Make sure that the air tube is connected to the compressor. Disconnect
the air tube from the nebulizer kit. Turn on the compressor and dry the
air tube by passing air through it.
Disinfecting the Nebulizer Kit, Mask, Mouthpiece and Adapter:
Always disinfect the parts before using the product for the first time and
after the product has not been used for an extended period of time as
well as after the last treatment of the day. If the parts are heavily stained,
replace them with new ones. The Nebulizer Kit, Mask, Mouthpiece,
Nosepiece and Adapter can be disinfected by
• using a chemical disinfectant such as ethanol, sodium
hypochlorite (Milton), Quaternary ammonium (Osvan),
Chlorhexidine (Hibitane) and Amphoteric Surfactant (Tego),
following product instructions and rinsing off with clean warm
water. Leave to air-dry. (Note: Never clean with benzene, thinner or
a flammable chemical. The (non-latex) elastic bands cannot be
disinfected. Remove them when disinfecting the masks.)
• boiling: The Nebulizer Kit, Mouthpiece and Adapter can also be
disinfected by boiling between 15 to 20 minutes in plenty of water.
After boiling, carefully remove the parts, shake off excess water and
allow to air-dry in a clean environment. (Note: Do not boil the Mask
and the Air Tube.)
Replacing the Air Filter:
If the air filter has changed colour, or has been used for more than 70
applications, replace it with a new one. To purchase the filters – see Other
Optional/Replacement Parts paragraph. Remove the air filter cover using
a flat screwdriver as shown in figure; remove the filter and fit the new
filter as shown. Put the air filter cover back in place. (Note: Do not wash
or clean the air filter. If the air filter Becomes wet, replace it. Damp air
filters can cause blockages.)
Maintenance
This device does not require any regular maintenance.