manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Omega
  6. •
  7. Transmitter
  8. •
  9. Omega TX787 User manual

Omega TX787 User manual

ConfigurConfigur
ConfigurConfigur
Configuraa
aa
ationtion
tiontion
tion
Configuration may be performed with a transmitter on-line or off-line.
When a transmitter is on-line, the upload and download icons will be
distinctly black. If no transmitter is on-line, those icons will be grey.
1.Upload Current configuration from the transmitter
2.Configure
•If the current “Sensor Type” does not match the application, click
on the “Sensor Selection” box then select the appropriate “tab” at the
top of the “Input Sensor Selection” screen.
•Make one selection from the “Input Connection” and “Type” boxes
(if offered) for the selected sensor type. Leave the “Mode” as “Stan-
dard” or the “Transfer Function” as “Linear” for now. Click the “OK”
buttonto returnto theConfiguration Screen.
•Select the Engineering Units (from the pull down menu), enter the
“Zero Scale” and “Full Scale” values and the “Burnout” mode
(transmitter to go above full scale or below zero scale upon input
failure).
•Click the "Set to Optimal" button to optimize the
transmitter'sfiltering.
3.Download New configuration to the transmitter.
CalibrCalibr
CalibrCalibr
Calibraa
aa
ationtion
tiontion
tion
To calibrate the transmitter, connect a 24VDC power supply, a 250Ωload
and multi-meter (in milliamp mode) in series to the transmitter's output.
1.Upload
2.Click on the Calibration Icon
3.Output Calibration: Select "Output 4mA", enter the mA reading from
the meter into the dialog box and click "Send to Device." Repeat for 20mA
calibration.
4A. Input Calibration Models XXXXX-X000 (e.g. TX787-0000):
Provide a precision 100.00mV source to the transmitter, select "Input mV"
on the Device Calibration screen, wait for the on-screen display to settle,
then click "Send to Device." Repeat the procedure, substituting a 300.00Ω
input (4-wire connection) and selecting "Input Ohms."
4B. Input Calibration Models XXXXX-X001(e.g. TX787-0001):
Providea precision 300.00mV source to the transmitter, select "Input
mV" on the Device Calibration screen, wait for the on-screen display to
settle, then click "Send to Device." Repeat procedure, substituting a 15.00
Volt input and selecting "Input Volts", and substituting a 50mA input,
selecting "Input mA." Note that the input terminals are different for each of
the signal levels.
5.To calibrate the cold junction compensation reference, select “Tem-
perature,”enter the nominal ambient temperature in the dialog box, set
the input simulation type to match the input simulator (i.e., type J thermo-
couple), verify that the simulator is set to the proper value, then click “Send
to Device.”
6. Click “Close.” A message in the lower left corner of the screen will
indicate the progress of the download and finally that the download has
beensuccessfully completed.
1. Install the Configuration and Calibration software
Run“setup.exe” on the distributiondisketteand follow the on-screeninstructions.
Note:thesoftware contained on the distribution diskette runs under Windows 95 and Windows NT.
Ifyou are running Windowsv3.1,contact Omega Engineering.
2. Connect the Communications Adapter to the computer and transmitter.
3. Identify the serial port to which the Adapter is connected.
Select“Options”fromthemenubar,then click “Communications” and select the correct COMM port from the pull-
down menu. Note that if you change the serial port you must exit the software and restart it with the correct setting.
4. Configure the transmitter.
What you will need:
Calibration
•PC
• Communications Adapter
• Transmitter
• Multimeter
• 24VDC power supply
Configuration
•PC
• Communications Adapter
• Transmitter
IntrIntr
IntrIntr
Introductionoduction
oductionoduction
oduction
Models TX787 and TX788 are microprocessor-based 2-wire transmitters that feature high accuracy and
high long-term stability. Setting the new standard in ease of setup, these transmitters feature PC-Only™
configuration technology. The unique Communications Adapter provides a fully isolated serial interface
and power source, allowing the transmitter to be configured with a “PC-Only” - no external power
supply, calibrator or meter is required!
QuicQuic
QuicQuic
Quick Stark Star
k Stark Star
k Startt
tt
t
Before beginning the process of setting up the transmitter, it is important to understand the difference between
calibration andconfiguration. The transmitter’s precision references (4 and 20 milliamps out; resistance,
voltage and current as appropriate; and a thermal reference) are calibrated at the factory and may be adjusted
