Onset InTemp CX450 User manual

InTemp® CX450 Temp/RH Logger Manual
25186-C
The InTemp CX450 logger measures temperature and relative humidity (RH) for storage and
transportation monitoring in the pharmaceutical, life science, and medical industries. This
Bluetooth® Low Energy-enabled logger is designed for wireless communication with a mobile
device. Using the InTemp app, you can easily configure the logger, download reports, and
monitor tripped alarms. Or, you can use InTempConnect® to configure and download the
logger via the CX5000 Gateway. InTempVerify™ app is also available to easily download loggers
and automatically upload reports to InTempConnect. Use the built-in LCD screen on the logger
to view the current temperature and humidity, logging status and battery usage, and also to
check and clear maximum and minimum readings. Once data is uploaded to InTempConnect,
you can track logger configurations and automatically upload logger data to build custom
reports for further analysis.
Specifications
Temperature Sensor
Range -30° to 70°C (-22° to 158°F)
Accuracy ±0.5°C from -30° to 70°C (±0.9°F from -22° to 158°F)
Resolution 0.024°C at 25°C (0.04°F at 77°F)
Drift <0.1°C (0.18°F) per year
RH Sensor
Range 0 to 95% (non-condensing)
Accuracy ±3% from 0% to 90% and ±3.5% from greater than 90% typical at 25°C
(77°F)
Resolution 0.01%
Drift <1% per year typical
Logger
Radio Power 1 mW (0 dBm)
Transmission Range Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard Bluetooth Low Energy (Bluetooth Smart)
Logger Operating Range -30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing)
NIST Calibration Two-point NIST calibration included
Logging Rate 1 second to 18 hours
Time Accuracy ±1 minute per month at 25°C (77°F)
Battery Type Two AAA 1.5 V alkaline or lithium batteries, user replaceable
Battery Life 1 year, typical with logging interval of 1 minute. Faster logging intervals,
remaining connected with the InTemp app, excessive report generation,
numerous audible alarms, and paging all impact battery life.
Memory 178 KB (103,400 total measurements or 51,700 temp & RH paired
measurements)
Full Memory Download
Time
Approximately 60 seconds; may take longer the farther the device is
from the logger
LCD LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or
go blank in temperatures outside this range
Dimensions 9.4 x 5.6 x 2.59 cm (3.7 x 2.21 x 1.02 inches)
Weight 90.2 g (3.18 oz)
Environmental Rating IP54
The CE Marking identifies this product as complying with all relevant
directives in the European Union (EU).
See last page
InTemp CX450 Temp/RH
Logger
Included Items:
•Two AAA 1.5 V alkaline
batteries
•Battery door and screw
•NIST Certificate of
Calibration
Required Items:
•InTemp app
•Device with iOS or Android™
and Bluetooth

