Blue light shows TTFields
are being delivered
Is there hair stubble beginning to grow?
Even if there is a small amount of hair growth, a new set of arrays
should be applied after shaving. See Guide to Scalp Care for more tips.
Are the arrays too warm?
Direct sunlight should be avoided. A breathable head covering could
be helpful, as well as a fan. If sweat causes arrays to shift, they should
be changed.
B ATTERIES
Is the yellow battery light on?
This is a great time to swap batteries for a full charge. When the battery
is low, you will hear an alarm.
Have you checked the battery gauge?
If the gauge shows only 1 dot, the battery should be changed.
C ONNECTION
Are any transducer array connectors loose?
All array cables need to be fully inserted into the connection box.
Is the connection cable plugged into the device?
The gray connection cable should be gently inserted into the device
with the arrows lined up. You should hear a click.
D EVICE
Are you hearing an alarm right after turning Optune on but BEFORE
pressing the TTFields button?
Device should be turned o, and the battery changed. Then the device can
be turned on.
Do you hear a rattling noise coming from the device, or has it been
damaged?
You can call nCompass® support for troubleshooting assistance.
ARRAYS
Are the arrays in good contact with your skin?
If not, using tape or Surgilast® can ensure good contact with the scalp.
When was the last time you changed your arrays?
Arrays should be changed at least 2 times per week (every 4 days at most).
Optune is designed to alert you if treatment is not being delivered.
The following items can cause your device to alarm.
What causes Optune®alarms?
If you hear an alarm, an issue
may be happening with ABCD
Contact nCompass®24/7
for any questions or technical support*
Call us toll free, any time of day:
1-855-281-9301 (toll free)
Or email:
• If you turn on Optune without pressing
the TTFields button, an alarm will
sound after 10 minutes and the blue
TTFields light will flash
• Press the TTFields button 1 time to
silence the alarm. Press the TTFields
button a second time to start therapy
* nCompass cannot provide you with medical advice.
Consult with your doctor for medical-related questions.
©2022 Novocure GmbH. All rights reserved. nCompass, Novocure, and Optune are registered trademarks of Novocure GmbH.
US-OPT-02683v2.0 January 2022
Did you know Optune®has a timer?
Optune®
Troubleshooting
Tips
Craig is an
Optune user
and Patient
Ambassador.
Patient images reflect the health status of the
patients at the time each photo was taken.
GUIDE TO UNDERSTANDING ALARMS
S:12.5"
S:16.5"
T:13"
T:17"
B:13.25"
B:17.25"
F:8.5"F:8.5"
F:6.5" F:6.5"

B
A
C
D
Blue light shows TTFields
are being delivered
Is there hair stubble beginning to grow?
Even if there is a small amount of hair growth, a new set of arrays
should be applied after shaving. See Guide to Scalp Care for more tips.
Are the arrays too warm?
Direct sunlight should be avoided. A breathable head covering could
be helpful, as well as a fan. If sweat causes arrays to shift, they should
be changed.
B ATTERIES
Is the yellow battery light on?
This is a great time to swap batteries for a full charge. When the battery
is low, you will hear an alarm.
Have you checked the battery gauge?
If the gauge shows only 1 dot, the battery should be changed.
C ONNECTION
Are any transducer array connectors loose?
All array cables need to be fully inserted into the connection box.
Is the connection cable plugged into the device?
The gray connection cable should be gently inserted into the device
with the arrows lined up. You should hear a click.
D EVICE
Are you hearing an alarm right after turning Optune on but BEFORE
pressing the TTFields button?
Device should be turned o, and the battery changed. Then the device can
be turned on.
Do you hear a rattling noise coming from the device, or has it been
damaged?
You can call nCompass® support for troubleshooting assistance.
ARRAYS
Are the arrays in good contact with your skin?
If not, using tape or Surgilast® can ensure good contact with the scalp.
When was the last time you changed your arrays?
Arrays should be changed at least 2 times per week (every 4 days at most).
Optune is designed to alert you if treatment is not being delivered.
The following items can cause your device to alarm.
What causes Optune®alarms?
If you hear an alarm, an issue
may be happening with ABCD
Contact nCompass®24/7
for any questions or technical support*
Call us toll free, any time of day:
1-855-281-9301 (toll free)
Or email:
• If you turn on Optune without pressing
the TTFields button, an alarm will
sound after 10 minutes and the blue
TTFields light will flash
• Press the TTFields button 1 time to
silence the alarm. Press the TTFields
button a second time to start therapy
* nCompass cannot provide you with medical advice.
Consult with your doctor for medical-related questions.
©2022 Novocure GmbH. All rights reserved. nCompass, Novocure, and Optune are registered trademarks of Novocure GmbH.
US-OPT-02683v2.0 January 2022
Did you know Optune®has a timer?
Optune®
Troubleshooting
Tips
Craig is an
Optune user
and Patient
Ambassador.
Patient images reflect the health status of the
patients at the time each photo was taken.
GUIDE TO UNDERSTANDING ALARMS
S:12.5"
S:16.5"
T:13"
T:17"
B:13.25"
B:17.25"
F:8.5"F:8.5"
F:6.5" F:6.5"

