oserio FTG-315 series User manual

www.oserio.com
Body Fat Monitor Instrucon Manual – FTG-315 series
Body Fat Monitor
Thank you for selecng oserio series of Health Management Products.
This Instrucon Manual will guide you through the setup procedures and outline the key features.
Please keep the instrucon manual at hand all the me for future reference.
Manufactured by:
Charder Electronic Co., Ltd.
No.103, Guozhong Rd., Dali Dist.,Taichung City, 412 Taiwan (R.O.C.)
PA-0051-V1.0

1
1
2
16
18
19
20
Table of Contents
Before Using the Monitor
1.1 Overview
1.2 Baery Installaon
1.3 Product Features
1.4 Introducon
1.5 Important Safety Notes
1.6 Used Symbol Descripon
1.7 EMC Guidance and
Manufacturer's Declaraon
1
2
3
4
5
8
9
Operang Instrucons
2.1 Measuring Body Composion
2.2 Error Message List
More Informaon
3.1 Care and Maintenance
3.2 Product Specificaon
FTG-315 series
Body Fat Monitor
Front Rear
Baery Cover
Feet
Plaorm
Bluetooth Indicator
LED display
Before Using the Monitor
1.1 Overview

NOTE:
Be sure that the polarity of the baery is set properly.
2
Please verify if the baeries are installed correctly
Install 4 ”AA” size baeries properly
Open the baery cover
Before Using the Monitor
1.2 Baery Installaon
Before Using the Monitor
1.3 Product Features
3
1. Measure weight, body fat, BMI, visceral fat, muscle rate and body age.
2. Data reference from Body fatness and risk for elevated blood pressure, total cholesterol, and serum
lipoprotein in children and adolescents American journal of public health. Aer compared measurement
value, indicator will shows “Low, Normal, High, Very High”
3. Large LCD display (76 x 36mm).
4. User friendly design with colorful and interesng face icons.
5. Compact design.
Body Fat References:Daniel P. Williams, Sco B. Going, Timothy G. Lohman, David W. Harsha, Satharnur R Srinivasan, Larry S.
Webber, Gerald S. Berenson, Body fatness and risk for elevated blood pressure, total cholesterol, and serum lipoprotein raons in
children and adolescents American journal of public health, March 1992, Vol. 82, No.3
21

45
Before Using the Monitor
1.4 Introducon
Product descripon: This unit is measuring the user’s body weight and to esmate the value of
the Bio-Impedance body composion.
Product use: Body Fat, Visceral Fat, BMI, Muscle Rate, Body Age and Body Weight.
Essenal Performance: To supply the accurate measuring results which defined at Product Specificaon
either by FTG-315 itself or via a compable smart phone.
Measuring Principle: Biological impedance Analysis (BIA) works together with such as gender,
height, weight, age, other relevant factors, conversion of the user’s body
composion to determine whether the body is obese.
(If input profile is incorrect, it would result in incorrect measurement values.)
Recommended Measurement Times: It is recommended to use this unit in the same
environment and daily circumstances.
1. People with an electronic medical implant, such as a pacemaker, should not use a Body Fat Monitor
as it passes a low-level electrical signal through the body, which may interfere with its operaon.
2. Pregnant women should use weight funcon only. All other funcons are not
intended for pregnant women.
3. Keep out of reach of young children.
4. Do not jumping on the unit plaorm.
5. Do not step on the edge of unit plaorm.
6. Person with disabilies or physically frail should be assisted by another
person when using this unit.
7. Do not disassemble or modify the unit.
8. Using the baeries specify by the product specificaons, polaries match
the polaries of the baery compartment as indicated.
1.5 Important Safety Notes
Before Using the Monitor

76
9. Remove the baeries if the unit not be used for three months or longer.
10. If the display not shows, check the baeries.
11. Do not mix old and new baeries or different types of baeries to extend the life of the unit.
12. Do not use the unit on slippery surfaces such as le floors or wet floors.
13. Place the unit in an area free from direct sunlight, heang equipment, high humidity or extreme
temperature change.
14. Avoid excessive impact or vibraon to the unit.
15. Never submerge in water. Use alcohol to clean the electrodes and glass cleaner (applied to a cloth first)
to keep them shiny.
16. Place the unit on a hard, flat surface where there is minimal vibraon
to ensure safe and accurate measurement.
1.5 Important Safety Notes
Before Using the Monitor
17. Always remove your socks or stockings, and be sure the soles of your feet are clean before stepping
on the measuring plaorm.
18. Properly dispose of baeries according to applicable local regulaons.
19. This unit is intended for home use only. It is not intended for professional use in hospitals or other
medical facilies.
20. These baeries should be kept away from small children.
If swallowed , promptly seek medical assistance.
21. Expected Service Life: 5 years

