lower 95% limits of
agreement. Each sample data point is
identified by subject from a clinical
study in non-motion conditions.
Accuracy Summary – Finger and Thumb
Range Specified Oxygen
Saturation (Arms)Thumb Oxygen
Saturation (Arms)
70 – 100% ± 2 ± 1.56
70 – 80% ± 2 ± 1.91
80 – 90% ± 2 ± 1.21
90 – 100% ± 2 ± 1.49
Finger Oxygen
Saturation (Arms)
± 1.31
± 1.65
± 1.05
± 1.18
Table 1: Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified in the Indications for Use section. The user of this
device should ensure that it is used in such an environment.
Emissions Test Compliance
RF Emissions
CISPR 11 Group 2, Class B
Table 2: Electromagnetic Immunity
Immunity Test Compliance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Power Frequency (50/60 Hz) Magnetic Field
IEC 61000-4-8 30 A/m
Radiated RF
IEC 61000-4-3
80 MHz – 2.7 GHz 10 V/m
380 – 390 MHz 27 V/m
430 – 470 MHz 28 V/m
704 – 787 MHz 9 V/m
800 – 960 MHz 28 V/m
1.7 – 1.99 GHz 28 V/m
2.4 – 2.57 GHz 28 V/m
5.1 – 5.8 GHz 9 V/m
Table 3: Not Applicable
Harmonic Emissions (IEC 61000-3-2), Voltage Flicker Emissions (IEC 61000-3-3), Electrical Fast Transients (IEC 61000-4-4),
Surge (IEC 61000-4-5), Voltage dips (IEC 61000-4-11), Conducted Immunity (IEC 61000-4-6)
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.

