
Section 2
12
CO
2
SMO
Plus
User’s Manual Rev. 02
• Periodically check the flow sensor and tubing for excessive moisture or secretion build up.
Although the
CO
2
SMO
Plus!
automatically purges the lines, excessive moisture or secretions may
still remain.
• While using the flow sensor, a system leak, such as that caused by uncuffed endotracheal tubes or
a damaged flow sensor may significantly effect flow related readings. These include flow,
pressure, dead space, CO2production and other respiratory mechanics parameters.
• Connect the line cord to a grounded hospital-grade outlet.
CO
2
SMO
Plus!
should be connected to
the same electrical circuit as other equipment in use on the patient. Outlets of the same circuit can
be identified by the hospital’s engineering department.
•The
CO
2
SMO
Plus!
has no protection against the ingress of water.
• Do not operate
CO
2
SMO
Plus!
when it is wet due to spills or condensation.
• Do not operate
CO
2
SMO
Plus!
if it appears to have been dropped or damaged.
• Keep
CO
2
SMO
Plus!
and its accessories clean.
• Never sterilize or immerse the monitor in liquids.
• Do not sterilize or immerse sensors except as directed in this manual.
• Do not apply excessive tension to any sensor cable or pneumatic tubing.
• Do not store the monitor or sensors at temperatures less than 14°F (-10°C) or above 131°F (55°C).
• Do not operate the monitor or sensors at temperatures below 50°F (10°C) or above 104°F (40°C).
• Federal (U.S.A.) law restricts this device to sale, distribution, or use by or on the order of a
licensed medical practitioner.
• DO NOT attach an SpO2sensor distal to a blood pressure cuff. Valid data CANNOT be processed
when the cuff is inflated. Attach the sensor to the limb opposite to the site used for the blood
pressure cuff.
• Excessive moisture in the flow sensor may affect the accuracy of the flow measurement.
• To avoidthe effects of excessive moisture in themeasurement circuit, insert the flow sensor in the
ventilator circuit with the tubes upright (as shown by arrows). The striped tube is positioned
closest to the patient. Improper placement will result in erroneous data.
• It is recommended that the CO2/Flow sensorbe removed from the circuit whenever an aerosolized
medication is delivered. This is due to the increased viscosity of the medications which may
contaminate the sensor windows, causing the sensor to fail prematurely.
• In case of interference with our equipment or another manufacturer’s equipment, notify your
Novametrix representative.
• Components of this product and its associated accessories which have patient contact are free of
latex.
• As with all flow measuring devices, adverse conditions may affect the accuracy of the flow
measurement. See page 10 for a discussion of flow measurement.
•IfCO
2is being measured while not using a flow sensor, certain rebreathing circuits, or the
presence of artifacts such as cardiogenic oscillations, may cause
CO
2
SMO
Plus!
to react to non-
respiratory CO2fluctuations as if they were breaths. This condition affects only the numerical
displays; the capnogram display continues to provide an accurate picture of the CO2waveform.
CAUTIONS
Indicates a condition that may lead to equipment damage or malfunction.
NOTES
Indicates points of particular interest or emphasis for more efficient or convenient operation.
!