8 9
Before Using the Monitor
1.7 EMC Guidance and Manufacturer's Declaraon
Guidance and manufacturer’s declaraon-electromagnec emissions
The FTG-315 is intended for use in the electromagnec environment specified below. The customer or the user of the
FTG-315 should assure that it is used in such an environment.
Emission test Compliance Electromagnec environment-guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Group 1
Class B
The FTG-315 uses RF energy only for its internal funcon. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment.
The FTG-315 is suitable for use in all establishments, including domesc
establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domesc
purposes.
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuaons/flicker
emissions
IEC 61000-3-3
Not applicable
Not applicable
Before Using the Monitor
1.6 Used Symbol Descripon
Explanaon of the graphic symbols:
SN-135-00001
Designaon of the serial number of every device,
applied at the device.
(Number as an example)
“Please note the accompanying documents“ or
“Observe operang instrucons”
Idenficaon of manufacturer of medical product
including address
Charder Electronic Co., Ltd.
No.103, Guozhong Rd.,
Dali Dist.,Taichung City,
412 Taiwan (R.O.C.)
“Electro-medical appliance“ with aachment for
type BF
Carefully read this operaon manual before
setup and commissioning, even if you are
already familiar with oserio scales.
Dispose of old appliances separately from your
household waste!Instead, take them to
communal collecon points.
Transport and storage temperature limit
indicang the upper and the lower limit.
(Transport and storage temperature on
packaging)
+60˚C
-20˚C

10 11
Before Using the Monitor
1.7 EMC Guidance and Manufacturer's Declaraon
The FTG-315 is intended for use in the electromagnec environment specified below. The customer or the user of the
FTG-315 should assure that it is used in such an environment.
Guidance and manufacturer’s declaraon-electromagnec immunity Guidance and manufacturer’s declaraon-electromagnec immunity
Immunity test IEC 60601 test level Compliance level Electromagnec environment-guidance
Immunity test IEC 60601 test level Compliance level Electromagnec environment-guidance
Electrostac
discharge(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
le. If floors are covered with synthec
material, the relave humidity should be at
least 30%.
± 2kV for power supply
lines + 1kV for
input/output lines
± 2kV for power
supply lines
Not applicable
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1kV line(s) to line(s)
± 2kV line(s) to earth
± 1kV differenal
mode
Not applicable
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage Dips,
short interrupt-
ions and voltage
variaons on
power supply
input lines
IEC 61000-4-11
<5% UT(>95% dip in UT)
for 0.5 cycle
40% UT(60% dip in UT)
for 5 cycles
70% UT(30% dip in UT)
for 25 cycles
<5% UT(>95% dip in UT)
for 5 s
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the FTG-315, requires
connued operaon during power mains
interrupons, it is recommended that the
FTG-315 be powered from an uninterrupble
power supply or a baery.
Not applicable
Not applicable
Not applicable
Not applicable
Power frequency
(50/60 Hz)
magnec field
IEC 61000-4-8
3 A/m 3 A/m The FTG-315 power frequency magnec
fields should be at levels characterisc of a
typical locaon in a typical commercial or
hospital environment.
NOTE: UT is the a.c. mains voltage prior to applicaon of the test level.

12 13
Before Using the Monitor
1.7 EMC Guidance and Manufacturer's Declaraon
Guidance and manufacturer’s declaraon-electromagnec immunity
Immunity test IEC 60601 test level Compliance level Electromagnec environment-guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 KHz to 80 MHz
Not applicable Portable and mobile RF communicaons equipment should be
used no closer to any part of the FTG-315 including cables, than
the recommended separaon distance calculated from the
equaon applicable to the frequency of the transmier.
NOTE1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2:
These guidelines may not apply in all situaons.
Electromagnec propagaon is affected by absorpon and
reflecon from structures, objects and people.
Field strengths from fixed transmiers, such as base
staons for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theorecally with
accuracy. To assess the electromagnec environment due
to fixed RF transmiers, an electromagnec site survey
should be considered. If the measured field strength in the
locaon in which the FTG-315 should be observed to
verify normal operaon. If abnormal performance is
observed, addional measures my be necessary, such as
re-orienng or relocang the FTG-315.Over the frequency
range 150 kHz to 80 MHz, field strengths should be les
than 3 V/m.
A.
B.
Radiated RF
IEC 61000-4-3
3 V/m
80MHz to 2.5 GHz
3 V/m
Where P is the maximum output power rang of the transmier in
was (W) according to the transmier manufacturer and d is the
recommended separaon distance in metres (m).
Interference may occur in the vicinity of
equipment marked with the following symbol:
Field strengths from fixed RF transmiers, as determined by an
electromagnec site survey , should be less than the compliance level in
each frequency range .
b
a
Recommended separaon distance:
d = 1.2 P
d = 1.2 P 80MHz to 800 MHz
d = 2.3 P 800MHz to 2.5 GHz