111604-001-05 02/2022
© 2022 Nonin Medical, Inc.
U.S. Patents 5,490,523; 5,792,052; other patents pending.
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, MN 55441, USA
www.nonin.com
E-mail: [email protected]
+1 (763) 553-9968
(800) 356-8874 (US and Canada)
Fax: +1 (763) 553-7807
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tiburg, Netherlands
Email: [email protected] (Europe)
+31 (0)13 - 45 87 130 (Europe) MPS, Medical Product Service GmbH
Borngasse 20
D-35619 Braunfels, Germany
EC
REP
Nonin, NoninConnect, and CorrectCheck are trademarks of Nonin Medical, Inc.
The Bluetooth® word mark and logo are registered trademarks owned by Bluetooth SIG, Inc.
Continua™, the Continua logo and Continua Certified™ are trademarks, service marks, or certification marks of the Continua Health Alliance.
Installing AAA Batteries
WARNING: Before changing
batteries, make sure the device is
off and is not applied to a digit.
1.
Hold the 3230 so you see the
back of the device and the
arrows on the battery door
point away from you.
.
2. Place your thumbs on the ovals.
3. Slide the battery door away
from you and off the 3230.
4. If applicable, remove the
old batteries from the 3230.
Properly dispose of the
batteries.
5. Insert two new 1.5 volt AAA-
size batteries. Carefully match
the polarity markings (+ and -).
The 3230 will not work if the
batteries are inserted the
wrong way.
6. Carefully slide the battery door
back onto the device.
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
Turning On the NoninConnect Model
3230
1. Insert a digit into the Model 3230
until it touches the built-in stop.
NOTE: Make sure the finger is
centered within the finger guide and
flat (not on its side). For best results,
keep the device at heart or chest
level.
2. If the CorrectCheck screen
(see Display Symbols table)
displays, slide finger further
into device. Correct positioning
of the finger is critical for
accurate measurements.
3. The 3230 begins sensing the
pulse and displaying readings.
4. View about 4 seconds of
readings before relying on the
displayed values. Continually
verify operation. It is common
for the displayed values to vary
slightly over a period of several
seconds. If the 3230 does not
turn on or if it shuts off
unexpectedly:
• Verify batteries are correctly
inserted.
• The batteries are depleted. Replace
batteries.
If the problem persists, remove the
batteries and contact Nonin Technical
Service.
NOTE: While on the finger, do
not press the device against any
surface and do not squeeze or hold
it together. The internal spring
provides the correct pressure;
additional pressure may cause
inaccurate readings.
Instructions for Use – English
NoninConnect™Model 3230
Bluetooth®Smart Pulse Oximeter
Indications for Use
The NoninConnect Model 3230 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in measuring and
displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate of patients who are well or poorly
perfused. It is intended for spot-checking of adult and pediatric patients with digits between 0.8 – 2.5 cm (0.3 – 1.0 inch) thick.
NOTE: Use Environment—Home healthcare environments under the supervision of qualified medical professionals. Users
include current/potential users of pulse oximetry in the home and caregivers/potential caregivers of such a user.
Clinical Benefits
Nonin pulse oximeters allow for the management of patients’ medical conditions by providing fast, accurate, real-time,
noninvasive oxygen measurement in order to meet patients’ medical needs.
Warnings
Do not use the device in an MR environment, in an explosive atmosphere, or on neonatal patients.
• This device is not defibrillation proof per IEC 60601-1.
• Use the Model 3230 within its designated range (approximately 10 m/32 ft, spherical radius, line of sight when connected to
a Bluetooth Smart Ready device). Moving outside this range may cause missing, lost, and/or inaccurate data.
• Inspect the sensor application site at least every 4 hours to ensure correct sensor alignment and skin integrity. Patient sensitivity
to sensor may vary due to medical status or skin condition.
• Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
• This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
• The device must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is
hindering the pulse measurement before relying on the SpO2 measurement.
• Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
• General operation of the device may be affected by the use of an electrosurgical unit (ESU).
• Keep the oximeter away from young children. Small items such as the battery door and battery are choking hazards.
• Before changing batteries, make sure the device is off and is not applied to a digit.
• Portable RF communications equipment such as cell phones or radios (including peripherals such as antenna cables and
external antennas) should be used no closer than 30cm (12 inches) to any part of the ME system, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Cautions
• This device has no audible alarms and is intended only for spot-checking.
•
T
his device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may
degrade pulse oximeter performance or affect the accuracy of the measurement include the followi
ng:
• The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
• The device is designed to be attached only to a digit.
• This device’s display will shut off after 30 seconds of no readings or poor readings.
• In some circumstances, the device will interpret motion as good pulse quality. Minimize patient motion as much as possible.
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
• This device is a precision electronic instrument and must be repaired by Nonin Technical Service. Field repair of the device is
not possible. Do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the
warranty.
• This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or systems.
This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However,
because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in health care and other
environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the
performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be
installed and put into service according to the EMC information specified in this manual.
• When device is connected via Bluetooth, other Wi-Fi devices within 6 meters (20 feet) could interrupt the Bluetooth
• connection.
• Portable and mobile RF communications equipment including CT, diathermy, RFID, and electronic article security systems
can affect medical electrical equipment.
• Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than 30 days. Do
not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time. These
actions may cause the batteries to leak.
• Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of the device
and device components, including batteries.
• In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not
dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor
regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your
distributor’s contact information.
Symbols
NOTE: Where applicable, an additional label bearing your country radio communications license information will appear
on the side of your device. This is not a serial number or device identifier.
Display Symbols
† These indicators only display when the associated feature has been activated by an integrator.
Using the NoninConnect Model 3230
Installing AAA Batteries
Use only alkaline batteries. When batteries are low, displays. Replace low batteries as soon as possible.
See the “Installing AAA Batteries” instructions and figures at left.
Turning On the NoninConnect Model 3230
See the “Turning on the NoninConnect Model 3230” instructions and figures at left.
Connection via Bluetooth Wireless Technology
When the Model 3230 is placed on the finger and turns on, it is ready for a Bluetooth wireless connection. The 3230 stays in this
mode until it is shut off or the Bluetooth radio turns off. The symbol is white when the Bluetooth radio is on, green when
the 3230 is connected, and flashes when there is a communication error.
The Bluetooth symbol is useful for the product installer.
Due to the wide variety of wireless environments, the Bluetooth connection between the 3230 and the host device must be tested
before using the 3230’s Bluetooth capabilities.
Turning Off the NoninConnect Model 3230
The Model 3230 will automatically turn off approximately 10 seconds after the digit is removed, or after a 2-minute period of poor
signals.
Cleaning the NoninConnect Model 3230
CAUTIONS:
!
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
1. To clean, wipe the device’s surfaces with a soft cloth dampened with one of the following:
• A 10% bleach solution (household bleach [5.25% sodium hypochlorite]).
• Warm, soapy water (hand dishwashing detergent – see note below), and then rinse the cleaned surfaces with a soft cloth
dampened with water (home use only).
2. Dry with a soft cloth, or allow to air dry. Ensure that all surfaces are completely dry.
NOTE: The hand dishwashing detergent that was tested includes these ingredients: Sodium Lauryl Sulfate, Sodium Laureth
Sulfate, Lauramine Oxide, Sodium Chloride, PPG-26, PEG-8 Propylheptyl Ether, and Phenoxyethanol.
Warranty
For warranty information please refer to: http://www.nonin.com/warranty.
All non-warranty work shall be done at Nonin's standard rates and charges in effect at the time
of delivery to Nonin. All repairs include a complete restest of the Nonin 3230 using factory test
fixtures.
Users and/or patients should report adverse events involving their Nonin device to Nonin Medical, Inc.
Specifications
Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 321 beats per minute (BPM)
Declared Accuracy*: The table below shows Arms values measured using the Model 3230 in a clinical study.