later if necessary. Configuration is the process of defining the sensor type and range in engineering units.
Configurationis performed by connecting the transmitter to a PC and running the configuration software.
Becausethe Communications Adapter provides power to the transmitter’smicroprocessor, itis not necessary to
provide any external power to the transmitter in order to configure it.
PC-Only Programmable Transmitter
AdAd
AdAd
Additional Configurditional Configur
ditional Configurditional Configur
ditional Configuraa
aa
ation Options and Ftion Options and F
tion Options and Ftion Options and F
tion Options and Feaea
eaea
eaturtur
turtur
tureses
eses
es
The TX787 and T788 is capable of performing many functions in addition to the simple example outlined in the Quick Start exercise.
Dynamic Input FilteringDynamic Input Filtering
Dynamic Input FilteringDynamic Input Filtering
Dynamic Input Filtering
Filtering is provided to optimize the stability and response time of the transmitter output signal. The filter is set by specifying a Filter Band and a Damping
Time Constant. To simplify the definition of the filter values, click on the “Set to Optimal” button on the configuration screen.
Filter BandFilter Band
Filter BandFilter Band
Filter Band
The Filter Band allows the transmitter to be configured to react quickly to significant changes in input while smoothing out small changes. This band
defines the range above and below the current input reading within which the Damping Time Constant will be applied. It is defined as a percent of
the configured span of the transmitter. For example, a 0.5% Filter Band on a transmitter configured for an input range of 200 to 700 deg F would mean
that the Damping Time Constant will be applied to input changes of less than 2.5 deg but that changes of more than 2.5 deg would be immediately
reflected in the output. In general, the larger the full scale range, the smaller the Filter Band should be. Conversely, a larger Filter Band is generally
appropriate for smaller input ranges. Clicking the “Set to Optimal” button triggers the calculation of an appropriate setting based on the currently defined
inputrange.
DampingDamping
DampingDamping
Damping
Damping defines the length of time, in seconds, that the input will be averaged over to determine the output value when the input change is less than
the Filter Band. Changes in the input that are greater than the Filter Band bypass the Damping filter and are immediately reflected in the output. The
default value of 15 seconds, when used with the optimal Filter Band yields a very stable output while maintaining a fast response to significant changes in
theinput.
Input/OutputInput/Output
Input/OutputInput/Output
Input/Output TT
TT
Trimmingrimming
rimmingrimming
rimming
Because of slight variations between sensing elements or other loop characteristics, it is sometimes beneficial to “trim” the transmitter so that its output
corrects for these effects. If the measurement error is known, the transmitter may be trimmed “off-line.” “ON-line” trimming allows the characteristics of an
individual sensor to measure and match to the transmitter to which it will be connected in the field. Trimming allows the basiccalibration of the transmitter to
be left intact.
Dynamic Sensor SimDynamic Sensor Sim
Dynamic Sensor SimDynamic Sensor Sim
Dynamic Sensor Simulaula
ulaula
ulationtion
tiontion
tion
This option allows a loop to be tested dynamically verifying control logic as well as wiring. The transmitter’s current loop output can be controlled via four
parameters: simulate an input value in raw units (ohm, mV, etc.); simulate an input value in engineering units (e.g. 450 DEG); set an ouput in percent; or set
an ouput in mA. After an initial simulation is selected, the ouput may be incremented or decremented in 1% or 10% steps.
OtherOther
OtherOther
Other TT
TT
Trr
rr
ransfansf
ansfansf
ansfer Functionser Functions
er Functionser Functions
er Functions
The transfer function is the “formula” that the transmitter uses to determine the appropriate output signal for a given input signal. A “Linear” function means
that output is changed in a straight line fashion as the input changes within the configured input zero and full scale range: 0% of range input yields 4 mA
output, 25% of range input yields 8 mA, etc. A “Standard” function means that the transmitter will apply a standard linearization curve for temperature inputs.
RR
RR
RTD Sensor MaTD Sensor Ma
TD Sensor MaTD Sensor Ma
TD Sensor Matctc
tctc
tchinghing
hinghing
hing