InTemp CX450 Temperature/RH Logger Manual
1-508-759-9500 (U.S. and International) 2 www.onsetcomp.com
1-800-LOGGERS (U.S. only)
Logger Components and Operation
Logger Front View
Logger Back View
Start Button: Press this button for 3 seconds to start the logger
when it is configured to start “on button push.” You can also
press this button for 1 second to bring the logger to the top of
the list in the app. Press and hold both buttons on the logger
simultaneously for 3 seconds to clear the minimum and
maximum values on the LCD (see Minimum and Maximum
Values) or for 10 seconds to reset a passkey (see Passkey
Protection).
Next or Mute Button: Press this button for 1 second to switch
between the temperature and humidity readings on the LCD.
You can also press this button for 1 second to mute a beeping
alarm (see Logger Alarms). Press and hold both buttons on the
logger simultaneously for 3 seconds to clear the minimum and
maximum values (see Minimum and Maximum Values).
Sensor Housing: Both the temperature and RH sensors are
located within the housing that protrudes from the case.
Audible Alarm Speaker: This is the speaker for the audible
alarm that beeps when an alarm is tripped. See Logger Alarms.
Alarm LED: This LED blinks every 5 seconds when an alarm is
tripped. See Logger Alarms.
Magnets: Use the four magnets on the back for mounting the
logger.
Vent: This is a membrane-covered vent for proper sensor
operation. Do not remove the protective membrane.
LCD: This screen shows the latest sensor readings and other
status information. The LCD screen refreshes at the same rate
as the logging interval. The example shows all symbols
illuminated on the LCD screen followed by a table with
descriptions of each symbol.
LCD Symbol Description
An alarm has tripped because the sensor reading
is outside the specified range. See Logger Alarms.
This indicates how much memory has been used
for the current configuration. In this example,
approximately 40 percent of the memory has been
used.
This shows the approximate battery power
remaining.
The logger is currently logging.
The logger is currently connected to a phone or
tablet via Bluetooth. The more bars there are, the
stronger the signal.
The logger is waiting to be started. Press and hold
the Start button for 3 seconds to start the logger.
This is an example of a temperature reading.
This is an example of an RH reading.
These are examples of minimum temperature and
RH readings. Press and hold both buttons on the
logger simultaneously for 3 seconds to clear the
minimum and maximum values on the LCD. See
Minimum and Maximum Values for additional
details.
These are examples of the maximum temperature
and RH readings. Press and hold both buttons on
the logger simultaneously for 3 seconds to clear
the minimum and maximum values on the LCD.
See Minimum and Maximum Values for additional
details.
or
MUTE indicates an alarm is beeping. Turn off the
beeping alarm by pressing the Mute button. The
LCD then changes to MUTED.
The audible alarm has been muted.
The logger has been configured to start logging on
a delay. The display will count down in days,
hours, minutes, and seconds until logging begins.
In this example, 5 minutes and 38 seconds remain
until logging will begin.
The profile settings are being loaded onto the
logger.
An error occurred while loading the profile
settings onto the logger. Try reconfiguring the
logger.
Next or Mute
button
Vent
Magnets
LCD
Alarm LED
Start button
Audible
alarm
speaker
Magnets
Sensor housing

InTemp CX450 Temperature/RH Logger Manual
1-508-759-9500 (U.S. and International) 3 www.onsetcomp.com
1-800-LOGGERS (U.S. only)
LCD Symbol Description
The logger was paged from the InTemp app.
The logger has been downloaded and stopped
with the InTemp app or because the memory is
full.
This is displayed while pressing the top button for
3 seconds to start the logger or when pressing
both buttons for 3 seconds to clear the minimum
and maximum values on the LCD.
The logger is being updated with new firmware.
Note: If the logger has stopped logging because the memory is
full, the LCD screen will remain on with “STOP” displayed until
the logger is downloaded to your mobile device. Once the
logger has been downloaded, the LCD will turn off
automatically after 2 hours. The LCD will turn back on the next
time the logger connects with your device.
Getting Started
InTempConnect is web-based software where you can monitor
CX450 logger configurations and view downloaded data online.
Using the InTemp app, you can configure the logger with your
phone or tablet and then download reports, which are saved in
the app and automatically uploaded to InTempConnect. Or,
anyone can download a logger using the InTempVerify app if
the loggers are enabled to be used with InTempVerify. See
www.intempconnect.com/help for details on both the gateway
and InTempVerify. If you do not need to access logged data via
the cloud-based InTempConnect software, then you also have
the option to use the logger with the InTemp app only
Follow these steps to begin using the logger:
1. Set up an InTempConnect account. Follow all steps if you are
a new administrator. If you already have an account and
roles assigned, follow step c.
If you are using the logger with the InTemp app only, skip to
step 2.
a. Go to www.intempconnect.com and follow the prompts to
set up an administrator account. You’ll receive an email to
activate the account.
b. Log into www.intempconnect.com and add roles for the
users you will be adding to the account. Click Settings and
then Roles. Click Add Role, enter a description, select the
privileges for the role and click Save.
c. Click Settings and then Users to add users to your account.
Click Add User and enter the email address and first and
last name of the user. Select the roles for the user and
click Save. New users will receive an email to activate their
user accounts.
2. Set up the logger. Insert two AAA batteries in the logger,
observing polarity. Insert the battery door in the back of the
logger making sure it is flush with the rest of the logger
case. Use the included screw and a Phillips-head
screwdriver to screw the battery door into place.
3. Download the InTemp app and log in.
a. Download InTemp to a phone or tablet from the App
Store® or Google Play™.
b. Open the app and enable Bluetooth in the device settings
if prompted.
c. InTempConnect users: Log in with your InTempConnect
user credentials. (Your device must be able to connect to
the internet.) Make sure to check the box that says “I am
an InTempConnect user” when signing in.
InTemp app only users: If you will not be using
InTempConnect, create a user account and log in when
prompted. DO NOT check the box that says “I am an
InTempConnect user” when signing in.
4. Set up a logger profile.
InTempConnect users (requires privileges):
a. In InTempConnect, select Loggers > Logger Profiles.
b. Click Add Logger Profile.
c. Type a profile name.
d. If you plan to use the logger with the InTempVerify app,
then select InTempVerify enabled.
e. Select CX450 for the Logger Family.
f. Select a Logging Interval, Start option, and any alarms.
g. Tap Save to add the new profile.
Note: Set up any optional trip information fields before
continuing (see www.intempconnect.com/help for details).
InTemp app only users:
a. In the app, tap the Settings icon and tap CX450 Logger.
b. Tap the plus in the upper right corner.
c. Type a profile name.
d. Tap Logging Interval. Select a logging interval and tap
Save.
e. Tap Start and select the option for when you want
logging to begin.
f. Set up alarms if desired.
g. Tap Save to add the new profile.
5. Configure the logger.
a. Tap the Devices icon in the app. Find the logger in the list
and tap it to connect to it.
b. Once connected, tap Configure. Swipe left and right to
select a logger profile. Type a name or label for the
logger. Tap Start to load the selected profile to the
logger. InTempConnect users: If the profile does not yet
appear in the app, close and then reopen it, and repeat
steps a and b. In addition, if trip information fields were
set up, you will be prompted to enter additional
information. Tap Start in the upper right corner when
done.
Note: You can also use the CX5000 Gateway to
automatically configure loggers. See
www.intempconnect.com/help for details.
If you are having trouble connecting:
•Make sure the logger is within range of your mobile
device. The range for successful wireless
communication is approximately 30.5 m (100 ft) with
full line-of-sight.
•If your device can connect to the logger intermittently
or loses its connection, move closer to the logger,
within sight if possible.
•Change the orientation of your phone or tablet to
ensure the antenna in your device is pointed toward
the logger. Obstacles between the antenna in the
device and the logger may result in intermittent
connections.