NOTE: If your national regulatory authority recognizes accuracy in motion, please contact
regulatory@nonin.com for accuracy data.
SpO2 Low Perfusion Accuracy (Arms)*: 70 to 100% ±2 digits
Pulse Rate Declared Accuracy Range (Arms)*: 20 to 250 BPM ±3 digits
Low Perfusion Pulse Rate Declared Accuracy Range (Arms)*: 40 to 240 BPM ±3 digits
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW max. average
Infrared: 910 nanometers @ 1.2 mW max. average
Temperature:
Operating: -5 °C to 40 °C / 23 °F to 104 °F
Storage/Transportation: -40 °C to 70 °C / -40 °F to 158 °F
Time (from storage) for device to be ready for its intended use:
3 minutes to warm from -40 °C to -5 °C
8 minutes to cool from 70 °C to 40 °C
Humidity:
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing
Altitude:
Operating: Up to 4,000 meters / 13,123 feet
Hyperbaric Pressure: Up to 4 atmospheres
Battery Life:
Operating: Approximately 2,200 spot checks (25 sec. per spot-check), within 10 meters/32 feet of
collector with streaming data
Storage: 1 month, with batteries installed. CAUTION: Remove batteries if the device will be
stored for more than 30 days.
* ±1 Arms represents approximately 68% of measurements at zero bias.
** This information is especially useful for clinicians performing photodynamic therapy.
Bluetooth Wireless Technology Information
Bluetooth Compliance: Version 4.0 single mode low energy
Operating Frequency: 2.4 to 2.4835 GHz
Output Power: TX: +3 dBM
Operating Range: 10 meter radius (line of sight)
Network Topology: Star - bus
Operation: Slave
Model 3230
Antenna Type: Integrated chip type antenna
Modulation Type: Frequency Hopping Spread Spectrum
Data Rate: 1 Mbit/second
Data Latency: 6 ms
Data Integrity: Adaptive Frequency Hopping
24-bit CRC (cyclic redundancy check)
32-bit message integrity check
Data Format: Nonin Proprietary: Sends data packets once per second. Includes a second counter
that allows the host to detect if packets are missing and the device to retransmit.
Bluetooth SIG Standard: Compliant with Bluetooth SIG Pulse Oximeter Profile
specifications adopted by Continua.
Quality of Service: This device uses Bluetooth Smart technology for wireless communications, which
allows for reliable communications in electrically noisy environments, and transmits
physiological data. If the connection is lost, the device will become available for a
connection in a few seconds.
Bluetooth Profiles Supported: GATT-based Nonin Proprietary Oximeter Profile; GATT-based Bluetooth SIG Pulse
Oximeter Profile
Authentication and Encryption: Supported
Encryption Key Size: 128 bits AES (advanced encryption standard)
Bluetooth Security
The Bluetooth radio contained in the 3230 is a Bluetooth Smart single-mode, low-energy radio. The 3230 supports an encryption
key size of 128 bits. While the 3230 is in a Bluetooth connection, it will be unavailable for other connections. Apart from the
standard Bluetooth security measures, Nonin has two non-standard security measures that are available.
For additional technical information, please see the insert, “NoninConnect Model 3230 Technical Description.”
WARNING: Portable RF communications equipment such as cell phones or radios
(including peripherals such as antenna cables and external antennas) should be used no
closer than 30cm (12 inches) to any part of the ME system, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Manufacturer’s DeclarationRefer to the following tables for specific information regarding this device’s compliance to 60601-1-2. Component compliance
level is determined by system compliance level.
Essential Performance
Essential performance of the 3230 is defined as SpO2 accuracy and pulse rate accuracy or an indication of abnormal
operation. Accuracies may be affected as a result of exposure to electromagnetic disturbances that are outside of the
environments listed in the Indications For Use. If issues are experienced, move the Nonin system away from the source of
electromagnetic disturbances.
Symbol Definition Symbol Definition
!
Caution! Not for continuous monitoring (no alarm for
SpO2)
Follow Instructions for Use. Battery orientation
Medical Device
Importer
MR
MR
MR unsafe
IP32
Protected against vertically falling water drops
when enclosure is tilted up to 15 degrees and
ingress of solid foreign objects greater than or
equal to 2.5 mm (0.1 in.) in diameter per
IEC 60529.
Type BF Applied Part (patient isolation
from electrical shock)
UL Mark for Canada and the United
States with respect to electric shock, fire,
and mechanical hazards only in
accordance with IEC 60601-1,
UL 60601-1 and CAN/CSA-C22.2
No. 601.1.
SN Serial Number
BDA Bluetooth Device Address
CE Marking indicating conformance to
EC Directive No. 93/42/EEC
concerning medical devices. Storage/shipping temperature range
Non-ionizing electromagnetic radiation.
Equipment includes RF transmitters.
Interference may occur in the vicinity of
equipment marked with this symbol.
Handle with care
Keep dry
UDIUDI
Unique Identifier
Date of Manufacture Federal law (USA) restricts this device to sale
by or on the order of a licensed practitioner.
Continua Certified™ signifies this device
meets Continua certification testing
requirements, which support
interoperability in personal health
devices (continuaalliance.org).
US
US
Country of Manufacture
Manufacturer
Indicates separate collection for electrical
and electronic equipment (WEEE)
EC
REP
Authorized Representative in the European
Community
REFREFREFREF
Catalogue number
QTYQTY
Quantity
Symbol Description
Nonin’s CorrectCheck™ senses that the finger has not been correctly inserted. If you see this symbol,
slide finger further into device.
The number next to this symbol is the amount of oxygen in your blood (functional oxygen saturation
of arterial hemoglobin).
MR
MR
• applying the pulse oximeter on the
same arm as a blood pressure cuff,
arterial catheter or infusion line(s)
(IVs)
• excessive light, such as sunlight or
direct home lighting
• excessive motion
• moisture in the device
• improperly applied device
• finger is outside recommended
size range
• poor pulse quality
• venous pulsations
• cardiogreen and other
intravascular dyes
• anemia or low hemoglobin
concentrations
• carboxyhemoglobin
• methemoglobin
• dysfunctional hemoglobin
• artificial nails or fingernail polish
• residue (e.g., dried blood, dirt,
grease, oil) in the light path
The number next to this animated symbol is your pulse rate. Pulse rate is the number of times your
heart beats per minute.
– – – Dashes replace the readings when the 3230 is unable to detect a usable signal.
White symbol – Radio is on.
Green symbol – 3230 is connected.
Flashing symbol – Connection error. The radio will reset.
No symbol – Radio is off.
Poor signal. Steady your hand, reposition finger, warm finger by rubbing, or select a different finger.
Low battery. Replace batteries.
Critical battery. Flashing indicator on full screen. The device will not work until the batteries are
replaced.
† Spot-check complete. While Spot-check is in progress, a clockwise spinning circular icon displays.
† Measurement complete (full screen).
Symbol Description
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, Minnesota 55441, USA
(800) 356-8874 (USA/Canada)
+1 (763) 553-9968 (outside USA and Canada)
E-mail: technicalservice@nonin.com
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tilburg, Netherlands
+31 (0)13 - 45 87 130 (Europe)
E-mail: technicalserviceintl@nonin.com
nonin.com
70 75 80 85 90 95 100
−6−4−20246
SaO2(%)
SpO2−SaO2(%)
Bias=0
Limits of Agreement
Linear Regression
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
22
2
2
22 2
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
44
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
6
6
6
6
66
6
6
6
66
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
7
7
7
7
7
7
7
7
7
7
7
7
77
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
9
9
9
9
9
9
9
9
9
9
99
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
10
10
10
1010
10
10
10
10
10
10
10
10
10
10
10
11
11
11
11
11
11
11
11
11
11
11
11
11 11
11
11
11
11
11
1111
11
11 11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11 12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
13
13
13
13
13
13
13
13 13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
14
14
1414
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14 14
14
14
SpO2−SaO2(%)
Finger
Thumb
This graph shows plots of the error
(SpO2 – SaO2) by SaO2 using the
3230 with a linear regression fit and
upper 95% and lower 95% limits of
agreement. Each sample data point is
identified by subject from a clinical
study in non-motion conditions.
Accuracy Summary – Finger and Thumb
Range Specified Oxygen
Saturation (Arms)Thumb Oxygen
Saturation (Arms)
70 – 100% ± 2 ± 1.56
70 – 80% ± 2 ± 1.91
80 – 90% ± 2 ± 1.21
90 – 100% ± 2 ± 1.49
Finger Oxygen
Saturation (Arms)
± 1.31
± 1.65
± 1.05
± 1.18
Table 1: Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified in the Indications for Use section. The user of this
device should ensure that it is used in such an environment.
Emissions Test Compliance
RF Emissions
CISPR 11 Group 2, Class B
Table 2: Electromagnetic Immunity
Immunity Test Compliance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Power Frequency (50/60 Hz) Magnetic Field
IEC 61000-4-8 30 A/m
Radiated RF
IEC 61000-4-3
80 MHz – 2.7 GHz 10 V/m
380 – 390 MHz 27 V/m
430 – 470 MHz 28 V/m
704 – 787 MHz 9 V/m
800 – 960 MHz 28 V/m
1.7 – 1.99 GHz 28 V/m
2.4 – 2.57 GHz 28 V/m
5.1 – 5.8 GHz 9 V/m
Table 3: Not Applicable
Harmonic Emissions (IEC 61000-3-2), Voltage Flicker Emissions (IEC 61000-3-3), Electrical Fast Transients (IEC 61000-4-4),
Surge (IEC 61000-4-5), Voltage dips (IEC 61000-4-11), Conducted Immunity (IEC 61000-4-6)
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.