The accuracy of RTD temperature sensing can be increased significantly when the temperature transmitter is “matched” to the individual sensor using a
method called Callender Van Dusen linearization. Doing so requires entering the sensor’s Alpha, Beta (measured for temperatures above zero and for
temperatures below zero), Sigma, and Ro values into the transmitters' configuration. These values can be requested from the manufacturer when the
sensor is purchased or may be measured by a testing lab.
SquarSquar
SquarSquar
Square Re R
e Re R
e Rootoot
ootoot
oot
Select this function if the transmitter’s output is to vary with the square root of the input.
Custom PCustom P
Custom PCustom P
Custom Polol
olol
olynomialynomial
ynomialynomial
ynomial
The transmitter can use up to a 10th order polynomial equation (y = A0+ A1x + A2x2+ … + A10x10) as the transfer function by inputting the coefficients
intothe transmitter’sconfiguration.
CustomCustom
CustomCustom
Custom TT
TT
Taa
aa
abb
bb
blele
lele
le
The transmitter can use a User Defined table by specifying a .tbl file. The data in the file should be two columns, tab delimited. Up to 140 data pairs may
be listed. The input values in column one must be equally spaced.
DefiningDefining
DefiningDefining
Defining TT
TT
Thehe
hehe
he TT
TT
Trr
rr
ransmitteransmitter
ansmitteransmitter
ansmitters User Fieldss User Fields
s User Fieldss User Fields
s User Fields
Each unit may be assigned an I.D. Tag name of up to eight characters, Job Number up to eight characters, and a “Message” of up to sixteen characters;
stored in the transmitter’s memory. Click on the desired field, enter the text, and click the download icon.
Postbus 8034, 1180 LA Amstelveen, The Netherlands
Tel: (31) 206418405
Toll Free in Benelux: 06 0993344
e-mail: [email protected]
Ostravska 767, 733 01 Karvina
Tel: 420 (69)6311627
e-mail: [email protected]
9, rue Denis Papin, 78190 Trappes
Tel: (33) 0130-621-400
Toll Free in France: 0800-4-06342
e-mail: [email protected]
Daimlerstrasse 26, D-75392
Deckenpfronn, Germany
Tel: 49 (07056)3017
Toll Free in Germany: 0130 11 21 66
e-mail: [email protected]
25 Swannington Road,
Broughton Astley, Leicestershire,
LE9 6TU, England
Tel: 44 (1455)285520
FAX: 44 (1455) 283912
SalesService: 1-800-826-6324 /1-800-TC-OMEGASM
CustomerService: 1-800-622-2378 /1-800-622-BESTSM
EngineeringServices: 1-800-872-9436 /1-800-USA-WHENSM
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA
Tel: (95) 800-TC-OMEGASM
En Espanol: (203) 359-1660 ext: 2203
Internete-mail
[email protected]
OMEGAnetSM On-Line Service
http://www.omega.com
One Omega Drive, Box 4047
Stamford, CT 06907-0047
Telephone: (203) 359-1660
e-mail: [email protected]
976Bergar
Laval (Quebec) H7L 5A1
Telephone: (514) 856-6928
e-mail: [email protected]
USA:
ISO 9001 Certified
Canada:
USA and Canada:
Mexico and
Latin America:
Benelux:
Czech Republic:
France:
Germany/Austria:
United Kingdom:
IS0 9002 Cerfified
FAX: (203) 359-7700
FAX: (514) 856-6886
FAX: (95) 203-359-7807
e-mail: [email protected]
FAX: (31) 20 6434643
FAX:420 (69) 6311114
FAX: (33) 130-699-120
FAX:49 (07056) 8540
P.O. Box 7, Omega Drive,
Irlam, Manchester,
M44 5EX, England
Tel: 44 (161) 777-6611
FAX: 44 (161) 777-6622
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply.OMEGA is constantly
pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every
appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts no liability
for any errors it contains, and reserves the right to alter specifications without notice.
WARNING:Theseproduct are not designed foruse in, and should not be used for,patient connected applications.
SaSa
SaSa
Savingving
vingving
ving TT
TT
The Configurhe Configur
he Configurhe Configur
he Configuraa
aa
ation Inftion Inf
tion Inftion Inf
tion Inforor
oror
ormama
mama
mationtion
tiontion
tion TT
TT
Too
oo
o A FileA File
A FileA File
A File
The transmitter’s configuration information may be saved to a file by clicking the diskette icon on the tool bar and specifying a file name and location.
Printing the ConfigurPrinting the Configur
Printing the ConfigurPrinting the Configur
Printing the Configuraa
aa
ation Inftion Inf
tion Inftion Inf
tion Inforor
oror
ormama
mama
mationtion
tiontion
tion
Clickingthe printer icon onthe tool bar willinitiate the printing ofa preconfigured summary ofthe transmitter’sconfigurationand calibration information.
The report includes the current time and date, the date of the last calibration, and date of the last configuration. Printing this report before and after
calibrating a unit provides "as found" and "as left" documentation. If a printer has been defined (in Windows) that prints to a file, this comprehensive report