InTemp CX450 Temperature/RH Logger Manual
1-508-759-9500 (U.S. and International) 4 www.onsetcomp.com
1-800-LOGGERS (U.S. only)
•If the logger appears in the list, but you cannot connect
to it, close the app, power down the mobile device,
and then turn it back on. This forces the previous
Bluetooth connection to close.
6. Deploy and start the logger. Place the logger in the location
where you will be monitoring the temperature and humidity.
Logging will begin based on the settings in the profile
selected.
Once logging begins, press the Next button on the logger to
switch between the latest temperature and humidity readings,
as well as the minimum and maximum values for both (see
Minimum and Maximum Values for more details). See
Downloading the Logger for details on how to access the logged
data.
Logger Alarms
You can set up an alarm to trip on the logger when a
temperature and/or humidity reading falls above or below a
specified value. Specific alarm settings are set in the logger
profile that you create in either InTempConnect or in the app.
When an alarm trips:
•The logger LED will blink every 5 seconds.
•The alarm icon will appear on the LCD and in the app.
•If audible alarms are enabled in the logger profile, the
logger will beep as follows every 15 seconds:
One quick beep for temperature alarms.
Two quick beeps for RH alarms.
One long beep if multiple alarms were tripped at the
same time.
•An Alarm Tripped event is logged.
To mute a beeping alarm, press the Mute button on the logger.
Once muted, you cannot turn the beeping back on.
Download the logger to view details about the tripped alarm
and to clear alarm indicators in the app and on the LCD.
Minimum and Maximum Values
The logger LCD displays the minimum and maximum
temperature and humidity readings for the entire logging
period. These values reset automatically when the logger is
downloaded and restarted or stopped and reconfigured.
You can also clear these values as needed while the logger is
logging by pressing both buttons on the logger simultaneously
for 3 seconds until HOLD disappears on the LCD. Dashes (--) will
then appear on the LCD for the minimum and maximum values
until the next logging interval. The values will then continue to
be updated for the remaining logging period or until they are
cleared again. Note: This clears the data on the screen only.
Actual logger and reporting data will not be cleared with this
reset.
Passkey Protection
The logger is protected by an encrypted passkey automatically
generated by the InTemp app for InTempConnect users and
optionally available if you are using the InTemp app only. The
passkey uses a proprietary encryption algorithm that changes
with every connection.
InTempConnect Users
Only InTempConnect users belonging to the same
InTempConnect account can connect to a logger once it is
configured. When an InTempConnect user first configures a
logger, it is locked with an encrypted passkey that is
automatically generated by the InTemp app. After the logger is
configured, only active users associated with that account will
be able to connect to it. If a user belongs to a different account,
that user will not be able to connect to the logger with the
InTemp app, which will display an invalid passkey message.
Administrators or users with the required privileges can also
view the passkey from the device configuration page in
InTempConnect and share them if needed. See
www.intempconnect.com/help for more details. Note: This
does not apply to InTempVerify. if the logger was configured
with a logger profile in which InTempVerify was enabled, then
anyone can download the logger with the InTempVerify app.
InTemp App Only Users
If you are using the InTemp app only (not logging in as an
InTempConnect user), you can create an encrypted passkey for
the logger that will be required if another phone or tablet
attempts to connect to it. This is recommended to ensure that
a deployed logger is not mistakenly stopped or purposely
altered by others.
To set a passkey:
1. Tap the Devices icon and connect to the logger.
2. Tap Set Logger Passkey.
3. Type a passkey up to 10 characters.
4. Tap Save.
5. Tap Disconnect.
Only the phone or tablet used to set the passkey can then
connect to the logger without entering a passkey; all other
mobile devices will be required to enter the passkey. For
example, if you set the passkey for the logger with your tablet
and then try to connect to the device later with your phone,
you will be required to enter the passkey on the phone but not
with your tablet. Similarly, if others attempt to connect to the
logger with different devices, then they would also be required
to enter the passkey. To reset a passkey, simultaneously press
both the top and bottom button on the logger for 10 seconds,
or connect to the logger in the app, tap Set Logger Passkey, and
select Reset to Factory Default.
Downloading the Logger
You can download the logger to a phone or tablet and generate
reports that include temperature and humidity readings,
events, user activity, alarm information, and more. Reports can
be shared immediately upon download or accessed later in the
InTemp app.
InTempConnect users: Privileges are required to download,
preview, and share reports in the InTemp app. Report data is
automatically uploaded to InTempConnect when you download
the logger. Log into InTempConnect to build custom reports
(requires privileges).
Note: InTempConnect users can also download CX loggers
automatically on a regular basis using the CX5000 Gateway. Or,
if the logger was configured with a logger profile in which
InTempVerify was enabled, then anyone can download the