111604-001-05 02/2022
© 2022 Nonin Medical, Inc.
U.S. Patents 5,490,523; 5,792,052; other patents pending.
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, MN 55441, USA
www.nonin.com
E-mail: [email protected]
+1 (763) 553-9968
(800) 356-8874 (US and Canada)
Fax: +1 (763) 553-7807
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tiburg, Netherlands
Email: [email protected] (Europe)
+31 (0)13 - 45 87 130 (Europe) MPS, Medical Product Service GmbH
Borngasse 20
D-35619 Braunfels, Germany
EC
REP
Nonin, NoninConnect, and CorrectCheck are trademarks of Nonin Medical, Inc.
The Bluetooth® word mark and logo are registered trademarks owned by Bluetooth SIG, Inc.
Continua™, the Continua logo and Continua Certified™ are trademarks, service marks, or certification marks of the Continua Health Alliance.
Installing AAA Batteries
WARNING: Before changing
batteries, make sure the device is
off and is not applied to a digit.
1.
Hold the 3230 so you see the
back of the device and the
arrows on the battery door
point away from you.
.
2. Place your thumbs on the ovals.
3. Slide the battery door away
from you and off the 3230.
4. If applicable, remove the
old batteries from the 3230.
Properly dispose of the
batteries.
5. Insert two new 1.5 volt AAA-
size batteries. Carefully match
the polarity markings (+ and -).
The 3230 will not work if the
batteries are inserted the
wrong way.
6. Carefully slide the battery door
back onto the device.
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
Turning On the NoninConnect Model
3230
1. Insert a digit into the Model 3230
until it touches the built-in stop.
NOTE: Make sure the finger is
centered within the finger guide and
flat (not on its side). For best results,
keep the device at heart or chest
level.
2. If the CorrectCheck screen
(see Display Symbols table)
displays, slide finger further
into device. Correct positioning
of the finger is critical for
accurate measurements.
3. The 3230 begins sensing the
pulse and displaying readings.
4. View about 4 seconds of
readings before relying on the
displayed values. Continually
verify operation. It is common
for the displayed values to vary
slightly over a period of several
seconds. If the 3230 does not
turn on or if it shuts off
unexpectedly:
• Verify batteries are correctly
inserted.
• The batteries are depleted. Replace
batteries.
If the problem persists, remove the
batteries and contact Nonin Technical
Service.
NOTE: While on the finger, do
not press the device against any
surface and do not squeeze or hold
it together. The internal spring
provides the correct pressure;
additional pressure may cause
inaccurate readings.
Instructions for Use – English
NoninConnect™Model 3230
Bluetooth®Smart Pulse Oximeter
Indications for Use
The NoninConnect Model 3230 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in measuring and
displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate of patients who are well or poorly
perfused. It is intended for spot-checking of adult and pediatric patients with digits between 0.8 – 2.5 cm (0.3 – 1.0 inch) thick.
NOTE: Use Environment—Home healthcare environments under the supervision of qualified medical professionals. Users
include current/potential users of pulse oximetry in the home and caregivers/potential caregivers of such a user.
Clinical Benefits
Nonin pulse oximeters allow for the management of patients’ medical conditions by providing fast, accurate, real-time,
noninvasive oxygen measurement in order to meet patients’ medical needs.
Warnings
Do not use the device in an MR environment, in an explosive atmosphere, or on neonatal patients.
• This device is not defibrillation proof per IEC 60601-1.
• Use the Model 3230 within its designated range (approximately 10 m/32 ft, spherical radius, line of sight when connected to
a Bluetooth Smart Ready device). Moving outside this range may cause missing, lost, and/or inaccurate data.
• Inspect the sensor application site at least every 4 hours to ensure correct sensor alignment and skin integrity. Patient sensitivity
to sensor may vary due to medical status or skin condition.
• Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
• This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
• The device must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is
hindering the pulse measurement before relying on the SpO2 measurement.
• Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
• General operation of the device may be affected by the use of an electrosurgical unit (ESU).
• Keep the oximeter away from young children. Small items such as the battery door and battery are choking hazards.
• Before changing batteries, make sure the device is off and is not applied to a digit.
• Portable RF communications equipment such as cell phones or radios (including peripherals such as antenna cables and
external antennas) should be used no closer than 30cm (12 inches) to any part of the ME system, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Cautions
• This device has no audible alarms and is intended only for spot-checking.
•
T
his device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may
degrade pulse oximeter performance or affect the accuracy of the measurement include the followi
ng:
• The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
• The device is designed to be attached only to a digit.
• This device’s display will shut off after 30 seconds of no readings or poor readings.
• In some circumstances, the device will interpret motion as good pulse quality. Minimize patient motion as much as possible.
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
• This device is a precision electronic instrument and must be repaired by Nonin Technical Service. Field repair of the device is
not possible. Do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the
warranty.
• This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or systems.
This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However,
because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in health care and other
environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the
performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be
installed and put into service according to the EMC information specified in this manual.
• When device is connected via Bluetooth, other Wi-Fi devices within 6 meters (20 feet) could interrupt the Bluetooth
• connection.
• Portable and mobile RF communications equipment including CT, diathermy, RFID, and electronic article security systems
can affect medical electrical equipment.
• Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than 30 days. Do
not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time. These
actions may cause the batteries to leak.
• Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of the device
and device components, including batteries.
• In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not
dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor
regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your
distributor’s contact information.
Symbols
NOTE: Where applicable, an additional label bearing your country radio communications license information will appear
on the side of your device. This is not a serial number or device identifier.
Display Symbols
† These indicators only display when the associated feature has been activated by an integrator.
Using the NoninConnect Model 3230
Installing AAA Batteries
Use only alkaline batteries. When batteries are low, displays. Replace low batteries as soon as possible.
See the “Installing AAA Batteries” instructions and figures at left.
Turning On the NoninConnect Model 3230
See the “Turning on the NoninConnect Model 3230” instructions and figures at left.
Connection via Bluetooth Wireless Technology
When the Model 3230 is placed on the finger and turns on, it is ready for a Bluetooth wireless connection. The 3230 stays in this
mode until it is shut off or the Bluetooth radio turns off. The symbol is white when the Bluetooth radio is on, green when
the 3230 is connected, and flashes when there is a communication error.
The Bluetooth symbol is useful for the product installer.
Due to the wide variety of wireless environments, the Bluetooth connection between the 3230 and the host device must be tested
before using the 3230’s Bluetooth capabilities.
Turning Off the NoninConnect Model 3230
The Model 3230 will automatically turn off approximately 10 seconds after the digit is removed, or after a 2-minute period of poor
signals.
Cleaning the NoninConnect Model 3230
CAUTIONS:
!
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
1. To clean, wipe the device’s surfaces with a soft cloth dampened with one of the following:
• A 10% bleach solution (household bleach [5.25% sodium hypochlorite]).
• Warm, soapy water (hand dishwashing detergent – see note below), and then rinse the cleaned surfaces with a soft cloth
dampened with water (home use only).
2. Dry with a soft cloth, or allow to air dry. Ensure that all surfaces are completely dry.
NOTE: The hand dishwashing detergent that was tested includes these ingredients: Sodium Lauryl Sulfate, Sodium Laureth
Sulfate, Lauramine Oxide, Sodium Chloride, PPG-26, PEG-8 Propylheptyl Ether, and Phenoxyethanol.
Warranty
For warranty information please refer to: http://www.nonin.com/warranty.
All non-warranty work shall be done at Nonin's standard rates and charges in effect at the time
of delivery to Nonin. All repairs include a complete restest of the Nonin 3230 using factory test
fixtures.
Users and/or patients should report adverse events involving their Nonin device to Nonin Medical, Inc.
Specifications
Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 321 beats per minute (BPM)
Declared Accuracy*: The table below shows Arms values measured using the Model 3230 in a clinical study.
