OSKA Air Overlay User manual

User Guide
Air Overlay Mattress

Air Overlay Mattress User Guide 3
Please read the following instructions carefully and observe the
warning instructions before using the system.
Contents
03 About your mattress
04 System set up
05 Operation
06 System maintenance
06 Cleaning
07 Technical specication
08 Troubleshooting guide
09 Waste disposal
09 Warranty
10 Safety
10 Standard parts list
Edward House, 5 Penner Road, Havant, PO9 1QZ
About your mattress
The system
OSKA Air Overlay is an overlay system which is
designed to be placed onto an ordinary mattress
or an OSKA underlay. It is designed to be used
for nursing care at home or in a nursing facility
environment.
Application - in regular bed types is possible,
however, preference is given to a patient prole
nursing bed.
The system consists of a control unit featuring a
membrane pump and a bed overlay with 17 air
cells arranged in a transverse manner.The entire
cell unit is equipped with a bi-elastic PU protection
cover.
Recommendations
OSKA Air Overlay system is recommended to
be used for pressure ulcers up to and including
category 2 (medium to high-risk). For higher risk
patients you should use the appropriate system
from OSKA.
Capacity
OSKA Air Overlay system is recommended for
patient weight up to 130kg. Heavier patients are
better accommodated with a more appropriate
OSKA pressure system.

Air Overlay Mattress User Guide 5
System set-up
01. Place pump on horizontal surface or hang by
means of the hook.
02. Open mattress on the bed, secure it with the
end aps or straps. Air tubes must be at foot
end of bed.
03. Connect air feed tubes to air outlet on the
mattress and pump (see 1), ensuring that the
tubes are not kinked or twisted. Allow space
for bed rails to drop freely.
04. Turn on the pump power (see 2).The power
switch indicator will come on.The pump will
start pumping air into the mattress.
05. The low pressure indicator (see 3 yellow or
red) will come on as ination takes place.
06. The audible alarm won’t alarm within 60
minutes and will be activated if the pressure
is still low inside the mattress. In case of low
pressure, check the connection and products.
Press the alarm reset button (see 4) to mute
the audible alarm, it’s indicator will ash.
07. The pump will automatically default to setting
‘5’ whilst inating. Leave the mattress to fully
inate for 60 minutes before attempting to
change the pump setting.
1
45 3
2
Operation
01. When the pressure reaches the pre-set
level (within approximately 30 minutes), the
normal pressure indicator will come on and
the low pressure indicator will turn off.
02. Turn the pressure dial (see 5) to set a
comfortable pressure level from 40kg –
130kg according to the approximate weight
of the patient.
03. Press the static button to activate static
mode, the static indicator will come on.The
static mode will start within approximately
6 minutes. Press the static button again to
switch back to alternating mode.
Bi elastic vapour
permeable cover
Manual pump
CPR valve
Anchor end
straps

Air Overlay Mattress User Guide 7
System maintenance
Maintenance at regular intervals (mandatory every
2 years) is necessary to preserve the function of
the OSKA Air Overlay alternating pressure system.
Used lters must be exchanged within this period.
Maintenance/inspection is carried out by the
producer or an authorised rm or engineer at the
client’s expense.
If the maintenance is not performed within the
rst 2 years of the products life or not conducted
by an authorised rm or engineer then the
warranty claims become invalid. Damage or any
other loss of function arising from the lack of
maintenance or interferences by an unauthorised
party will also invalidate the product’s warranty.
Maintenance/inspection includes the following
procedures:
01. Replace membrane unit
02. Replace air lter
03. Replace synchronous motor
04. Conduct function test/function check
05. Sealing of case and attaching an inspection
sticker
06. Protective conductor check in accordance
with DIN VDE 0751
Cleaning
Dirt or stains should be removed using warm
water with a mild detergent and a cloth. Do not
use aggressive cleaning agents or solvents.
Regular wipe disinfection is recommended using
a suitable disinfectant cleaner and following
manufacturer’s guidelines.
The Top covers are machine washable. Care must
be taken, in order to maintain the integrity of the
covers.
Light Blue Cover – This cover can be machine
washed up to 70˚C for disinfection, following
manufacturer’s dosing directions.Tumble dry on
low heat and ensure cover is thoroughly dry
before replacing.
Dark Blue Cover – This cover can be machine
washed up to 95˚C for disinfection, following
manufacturers dosing directions.Tumble dry on
low heat and ensure cover is thoroughly dry
before replacing.
Technical specification
Pump
OSKA Air Overlay
Power supply: 220~240V/50Hz, 1A
Air output: 8 litre/min.
Pressure range: 30-60 mmHg
Cycle time: 12 min.
Static mode
Alarm reset mode
With normal/low pressure light
Size: 28cm(L) x 20cm(W) x 10cm(H)
Weight: 2.2 kg
Mattress
Size: 86.5cm(W) x 200cm(L) x 12.5cm(H)
Top cover type 1 – Light blue bi-elastic cover with
PU/Polyester coating and zipper.
Top cover type 2 – Dark blue bi-elastic cover with
PU/Polyester coating and zipper.
Air cells: Nylon / PVC
Base: Nylon / PVC
17 air cells
CPR valve for emergency procedures
Pillow function
Quick connectors