InTemp CX450 Temperature/RH Logger Manual
1-508-759-9500 (U.S. and International) 5 www.onsetcomp.com
1-800-LOGGERS (U.S. only)
logger with the InTempVerify app. For details on the gateway
and InTempVerify, see www.intempconnect/help.
To download the logger with the InTemp app:
1. Tap the Devices icon and connect to the logger.
2. Tap Download.
3. Choose a download option:
•Download & Continue. The logger will continue logging
once the download is complete.
•Download & Restart. The logger will start a new data set
using the same profile once the download is complete.
•Download & Stop. The logger will stop logging once the
download is complete.
A report of the download is generated and is also uploaded to
InTempConnect if you are logged into the InTemp app with
your InTempConnect user credentials.
In the app, tap Settings to change the default report type and
report sharing options. The report is also available in Secure
PDF and XLSX formats for sharing at a later time. Tap the
Reports icon to access previously downloaded reports. See
www.intempconnect.com/help for details on working with
reports in both the InTemp app and InTempConnect.
Logger Events
The logger records the following events to track logger
operation and status. These events are listed in reports
downloaded from the logger.
Event Name Definition
Started The logger started logging.
Stopped The logger stopped logging.
Connected The logger connected to the InTemp app.
Downloaded The logger was downloaded.
Alarm
Tripped/Cleared
An alarm has tripped because the reading
was outside the alarm limits or has cleared.
Safe Shutdown The battery level dropped below 1.85 V; the
logger performs a safe shutdown.
Deploying and Protecting the Logger
Use the four magnets on the back of the logger case to mount it
to a magnetic surface.
The logger is designed for indoor use and can be permanently
damaged by corrosion if it gets wet. Protect it from direct water
contact. If the logger gets wet, remove the battery immediately
and dry the circuit board.
Note: Static electricity may cause the logger to stop logging.
The logger has been tested to 8 KV, but avoid electrostatic
discharge by grounding yourself to protect the logger. For more
information, search for “static discharge” on onsetcomp.com.
Battery Information
The logger requires two user-replaceable AAA 1.5 V alkaline or
optional lithium batteries for operation at the extreme ends of
the logger operating range. Expected battery life varies based
on the ambient temperature where the logger is deployed, the
frequency of connecting to the phone or tablet and
downloading reports, the duration of audible alarms, and
battery performance. New batteries typically last 1 year with
logging intervals 1 minute or greater. Deployments in extremely
cold or hot temperatures or a logging interval faster than 1
minute can impact battery life. Estimates are not guaranteed
due to uncertainties in initial battery conditions and operating
environment.
Note: Make sure the batteries installed have flat negative
terminals. There should be no indent in the bottom of the
batteries. Batteries with indents in the negative terminals may
become loose and prevent proper operation.
To install or replace the batteries:
1. If the battery door is already installed on the back of the
logger, use a Phillips-head screwdriver to remove it.
2. Remove any old batteries.
3. Insert two new batteries observing polarity.
4. Screw the battery door into place.
WARNING: Do not cut open, incinerate, heat above 85°C
(185°F), or recharge lithium batteries. The batteries may
explode if the logger is exposed to extreme heat or conditions
that could damage or destroy the battery case. Do not dispose
of the logger or batteries in fire. Do not expose the contents of
the batteries to water. Dispose of the batteries according to
local regulations for lithium batteries.
Only use batteries with
flat negative terminals
Do not use batteries with an
indent in the negative terminal