NOTE: If your national regulatory authority recognizes accuracy in motion, please contact
regulatory@nonin.com for accuracy data.
SpO2 Low Perfusion Accuracy (Arms)*: 70 to 100% ±2 digits
Pulse Rate Declared Accuracy Range (Arms)*: 20 to 250 BPM ±3 digits
Low Perfusion Pulse Rate Declared Accuracy Range (Arms)*: 40 to 240 BPM ±3 digits
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW max. average
Infrared: 910 nanometers @ 1.2 mW max. average
Temperature:
Operating: -5 °C to 40 °C / 23 °F to 104 °F
Storage/Transportation: -40 °C to 70 °C / -40 °F to 158 °F
Time (from storage) for device to be ready for its intended use:
3 minutes to warm from -40 °C to -5 °C
8 minutes to cool from 70 °C to 40 °C
Humidity:
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing
Altitude:
Operating: Up to 4,000 meters / 13,123 feet
Hyperbaric Pressure: Up to 4 atmospheres
Battery Life:
Operating: Approximately 2,200 spot checks (25 sec. per spot-check), within 10 meters/32 feet of
collector with streaming data
Storage: 1 month, with batteries installed. CAUTION: Remove batteries if the device will be
stored for more than 30 days.
* ±1 Arms represents approximately 68% of measurements at zero bias.
** This information is especially useful for clinicians performing photodynamic therapy.
Bluetooth Wireless Technology Information
Bluetooth Compliance: Version 4.0 single mode low energy
Operating Frequency: 2.4 to 2.4835 GHz
Output Power: TX: +3 dBM
Operating Range: 10 meter radius (line of sight)
Network Topology: Star - bus
Operation: Slave
Model 3230
Antenna Type: Integrated chip type antenna
Modulation Type: Frequency Hopping Spread Spectrum
Data Rate: 1 Mbit/second
Data Latency: 6 ms
Data Integrity: Adaptive Frequency Hopping
24-bit CRC (cyclic redundancy check)
32-bit message integrity check
Data Format: Nonin Proprietary: Sends data packets once per second. Includes a second counter
that allows the host to detect if packets are missing and the device to retransmit.
Bluetooth SIG Standard: Compliant with Bluetooth SIG Pulse Oximeter Profile
specifications adopted by Continua.
Quality of Service: This device uses Bluetooth Smart technology for wireless communications, which
allows for reliable communications in electrically noisy environments, and transmits
physiological data. If the connection is lost, the device will become available for a
connection in a few seconds.
Bluetooth Profiles Supported: GATT-based Nonin Proprietary Oximeter Profile; GATT-based Bluetooth SIG Pulse
Oximeter Profile
Authentication and Encryption: Supported
Encryption Key Size: 128 bits AES (advanced encryption standard)
Bluetooth Security
The Bluetooth radio contained in the 3230 is a Bluetooth Smart single-mode, low-energy radio. The 3230 supports an encryption
key size of 128 bits. While the 3230 is in a Bluetooth connection, it will be unavailable for other connections. Apart from the
standard Bluetooth security measures, Nonin has two non-standard security measures that are available.
For additional technical information, please see the insert, “NoninConnect Model 3230 Technical Description.”
WARNING: Portable RF communications equipment such as cell phones or radios
(including peripherals such as antenna cables and external antennas) should be used no
closer than 30cm (12 inches) to any part of the ME system, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Manufacturer’s DeclarationRefer to the following tables for specific information regarding this device’s compliance to 60601-1-2. Component compliance
level is determined by system compliance level.
Essential Performance
Essential performance of the 3230 is defined as SpO2 accuracy and pulse rate accuracy or an indication of abnormal
operation. Accuracies may be affected as a result of exposure to electromagnetic disturbances that are outside of the
environments listed in the Indications For Use. If issues are experienced, move the Nonin system away from the source of
electromagnetic disturbances.
Symbol Definition Symbol Definition
!
Caution! Not for continuous monitoring (no alarm for
SpO2)
Follow Instructions for Use. Battery orientation
Medical Device
Importer
MR
MR
MR unsafe
IP32
Protected against vertically falling water drops
when enclosure is tilted up to 15 degrees and
ingress of solid foreign objects greater than or
equal to 2.5 mm (0.1 in.) in diameter per
IEC 60529.
Type BF Applied Part (patient isolation
from electrical shock)
UL Mark for Canada and the United
States with respect to electric shock, fire,
and mechanical hazards only in
accordance with IEC 60601-1,
UL 60601-1 and CAN/CSA-C22.2
No. 601.1.
SN Serial Number
BDA Bluetooth Device Address
CE Marking indicating conformance to
EC Directive No. 93/42/EEC
concerning medical devices. Storage/shipping temperature range
Non-ionizing electromagnetic radiation.
Equipment includes RF transmitters.
Interference may occur in the vicinity of
equipment marked with this symbol.
Handle with care
Keep dry
UDIUDI
Unique Identifier
Date of Manufacture Federal law (USA) restricts this device to sale
by or on the order of a licensed practitioner.
Continua Certified™ signifies this device
meets Continua certification testing
requirements, which support
interoperability in personal health
devices (continuaalliance.org).
US
US
Country of Manufacture
Manufacturer
Indicates separate collection for electrical
and electronic equipment (WEEE)
EC
REP
Authorized Representative in the European
Community
REFREFREFREF
Catalogue number
QTYQTY
Quantity
Symbol Description
Nonin’s CorrectCheck™ senses that the finger has not been correctly inserted. If you see this symbol,
slide finger further into device.
The number next to this symbol is the amount of oxygen in your blood (functional oxygen saturation
of arterial hemoglobin).
MR
MR
• applying the pulse oximeter on the
same arm as a blood pressure cuff,
arterial catheter or infusion line(s)
(IVs)
• excessive light, such as sunlight or
direct home lighting
• excessive motion
• moisture in the device
• improperly applied device
• finger is outside recommended
size range
• poor pulse quality
• venous pulsations
• cardiogreen and other
intravascular dyes
• anemia or low hemoglobin
concentrations
• carboxyhemoglobin
• methemoglobin
• dysfunctional hemoglobin
• artificial nails or fingernail polish
• residue (e.g., dried blood, dirt,
grease, oil) in the light path
The number next to this animated symbol is your pulse rate. Pulse rate is the number of times your
heart beats per minute.
– – – Dashes replace the readings when the 3230 is unable to detect a usable signal.
White symbol – Radio is on.
Green symbol – 3230 is connected.
Flashing symbol – Connection error. The radio will reset.
No symbol – Radio is off.
Poor signal. Steady your hand, reposition finger, warm finger by rubbing, or select a different finger.
Low battery. Replace batteries.
Critical battery. Flashing indicator on full screen. The device will not work until the batteries are
replaced.
† Spot-check complete. While Spot-check is in progress, a clockwise spinning circular icon displays.
† Measurement complete (full screen).
Symbol Description
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, Minnesota 55441, USA
(800) 356-8874 (USA/Canada)
+1 (763) 553-9968 (outside USA and Canada)
E-mail: [email protected]
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tilburg, Netherlands
+31 (0)13 - 45 87 130 (Europe)
E-mail: [email protected]
nonin.com
70 75 80 85 90 95 100
−6−4−20246
SaO2(%)
SpO2−SaO2(%)
Bias=0
Limits of Agreement
Linear Regression
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
22
2
2
22 2
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
44
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
6
6
6
6
66
6
6
6
66
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
7
7
7
7
7
7
7
7
7
7
7
7
77
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
9
9
9
9
9
9
9
9
9
9
99
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
10
10
10
1010
10
10
10
10
10
10
10
10
10
10
10
11
11
11
11
11
11
11
11
11
11
11
11
11 11
11
11
11
11
11
1111
11
11 11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11 12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
13
13
13
13
13
13
13
13 13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
14
14
1414
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14 14
14
14
SpO2−SaO2(%)
Finger
Thumb
This graph shows plots of the error
(SpO2 – SaO2) by SaO2 using the
3230 with a linear regression fit and
upper 95% and lower 95% limits of
agreement. Each sample data point is
identified by subject from a clinical
study in non-motion conditions.
Accuracy Summary – Finger and Thumb
Range Specified Oxygen
Saturation (Arms)Thumb Oxygen
Saturation (Arms)
70 – 100% ± 2 ± 1.56
70 – 80% ± 2 ± 1.91
80 – 90% ± 2 ± 1.21
90 – 100% ± 2 ± 1.49
Finger Oxygen
Saturation (Arms)
± 1.31
± 1.65
± 1.05
± 1.18
Table 1: Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified in the Indications for Use section. The user of this
device should ensure that it is used in such an environment.
Emissions Test Compliance
RF Emissions
CISPR 11 Group 2, Class B
Table 2: Electromagnetic Immunity
Immunity Test Compliance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Power Frequency (50/60 Hz) Magnetic Field
IEC 61000-4-8 30 A/m
Radiated RF
IEC 61000-4-3
80 MHz – 2.7 GHz 10 V/m
380 – 390 MHz 27 V/m
430 – 470 MHz 28 V/m
704 – 787 MHz 9 V/m
800 – 960 MHz 28 V/m
1.7 – 1.99 GHz 28 V/m
2.4 – 2.57 GHz 28 V/m
5.1 – 5.8 GHz 9 V/m
Table 3: Not Applicable
Harmonic Emissions (IEC 61000-3-2), Voltage Flicker Emissions (IEC 61000-3-3), Electrical Fast Transients (IEC 61000-4-4),
Surge (IEC 61000-4-5), Voltage dips (IEC 61000-4-11), Conducted Immunity (IEC 61000-4-6)
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.

