
18
Instructions for Use
Figure 41
F. Performing IV Infusions
1. Procure a 3rd IV bag (sold separately, LF01130U).
2. Begin with a fully pressurized system.
3. Insert the IV needle or butterfly into a vein. Flashback will indicate a proper insertion.
4. Attach the needle or butterfly to the tubing from the infusion solution or fluid supply Bag C using the small latex piece.
(See figure 39.)
5. Open the clamp on fluid supply Bag B.
6. Open the clamp from the infusion solution.
7. Proof of proper procedure will be evidenced by the flow of fluid from the infusion solution or fluid supply Bag C.
Note: Simulate cleansing IV, injection, and withdrawal sites using distilled water. Multiple or prolonged exposure to
harsh antiseptics such as alcohol or iodine could damage or stain the simulator. Performing infusions can dilute your
simulated blood solution; ensure Bag B is mostly empty before infusing.
8. Always drain the blood and run clear water through the system at the end of each use.
9. Allow all components to air dry completely after each use and before re-assembly or storage.
G. Chest Tube
A chest tube may be inserted into the site in the left mid-axial line. This is a nonfunctioning site that may be used for prac-
tice in chest tube care. (See figure 40.)
1. Apply a small amount of the provided lubricant to chest tube appliances.
2. Use an alternative method to secure. Most adhesives do not stick well to the infant simulator’s skin.
3. Remove all devices from the infant simulator after each use.
4. Ensure the infant simulator and chest tube site are clean and dry after each use.
H. Umbilicus
The umbilicus may be catheterized through the vein or the artery. Always lubricate catheters before insertion.
1. To take blood samples, pull out the umbilicus, using caution to not disconnect the pulse system tubing.
2. Fill the reservoir with approximately 2-4 cc of blood mixture, and replace umbilicus. (See figures 41 and 42.)
3. To infuse fluids, leave the reservoir empty. (See figure 43.)
Figure 43
Figure 42
Figure 40
15
Figure 39
HC190001INM26.indd 15 2/19/19 1:55 PM
1. Apply a small amount of the provided lubricant to chest tube
appliances.
2. Use an alternative method to secure. Most adhesives do not stick
well to the infant simulator’s skin.
3. Remove all devices from the infant simulator after each use.
4. Ensure the infant simulator and chest tube site are clean and dry
after each use.
H. Umbilicus
The umbilicus may be catheterized through the vein or the artery.
Always lubricate catheters before insertion.
1. To take blood samples, pull out the umbilicus, using caution to not
disconnect the pulse system tubing.
2. Fill the reservoir with approximately 2-4 cc of blood mixture, and
replace umbilicus. (See figures 41 and 42.)
3. To infuse fluids, leave the reservoir empty. (See figure 43.)
4. Maximum capacity is approximately 5 cc.
5. After each use, remove the umbilicus, using caution to not
disconnect the pulse system tubing, and empty the reservoir with
the syringe.
6. Wipe out remaining fluid with a paper towel.
7. Leave open to air dry.
Figure 41
F. Performing IV Infusions
1. Procure a 3rd IV bag (sold separately, LF01130U).
2. Begin with a fully pressurized system.
3. Insert the IV needle or butterfly into a vein. Flashback will indicate a proper insertion.
4. Attach the needle or butterfly to the tubing from the infusion solution or fluid supply Bag C using the small latex piece.
(See figure 39.)
5. Open the clamp on fluid supply Bag B.
6. Open the clamp from the infusion solution.
7. Proof of proper procedure will be evidenced by the flow of fluid from the infusion solution or fluid supply Bag C.
Note: Simulate cleansing IV, injection, and withdrawal sites using distilled water. Multiple or prolonged exposure to
harsh antiseptics such as alcohol or iodine could damage or stain the simulator. Performing infusions can dilute your
simulated blood solution; ensure Bag B is mostly empty before infusing.
8. Always drain the blood and run clear water through the system at the end of each use.
9. Allow all components to air dry completely after each use and before re-assembly or storage.
G. Chest Tube
A chest tube may be inserted into the site in the left mid-axial line. This is a nonfunctioning site that may be used for prac-
tice in chest tube care. (See figure 40.)
1. Apply a small amount of the provided lubricant to chest tube appliances.
2. Use an alternative method to secure. Most adhesives do not stick well to the infant simulator’s skin.
3. Remove all devices from the infant simulator after each use.
4. Ensure the infant simulator and chest tube site are clean and dry after each use.
H. Umbilicus
The umbilicus may be catheterized through the vein or the artery. Always lubricate catheters before insertion.
1. To take blood samples, pull out the umbilicus, using caution to not disconnect the pulse system tubing.
2. Fill the reservoir with approximately 2-4 cc of blood mixture, and replace umbilicus. (See figures 41 and 42.)
3. To infuse fluids, leave the reservoir empty. (See figure 43.)
Figure 43
Figure 42
Figure 40
15
Figure 39
HC190001INM26.indd 15 2/19/19 1:55 PM
Figure 41
F. Performing IV Infusions
1. Procure a 3rd IV bag (sold separately, LF01130U).
2. Begin with a fully pressurized system.
3. Insert the IV needle or butterfly into a vein. Flashback will indicate a proper insertion.
4. Attach the needle or butterfly to the tubing from the infusion solution or fluid supply Bag C using the small latex piece.
(See figure 39.)
5. Open the clamp on fluid supply Bag B.
6. Open the clamp from the infusion solution.
7. Proof of proper procedure will be evidenced by the flow of fluid from the infusion solution or fluid supply Bag C.
Note: Simulate cleansing IV, injection, and withdrawal sites using distilled water. Multiple or prolonged exposure to
harsh antiseptics such as alcohol or iodine could damage or stain the simulator. Performing infusions can dilute your
simulated blood solution; ensure Bag B is mostly empty before infusing.
8. Always drain the blood and run clear water through the system at the end of each use.
9. Allow all components to air dry completely after each use and before re-assembly or storage.
G. Chest Tube
A chest tube may be inserted into the site in the left mid-axial line. This is a nonfunctioning site that may be used for prac-
tice in chest tube care. (See figure 40.)
1. Apply a small amount of the provided lubricant to chest tube appliances.
2. Use an alternative method to secure. Most adhesives do not stick well to the infant simulator’s skin.
3. Remove all devices from the infant simulator after each use.
4. Ensure the infant simulator and chest tube site are clean and dry after each use.
H. Umbilicus
The umbilicus may be catheterized through the vein or the artery. Always lubricate catheters before insertion.
1. To take blood samples, pull out the umbilicus, using caution to not disconnect the pulse system tubing.
2. Fill the reservoir with approximately 2-4 cc of blood mixture, and replace umbilicus. (See figures 41 and 42.)
3. To infuse fluids, leave the reservoir empty. (See figure 43.)
Figure 43
Figure 42
Figure 40
15
Figure 39
HC190001INM26.indd 15 2/19/19 1:55 PM