B
A
C
D
Blue light shows TTFields
are being delivered
Is there hair stubble beginning to grow?
Even if there is a small amount of hair growth, a new set of arrays
should be applied after shaving. See Guide to Scalp Care for more tips.
Are the arrays too warm?
Direct sunlight should be avoided. A breathable head covering could
be helpful, as well as a fan. If sweat causes arrays to shift, they should
be changed.
B ATTERIES
Is the yellow battery light on?
This is a great time to swap batteries for a full charge. When the battery
is low, you will hear an alarm.
Have you checked the battery gauge?
If the gauge shows only 1 dot, the battery should be changed.
C ONNECTION
Are any transducer array connectors loose?
All array cables need to be fully inserted into the connection box.
Is the connection cable plugged into the device?
The gray connection cable should be gently inserted into the device
with the arrows lined up. You should hear a click.
D EVICE
Are you hearing an alarm right after turning Optune on but BEFORE
pressing the TTFields button?
Device should be turned o, and the battery changed. Then the device can
be turned on.
Do you hear a rattling noise coming from the device, or has it been
damaged?
You can call nCompass® support for troubleshooting assistance.
ARRAYS
Are the arrays in good contact with your skin?
If not, using tape or Surgilast® can ensure good contact with the scalp.
When was the last time you changed your arrays?
Arrays should be changed at least 2 times per week (every 4 days at most).
Optune is designed to alert you if treatment is not being delivered.
The following items can cause your device to alarm.
What causes Optune®alarms?
If you hear an alarm, an issue
may be happening with ABCD
Contact nCompass®24/7
for any questions or technical support*
Call us toll free, any time of day:
1-855-281-9301 (toll free)
Or email:
• If you turn on Optune without pressing
the TTFields button, an alarm will
sound after 10 minutes and the blue
TTFields light will flash
• Press the TTFields button 1 time to
silence the alarm. Press the TTFields
button a second time to start therapy
* nCompass cannot provide you with medical advice.
Consult with your doctor for medical-related questions.
©2022 Novocure GmbH. All rights reserved. nCompass, Novocure, and Optune are registered trademarks of Novocure GmbH.
US-OPT-02683v2.0 January 2022
Did you know Optune®has a timer?
Optune®
Troubleshooting
Tips
Craig is an
Optune user
and Patient
Ambassador.
Patient images reflect the health status of the
patients at the time each photo was taken.
GUIDE TO UNDERSTANDING ALARMS
S:12.5"
S:16.5"
T:13"
T:17"
B:13.25"
B:17.25"
F:8.5"F:8.5"
F:6.5" F:6.5"

TTFIELDS LIGHT: goes o or is flashing
ERROR LIGHT: blinks once every 3 beeps
SOUND: 3 high-pitched quick beeps in
less than 1 second
Change the battery and power
on the device again.
If alarm persists, call nCompass
for troubleshooting assistance.
Press TTFields button once to
silence alarm. Press TTFields
button a second time to
restart treatment.
What does the alarm sound and look like?
When did the alarm occur or what
lights do you see on the device?
Alarm occurs immediately after
powering on but before
pressing the TTFields button
Blue TTFields light flashing Battery light is yellow Battery light is green or o;
TTFields light o
If there are 1-2 lights on
the battery gauge, change to a
fully charged battery.
If there are 3-4 lights on the
battery gauge, call nCompass
for troubleshooting assistance.
Ensure array cables are
properly connected to the
connection cable.
BATTERY
CONNECTIONS
BATTERYTHERAPY TIMER CONNECTION
CONNECTIONS
SOUND: 3 moderately paced
beeps in 1 second
ERROR LIGHT: flashes on for 1
second; flashes o for 1 second
TTFIELDS LIGHT: goes o
Ensure good contact by pressing
discs gently to scalp or use tape to
secure the edges.
Sweat, moisture, or hair growth may
cause arrays to lift. Shave head and
change arrays as needed.
If outside, avoid direct sunlight. Wear a
breathable head covering and/or use an
umbrella to shade the array.
If sleeping or napping, roll onto a
dierent side to release trapped warmth.
A fan or cooling pillow may help.
What does the alarm sound and look like?
When does the alarm occur?
15+ seconds after the TTFields
button is pressed
After turning the device on and right
after pressing the TTFields button
CONNECTIONS
ARRAY CONTACT
SOUND: beeps in a repeating 3-2 pattern
ERROR LIGHT: flashes rapidly. Flashes
on for a quarter-second; flashes o for a
quarter-second
TTFIELDS LIGHT: goes o
Ensure good contact by pressing
discs gently to scalp or use tape to
secure the edges.
Sweat, moisture, or hair growth may
cause arrays to lift. Shave head and
change arrays as needed.
ARRAY TEMPERATURE TOO WARM
If outside, avoid direct sunlight. Wear a
breathable head covering and/or use an
umbrella to shade the array.
If sleeping or napping, roll onto a
dierent side to release trapped warmth.
A fan or cooling pillow may help.
What does the alarm sound and look like?
What to do if your
Optune®device alarms
This guide may help you if Optune alarms.
It does not replace 24/7 nCompass® support.
For troubleshooting assistance, complete a
MyLink™ upload and call nCompass.
1
2
3
S:12.5"
S:16.5"
T:13"
T:17"
B:13.25"
B:17.25"
F:8.5"F:8.5"
F:6.5" F:6.5"
Other manuals for Optune
1
Other novocure Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Merivaara
Merivaara OPTIMA MINOR Use and maintenance instructions

Dentsply Sirona
Dentsply Sirona Cerec Guide 2 operating instructions

Novacor
Novacor R.TEST Evolution 4 user manual

COOK Medical
COOK Medical EchoTip Ultra manual

Seers Medical
Seers Medical SM8583D operating instructions

Comdek
Comdek MD-680P owner's manual