111604-001-05 02/2022
© 2022 Nonin Medical, Inc.
U.S. Patents 5,490,523; 5,792,052; other patents pending.
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, MN 55441, USA
www.nonin.com
E-mail: [email protected]
+1 (763) 553-9968
(800) 356-8874 (US and Canada)
Fax: +1 (763) 553-7807
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tiburg, Netherlands
Email: [email protected] (Europe)
+31 (0)13 - 45 87 130 (Europe) MPS, Medical Product Service GmbH
Borngasse 20
D-35619 Braunfels, Germany
EC
REP
Nonin, NoninConnect, and CorrectCheck are trademarks of Nonin Medical, Inc.
The Bluetooth® word mark and logo are registered trademarks owned by Bluetooth SIG, Inc.
Continua™, the Continua logo and Continua Certified™ are trademarks, service marks, or certification marks of the Continua Health Alliance.
Installing AAA Batteries
WARNING: Before changing
batteries, make sure the device is
off and is not applied to a digit.
1.
Hold the 3230 so you see the
back of the device and the
arrows on the battery door
point away from you.
.
2. Place your thumbs on the ovals.
3. Slide the battery door away
from you and off the 3230.
4. If applicable, remove the
old batteries from the 3230.
Properly dispose of the
batteries.
5. Insert two new 1.5 volt AAA-
size batteries. Carefully match
the polarity markings (+ and -).
The 3230 will not work if the
batteries are inserted the
wrong way.
6. Carefully slide the battery door
back onto the device.
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
MODEL 3230 OXIMETER
NONIN MEDICAL, INC.
PLYMOUTH, MN USA
IP32
R-NZ
Turning On the NoninConnect Model
3230
1. Insert a digit into the Model 3230
until it touches the built-in stop.
NOTE: Make sure the finger is
centered within the finger guide and
flat (not on its side). For best results,
keep the device at heart or chest
level.
2. If the CorrectCheck screen
(see Display Symbols table)
displays, slide finger further
into device. Correct positioning
of the finger is critical for
accurate measurements.
3. The 3230 begins sensing the
pulse and displaying readings.
4. View about 4 seconds of
readings before relying on the
displayed values. Continually
verify operation. It is common
for the displayed values to vary
slightly over a period of several
seconds. If the 3230 does not
turn on or if it shuts off
unexpectedly:
• Verify batteries are correctly
inserted.
• The batteries are depleted. Replace
batteries.
If the problem persists, remove the
batteries and contact Nonin Technical
Service.
NOTE: While on the finger, do
not press the device against any
surface and do not squeeze or hold
it together. The internal spring
provides the correct pressure;
additional pressure may cause
inaccurate readings.
Instructions for Use – English
NoninConnect™Model 3230
Bluetooth®Smart Pulse Oximeter
Indications for Use
The NoninConnect Model 3230 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in measuring and
displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate of patients who are well or poorly
perfused. It is intended for spot-checking of adult and pediatric patients with digits between 0.8 – 2.5 cm (0.3 – 1.0 inch) thick.
NOTE: Use Environment—Home healthcare environments under the supervision of qualified medical professionals. Users
include current/potential users of pulse oximetry in the home and caregivers/potential caregivers of such a user.
Clinical Benefits
Nonin pulse oximeters allow for the management of patients’ medical conditions by providing fast, accurate, real-time,
noninvasive oxygen measurement in order to meet patients’ medical needs.
Warnings
Do not use the device in an MR environment, in an explosive atmosphere, or on neonatal patients.
• This device is not defibrillation proof per IEC 60601-1.
• Use the Model 3230 within its designated range (approximately 10 m/32 ft, spherical radius, line of sight when connected to
a Bluetooth Smart Ready device). Moving outside this range may cause missing, lost, and/or inaccurate data.
• Inspect the sensor application site at least every 4 hours to ensure correct sensor alignment and skin integrity. Patient sensitivity
to sensor may vary due to medical status or skin condition.
• Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
• This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
• The device must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is
hindering the pulse measurement before relying on the SpO2 measurement.
• Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
• General operation of the device may be affected by the use of an electrosurgical unit (ESU).
• Keep the oximeter away from young children. Small items such as the battery door and battery are choking hazards.
• Before changing batteries, make sure the device is off and is not applied to a digit.
• Portable RF communications equipment such as cell phones or radios (including peripherals such as antenna cables and
external antennas) should be used no closer than 30cm (12 inches) to any part of the ME system, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Cautions
• This device has no audible alarms and is intended only for spot-checking.
•
T
his device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may
degrade pulse oximeter performance or affect the accuracy of the measurement include the followi
ng:
• The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
• The device is designed to be attached only to a digit.
• This device’s display will shut off after 30 seconds of no readings or poor readings.
• In some circumstances, the device will interpret motion as good pulse quality. Minimize patient motion as much as possible.
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
• This device is a precision electronic instrument and must be repaired by Nonin Technical Service. Field repair of the device is
not possible. Do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the
warranty.
• This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or systems.
This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However,
because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in health care and other
environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the
performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be
installed and put into service according to the EMC information specified in this manual.
• When device is connected via Bluetooth, other Wi-Fi devices within 6 meters (20 feet) could interrupt the Bluetooth
• connection.
• Portable and mobile RF communications equipment including CT, diathermy, RFID, and electronic article security systems
can affect medical electrical equipment.
• Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than 30 days. Do
not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time. These
actions may cause the batteries to leak.
• Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of the device
and device components, including batteries.