can be saved on disk for future reference.
WW
WW
Wherher
herher
here Do I Find Eve Do I Find Ev
e Do I Find Eve Do I Find Ev
e Do I Find Everer
erer
erything I Need fything I Need f
ything I Need fything I Need f
ything I Need foror
oror
or
PrPr
PrPr
Process Measurocess Measur
ocess Measurocess Measur
ocess Measurement and Contrement and Contr
ement and Contrement and Contr
ement and Control?ol?
ol?ol?
ol?
OmeOme
OmeOme
Omegg
gg
ga...Ofa...Of
a...Ofa...Of
a...Of CourCour
CourCour
Course!se!
se!se!
se!
TEMPERATURE
þThermocouple, RTD & Thermistor
Probes, Connectors, Panels & Assemblies
þWire: Thermocouple, RTD & Thermistor
þCalibrators & Ice Point References
þRecorders, Controllers & Process Monitors
þInfraredPyrometers
PRESSURE/STRAIN FORCE
þTranducers & Strain Gauges
þLoad Cells & Pressure Gauges
þDisplacement Transducers
þInstrumentation& Accessories
FLOW/LEVEL
þRotameters, Gas Mass Flowmeters
& Flow Computers
þAirVelocityIndicators
þTurbine/PaddlewheelSystems
þTotalizers & Batch Controllers
pH/CONDUCTIVITY
þpH Electrodes, Testers & Accessories
þBenchtop/Laboratory Meters
þControllers,Calibrators, Simulators &Pumps
þIndustrial pH & Conductivity Equipment
DATA ACQUISITION
þDataAcquisition &
Engineering Software
þCommunications-Based
AcquisitionSystems
þPlug-in Cards for Apple, IBM
&Compatibles
þDataloggingSystems
þRecorders, Printers & Plotters
HEATERS
þHeating Cable
þCartridge & Strip Heaters
þImmersion & Band Heaters
þFlexibleHeaters
þLaboratory Heaters
ENVIROMENTAL
MONITORING AND CONTROL
þMetering& Control Instrumentation
þRefractometers
þPumps & Tubing
þAir, Soil & Water Monitors
þIndustrialWater& Wastewater Treatment
þpH,Conductivity & Dissolved
OxygenInstruments
WARRANTY/DISCLAIMER
RETURN REQUEST/ INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY
PRODUCT(S)TOOMEGA, PURCHASER MUSTOBTAINANAUTHORIZED RETURN (AR)NUMBER FROMOMEGA'S CUSTOMER
SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on
the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following
information available BEFORE contacting OMEGA:
1. P.O. number under which the product was
PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems relative
to the product
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for the a period of
37 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal
three (3) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive
maximum coverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA's Customer Service Depart-
ment will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by
OMEGA, if the unit is found to be defective it will be repaired or replaced at no charge. OMEGA's WARRANTY does
not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper
interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is
VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of
excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other
operating conditions outside of OMEGA's control. Components which wear are not warranted, including but not limited
to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes
responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its
products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only
that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRAN-
TIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of pur-
chaser set forth herein are exclusive and the total liability of OMEGA with respect to this order, whether based
on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase
price of the component upon which liability is based. In no event shall OMEGA be liable for consequential,
incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a "Basic Compo-
nent" under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used
on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on
humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/DIS-
CLAIMER language, and additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability
or damage whatsoever arising out of the use of the Product(s) in such a manner.
FOR NON-WARRANTY REPAIRS, consult OMEGA for
current repair charges. Have the following information
available BEFORE contacting OMEGA:
1. P.O. number to cover the COST of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems relative to the
product.
OMEGA's policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers
the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
©Copyright1998OMEGA ENGINEERING, INC. All rights reserved. This documentationmaynot be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without
prior written consent of OMEGA ENGINEERING, INC.
M2191
721-0765-00A 8/99