Air Overlay Mattress User Guide 9
Troubleshooting guide
The pump appears
not to be on/running
The mains supply is
plugged in.The switch
is in the on position.
The fuse in the plug is
not blown
Connect the appliance
to the nearest mains
socket and ensure that
the power indicator light
is on
Contact OSKA Service
The pump appears
to be on running
but is not reaching
the desired pressure
setting
The hoses are
connected correctly
Disconnect the hoses
and re-connect them in
to the outlet connectors
on the side of the pump
Contact OSKA Service
The pump is running
but is not inating
the mattress
The hoses are
connected correctly
Disconnect the hoses
and re-connect them in
to the outlet connectors
on the side of the pump
Contact OSKA Service
There are no leaks in
the mattress
Contact OSKA Service Contact OSKA Service
There are no kinks in
the tubing down the
side of the mattress
Carefully untwist any
kinks
Contact OSKA Service
Check that the
pressure has been
set in the correct
range for the patient’s
weight
Adjust the pressure
setting accordingly
Contact OSKA Service
The system does
not appear to be
alternating
There are no kinks in
the tubing down the
side of the mattress
Carefully untwist any
kinks
Contact OSKA Service
The operation pump
is noisy
The pump is not
resting against a solid
surface
Re-position the pump to
a suitable surface
Contact OSKA Service
Fault occurring Ensure that Possible solution If your product is still
not working
Waste disposal
The materials used to manufacture the OSKA Air
Overlay mattress do not cause any danger to the
environment.They may be disposed of with the
household waste.
The pump control unit is considered as electrical
equipment waste. Please contact the proper waste
disposal ofcials.
Warranty
Your OSKA product comes with a 1 year parts
and labour warranty.
The warranty is based on statutory regulations.
The warranty does not apply if the goods supplied
by OSKA are processed, handled or modied by
other parties without prior consent from OSKA
or if these operating instructions are not followed.
For medical devices as understood by the EU -
Directive 93/94, whose usage (also operations)
is subject to regular service and maintenance,
a warranty can be offered only if the service
intervals prescribed by OSKA are adhered to.
If during inspection it is found that damages
are due to wear and tear or are not subject to
warranty, then OSKA are empowered to claim the
expenses (inspection, transport costs etc) from
the client.
The use of parts or individual components of
other systems or products or combinations
thereof are not permissible.Warranty claims
become invalid should any damage result from
this kind of conduct and therefore remain the
responsibility of the user/owner.
WARNING!
For private non-professional operators. In the
case of system failure please check:
01. Is there sufcient power supply?
02. Is the unit turned on?
03. Is the power cord connected properly
and is the power plug inserted correctly?
04. Is the mattress connection tube
connected properly?
If all these primary checks have been carried
out and the system continues not to function
then you must contact the responsible
supplier or user who is authorised to effect
the maintenance and/or repair of this
equipment.
Never attempt to repair the system yourself.
Fuses may only be exchanged by qualied
and authorised personnel.

Air Overlay Mattress User Guide 11
Cells (Individual) WP82-088
Mattress (with pump) PR13-864
CPR Valve WP82-088
Inlet/outlet connector WP82-083
Air Overlay pump lter WP83-027
Safety
In the interest of your safety please be informed
that all OSKA products are certied in accordance
to the following regulations:
Testing for DIN EN 60601-1
CPR - emergency valve
ISO 13485
For your safety and the safety of those around
you, please adhere to all safety and warning as
detailed in this manual.
Standard parts list
Part description Part No.
WARNING!
You MUST be aware of the susceptibility of
this air mattress to the incidence of re. In
the case of a re the unavoidable release of
air acts as a fan effect to assist the ames.
You must avoid smoking at all times in the
immediate vicinity of this product. Other
hazards can include:The use of petroleum
based creams and gels such as Vaseline etc.
The manufacturer of this product takes no
responsibility for actions that are in breach
of this re hazard warning which includes
all environments: Hospitals, care homes and
private/domestic users.

Air Overlay Mattress User Guide
Edward House, 5 Penner Road, Havant, PO9 1QZ
oska.uk.com
Pressure Care Experts
Table of contents
Other OSKA Indoor Furnishing manuals
Popular Indoor Furnishing manuals by other brands

Invacare
Invacare H440 Omega Eco user manual

Homelegance
Homelegance 5686-32 Assembly instruction

Article
Article JOKUNA Assembly instructions

Sunset Furniture
Sunset Furniture Sunset Trading DLU-TPD-4242-B Assembly instructions

Mocka
Mocka Taylor Assembly instructions

NEXERa
NEXERa 609072 Assembly instructions

Ameriwood HOME
Ameriwood HOME 2456339COM manual

Powell
Powell 149-534A Assembly instructions

Wind Mill Slatwall
Wind Mill Slatwall RTA CUBE Assembly instructions

The Land of Nod
The Land of Nod Cube 6-cube Bookcase Assembly instructions

Holland House
Holland House 2678-30F Assembly instructions

Qaio
Qaio Flex Series installation manual