ScoutTM
SJX Balances Quick Start Guide
ScoutTM SJX Balances iv Corporation www.ohaus.com
CALIBRATION
When the balance is operated for the first time, a calibration is recommended to
ensure accurate weighing results.
External Span Calibration
With balance turned on:
Press and hold Menu until [mMeNU] (Menu) is displayed.
Release the button, the display will show [C.A.L].
Press Yes to accept. [SpaN] will then be shown.
Press Yes to begin the span calibration. [--C--] will blink and the
Zero reading is stored automatically.
The specified mass blinks on the display.
Place the specified span calibration mass on the pan or press
No to select an alternate weight.
[--C--] will blink while the reading is stored.
The display will show [done] if the calibration was successful. The
balance returns to the previous application mode and is ready for
use.
Internal Calibration (not available for SJX/E models)
With balance turned on:
Press and hold Menu until [mMeNU] (Menu) is displayed.
Release the button, the display will show [C.A.L].
Press Yes to accept. [In.CAL] will then be shown.
Press Yes to begin the Internal calibration. Zero reading is
stored automatically.
Pull the InCal handle while the
display show [PULL].
Push the InCal handle while the
display show [PUSH].
Repeat the step-5, and step-6 one
more time.
The display will show [done] if the calibration was successful.
The balance returns to the previous application mode and is
ready for use.