This manual suits for next models

1

Other Omega Transmitter manuals

Omega MT12TX Series User manual

Omega

Omega MT12TX Series User manual

Omega CDTX-90 Series User manual

Omega

Omega CDTX-90 Series User manual

Omega XW-ED User manual

Omega

Omega XW-ED User manual

Omega UWTC Series User manual

Omega

Omega UWTC Series User manual

Omega TX12 User manual

Omega

Omega TX12 User manual

Omega ZW Series User manual

Omega

Omega ZW Series User manual

Omega TX93 User manual

Omega

Omega TX93 User manual

Omega TX93A User manual

Omega

Omega TX93A User manual

Omega OM-CP-RF Series User manual

Omega

Omega OM-CP-RF Series User manual

Omega OS137 SERIES User manual

Omega

Omega OS137 SERIES User manual

Omega DRST-UR User manual

Omega

Omega DRST-UR User manual

Omega LVU800 Series User manual

Omega

Omega LVU800 Series User manual

Omega OS4000 Series User manual

Omega

Omega OS4000 Series User manual

Omega FMG-1000-MAW User manual

Omega

Omega FMG-1000-MAW User manual

Omega UWPC-2-NEMA User manual

Omega

Omega UWPC-2-NEMA User manual

Omega EWS-BP-A User manual

Omega

Omega EWS-BP-A User manual

Omega TX31 User manual

Omega

Omega TX31 User manual

Omega TX90A Series User manual

Omega

Omega TX90A Series User manual

Omega UWBT Series User manual

Omega

Omega UWBT Series User manual

Omega TX94A User manual

Omega

Omega TX94A User manual

Omega DRST-CM User manual

Omega

Omega DRST-CM User manual

Omega W Series User manual

Omega

Omega W Series User manual

Omega FP90 Series User manual

Omega

Omega FP90 Series User manual

Omega DMD4380 User manual

Omega

Omega DMD4380 User manual

Popular Transmitter manuals by other brands

WIKA OC-1 operating instructions

WIKA

WIKA OC-1 operating instructions

Sharp GP1A173LCS3F manual

Sharp

Sharp GP1A173LCS3F manual

GE VT6010 Installation & operation instructions

GE

GE VT6010 Installation & operation instructions

Emerson Micro Motion 3500 installation manual

Emerson

Emerson Micro Motion 3500 installation manual

PR electronics 5335 manual

PR electronics

PR electronics 5335 manual

Rinstrum 1203 Reference manual

Rinstrum

Rinstrum 1203 Reference manual

Kramer Cobra TS2 user manual

Kramer

Kramer Cobra TS2 user manual

RHEONIK RHE45 Installation & start?up guide

RHEONIK

RHEONIK RHE45 Installation & start?up guide

Vaisala WXT520 user guide

Vaisala

Vaisala WXT520 user guide

Black Widow Security BW FM 5 user manual

Black Widow Security

Black Widow Security BW FM 5 user manual

MuxLab 500060 installation guide

MuxLab

MuxLab 500060 installation guide

SEAANGEL SA15 AIS FLARE user manual

SEAANGEL

SEAANGEL SA15 AIS FLARE user manual

Endress+Hauser Liquiline CM442R operating instructions

Endress+Hauser

Endress+Hauser Liquiline CM442R operating instructions

Teufel Subconnect TX Technical description and instruction manual

Teufel

Teufel Subconnect TX Technical description and instruction manual

Akura ATW-GA2929 instruction manual

Akura

Akura ATW-GA2929 instruction manual

Elta ADT406 AF/AP User handbook

Elta

Elta ADT406 AF/AP User handbook

E+E Elektronik EE360 Operation manual

E+E Elektronik

E+E Elektronik EE360 Operation manual

RKI Instruments 65-2640RK-05-04 Operator's manual

RKI Instruments

RKI Instruments 65-2640RK-05-04 Operator's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.