
9
2.0. EQUIPMENT AND INSTALLATION
2.1. EQUIPMENT SUPPLIED
1. Cabinet containing filter, flowmeter and pump, plus all associated electronics, displays and
ionization chamber/electrometer assembly
2. Detachable AC line cord, 10ft (3m) long
3. Manual in printed and electronic (CD-ROM) formats.
2.2. INSTALLATION
NOTE: NOT SUITABLE FOR USE IN A WET LOCATION. NOT SUITABLE FOR USE IN
EXPLOSION HAZARD ENVIRONMENTS
The following information is provided to the user to ensure stable and accurate performance.
The cabinet can be located on any flat surface, such as a table top, or, it can be mounted to a wall
bracket, or on a small moveable cart. In all cases, the instrument must be protected against
vibration, shock, moisture and dirt.
Gamma compensation of this instrument has been designed to correct for normal terrestrial
background, and small artificial gamma background. In the event that the instrument is to be used
in the presence of somewhat higher gamma background, the user may wish to construct a lead
shield to reduce radiation incident upon the monitor. Of course, there is provision for the instrument
to be used for the direct measurement of gamma. In this instance, the pump should be turned off,
after first flushing the sampling system with radio-logically clean air.
ELECTRICAL GROUNDING
The electrical and electronic equipment grounding is often considered only form the viewpoint of
hazard and safety. Indiscriminate or excessive grounding may actually enhance the potential of
danger and disturb the proper internal operation of the instrument. The electronic circuitry, including
logic, adjustment controls, and local and remote displays, are centrally and all inclusively grounded
at the ionization chamber module. The circuit system common line is electrically connected to the
metal frame or housing of the electrometer module. When signal outputs are connected to remote
displays, computer interfaces, or similar devices, it is necessary that no significant ground potential
differences exist between the monitor and other equipment. If significant potential ac or dc
differences exist, shifts in the instrument “zero” can appear.
THE FOLLOWING IS RECOMMENDED:
1. Make all interconnections. The AC power shall be connected using the flexible line cord
provided. Activate total instrument. Allow ten minute “warm-up”. Adjust zero if needed.
2. Attach remote connections (devices) and verify absence of change in zero.
If the zero has changed, check for ground loops and spurious ac or dc potential differences from
one location to the other.
Select a site for the instrument, turn off the main power switch and the switch for the pump.
Attach all remote equipment with the wiring as supplied by OTC, or as designed locally for interface
or remote alarms. Refer to 2.3.1. for important information.
CAUTION: Do NOT replace the supply cord with an improperly rated one, for additional
information, refer to the Safety Notice at the beginning of this manual