• In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not
dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor
regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your
distributor’s contact information.
Symbols
NOTE: Where applicable, an additional label bearing your country radio communications license information will appear
on the side of your device. This is not a serial number or device identifier.
Display Symbols
† These indicators only display when the associated feature has been activated by an integrator.
Using the NoninConnect Model 3230
Installing AAA Batteries
Use only alkaline batteries. When batteries are low, displays. Replace low batteries as soon as possible.
See the “Installing AAA Batteries” instructions and figures at left.
Turning On the NoninConnect Model 3230
See the “Turning on the NoninConnect Model 3230” instructions and figures at left.
Connection via Bluetooth Wireless Technology
When the Model 3230 is placed on the finger and turns on, it is ready for a Bluetooth wireless connection. The 3230 stays in this
mode until it is shut off or the Bluetooth radio turns off. The symbol is white when the Bluetooth radio is on, green when
the 3230 is connected, and flashes when there is a communication error.
The Bluetooth symbol is useful for the product installer.
Due to the wide variety of wireless environments, the Bluetooth connection between the 3230 and the host device must be tested
before using the 3230’s Bluetooth capabilities.
Turning Off the NoninConnect Model 3230
The Model 3230 will automatically turn off approximately 10 seconds after the digit is removed, or after a 2-minute period of poor
signals.
Cleaning the NoninConnect Model 3230
CAUTIONS:
!
• Clean the device before applying it to a new patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.
• Do not use cleaning solutions other than those recommended here, as permanent damage could result.
1. To clean, wipe the device’s surfaces with a soft cloth dampened with one of the following:
• A 10% bleach solution (household bleach [5.25% sodium hypochlorite]).
• Warm, soapy water (hand dishwashing detergent – see note below), and then rinse the cleaned surfaces with a soft cloth
dampened with water (home use only).
2. Dry with a soft cloth, or allow to air dry. Ensure that all surfaces are completely dry.
NOTE: The hand dishwashing detergent that was tested includes these ingredients: Sodium Lauryl Sulfate, Sodium Laureth
Sulfate, Lauramine Oxide, Sodium Chloride, PPG-26, PEG-8 Propylheptyl Ether, and Phenoxyethanol.
Warranty
For warranty information please refer to: http://www.nonin.com/warranty.
All non-warranty work shall be done at Nonin's standard rates and charges in effect at the time
of delivery to Nonin. All repairs include a complete restest of the Nonin 3230 using factory test
fixtures.
Users and/or patients should report adverse events involving their Nonin device to Nonin Medical, Inc.
Specifications
Oxygen Saturation Display Range: 0% to 100% SpO2
Pulse Rate Display Range: 18 to 321 beats per minute (BPM)
Declared Accuracy*: The table below shows Arms values measured using the Model 3230 in a clinical study.
NOTE: If your national regulatory authority recognizes accuracy in motion, please contact
regulatory@nonin.com for accuracy data.
SpO2 Low Perfusion Accuracy (Arms)*: 70 to 100% ±2 digits
Pulse Rate Declared Accuracy Range (Arms)*: 20 to 250 BPM ±3 digits
Low Perfusion Pulse Rate Declared Accuracy Range (Arms)*: 40 to 240 BPM ±3 digits
Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW max. average
Infrared: 910 nanometers @ 1.2 mW max. average
Temperature:
Operating: -5 °C to 40 °C / 23 °F to 104 °F
Storage/Transportation: -40 °C to 70 °C / -40 °F to 158 °F
Time (from storage) for device to be ready for its intended use:
3 minutes to warm from -40 °C to -5 °C
8 minutes to cool from 70 °C to 40 °C
Humidity:
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing
Altitude:
Operating: Up to 4,000 meters / 13,123 feet
Hyperbaric Pressure: Up to 4 atmospheres
Battery Life:
Operating: Approximately 2,200 spot checks (25 sec. per spot-check), within 10 meters/32 feet of
collector with streaming data
Storage: 1 month, with batteries installed. CAUTION: Remove batteries if the device will be
stored for more than 30 days.
* ±1 Arms represents approximately 68% of measurements at zero bias.
** This information is especially useful for clinicians performing photodynamic therapy.
Bluetooth Wireless Technology Information
Bluetooth Compliance: Version 4.0 single mode low energy
Operating Frequency: 2.4 to 2.4835 GHz
Output Power: TX: +3 dBM
Operating Range: 10 meter radius (line of sight)
Network Topology: Star - bus
Operation: Slave
Model 3230
Antenna Type: Integrated chip type antenna
Modulation Type: Frequency Hopping Spread Spectrum
Data Rate: 1 Mbit/second
Data Latency: 6 ms
Data Integrity: Adaptive Frequency Hopping
24-bit CRC (cyclic redundancy check)
32-bit message integrity check
Data Format: Nonin Proprietary: Sends data packets once per second. Includes a second counter
that allows the host to detect if packets are missing and the device to retransmit.
Bluetooth SIG Standard: Compliant with Bluetooth SIG Pulse Oximeter Profile
specifications adopted by Continua.
Quality of Service: This device uses Bluetooth Smart technology for wireless communications, which
allows for reliable communications in electrically noisy environments, and transmits
physiological data. If the connection is lost, the device will become available for a
connection in a few seconds.
Bluetooth Profiles Supported: GATT-based Nonin Proprietary Oximeter Profile; GATT-based Bluetooth SIG Pulse
Oximeter Profile
Authentication and Encryption: Supported
Encryption Key Size: 128 bits AES (advanced encryption standard)
Bluetooth Security
The Bluetooth radio contained in the 3230 is a Bluetooth Smart single-mode, low-energy radio. The 3230 supports an encryption
key size of 128 bits. While the 3230 is in a Bluetooth connection, it will be unavailable for other connections. Apart from the
standard Bluetooth security measures, Nonin has two non-standard security measures that are available.
For additional technical information, please see the insert, “NoninConnect Model 3230 Technical Description.”
WARNING: Portable RF communications equipment such as cell phones or radios
(including peripherals such as antenna cables and external antennas) should be used no
closer than 30cm (12 inches) to any part of the ME system, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Manufacturer’s DeclarationRefer to the following tables for specific information regarding this device’s compliance to 60601-1-2. Component compliance
level is determined by system compliance level.
Essential Performance
Essential performance of the 3230 is defined as SpO2 accuracy and pulse rate accuracy or an indication of abnormal
operation. Accuracies may be affected as a result of exposure to electromagnetic disturbances that are outside of the
