ORTEC ICS Stand II User manual

ORTEC® Part No. 1108265
Manual Revision B, ORT-X4548
ICS™ Stand II
User Manual


ICS Stand 2 1108265B
1
Advanced Measurement Technology, Inc.
a/k/a/ ORTEC®, a subsidiary of AMETEK®, Inc.
WARRANTY
ORTEC* warrants that the items will be delivered free from defects in material or workmanship. ORTEC makes no
other warranties, express or implied, and specifically NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR
A PARTICULAR PURPOSE.
ORTEC’s exclusive liability is limited to repairing or replacing at ORTEC’s option, items found by ORTEC to be
defective in workmanship or materials within one year from the date of delivery. ORTEC’s liability on any claim of
any kind, including negligence, loss, or damages arising out of, connected with, or from the performance or breach
thereof, or from the manufacture, sale, delivery, resale, repair, or use of any item or services covered by this
agreement or purchase order, shall in no case exceed the price allocable to the item or service furnished or any part
thereof that gives rise to the claim. In the event ORTEC fails to manufacture or deliver items called for in this
agreement or purchase order, ORTEC’s exclusive liability and buyer’s exclusive remedy shall be release of the buyer
from the obligation to pay the purchase price. In no event shall ORTEC be liable for special or consequential
damages.
Quality Control
Before being approved for shipment, each ORTEC instrument must pass a stringent set of quality control tests
designed to expose any flaws in materials or workmanship. Permanent records of these tests are maintained for use
in warranty repair and as a source of statistical information for design improvements.
Repair Service
If it becomes necessary to return this instrument for repair, it is essential that Customer Services be contacted in
advance of its return so that a Return Authorization Number can be assigned to the unit. Also, ORTEC must be
informed, either in writing, by telephone [(865) 482-4411] or by facsimile transmission [(865) 483-2133], of the nature
of the fault of the instrument being returned and of the model, serial, and revision (“Rev” on rear panel) numbers.
Failure to do so may cause unnecessary delays in getting the unit repaired. The ORTEC standard procedure requires
that instruments returned for repair pass the same quality control tests that are used for new-production instruments.
Instruments that are returned should be packed so that they will withstand normal transit handling and must be
shipped PREPAID via Air Parcel Post or United Parcel Service to the designated ORTEC repair center. The address
label and the package should include the Return Authorization Number assigned. Instruments being returned that
are damaged in transit due to inadequate packing will be repaired at the sender's expense, and it will be the sender's
responsibility to make claim with the shipper. Instruments not in warranty should follow the same procedure and
ORTEC will provide a quotation.
Damage in Transit
Shipments should be examined immediately upon receipt for evidence of external or concealed damage. The carrier
making delivery should be notified immediately of any such damage, since the carrier is normally liable for damage
in shipment. Packing materials, waybills, and other such documentation should be preserved in order to establish
claims. After such notification to the carrier, please notify ORTEC of the circumstances so that assistance can be
provided in making damage claims and in providing replacement equipment, if necessary.
Copyright © 2014, Advanced Measurement Technology, Inc. All rights reserved.
*ORTEC®is a registered trademark of Advanced Measurement Technology, Inc. All other trademarks used herein are the property
of their respective owners.
NOTICE OF PROPRIETARY PROPERTY — This document and the information contained in it are the proprietary property of
AMETEK Inc., ORTEC Business Unit. It may not be copied or used in any manner nor may any of the information in or upon it be
used for any purpose without the express written consent of an authorized agent of AMETEK Inc., ORTEC Business Unit.

1108265B ICS Stand 2
2
SAFETY INSTRUCTIONS AND SYMBOLS
This manual contains up to three levels of safety instructions that must be observed in order to avoid personal
injury and/or damage to equipment or other property. These are:
DANGER Indicates a hazard that could result in death or serious bodily harm if the safety instruction is not
observed.
WARNING Indicates a hazard that could result in bodily harm if the safety instruction is not observed.
CAUTION Indicates a hazard that could result in property damage if the safety instruction is not observed.
Please read all safety instructions carefully and make sure you understand them fully before attempting to use
this product.