14 15
Before Using the Monitor
1.7 EMC Guidance and Manufacturer's Declaraon
0.01 0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
N/A
N/A
N/A
N/A
N/A
0.1
1
10
100
Recommended separaon distance between portable and mobile RF communicaons equipment and the FTG-315
The FTG-315 is intended for use in an electromagnec environment in which radiated RF disturbances are controlled. The customer or
the user of the FTG-315 can help prevent electromagnec interference by maintaining a minimum distance between portable and
mobile RF communicaons equipment (transmiers) and the FTG-315 as recommended below, according to the maximum output
power of the communicaons equipment.
Rated maximum output
power of transmier
(W)
For transmiers rated at a maximum output power not listed above, the recommended separaon distance d in meters (m) can be
esmated using the equaon applicable to the frequency of the transmier, where p is the maximum output power rang of the
transmier in was (W) according to the transmier manufacturer.
NOTE1:
At 80 MHz and 800 MHz, the separaon distance for the higher frequency range applies.
NOTE2:
These guidelines may not apply in all situaons. Electromagnec propagaon is affected by absorpon and reflecon from
structures, objects and people.
Separaon distance according to frequency of transmier
(m)
150 kHz to 80 MHz
d =1.2 P
80 MHz to 800 MHz
d =1.2 P
800 MHz to 2.5 GHz
d =2.3 P

6. LED bar running during connecng with App.5. The measurement completed. (show weight reading)
7. Turning off automacally aer transming .
Operang Instrucons
Note:
The unit will shut down automacally if the weighing data can’t be transmied in 30 seconds .
1. Place the unit on a hard, flat surface .
3. Step on the scale, LED bar will flash twice then
display user’s weight.
2. Run o’care APP and enter “ ” bubble .
4. Measuring Body Composion .
2.1 Measuring Body Composion
Note:
If you didn’t step on the scale but LED bar is flashing , please await ll the LED turn off or re-install the baeries. Such occasional display
might be caused by the interference of environment; however, it won’t cause the FTG-315 permanently damage / malfuncon and, is no
related with the basic safety.
Place the unit on a hard, flat surface to avoid inaccurate measurement.
16 17
5
66
5 SEC. OFF
---
kcal
--- kcal
---
LV
--- %
---
kg
Today
Weight
2
1 3
STEP ON
4

Operang Instrucons
2.7 Error Message List
The product is low power but sll
workable while “LO” flickers.
Connecng error.
Please try again or contact our service line .
Connecng error.
Please try again or contact our service line .
Replace baeries immediately while
“LO” appears.
Overloading or loading is unbalanced
within 1 minute. Please stand sll
during measurement.
1918
More Informaon
3.1 Care and Maintenance
1. Store the unit in a safe and dry locaon.
2. Avoid place on the aisle, to avoid accidentally tripped over
pedestrians.
3. Never submerge in water.
4. Do not expose the unit to direct sunlight, extreme hot or cold temperatures
or high humidity.
5. Avoid contact with corrosive chemicals or gas.
6. Clean the unit with a so dry cloth for daily use.
7. Wipe the plaorm and display using clean so cloth.
8. Do not use benzene, detergents or gasoline to clean the unit.
9. Avoid rough, scking cloth, to prevent scratching the plaorm and screen panels.
10. Prevent water infiltraon affecng the measurement accuracy.

0~40kg:±0.4 kg , 40~150kg:±1%
More Informaon
3.2 Product Specificaon
Model Body Fat Monitor FTG-315
Note: 1. Body fat measurement is for users age 10~80.
2. Specificaons may be changed without prior noce.
External Dimensions Main unit: 320mm(d) x 320mm(w) x 39mm(h) LCD: 76mm(d) x 36mm(w)
Weight Accuracy
Seng Range Age:10~80 Gender: Male/Female Height: 100~199.5cm
Power Supply Weight Approx. 1.8kg4 AA baeries
Operang Environment
Limit
Storage Environment
Limit
Temperature 5˚C to 35˚C
Humidity 30% to 85% RH
Temperature -20˚C to 60˚C
Humidity 10% to 95% RH
Maximum atmospheric
pressure range
Maximum operaon
altude
1013hpa – 800hpa 2000m
Scale Display Content
APP Display Content
Weight: 5 ~ 150kg, Graduaon: 0.1kg
Body Fat: 5.0 ~ 50.0%, Graduaon: 0.1%
Accessories
20
Body Fat Status Category: Low/Normal/High/Very High
Visceral Fat Status Category: 1~3(Normal)/4~6(High)/7~9(Very High)
BMI: 0.0~150.0, Graduaon: 0.1
Muscle Rate: 0.0~100.0%, Graduaon: 0.1%
Body Age: 18~80
Visceral Fat Level: Level 1~9 with 1 level increment
Manual, Warranty card, Connecng instrucons
Health conscious‧Professional‧Innovave
Table of contents
Other oserio Scale manuals

oserio
oserio BTG-365 Series User manual

oserio
oserio BBP-703 Series User manual

oserio
oserio FFP-331 SERIES User manual

oserio
oserio FTG-588 User manual

oserio
oserio BFP-219 User manual

oserio
oserio FTG-168 User manual

oserio
oserio MEG-213 User manual

oserio
oserio BHG-208 User manual

oserio
oserio FSC-330 User manual

oserio