InTemp CX450 Temperature/RH Logger Manual
1-508-759-9500 (U.S. and International)
1-800-LOGGERS (564-4377) (U.S. only)
www.onsetcomp.com/support/contact
© 2020 Onset Computer Corporation. All rights reserved. Onset, InTemp, InTempConnect, and InTempVerify
are trademarks or registered trademarks of Onset Computer Corporation. App Store is a service mark of Apple
Inc. Google Play is a trademark of Google Inc. Bluetooth is a registered trademark of Bluetooth SIG, Inc.
Bluetooth and Bluetooth Smart are registered trademarks of Bluetooth SIG, Inc. All other trademarks are the
property of their respective companies.
Patent #: 8,860,569 25186-C
Federal Communication Commission Interference Statement
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user
is encouraged to try to correct the interference by one of the following measures:
•Reorient or relocate the receiving antenna
•Increase the separation between the equipment and receiver
•Connect the equipment into an outlet on a circuit different from that to which the receiver is connected
•Consult the dealer or an experienced radio/TV technician for help
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device
must accept any interference received, including interference that may cause undesired operation.
FCC Caution: Any changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate this equipment.
Industry Canada Statements
This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and
(2) this device must accept any interference, including interference that may cause undesired operation of the device.
Avis de conformité pour l’Industrie Canada
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes : (1)
l'appareil ne doit pas produire de brouillage, et (2) l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le
fonctionnement.
To comply with FCC and Industry Canada RF radiation exposure limits for general population, the logger must be installed to provide a separation distance of at least 20cm from
all persons and must not be co-located or operating in conjunction with any other antenna or transmitter.
Pour se conformer aux limites d'exposition aux rayonnements RF de la FCC et d'Industrie Canada pour la population générale, l'enregistreur doit être installé de manière à
fournir une distance de séparation d'au moins 20 cm de toutes les personnes et ne doit pas être colocalisé ou fonctionner avec une autre antenne ou un autre émetteur.
Table of contents
Other Onset Data Logger manuals

Onset
Onset EG4100 Series User manual

Onset
Onset HOBO U30 Series User manual

Onset
Onset MX1102A User manual

Onset
Onset HOBO U30 NRC User manual

Onset
Onset HOBO MX100 User manual

Onset
Onset HOBO ZW-008 User manual

Onset
Onset HOBO MX CO2 Logger User manual

Onset
Onset HOBO Micro Station User manual

Onset
Onset HOBO UX100-023 User manual

Onset
Onset HOBO UX90-002 User manual

Onset
Onset HOBO MX User manual

Onset
Onset InTemp CX40 Series User manual

Onset
Onset EG4 Series User manual

Onset
Onset HOBO User manual

Onset
Onset U12-015-03 User manual

Onset
Onset HOBO UX100-014M User manual

Onset
Onset MX2501 User manual

Onset
Onset HOBO UX120-017 User manual

Onset
Onset HOBO U24 User manual

Onset
Onset HOBO UX100-011A User manual