ICS Stand 2 1108265B
3
TABLE OF CONTENTS
1. INTRODUCTION........................................................................................................................................................... 4
1.1. SPECIFICATIONS ............................................................................................................................................... 4
1.1.1. Performance ....................................................................................................................................... 4
1.1.2. Mechanical.......................................................................................................................................... 4
1.2. INSPECTION FOR SHIPPING DAMAGE .............................................................................................................. 4
1.3. CONTENTS OF THE SHIPPING CARTON............................................................................................................. 5
1.4. UPACKING THE ICS STAND ............................................................................................................................... 5
1.5. THE ICS STAND ................................................................................................................................................. 6
2. ICS INSTALLATION ONTO THE ICS STAND ................................................................................................................... 8
2.1. THE ICS AND ICS STAND ASSEMBLY ................................................................................................................. 8
3. INSTALLATION UNDER A TYPICAL 4” TOP LOADING LEAD SHIELD............................................................................ 12
3.1. Preparing the lead shield ............................................................................................................................... 12
3.2. Positioning the ICS and ICS Stand with Adequate Shield Clearance .............................................................. 12
3.3. Positioning the ICS and ICS Stand with Inadequate Shield Clearance............................................................ 15
4. LOCKING THE ICS STAND 2 IN POSITION................................................................................................................... 18

1108265B ICS Stand 2
4
ICS™ Stand Manual
1. INTRODUCTION
The ICS Stand is designed to operate only with the Integrated Cryocooling System (ICS). This stand is
designed to support the ICS in the vertical orientation and is useful for mounting the ICS under standard lead
shields.
For a standard 10cm (4 inch) thick lead shield, the ICS detector package is installed from the bottom of the
shield. This is different than a regular dewar cooled detector that is installed in the shield from the top.
Therefore, before installing the ICS detector and ICS, you must ensure that the detector endcap and metal
seals will easily fit through the shield’s opening from the bottom.
To decrease background streaming into the shield, the cooling rod neck length of the ICS (reference the “LL”
dimension in the ICS brochure) must be at least 10cm (4 inches) long and optional custom split plugs may be
needed.
For cooling rod neck lengths greater than 10cm (4 inches), the shield table may need to be raised or the ICS
Stand may require tilting when positioning under the shield table.
1.1. SPECIFICATIONS1
1.1.1. Performance
●Maintenance The ICS Stand doesn’t require periodic maintenance. Consult the detector and ICS
manuals for regular maintenance of the HPGe detector and the cryocooler.
1.1.2. Mechanical
●Stand Weight 26lbs.
●Maximum load 100lbs. The ICS Stand is designed to support the weight of any ICS configured HPGe
detector. It can also support the additional weight of the shield’s lead plugs which can be positioned
on top of the ICS Stand up to a combined total of 100lbs.
●Operating Temperature and HumidityNone specified for the ICS Stand, however, please consult
the ICS brochure for details on the ICS and associated detectors.
1.2. INSPECTION FOR SHIPPING DAMAGE
If the shipping carton(s) arrives with externally visible damage, do not unpack it. Notify the carrier and make
arrangements to file a damage claim. In all cases of shipping damage, it is the customer's responsibility
to file a damage claim.
If, during unpacking, you find concealed damage, notify the carrier and file a claim. Be sure to save all packing
materials, waybills, and other such documentation to establish claims.
Contact our Global Service Center or your ORTEC representative for further instructions.
1All system components and specifications subject to change without notice.

ICS Stand 2 1108265B
5
ICS STAND OVERVIEW
1.3. CONTENTS OF THE SHIPPING CARTON
When you receive the ICS Stand, its shipping carton will contain the following components:
●The ICS Stand main assembly.
●User manual and documentation package.
●Pneumatic air pump (for units shipped after November 2015)
1.4. UPACKING THE ICS STAND
1) Unpack the ICS Stand and make sure the stand is empty of any packing materials.
2) As shipped from the factory, the ICS Stand is fully assembled and only requires mounting of the HPGe
detector with the ICS.

ICS Stand 2 1108265B
7
Figure 1. ICS Stand 2 fully compressed.
Mount panel
Air piston
Schrader air valve
Directional
selection lever
Adjustable feet (4)
Plug support
plate

1108265B ICS Stand 2
8
2. ICS INSTALLATION ONTO THE ICS STAND
Tools required:
•One hex wrench (included)
•One pneumatic air pump with a Schrader valve, such as a bicycle pump (included)
2.1. THE ICS AND ICS STAND ASSEMBLY
1) Remove the plug support plate by lifting the plate off the main bracket (no tools necessary).
2) Connect the air pump to the Schrader valve.Lift the lever to lock it in place.

ICS Stand 2 1108265B
9
3) Flip the directional valve to the up position.
4) Use the air pump to raise the platform about 6 inches (150 mm).