environments listed in the Indications For Use. If issues are experienced, move the Nonin system away from the source of
electromagnetic disturbances.
Symbol Definition Symbol Definition
!
Caution! Not for continuous monitoring (no alarm for
SpO2)
Follow Instructions for Use. Battery orientation
Medical Device
Importer
MR
MR
MR unsafe
IP32
Protected against vertically falling water drops
when enclosure is tilted up to 15 degrees and
ingress of solid foreign objects greater than or
equal to 2.5 mm (0.1 in.) in diameter per
IEC 60529.
Type BF Applied Part (patient isolation
from electrical shock)
UL Mark for Canada and the United
States with respect to electric shock, fire,
and mechanical hazards only in
accordance with IEC 60601-1,
UL 60601-1 and CAN/CSA-C22.2
No. 601.1.
SN Serial Number
BDA Bluetooth Device Address
CE Marking indicating conformance to
EC Directive No. 93/42/EEC
concerning medical devices. Storage/shipping temperature range
Non-ionizing electromagnetic radiation.
Equipment includes RF transmitters.
Interference may occur in the vicinity of
equipment marked with this symbol.
Handle with care
Keep dry
UDIUDI
Unique Identifier
Date of Manufacture Federal law (USA) restricts this device to sale
by or on the order of a licensed practitioner.
Continua Certified™ signifies this device
meets Continua certification testing
requirements, which support
interoperability in personal health
devices (continuaalliance.org).
US
US
Country of Manufacture
Manufacturer
Indicates separate collection for electrical
and electronic equipment (WEEE)
EC
REP
Authorized Representative in the European
Community
REFREFREFREF
Catalogue number
QTYQTY
Quantity
Symbol Description
Nonin’s CorrectCheck™ senses that the finger has not been correctly inserted. If you see this symbol,
slide finger further into device.
The number next to this symbol is the amount of oxygen in your blood (functional oxygen saturation
of arterial hemoglobin).
MR
MR
• applying the pulse oximeter on the
same arm as a blood pressure cuff,
arterial catheter or infusion line(s)
(IVs)
• excessive light, such as sunlight or
direct home lighting
• excessive motion
• moisture in the device
• improperly applied device
• finger is outside recommended
size range
• poor pulse quality
• venous pulsations
• cardiogreen and other
intravascular dyes
• anemia or low hemoglobin
concentrations
• carboxyhemoglobin
• methemoglobin
• dysfunctional hemoglobin
• artificial nails or fingernail polish
• residue (e.g., dried blood, dirt,
grease, oil) in the light path
The number next to this animated symbol is your pulse rate. Pulse rate is the number of times your
heart beats per minute.
– – – Dashes replace the readings when the 3230 is unable to detect a usable signal.
White symbol – Radio is on.
Green symbol – 3230 is connected.
Flashing symbol – Connection error. The radio will reset.
No symbol – Radio is off.
Poor signal. Steady your hand, reposition finger, warm finger by rubbing, or select a different finger.
Low battery. Replace batteries.
Critical battery. Flashing indicator on full screen. The device will not work until the batteries are
replaced.
† Spot-check complete. While Spot-check is in progress, a clockwise spinning circular icon displays.
† Measurement complete (full screen).
Symbol Description
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, Minnesota 55441, USA
(800) 356-8874 (USA/Canada)
+1 (763) 553-9968 (outside USA and Canada)
E-mail: [email protected]
Nonin Medical B.V.
Doctor Paul Janssenweg 150
5026 RH Tilburg, Netherlands
+31 (0)13 - 45 87 130 (Europe)
E-mail: [email protected]
nonin.com
70 75 80 85 90 95 100
−6−4−20246
SaO2(%)
SpO2−SaO2(%)
Bias=0
Limits of Agreement
Linear Regression
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
22
2
2
22 2
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
33
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
33
3
3
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
44
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
6
6
6
6
66
6
6
6
66
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
7
7
7
7
7
7
7
7
7
7
7
7
77
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
88
8
9
9
9
9
9
9
9
9
9
9
99
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
9
10
10
10
1010
10
10
10
10
10
10
10
10
10
10
10
11
11
11
11
11
11
11
11
11
11
11
11
11 11
11
11
11
11
11
1111
11
11 11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11
11 12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
12
13
13
13
13
13
13
13
13 13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
13
14
14
1414
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14
14 14
14
14
SpO2−SaO2(%)
Finger
Thumb
This graph shows plots of the error
(SpO2 – SaO2) by SaO2 using the
3230 with a linear regression fit and
upper 95% and lower 95% limits of
agreement. Each sample data point is
identified by subject from a clinical
study in non-motion conditions.
Accuracy Summary – Finger and Thumb
Range Specified Oxygen
Saturation (Arms)Thumb Oxygen
Saturation (Arms)
70 – 100% ± 2 ± 1.56
70 – 80% ± 2 ± 1.91
80 – 90% ± 2 ± 1.21
90 – 100% ± 2 ± 1.49
Finger Oxygen
Saturation (Arms)
± 1.31
± 1.65
± 1.05
± 1.18
Table 1: Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified in the Indications for Use section. The user of this
device should ensure that it is used in such an environment.
Emissions Test Compliance
RF Emissions
CISPR 11 Group 2, Class B
Table 2: Electromagnetic Immunity
Immunity Test Compliance
Electrostatic Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
Power Frequency (50/60 Hz) Magnetic Field
IEC 61000-4-8 30 A/m
Radiated RF
IEC 61000-4-3
80 MHz – 2.7 GHz 10 V/m
380 – 390 MHz 27 V/m
430 – 470 MHz 28 V/m
704 – 787 MHz 9 V/m
800 – 960 MHz 28 V/m
1.7 – 1.99 GHz 28 V/m
2.4 – 2.57 GHz 28 V/m
5.1 – 5.8 GHz 9 V/m
Table 3: Not Applicable
Harmonic Emissions (IEC 61000-3-2), Voltage Flicker Emissions (IEC 61000-3-3), Electrical Fast Transients (IEC 61000-4-4),
Surge (IEC 61000-4-5), Voltage dips (IEC 61000-4-11), Conducted Immunity (IEC 61000-4-6)
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects, and people.
Other manuals for 3230
4
Table of contents
Other Nonin Medical Equipment manuals

Nonin
Nonin NoninConnect Elite 3240 User manual

Nonin
Nonin CO-Pilot H500 User manual

Nonin
Nonin GO2 LED Achieve User manual

Nonin
Nonin 9570 User manual

Nonin
Nonin WristOx2 3150 User manual

Nonin
Nonin Onyx II 9550 User manual

Nonin
Nonin Avant 9600 User manual

Nonin
Nonin 2500A PalmSAT User manual

Nonin
Nonin 7500 User manual

Nonin
Nonin 9847 User manual

Nonin
Nonin 7500 Troubleshooting guide

Nonin
Nonin 9550 User manual

Nonin
Nonin 7600 User manual

Nonin
Nonin 9847V User manual

Nonin
Nonin 2120 User manual

Nonin
Nonin 7500FO User manual

Nonin
Nonin Onyx 9500 User manual

Nonin
Nonin Avant 9700 User manual

Nonin
Nonin GO2 Achieve User manual

Nonin
Nonin Avant 9600 User manual