1108265B ICS Stand 2
10
5) Place ICS on the platform. (ICS-P4 shown)

ICS Stand 2 1108265B
11
6) If a secondary plug is used in the shield installation, re-install the plug support plate. No tools are
necessary. The plate slides into position. Installation of the plate is meant to be snug. If a secondary
plug is not used, it is not necessary to reinstall the plug plate.
NOTE: The ICS and plug plate must be installed in the orientation shown. The handle(s) on the ICS
should be facing out as shown.

1108265B ICS Stand 2
12
3. INSTALLATION UNDER A TYPICAL 4” TOP LOADING LEAD SHIELD
3.1. Preparing the lead shield
1) The installation is for a typical top loading, 4” lead shield.
2) Remove any shield plugs that may be installed so that there is a clear opening in the bottom of the
shield large enough to easily accommodate the size of the detector capsule and endcap.
3) Verify the height of the shield bottom and/or lip of the shield stand and whether or not the ICS Stand
in the fully lowered position will have sufficient clearance to roll underneath, or if there will be
inference.
3.2. Positioning the ICS and ICS Stand with Adequate Shield Clearance
1) With the ICS Stand fully lowered, carefully slide the ICS underneath the lead shield. The adjustable
feet can be used to level the instrument when it is in place.

ICS Stand 2 1108265B
13
2) Continue to adjust the position of the ICS Stand to align the detector capsule with the center of the
access hole in the shield bottom.
3) Once centered, attach the pneumatic air pump, move the directional selection lever on the front of
the ICS Stand to the UP () position and slowly raise the ICS detector into the bottom opening of the
lead shield.

1108265B ICS Stand 2
14
4) Continue to raise the ICS detector until the proper position is reached. Disconnect the pneumatic air
pump and move the directional selection lever to the LOCK ( –) position.
CAUTION: The shield should not electrically contact the cryostat. Plastic sheeting or tape may
be used to insulate.
5) Reinstall any shield plugs as necessary.

ICS Stand 2 1108265B
15
3.3. Positioning the ICS and ICS Stand with Inadequate Shield Clearance
CAUTION: This procedure requires careful handling and alignment to prevent possible damage to the detector. It is
recommended to keep the protective cover on the endcap when installing. Additionally, this installation is NOT
recommended for a single person, and multiple individuals should be present to assist.
1) With the ICS Stand fully lowered, carefully roll the ICS assembly to the front of the lead shield.
2) Carefully lift and tilt the ICS stand and ICS detector towards off the floor and towards its back as if to
lay it on its back.
3) Now rotate the entire assembly underneath the shield while simultaneously aligning the detector
endcap into the access hole in the bottom of the shield.
4) Once centered tilt the assembly upright so that the detector endcap is positioned through the bottom
shield opening and the ICS assembly is resting on the floor.

1108265B ICS Stand 2
16
5) If necessary, adjust the position of the ICS Stand to align the detector capsule with the center of the
access hole in the shield bottom.
6) Once centered, attach the pneumatic air pump, move the directional selection lever on the front of
the ICS Stand to the UP () position and slowly raise the ICS detector into the bottom opening of the
lead shield.

ICS Stand 2 1108265B
17
7) Continue to raise the ICS detector until the proper position is reached. Disconnect the pneumatic air
pump and move the directional selection lever to the LOCK ( –) position.
CAUTION: The shield should not electrically contact the cryostat. Plastic sheeting or tape
may be used to insulate.
8) Reinstall any shield plugs as necessary.

1108265B ICS Stand 2
18
4. LOCKING THE ICS STAND 2 IN POSITION.
This is done after the ICS is installed on the stand and the ICS Stand II is in its desired
location in the shield. Using the hex wrench provided, torque the lock nut to a maximum of 130 in-
lbs (15 N-m) to lock the stand in position.
Table of contents
Other ORTEC Measuring Instrument manuals

ORTEC
ORTEC Micro-Detective-HX User manual

ORTEC
ORTEC 550A Service manual

ORTEC
ORTEC Detective-EX-100T User manual

ORTEC
ORTEC DSPEC Pro Instructions for use

ORTEC
ORTEC 935 Service manual

ORTEC
ORTEC 455 User manual

ORTEC
ORTEC 439 User manual

ORTEC
ORTEC RADEAGLET User manual

ORTEC
ORTEC 551 Service manual

ORTEC
ORTEC 418A Service manual